Thank you for purchasing counting electronic scale, in order to effectively help you to use our products correctly,
please read the manual carefully, which will help you to operate and extend the life of the products, and reduce
inconvenience.
1. Safety requirements:
1. Check if ac input power is matched with local ac power
2. Do not drop items on the scale plate when storing or using
3. Do not put any object under the weighing plate
4. Power must be cut off before cleaning and maintenance
5. Please weigh within the specified range when weighing
6. Shall not be used in bad environment
7. When taking the electronic scale, do not force on the bracket or weighing plate
8. The maintenance of electronic scale must be maintained by authorized professional personnel
2. Packing list
1
2
3
4
5
Electronic scale
Scale plate
Power cord
Operating
instruction
Certificate of
quality
3. Technical parameters
Rated voltage: AC/220V 50Hz
Power consumption: 3W
Working temperature: -10
℃
to 40
℃
Storage temperature: -20
℃
to 50
℃
Operating humidity: ≤85%R.H
Storage temperature ≤70%R.H
Range of zero setting: ≤±3%FS
Less than or equal to 3% maximum weighting
Peeling range: derest is the maximum weight (derest + net weight = maximum weight)

4
4. Specification
5. Preparation
1. Please place the electronic balance on a stable and flat table for use, do not place it on a shaking or
vibrating table, and use the four adjusting feet to keep the balance body stable. Please note that the
spirit bubble should be located in the center of the circle.
2. Please use a separate power outlet to avoid interference from other electrical appliances. The first
use should be charged for more than 12 hours.
3. When the power is on, please do not put anything on the scale plate.
4. When used, the center of gravity of the weighing object must be located in the center of the scale
plate, and the weighing object should not exceed the scope of the scale plate, to ensure its accuracy.
5. Before using the electronic balance, please heat up the engine for 15 ~ 20 minutes.
Maximum capacity
(g)
3000
6000
15000
30000
Minimum capacity
(g)
20
40
100
200
verification scale
intervale(g)
1
2
5
10
Division (g)
0.1
0.2
0.5
1
Minimum sample
weight(g)
2
4
10
20
Minimum single weight
(g)
0.15
0.3
0.75
1.5
Minimum sampling
quantity(PCS)
10
10
10
10
repeatability(g)
±0.2
±0.4
±1.0
±2.0
linearity(g)
±0.2
±0.4
±1.0
±2.0
Display
3*LCD backlight display
Keys
20 touch keys
Accuracy
Ⅲ
Plate size(mm)
295*230

5
6. Display and key panel description
7. Keyboard Functions
7.1. Keys
1. [PCS SET]: place a sample on the scale plate, input the number of samples through the digit keyboard
[also can input the number of completed samples in advance], press [PCS SET] the ‘UNIT WEIGHT’
window display ‘SnnP’ and the ‘QUANTITY(pcs)’ window display ‘------’, and when the number in the
‘QUANTITY(pcs)’ window flashes, start to sample the number of single volume, after the sample is
finished, the buzzer will make a sound, and the display screen will display the set number and average
single volume.
2. [U.W. SET]: Input a known sample single weight through the keyboard, press [U.W. SET] when the
number window flashes, and then place the sample to be counted on the scale plate.
3. [ALARM]: By entering the number of alarm upper limit, press the [ALARM] to light up the ‘PSET’ sign,
when the count amount exceeds the preset value, the ‘UNIT WEIGHING’ window displays alternately -
Oty- and single weight with a buzzer sound.
4. [AC]: If you need to cancel the number alarm setting, press [AC] the ‘PSET’ symbol will go off.

6
5. [M+]: used for the accumulative number of counting. Under the accumulative state, the ‘M+’ is lit up,
and the accumulative number is not more than 999999 PCS, and the accumulative number is not more
than 99 times.
6. [MC]: clear cumulative total, cumulative symbol off.
7. [TARE]: it is used to remove the preset tare weight. In the ‘TARE’ state, the ‘TARE’ sign is lit up.
Remove the preset tare weight and press the [TARE] to turn off the ‘TARE’ sign.
8. [ZERO]: When the ‘WEIGHT’ window ‘ZERO’ sign is unstable or there is a small weight does not
return to zero, press [ZERO] key to set zero.
9. [C]: Press the [C] key when entering quantity or single error to clear and re-enter immediately, or wait
5 seconds to enter again, the previous error will be automatically updated.
7.2. Indication sign description
UDSP
Sample number less than 10 PCS (the more sample number, the better count)
UDWT
Count sample single weight less than 1.5e (show 1.5 times the scale value)
AUTO
In the process of loading counting samples, the average weight of samples is intelligently
optimized according to the single weight and number of samples
PSET
Indicating symbol lighting, if the number of counting exceeds the preset value single weight,
the window will display -Oty- and single weight alternately with a buzzer
9. Menu Settings
8.1 When starting up self-check 0~9 hop number, quickly enter the user's calibration password 9011

7
8.2. Description of submenu Settings
WEIGHT
window
UNIT WEIGHT
window
FUN01-DIV
Indexing value
setting
WEIGHT
window
UNIT WEIGHT
window
FUN02-PNT
Decimal position
setting
DIV
1
PNT
0
DIV
2
PNT
0
DIV
5
PNT
0
DIV
10
PNT
0
DIV
20
PNT
0
DIV
50
PNT
0
WEIGHT
window
UNIT WEIGHT
window
FUN03-CAP
Full range
setting
WEIGHT
window
UNIT WEIGHT
window
FUN04-CAL
Factory linear
calibration
CAP
0
CAL
1
Enter the maximum weight directly
through the numeric key.
CAL
2
WEIGHT
window
UNITWEIGHT
window
FUN05-TRA
Zero tracking
setting
CAL
3
TRA
0
CAL
4
TRA
0.5
CAL
5
TRA
0.1
WEIGHT
window
UNIT WEIGHT
window
QUANTITY
window
TRA
1.5
0
PUT
AU-CAL
TRA
2
Add the maximum weight to
calibrate,the weight and stabilize it
for 3 seconds,After completion of
automatic calibration exit
FUN06-USR
User automatic
calibration
TRA
2.0
TRA
2.5

8
TRA
3
Direct input needs to load the value
of calibration weight,Load the
calibration weight,press
[ALARM],Exit after completion of
calibration.
FUN06-USR
User manual
calibration
TRA
3.5
TRA
4
WEIGHT
window
UNIT WEIGHT
window
FUN07-LIT
Backlight mode
setting
WEIGHT
window
UNIT WEIGHT
window
FUN08-BUD
Baud rate Setting
(this function is
temporarily
unavailable)
LIT
Au
BUD
1200
LIT
On
BUD
2400
LIT
Off
BUD
4800~19200
WEIGHT
window
UNIT WEIGHT
window
FUN09-ODE
Communication
format setting
[this function is
temporarily
unavailable]
WEIGHT
window
UNIT WEIGHT
window
FUN10-UPT
AUTO setting
ODE
0
UPT
On
ODE
1
UPT
On
ODE
2
WEIGHT
window
UNIT WEIGHT
window
FUN11-ALR
Alarm setting
ODE
3
ALR
Cnt
ALR
Elt
Description: in the above steps, press [warning preset] to enter the sub-menu, press [quantity setting] to
select, press [warning preset] to save and exit, and then press [quantity setting] to enter the next step.
10. Quick calibration by users (not applicable for the initial calibration of replacement
mainboard or sensor)
When starting up self-check 0~9 hop number, quickly enter the user's calibration password [9158].
WEIGHT
UNIT WEIGHT
QUANTITY
0.0
put
Au-cal
After loading the calibration weights above the half-range and stabilizing for 3 seconds,
the automatic calibration completes and exits

9
11. Trouble Removal
The following table shows the possible causes and solutions of common problems and failures. It is for
reference only. If you still have any questions, please contact the authorized dealer.
Symptoms
Possible reasons
Check solution for reference
Can not turn on
No power connection, no battery
Check that the power connection and
battery voltage are correct
Instability
[hop number]
The use environment is unstable
Move the electronic counter to the appropriate
environment
Accuracy is poor
Improper calibration
Calibrate
Can't calibration
Unstable environment
Move the electronic counter to the
appropriate environment
The backlight
doesn't work
Mode off
Enter the menu Settings activation mode
Alarm is invalid
Mode off
Enter the menu Settings activation mode
Battery symbol flicker
Low battery
Charge and replace the battery in time
Battery failure
Battery out of service
Contact the company authorized dealer to
replace the battery
The scale is not full
The zero of the sensor is too high
and the preset tare is too heavy
Replace the sensor and remove the preset tare
weight on the scale
-------
overload
Reduce the load on the scale

10