Accutrol Iso-Tek ISO User manual

© 2023 ACCUTROL LLC
Manufactured by: Accutrol Representative:
ISO-TEK MANUAL REV B 11-30-2023
ECN2918
Room Pressure Monitor
21 Commerce Drive
Danbury, CT 06810 USA
WWW.ACCUTROLLLC.COM
NORMAL
Isolation Room
August 15 2022 4:44
NEGATIVE
i
OCCUPIED
0.05" WC
24% RH
2 ACH
73F

Installation & Operation Manual - Model ISO
This Page Left Intentionally Blank

Installation & Operation Manual - Model ISO
ii
LIMITED WARRANTY
Accutrol LLC, having its principal place of business at 21 Commerce Drive, Danbury, CT USA
("Manufacturer") warrants its Iso-Tek® Room Pressure Monitor, as follows:
1. Limited Warranty.
Manufacturer warrants that the Products sold hereunder will be free from defects in material
and workmanship for a period of twelve (12) months from the date of purchase. If the Products
do not conform to this Limited Warranty during the warranty period (as herein above specified),
Buyer shall notify Manufacturer in writing of the claimed defects and demonstrate to
Manufacturer’s satisfaction that said defects are covered by this Limited Warranty. If the defects
are properly reported to Manufacturer within the warranty period, and the defects are of such
type and nature as to be covered by this warranty, Manufacturer shall, at its own expense,
furnish, replacement Products or, at Manufacturer's option, replacement parts or services for
the defective Products. Shipping and installation of the replacement Products or replacement
parts shall be at Buyer's expense.
2. Other Limits.
THE FOREGOING IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT
NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. Manufacturer does not warrant against damages or defects arising out
of improper or abnormal use or handling of the Products; against defects or damages arising from
improper installation (where installation is by persons other than Manufacturer), against defects
in products or components not manufactured by Manufacturer, or against damages resulting
from such non-Manufacturer made products or components. Manufacturer passes on to Buyer
the warranty it received (if any) from the maker thereof of such non-Manufacturer made
products or components. This warranty also does not apply to Products upon which repairs have
been affected or attempted by persons other than Manufacturer or pursuant to written
authorization by Manufacturer. This warranty also does not apply to any product provided by
the Buyer and mounted by the Manufacturer to Products.
3. Exclusive Obligation.
THIS WARRANTY IS EXCLUSIVE. The sole and exclusive obligation of Manufacturer shall be to
repair or replace the defective Products in the manner and for the period provided above.
Manufacturer shall not have any other obligation with respect to the Products or any part
thereof, whether based on contract, tort, and strict liability or otherwise. Under no
circumstances, whether based on this Limited Warranty or otherwise, shall Manufacturer be
liable for incidental, special, or consequential damages.
4. Other Statements.
Manufacturer's employees or representatives' ORAL OR OTHER WRITTEN STATEMENTS DO NOT
CONSTITUTE WARRANTIES, shall not be relied upon by Buyer, and are not a part of the contract
for sale or this limited warranty.
5. Entire Obligation.
This Limited Warranty states the entire obligation of Manufacturer with respect to the Products.
If any part of this Limited Warranty is determined to be void or illegal, the remainder shall remain
in full force and effect.

Installation & Operation Manual - Model ISO
iii
TABLE OF CONTENTS
LIMITED WARRANTY................................................................................................................ii
SECTION 1 –INTRODUCTION .................................................................................................. 1
SECTION 2 –SPECIFICATIONS................................................................................................. 3
SECTION 3 –PHYSICAL DIMENSIONS ................................................................................... 4
SECTION 4 –INSTALLATION.................................................................................................... 5
SECTION 5 –WIRING.................................................................................................................. 7
SECTION 6 –ISO-TEK INSIGHT CONFIGURATION SOFTWARE INSTALLATION.......... 8
SECTION 7 –CONFIGURING ISO-TEK USING INSIGHT .................................................... 11
SECTION 8 –TOUCHSCREEN NAVIGATION ....................................................................... 19
SECTION 9 –MAINTENANCE ................................................................................................. 23
APPENDIX A –BACnet MS/TP................................................................................................. 24
APPENDIX B: BLUETOOTH..................................................................................................... 27

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 1
SECTION 1 –INTRODUCTION
The Accutrol® Iso-Tek®is comprised of a high-resolution 7-inch color touch-screen display, a
robust IO Module, and an advanced pressure sensor kit that is designed to measure ultra-low-
differential pressures with a high degree of precision and excellent long-term stability. Field
configuration of the Iso-Tek is made simple using the Accutrol Insight™configuration software.
1.1 Theory of Operation
The Iso-Tek uses a High Pneumatic Impedance Sensor (HPIS®) which refers to the sensor’s
ability to avoid contamination due to the extremely small amount of air that flows through it.
What makes the HPIS unique is that it incorporates a micro-flow-channel etched into the silicon
wafer. This architecture makes the flow channel substantially smaller than other thermal-
based-flow-through sensors, thereby eliminating the disadvantages normally associated with
this type of sensor (see Figure 1).
Top
Wafer
Bottom
Wafer
Micro-flow
Channel Measuring Element
Flow
Flow
Heating
Element
Temp
Sensitive
Resistor
Temp
Sensitive
Resistor
Measuring Element Detail
Figure 1 –HPIS Sensing Element Used in the Iso-Tek
Advantages of HPIS®
The HPIS combines the best performance features of the diaphragm-based-dead-ended and
thermal-based-flow-through type sensors. The disadvantages normally associated with flow-
through sensors have been addressed through the high input pneumatic impedance of the
micro-flow channel which virtually eliminates airflow through the sensor. The airflow through
the microchannel with a DP = 0.10” WC (25 Pa) is only 0.000005 CFM resulting in an inlet
velocity so low that it is not possible for contaminants to be drawn inside.

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 2
1.2 Model Code
DISPLAY TYPE
1 = 7" Touch Screen
DISPLAY CABLE LENGTH
1 = 25' (7.6m)
2 = 50' (15.2m)
3 = 75' (22.9m)
4 = 100' (30.5m)
ISO -OPTIONS
BLANK = No Options
M = BACnet MS/TP
W = Wireless Bluetooth
SENSOR RANGE
1 = -0.1" TO +0.1" wc
NUMBER OF ROOM SENSOR KITS
1 = 1 Room Sensor Kit
2 = 2 Room Sensor Kits
3 = 3 Room Sensor Kits
4 = 4 Room Sensor Kits
1.3 Parts Included in the Iso-Tek System
IO Module w/Options Display w/(4) #6-32x1” Screws Display/IO Cable
Room Sensor Kit with Mounting Hardware
Sensor Probe Reference Probe
(2) Sets of #6-32x1”
Screws w/Gaskets

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 3
SECTION 2 –SPECIFICATIONS
PRESSURE SENSOR
Pressure Range
+/- 0.100”wc (+/- 25Pa)
Pressure Measurement Resolution
0.00000335”wc (0.00083Pa)
Zero Pressure Offset Accuracy
Typical: +/- 0.00016”wc (+/- 0.04Pa)
Max: +/- 0.0004”wc (+/- 0.1Pa)
Span Accuracy
Typical: +/- 0.75% of reading
Max: +/- 1.5% of reading
Thermal Effects -4 to 185 deg F (-20 to 85 deg C)
Offset: Max +/- 0.0004”wc (+/- 0.1Pa)
Span: Typical +/- 0.2% of reading per 18 deg F (10 deg C)
Max +/- 0.5% of reading per 18 deg F (10 deg C)
Long Term Stability
+/- 0.00008”wc (+/- 0.02Pa) per year max
Status Indicators
Green LED to indicate status
Connections
4-pos. pluggable screw terminal
Note: connect to IO Module using Windy City Wire # 042003 or equivalent
Operating Temp Range
-4 to 176 deg F (-20 to 80 deg C)
Storage Temp Range
-40 to 185 deg F (-40 to 85 deg C)
Humidity
0 to 95% noncondensing
DISPLAY (Local HMI)
Display Type
7” LCD-TFT Resistive touch display with LED backlight
Color
RGB 65K colors
Active Area
6.49” x 3.937” (164.9mm x 100mm)
Resolution
800 x 480 pixel
Speaker
audible notifications, 550~18KHz, 83dBA SPL @ 1W/0.5M
Connections
RJ-45 connector
Note: connect to IO Module using factory-provided cable only
Configuration Port
USB 2.0, Isolated, Type C Conn., connect PC with Iso-Tek Insight
Operating Temp Range
-4 to 150 deg F (-20 to 65 deg C)
Storage Temp Range
-40 to 185 deg F (-40 to 85 deg C)
Humidity
10 to 90% noncondensing
IO MODULE
System Input Power
24VAC +20/-10%, 22VA, 50-60Hz or 24 VDC +/- 10%, 10W
Max Power requirements for fully loaded system: 15VA or 9 Watts
Relay Outputs
2 Independent SPST Relay Contacts (Software Configurable)
Contact Rating; 2A @ 30VDC, 0.5A @ 60VDC, 0.3A @ 125VAC
Analog Outputs
2 Analog Outputs for Room Pressure (Software Configurable)
0-5v, 1-5v, 0-10v, 2-10v, 0-20mA or 4-20mA
capable of driving 1 K-ohm load
Door Switch Input
Dry Contact Input for Primary Room Door Switch (software configurable)
Accutrol Sensor Port
5-pos. pluggable screw terminal
Note: connect to Pressure Sensor Network using Windy City Wire # 042003 only
Accutrol Display Port
RJ-45 connector
Note: Connect to Display Module using factory-provided cable only
BACnet MS/TP (Optional)
EIA 485 2-wire, BACnet MS/TP, Galvanically Isolated
1/8 unit load transceiver impedance
Full Master Node State Machine
Data Rates 9600, 19200, 38400, 76800 and 115200
MAC address is software configurable
BACnet IP (Optional)
Future
Status Indicators
LED Indicators for Power, Pressure Sensor, Display, BACnet, & Bluetooth
I/O Terminal Blocks
Removable Vertical Plugs, Wire Size Range 12-30 AWG
Bluetooth (Optional)
Bluetooth Version 4.2 or later, connect to PC with Iso-Tek Insight
Operating Temp Range
-20 to 176 deg F (-29 to 80 deg C)
Storage Temp Range
-40 to 185 deg F (-40 to 85 deg C)
Humidity
0 to 95% noncondensing
COMPLIANCE
RoHS, UL94-V0

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 4
SECTION 3 –PHYSICAL DIMENSIONS
3.1 Display Module
8.50"
(216mm)
7.00"
(178mm)
1.40"
(102mm)
1.05"
(27mm)
3.65"
(92.7mm)
3.2 IO Module
8.08" (205mm)
7.63" (194mm)
6.23"
(158mm)
5.26"
(134mm)
1.15"
(30mm)
Antenna
(wireless option)
3.3 Sensor Probe
3.25"
(82.6mm)
5.0"
(127mm)
0.73"
(18.5mm)
1.42"
(36.1mm)
Flex-tube
2.5"
(63.5mm)
3.28"
(83.3mm)
3.4 Reference Probe
Reference Probe
3.25"
(82.6mm)
5.0"
(127mm) 3.28"
(83.3mm)
0.73"
(18.5mm)
Flex-tube
2.5"
(63.5mm)
2.62"
(67mm)
3.05"
(78mm)
1.40"
(102mm)
1.05"
(27mm)
3.65"
(92.7mm)

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 5
SECTION 4 –INSTALLATION
4.1 Display Module
The Iso-Tek Display Module is designed to be installed onto a double or triple-gang e-box
(provided by others) typically located at eye level directly outside of the room being monitored.
4.1.1 Installation Steps
1. Insert Display Cable from e-box through the hole in
the Mounting Plate.
2. Position Mounting Plate with the arrow UP.
3. Secure Mounting Plate to e-box using the (4) #6-32
x 1" screws provided.
CAUTION: Do Not Overtighten Screws.
Overtightening may deform Mounting Plate.
4. Plug the Display Cable into the RJ45 connector located
at the back of the Display.
5. Align the three nubs on the top of Display Bezel with
the three slots on the top of the Mounting Plate and
rotate the Display into position over the Mounting Plate.
6. Secure the Display to the Mounting Plate by tightening
the (2) screws located at the bottom of the Mounting
Plate.
4.2 IO Module
The Iso-Tek IO Module is typically located in the vicinity of the room above the ceiling or inside
a control panel located within reach of the Display Cable.
4.2.1 Installation Steps
1. Select a location that is easily accessible with enough clearance to remove the cover
and access the IO terminal blocks for wiring.
NOTE: Iso-Tek IO Modules with the wireless option (-W) shall NOT be installed
inside of a metal control panel or enclosure because the RF signal will be blocked.
2. The IO Module enclosure has four mounting flanges
with 0.2” diameter (5mm) holes. Secure IO Module to
mounting surface using (4) #8 or #10 screws (provided by
others). For installation over sheetrock walls, use
appropriate wall anchors (provided by others).

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 6
4.3 Sensor Probe and Reference Probe
The Sensor and Reference Probes are designed to be installed onto a single-gang electrical box
(e-box). The Sensor Probe shall be installed inside the pressurized space and the Reference
Probe shall be installed outside the pressurized space in a location designated as the reference
pressure zone for the space. Both probes shall be installed in areas that will not be influenced
by air currents from supply diffusers, fans, personnel, etc.
4.3.1 Installation Steps
1. Install the e-box level & flush with finished wall surface.
2. Run field tubing (by others) from Sensor Probe to the Reference Probe.
3. Each probe is provided with a short length of kink-proof tubing with a ¼” barb fitting.
Connect the field tubing to the barb fitting on the Sensor Probe and Reference Probe.
Factory Supplied
kink-proof tubing
Sensor
Probe
Field tubing: 0.25" (6mm) O.D., .040"
(1mm) wall, NFPA 90 flame retardant
polyethylene control tubing, such as
Chevron Plexco or equivalent
approved by local building codes.
4. Remove the diamond shaped airflow deflector
plates from both probes to reveal mounting holes
by pulling outwards on the side tabs.
5. Position the Probe with the arrow UP and secure
to e-box using the (2) #6-32 x 1" long screws with
rubber seals provided.
CAUTION: Be sure to use
the screws provided with seals
to prevent airflow leakage.
CAUTION: Do not overtighten mounting
screws. Overtightening may deform the mounting
plate.
6. After installation is complete, reattach the airflow deflector plate.

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 7
SECTION 5 –WIRING
One Iso-Tek can monitor up to four rooms. The diagram below shows a 4-room application.
Pnet
Sensor
Network
BACnet
MS/TP
Network
BMS Ctrl.
+
-
Analog
Inputs
Dry
Contact
Inputs
OR
24 VAC
CAUTION:
Maintain power
polarity otherwise
equipment may
be damaged.
DI
DI
AI
AI+
-
BACnet
MS/TP +
-
+
-
End of
MS/TP
Network
Segment
To Next
Device
Factory
Cable
VCC
COM
Pnet (-)
Pnet (+)
RPM
CONNECT EOL
DOOR
SW
ADDR
2 3 4
Iso-Tek
Room
Pressure
Sensor
Room 2 Sensor
VCC
COM
Pnet (-)
Pnet (+)
RPM
CONNECT EOL
DOOR
SW
ADDR
2 3 4
Iso-Tek
Room
Pressure
Sensor
Room 1 Sensor
T
T
24 VDC
Note 1
Note 1
Note 1
Note 4
Note 4
Note 3
Note 3
Note 1
Note 1
Room 1
Door
Door SW
Room 2
Door
Door SW.
Analog
Output 1
Shield
24 VAC 50/60
Hz or
24 VDC +/-10%
Analog
Output 2
+
-
+
-
Relay 1
Output
Relay 2
Output
+
-
Pnet -
ORG
Pnet +
WHT/ORG
Shield
VCC
BLU
COM
WHT/BLU
ACCUTROL
SENSOR
BACnet
MS/TP
Module
+
-
DRY
CONTACT
INPUT
ACCUTROL
DEVICE
NETWORK
ACCUTROL
DISPLAY
+
-
Shield
Iso-Tek IO Module
Input Power
Iso-Tek Display VCC
COM
Pnet (-)
Pnet (+)
RPM
CONNECT EOL
DOOR
SW
ADDR
2 3 4
Iso-Tek
Room
Pressure
Sensor
Room 3 Sensor
T
Note 3
Note 4
Room 3
Door
Door SW.
VCC
COM
Pnet (-)
Pnet (+)
RPM
CONNECT EOL
DOOR
SW
ADDR
2 3 4
Iso-Tek
Room
Pressure
Sensor
Room 4 Sensor
T
Note 3
Note 4
Room 4
Door
Door SW.
Cable Shields: Connect cable shields at each device & terminate
shield to ground at one end of network segment only.
1.
WIRING NOTES:
3.
2.
4.
Pnet Address: Set Room 1 Sensor address to 1 by removing the ADDR
jumper or parking it on one pin. Set additional ADDR jumpers in
sequence; 2, 3, 4.
Pnet End of Line: Set the EOL Jumper to the terminate (T) position on
the last sensor only. All other sensors should have the jumper
removed.
EOL Termination & Bias: BACnet network bias & end-of-line
terminations are NOT provided by the Iso-Tek BACnet MS/TP module.
BACnet MS/TP bias and EOL termination shall be provided by others.
Iso-Tek I/O Module Wiring Table
24v Input Power
Twisted-pair, 18-22AWG
BACnet MS/TP
Twisted-shielded-pair, low capacitance com cable
Accutrol Pnet Sensor Network
Windy City Wire # 042003 or Equivalent
Relay Outputs & Dry Contact Inputs
Twisted-pair, 18-22AWG
Analog Outputs
Twisted-shielded-pair cable, 18-22 AWG

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 8
SECTION 6 –ISO-TEK INSIGHT CONFIGURATION SOFTWARE INSTALLATION
Iso-Tek Insight is a software configuration tool that provides a means to communicate directly
with the Iso-Tek Room Pressure Monitor. All that’s required is a Windows PC or tablet with a
USB Port, USB cable, and the Iso-Tek Insight Software. For wireless communication, the Iso-Tek
can be ordered with an optional Bluetooth module.
6.1 Minimum Requirements for PC
•Windows PC or Tablet with Windows Version 10 or greater
•USB 2.0 Port and USB to USB-Type C Cable
•Bluetooth v4.2 or later required for Iso-Tek with “-W” wireless option.
6.2 Installing the Iso-Tek Insight Software
1. If the Accutrol Software Manager (ASM) is already installed on your Windows Device,
skip to Step 3. Otherwise, download the ASM from using the following link:
https://www.accutrolllc.com/accuvalve-software-downloads/
2. Go to the downloads folder in the File Explorer app of your PC and Extract All contents
from the folder. Once done, install the ASM by executing the Setup File within the
extracted AccutrolSoftwareSuite.zip file. A desktop icon will appear.
3. Once installed, launch the ASM using the Desktop Icon shown here.
4. Now select the red ISO icon to download the Iso-Tek Insight install folder to
your PC. This will launch the Default Browser on your device.
5. Extract All contents from the folder and execute the Setup File to install the program.
6.3 Registering the Iso-Tek Insight Software
1. After Iso-Tek Insight has been installed on your device, open the ASM and select the ISO
Insight icon. The yellow registration window shown below will appear.
2. Copy the Insight Serial Number then select “Click to Go to Website” button.
3. Complete the product registration form on the
Accutrol website and select “submit” when
completed.
4. The Accutrol customer service team will evaluate
the key request and provide the software key via
email within 1 business day if approved.
Serial Number
Web Registration Form
Link to Web Registration Form

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 9
6.4 Activate Iso-Tek Insight Software Key
After receiving the Product Key, verify your device is connected to the internet and perform
the following steps to unlock the software.
1. Open the Accutrol Software Manager and select the ISO Insight application.
2. The yellow registration window shown below will appear.
3. Copy the Product Key provided in the email from Accutrol and paste it into the
“Enter Product Key” field provided and select the “Click to Activate” button.
4. You will now be able to use the ISO Insight Software on your computer. If you have
any problems, contact Accutrol LLC at 203-445-9991.
NOTE: The Product Key is only valid for the computer which it was registered for.
6.5 Connecting to the Iso-Tek with a USB Cable
6.5.1 Verify the Iso-Tek system is powered, then connect the PC to the USB port located
at the bottom of the Iso-Tek display using a USB cable.
6.5.2 Open ISO Insight and select “Locate USB Devices” from the connection window.
NOTE: If you receive the “No Accutrol Device Located” message shown above, the
computer is not able to communicate with the Iso-Tek over the USB port. First check
your USB cable connections. If the USB connections are good, it’s likely your computer
does not have the appropriate USB driver package installed. Refer to Section 6.5.2.1 for
instructions on installing the required USB drivers.
Windows Device with
ISO Insight Software
USB Cable
with C-type
Connector
Iso-Tek Display

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 10
6.5.2.1 Installing USB Drivers
1. Navigate to the file folder titled “ISO Insight”which is included in the Iso-Tek
Install Zip file previously downloaded.
2. Within this folder is another Zip file called CP210x_Windows_Drivers.
3. Extract, then view the contents of this .zip file.
4. Find the .exe files called “CP210xVCPInstaller” and run the x64 version.
5. After installing the CP210 USB drivers, open ISO Insight and select “Locate USB
Devices”. Select the COMx port that appears to connect to the Iso-Tek.
6.6 Connecting to the Iso-Tek with Wireless Bluetooth
6.6.1 Verify the Iso-Tek system is powered and make note of the device serial number
that appears on the Iso-Tek Display which you are attempting to connect to.
ISO000671380012
6.6.2 Select “Locate BT Devices” then select the discovered device SN associated with
the Iso-Tek SN that you would like to connect to. Reference Appendix B for details on
the Bluetooth connection.
NOTE: You will be unable to connect using wireless if the USB cable is plugged
into the device.
Device SN
Device SN
ISO000671380012

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 11
SECTION 7 –CONFIGURING ISO-TEK USING INSIGHT
7.1 Power Up
7.2 Configure Iso-Tek
A. Select the number of rooms to be monitored by this Iso-Tek system.
B. Enter a name for each room associated with its corresponding room sensor.
NOTE: The LED on the Insight screen will flash at the same rate as the LED on the room
sensor that it is associated with.
C. Select the room pressure polarity for each room.
D. Select the room display position on the home screen.
E. Enable a mode by selecting a color and entering a mode name associated with that
color.
F. Select the type of room pressure alarm for each mode.
G. Select which rooms each mode applies to.
H. Mode Timeout will revert the currently selected mode to the default mode after a set
time.
I. Select Update Device when done.
7.3 Iso-Tek Dashboard
After the initial configuration has been completed and the device has been updated, Insight
will display a dashboard based on the number of rooms previously configured.

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 12
7.4 Room Configuration
At the bottom of each room panel on the dashboard you will find a room configuration
button which provides access to the Display, Door Switch and Alarm Configuration screens
for each room.
7.4.1 Display Configuration
7.4.2 Door Switch Configuration
-Prior to configuring the door switch, verify the door switch has been properly
installed and has been wired to the Iso-Tek IO Module.
-Check the Enable box and follow the on-screen instructions to configure the door
switch.
-After completing the configuration process, open and close the door while
observing the door position indicator to verify proper indication.
7.4.2 Alarm Configuration
The alarm configuration screen is used enable and set the room alarms and the
associated alarm parameters, each of which is described in this section.
-Set switch to right to show value on Display.
-Set switch to left to hide value on Display.
Note: Humidity, ACH, and Temperature values are
only available if the Iso-Tek has been configured to
receive them over BACnet.
-Select UPDATE DEVICE to save changes.

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 13
7.4.2.1 Room Pressure Alarm Setpoints
7.4.2.2 Room Pressure Alarm Parameters
Alarm in Delay is used to delay the transition period from the normal pressure state to
alarm state based on the value entered. The Alarm in Delay can be used to prevent
nuisance alarms caused by doors being opened for applications that do not have door
switches.
Alarm out Delay is used to delay the transition period from the alarm pressure state to
normal state based on the value entered.
Alarm Mute Reminder Interval is used to provide an audible reminder that an alarm
condition exists however the audible alarm has been muted.
Enable Alarm Mute Expiration is used to put an expiration time on how long an alarm
will remain muted after the mute button is pressed.
Mute the Audile Room Pressure Alarm when Door is Open is used to silence the
audible alarm when the door is open. This will not delay the unit from entering into an
alarm condition, it will only mute the audible alarm. The room must have a door switch
configured to use this feature.
Mute the Audile Room Pressure Alarm for ____ Sec after Door is Opened is used to
silence the audible alarm for a period of time after the door is opened. The room must
have a door switch configured to use this feature.
7.4.2.3 Room Door Open Alarm
Select the Enable box to enable/disable the Door Open Alarm. A time delay can be used
to hold off the door alarm for the programmed interval based on the application
requirements. The room must have a door switch configured to use this feature.
-Click Enable box to enable/disable alarm.
-If enabled, enter the alarm setpoint in the
box provided.
-The example shown is a Positive Room
with a low pressure alarm setpoint of
0.02”wc and a high pressure alarm setpoint
of 0.08”wc.
Alarm Warning Band is based on the
% of the alarm setpoint. In this
example, a high alarm setpoint of
0.08”wc with a 10% Warning Band
would result in a warning when the
pressure exceeds 0.072”wc.

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 14
7.5 Analog Output Configuration
The Iso-Tek has two independent analog outputs which can be configured to provide an
analog signal representing room pressure for any two rooms connected to the Iso-Tek. The
analog signal can be configured for voltage or current and scaled per the application
requirements.
Analog Output Configuration Window Analog Output Dashboard Gauges
7.6 Configuration Pull-Down Menu
-Set Clock: Set the Iso-Tek’s real-time-clock (RTC) to your local time. The accuracy
of the RTC is critical for accurate Event Log performance.
-Altitude Configuration: The altitude of the installed Iso-Tek is important due to
the ultra-low differential pressure that is being measured. To optimize the
accuracy of the differential pressure measurement as installed, enter the altitude
above sea level of the installed Iso-Tek system. The factory default of 300 ft is
based on the location where the differential sensor was calibrated.
-Configure Device: Use to configure/reconfigure
device. Reference Section 7.2 herein for description
of the Configuration Screen.

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 15
-BACnet MSTP: The BACnet selection is only available when a BACnet Card is
installed in the IO Module. The BACnet window shown below is used to configure
the BACnet communication settings for the unit.
oDevice Configuration: Enter the BACnet Instance Number, MAC Address
and Baud Rate. A BACnet Baud Rate of zero (0) will disable BACnet
Communications preventing conflicts on a working BACnet Network.
oPressure Units: Using the drop-down menu, select the pressure units
required for the application.
oDevice Object Name & Device Object Description: The default name and
description can be change to meet the requirements of the BMS network.
oConfigure External Data is only used for applications that will display data
points such as room temp, room RH, etc…from another device that is
servicing the same room on the local MSTP network.
-Selecting “Enable External
Data” will provide access to
the configuration window
shown to the left.
-For each parameter listed
(Humidity, ACH, Temp…)
enter the Network Device ID,
Network Object ID, and select
Point Type from the
dropdown list.

Installation & Operation Manual - Model ISO
Contents of this Manual are Subject to Change Without Notification Page 16
-Password / Pin Configuration: The Iso-Tek can be configured with an Admin
Password with up to 10 unique User Pins that can be assigned to individual users
by the Administrator.
oAdmin Password: Every Iso-Tek is shipped with a default Admin Password
of “Accutrol”. When the Admin Password is set to “Accutrol”, anyone with
ISO Insight will have open access to enable, view and assign user pins.
Changing the Admin Password to anything other than “Accutrol” will reset
the Admin Password when the “update device” button is selected. If you
change the Admin Password, be sure to update flash to save change to
nonvolatile memory. After the Admin Password has been changed, only
users with the Admin Password will be able to access the user PIN codes.
oUser Pins: The User Pins can be configured to allow specific users access to
initiating selected Mode changes from the Iso-Tek touchscreen display and
also provide access to setting the date/time of the Iso-Tek RTC from the
touchscreen display. If the Admin Password is set to anything other than
“Accutrol”, only the Admin with the proper password can access the User
Pins.
-Sounds: The sounds for Pressure Alarm, PIN entry
and Door Alarm can be enabled/disabled from this
screen. There are multiple options for the Pressure
Alarm sound which can be selected using the drop-
down menu, however the sounds for Pin keypad
entry and Door Open Alarm are not selectable.
Table of contents
Popular Measuring Instrument manuals by other brands

LDG
LDG FTL-Meter Operation manual

Teledyne Lecroy
Teledyne Lecroy HVD3000 Series Operator's manual

Thermo Scientific
Thermo Scientific Nicolet iS5 LS user guide

Pyle
Pyle SereneLife SLGPF30SL user manual

Benstone
Benstone Impaq Elite user manual

Vanguard Instruments Company
Vanguard Instruments Company WRM-10 user manual