Activox LifeChoice XYC105B User manual

1.800.220.0977
www.InovaLabs.com
Manufactured & Distributed by
|3500 Comsout Drive, Suite 100, Austin, Texas 78744
Corporate Office 1.512.617.1700
© 2013 I ova Labs I c. All rights reserved. Assembled i the USA. XYC276B REV C Ja 2013
1.800.220.0977
www.InovaLabs.com
Manufactured & Distributed by
LifeC oice Activox™
Portable Oxygen Concentrator
Operator’s Manual
NEW!

Inova Labs Inc.
3500 Comsouth Drive
Suite 100
Austin, TX 78744
Phone: 1.512.617.1700
Toll-Free: 1.800.220.0977
www.InovaLabs.com
0459
Represented in Europe by:
QNET B
Hommerterweg 286
6436 AM Amstenrade
The Netherlands
Copyright © 2013 Inova Labs, Inc. All rights reserved.
No part of this document may be reproduced or
transmitted in any form or by any means, electronic,
mechanical, photocopying, recording, or otherwise,
without prior written permission from Inova Labs Inc.
TAB E OF CONTENTS
Introduction 2
Application 2
Symbol Descriptions 3
Warnings 3
Life hoice Activox PO Standard Package 5
User ontrols 6
Part Names 8
Operating Instructions 9
Operating Procedure 12
Alarms/Light Indicators 14
Routine Maintenance 16
Specifications 17
Accessories 22
Technical Support 22
Warranty 23
Disposal 26

REF
General Warning,
aution,
Risk of Danger
Danger,
No Smoking
or Open Flame
Waste Disposal
Humidity
Limitation
Temperature
Limitation
Keep Dry
!
Type BF
Applied Part
Device that has
conductive contact
with patient
lass II Equipment
D Power
Radio
Frequency
Manufacturer
atalogue Number E Marking
of onformity
EU Authorized
Representative
External
Rechargeable
Battery
Gas
Flow
INTRODUCTION
The Operator’s Manual will provide familiarity with the Life hoice
Activox Portable Oxygen oncentrator (PO ) and its accessories. Be
sure to read all of the enclosed information in its entirety before using
the Life hoice Activox PO . The Life hoice Activox PO is an
internally powered, Type BF device when powered by the internal
battery and a lass II, Type BF device when connected to the external
A Adapter/ harger, D Adapter/ harger or rechargeable external
battery. The essential performance of the Life hoice Activox PO is
to provide oxygen at a volume that remains within tolerance (the
tolerance was defined based on technical judgment from within the
manufacturer’s expertise in this specific medical application). In
addition, the device’s ability to detect certain error conditions (such
as low concentration or no breath) and create an alarm is also considered
a part of essential performance.
APP ICATION
This manual applies to the Life hoice Activox PO XY 100B.
INDICATIONS FOR USE: The Life hoice Activox Oxygen oncentrator
is used on a prescriptive basis by patients who are diagnosed as
requiring supplemental oxygen. This oxygen concentrator will provide
supplemental, high concentration oxygen to these patients. It is not life-
supporting nor life-sustaining. It may be used continuously in a home,
institution or travel environment. The Life hoice Activox is also portable.
The Life hoice Activox PO is FAA-approved for air travel. The
required FAA label location is on the bottom of the Life hoice Activox
PO . The FAA-approved PO list shows Inova La s LifeChoice Activox.
Every airline has specific information required for traveling with
oxygen. Before traveling, review your airline’s specific requirements.
The Inova Labs website contains additional information at
www.Inova abs.com/resources.
32
WARNINGS
1. This device is not intended to be life-sustaining or life-supporting.
2. This device should be used only when prescribed by a physician.
3. In certain circumstances, oxygen therapy can be hazardous. Please
seek medical advice before using the Life hoice Activox PO .
4. U.S. Federal law restricts this device to sale by or on the order of a
physician.
5. The Life hoice Activox PO is designed to provide a flow of high
purity oxygen. Patients with a fast breathing rate or requiring high
oxygen flow may require more oxygen than the Life hoice Activox
PO can produce and, therefore, may not be suitable candidates.
6. This device is not suitable for patients who are unable to hear or
see alarms or communicate discomfort. If the patient shows any
signs of discomfort, a physician should be consulted immediately.
7. It is the responsibility of the patient and/or provider to make back-up
arrangements for an alternative oxygen supply when traveling;
Inova Labs Inc. assumes no liability for persons choosing not to
adhere to manufacturer recommendations.
8. Availability of an alternate source of oxygen is required in case of
power outage or mechanical failure. onsult your physician or medical
equipment supplier for the type of back-up system required.
9. The Life hoice Activox PO is designed to be used in the provided/
approved carry case.
REF
!
SYMBO DESCRIPTIONS
Symbol Description Symbol Description Symbol Description

54
IFECHOICE ACTIVOX POC STANDARD PACKAGE
The Life hoice Activox PO can be operated using A power, D
power or internal battery power while inside the approved carry case.
All necessary components are included in the Standard Package. The
Life hoice Activox PO Standard Package includes the following:
(1) 4-Way arry ase with Adjustable Strap System
(1) Standard Single Lumen 7-Foot (2-meter) Long Nasal annula*
(1) Life hoice Activox PO Accessory Bag
(1) Life hoice Activox Portable Oxygen oncentrator with 12-hour
Internal Battery at 1 LPMeq (Reference Battery Life table since
actual time will vary depending upon breath rate, dose,
and age of battery.)
(1) A Adapter/ harger (universal 100v – 240v; 50hz – 60hz)
(1) D Adapter/ harger (12v) with standard “car” adapter plug
An External Battery is available for purchase separately which provides 3
hours of additional portable time. (Reference Battery Life table since actual
time will vary depending upon breath rate, dose and age of battery.)
*Nasal cannula may not be included in XY 100B package in all countries based on local in-
ternational requirements.
10. The Life hoice Activox PO should be located so as to avoid
pollutants or fumes.
11. The Life hoice Activox PO does not require the use of any
lubricants. DO NOT use oil, grease or petroleum-based products
on or near the Life hoice Activox PO . DO NOT use power
supplies/adapters, carry bags or accessories other than those
which come with the Life hoice Activox PO . The use of
non-specified accessories may impair equipment performance
and will void all warranties.
12. The Life hoice Activox PO produces enriched oxygen gas
which accelerates combustion. DO NOT allow smoking or open
flames within 10 feet (3 meters) of this device while in use.
igarette/cigar/pipe smoking while using the Life hoice Activox
PO will void all warranties.
13. Never operate the Life hoice Activox PO in the accessory bag
nor any other enclosed area (e.g. cabinets, other bags, etc.). This
is provided for accessories only.
14. DO NOT submerge the Life hoice Activox PO or any of the
accessories in liquid. DO NOT expose to water or precipitation.
DO NOT operate in rain as this could lead to damage.
15. Portable and mobile RF communications equipment can affect
medical electrical equipment.
16. The Life hoice Activox PO should not be used adjacent to or
stacked with other equipment. If adjacent or stacked use is
necessary, the Life hoice Activox PO should be observed to verify
normal operation in the configuration in which it will be used.
17. The Life hoice Activox PO and external battery pack contains
lithium-ion batteries that are subject to special shipping
regulations. If shipping either the PO or battery pack, notify the
shipper that the shipment will contain lithium-ion batteries.
18. When traveling by air, the Life hoice Activox PO and external
battery must be transported as carry-on (NOT checked) baggage.
19. DO NOT operate the unit outside of specified operational
environmental conditions.
WARNINGS
continued
AC Power Adapter
(universal 100v – 240v; 50hz – 60hz)
DC Power Adapter
(12v) with standard “car” adapter plug
(1) Standard Single umen
7-foot ong Nasal Cannula
Adjustable Straps
For use with 4-Way Carry Case
Accessory Bag
ifeChoice Activox POC
4-Way Carry Case
Use as a ackpack, shoulder
strap, waist pack or riefcase
1
1
2
4
5
6
7
3
2
3
7
5
4
6

7
USER CONTRO S
The user controls on the Life hoice Activox PO are specifically
designed to be pressed by a finger. Never press any of the buttons
with a pen, screwdriver, stylus or any other object. Pressing the buttons
with any hard, sharp and/or small object can damage the user controls
and void all warranties. The following items are essential parts of the
user controls for the Life hoice Activox PO :
Indicator ights: olored LED lights indicate either a change in
operating status or a condition that may need response (alarm).
Audible Signals: An audible signal (beep) indicates either a change
in operating status or a condition that may need response (alarm).
Subjective audio response will vary among different people due to
different hearing abilities.
Power Button: To turn the Life hoice Activox PO on or off, press the
Power Button.
Auto Mode EDs: There are two inhalation sensitivity modes on the
control panel: Active-High and Sleep-Low, which automatically adjust
based on oxygen demand (e.g breathing patterns) requirements.
(Reference ontrol Panel diagram.)
Mute Button: The Life hoice Activox PO features a mute function to
disable audible alarms during operation. If an alarm has been muted,
the amber Mute Button LED will be illuminated. (Reference ontrol
Panel diagram.) Please use this function appropriately since it silences
information about the health condition of the device (important audio
signals).
INT and EXT ED: These LED lights will indicate which power
source the unit is using or charging. The INT LED represents the
internal battery, while the EXT LED represents the external battery.
Battery Check Button: The Battery heck Button causes the corre-
sponding battery charge level (e.g. internal or external battery) to be
displayed. (Reference ontrol Panel diagram.)
Battery Charge Indicator: The Battery harge Indicator consists of
four LEDs labeled “25%”, “50%”, “75%”, and “100%”. (Reference
ontrol Panel diagram.) harge-level indication is approximate
except at full charge when all four LEDs are solid green. Use the
following examples to interpret other charge levels:
Pulse Setting Control Buttons: There are three individual Pulse
Setting ontrol Buttons on the control panel: 1, 2, and 3 which are pulse
equivalent to 1, 2, or 3 liters per minute (LPMeq) settings on a
continuous flow device. (Reference ontrol Panel diagram.)
Hour Meter Button: The Life hoice Activox PO will read and
display the number of hours that the unit has been used. (Refer to
Routine Maintenance section.)
Nasal Cannula: A standard single lumen nasal cannula or equivalent
must be used with the Life hoice Activox PO to provide oxygen from
the concentrator. A standard single lumen type cannula, 7-foot
(2-meter), is recommended for use. An Adult Nasal annula like
#1107 by Ventlab orporation is supplied with each Life hoice
Activox PO .* For replacement cannulas, please contact your local
medical equipment supplier. Follow cleaning and care instructions
provided with the nasal cannula.
CAUTION: Use of some administration accessories not specified for
use with this oxygen concentrator may impair the performance.
*Nasal cannula may not be included in XY 100B package in all countries based on local in-
ternational requirements.
6
!
Charge evel 25% ED 50% ED 75% ED 100% ED<
>0% and <13% Flashing Green Off Off Off
>13% and
< 26% Solid Green Off Off Off
>26% and
<37% Solid Green Flashing Green Off Off
>37% and
<51% Solid Green Solid Green Off Off
>51% and
<63% Solid Green Solid Green Flashing Green Off
>63% and
<76% Solid Green Solid Green Solid Green Off
>76% and
< 87% Solid Green Solid Green Solid Green Flashing Green
>87% Solid Green Solid Green Solid Green Solid Green
Battery Charge
Indicator
Alarm Indicator
Hour Meter
Button
For use y trained
service personnel
only
Pulse Setting
Control Buttons
1, 2, 3 LPMeq
Power Button
Auto Mode EDs
Control Panel
Mute Button Battery Check
Button
INT & EXT
Indicator

9
PART NAMES
Cannula Nozzle Fitting: The nasal cannula connects to the oxygen
output nozzle at the top of the Life hoice Activox PO . Do not place
the cannula into nasal passages until the Life hoice Activox PO is
fully powered on. Breathing through the cannula during start-up will
prevent the system from powering on.
Control Panel: (See User ontrols section)
Fan Inlet Vent: ooling air is drawn in through this opening.
(See Maintenance section for cleaning instructions.)
Fan Outlet Vent: Processed air is exhausted through this opening.
External Battery Port: onnect external battery (if used) to this port.
(This connection is to be used solely with the specified external
battery.) Precaution: Please pay attention to the orientation of the
connector. Wrong alignment could cause damage to the connector.
FAA Approval Sticker: Located on the bottom of the unit.
OPERATING INSTRUCTIONS
BEFORE USING THE IFECHOICE ACTIVOX POC
FOR THE FIRST TIME
1. The Life hoice Activox PO should be charged for a minimum of
four hours before initial use. Begin charging by connecting the A
Adapter/ harger to the Life hoice Activox PO and plugging it
into an electrical outlet.
2. Make sure to verify that the Battery heck LED labeled INT is solid
green. This indicates that the system recognizes the external power
source and is charging the internal battery.
If an external battery was purchased with the Life hoice Activox
PO , connect it to the External Battery Port. It should take
approximately three hours to fully charge the external battery (with
internal battery already fully charged), and seven hours to charge
both the internal and external batteries together. (See Battery
harging section for more details.)
Unless an optional external battery charger has been purchased
separately, the external battery can only be charged when
connected directly to the Life hoice Activox PO . ( ontact your
PO provider or customer support for information on purchasing
the optional charger.) Never use an external battery that is not
authorized by Inova Labs Inc.
3. The Life hoice Activox PO is recommended and designed to be
operated and stored in its carrying case at all times.
4. onnect the nasal cannula securely to the Life hoice Activox PO
cannula nozzle fitting.
POWERING THE IFECHOICE ACTIVOX POC
The Life hoice Activox PO can be powered in four different ways:
(1) internal battery, (2) A Adapter/ harger, (3) D Adapter/ harger,
or (4) external battery (optional, sold separately from the Standard
Package). The rechargeable batteries are maintenance-free and can
only be purchased at the factory or through an authorized repair facility.
The following specifics apply to each of the four modes (listed above)
of powering the Life hoice Activox PO :
1. Internal Battery: A rechargeable
internal battery is located within
each Life hoice Activox PO unit.
When fully charged, it supplies
power to the Life hoice Activox
PO for 12 hours. (Reference
Battery Life table since actual time
will vary depending upon breath
rate and Pulse Setting.)
8
Fan Inlet Vent
FAA Approval
Sticker (located
on ottom)
Cannula
Nozzle Fitting
Control Panel
Fan Outlet
Vent
External
Battery Port
Part Names

11
When the Life hoice Activox PO is being powered from the
internal battery, the Battery heck LED labeled INT will be
slowly flashing. Pushing the Battery heck Button will display
the charge level of the internal battery.
If not connected to the A or D Adapter/ harger, the internal
battery will slowly discharge over time even when not in use. Always
check the battery level prior to use to ensure adequate charge level.
2. AC Adapter/Charger: An A Adapter/ harger allows the Life hoice
Activox PO to be powered from a
standard 100-240 VA , 50/60 Hz
outlet. When the unit is operated from
an A Adapter/ harger, power from
the A outlet powers the unit and
recharges the internal and/or external
battery. (See Battery harging
section for details.)
Using an A Adapter/ harger that is not specified by Inova Labs
Inc. for use with the Life hoice Activox PO may damage the unit
and will void all warranties.
3. DC Adapter/Charger: A D Adapter/ harger can be connected
from the Life hoice Activox PO unit to a vehicle’s (car, boat, motor
home, etc.) standard 12 VD outlet.
When the system is connected to the
D outlet, power from the vehicle battery
powers the Life hoice Activox PO unit
and recharges the internal and/or
external battery. (See Battery harging
section for details.)
The 12 VD outlets in some pre-2000
model cars will not provide sufficient current to operate the PO
and charge the battery at the same time. Verify that any D outlet
used can provide at least 10 amps at 12 VD . ( ar auxiliary power
varies significantly; therefore the rate of charging will vary.)
4. External Battery: To power the Life hoice Activox PO from an external
battery, connect the external battery to the external battery connector
port. A fully charged external battery will supply power to the Life hoice
Activox PO for 3 hours. (Reference Battery Life table since actual time
will vary depending upon breath rate and Pulse Setting.)
When the Life hoice Activox PO is being powered from the
external battery, the Battery heck LED labeled EXT will be slowly
flashing. Pushing the Battery heck Button will display the charge
level of the external battery.
NOTE: After turning on the Life hoice Activox PO , allow 30 seconds
for the unit to reach stated performance.
BATTERY CHARGING
To charge the internal and/or external battery, connect the Life hoice
Activox PO to either an A or D Adapter/ harger and a suitable
outlet. If the external battery is being charged, also connect the external
battery to the External Battery Port. onfirm that one of the Battery
heck LEDs, either INT or EXT, is solid green. This indicates that an
A or D external power source has been recognized and that battery
charging is in progress.
If an external battery is connected to the Life hoice Activox PO , it is
important to understand the battery charging sequence. By design,
the internal battery is charged first. That means that the external battery
will not start charging until the internal battery is charged to a 100%
level. Use the INT and EXT Battery heck LEDs to determine which (if
any) batteries are charging.
To check the battery charge level while charging is in progress, press
the Battery heck Button. The charge level will be displayed as indicated
in the Battery harge Indicator section (See User ontrols). When a
battery is fully charged, the corresponding Battery heck LED (INT or
EXT) will be off.
It takes a minimum of four hours for a fully discharged internal battery
to reach a 100% charge level. If both internal and external batteries
are fully discharged, it can take up to seven hours to achieve a 100%
charge level in both batteries. It is recommended to recharge the
batteries, even if only partially depleted, as often as possible. The
batteries cannot be overcharged, and it is recommended to leave the
Life hoice Activox PO plugged in via the A or D Adapter/ harger
as often as possible.
In the event the power supply is interrupted to the Life hoice Activox
PO when the A or D charger is being used, a two-second audible
alarm will sound to inform you that power has been lost. heck that
the charger has been properly connected to the power supply.
The rechargeable batteries are maintenance-free and can only be
exchanged at the factory.
NOTE: If the Life hoice Activox PO is powered off and either the A
or D Adapter/ hargers are connected to the Life hoice Activox PO
after charging is complete, the internal fans may continue to operate
for an additional period of time until the temperature inside the unit is
sufficiently cool.
REPRESSURIZATION
If the unit has not been used for a period of one or more days, it will
automatically repressurize itself. During this time, you may hear the
compressors running for a few seconds.
10

13
OPERATING PROCEDURE
1. Locate and position the Life hoice Activox PO so that the Fan Inlet
Vent and Fan Outlet Vent are not obstructed. The unit, particularly
the air inlet slots and the air outlet areas, should be positioned in a
well-ventilated space.
2. Power the Life hoice Activox PO
using one of the four modes described
in the Powering the Life hoice
Activox PO section above.
3. Press the Power Button once to turn the unit on. An audible beep will
occur and the Power LED, the Auto
Mode Active-High LED and the Pulse
Setting of 1 LED will illuminate solid
green. This indicates that the Life-
hoice Activox PO is powered on,
is ready for use, and that settings
may now be changed. Pressing the
Power Button for an extended time
after the audible beep will cause the
system to turn back off.
NOTE: After the initial power on, the Pulse Setting LED corresponding
to the last flow rate used will illuminate.
CAUTION: DO NOT place the cannula into nasal passages until the
Life hoice Activox PO is fully powered on. Breathing through the
cannula during start-up will prevent the system from powering on.
4. There are two inhalation sensitivity
modes on the control panel:
Active-High and Sleep-Low which
automatically adjust based on oxygen
demand (e.g. breathing pattern)
requirements. No action is required
to start the delivery of oxygen.
12
!
5. Select a Pulse Setting of 1, 2, or 3 as prescribed by your doctor.
These settings are equivalent to 1, 2, or 3 LPM settings on a
continuous flow device. The green LED above the selected setting
illuminates and will flash each time inhalation is detected indicating
that a pulse-wave of oxygen has been delivered.
6. Place the nasal cannula in the nasal passages and breathe normally.
When the Life hoice Activox PO senses inhalation, oxygen is
supplied through the cannula. The duration and size of the pulse
is dependent upon the Pulse Setting selected.
7. heck to make sure your unit’s
battery is fully charged before
venturing out with the Life hoice
Activox PO for the first time. (See
Battery harge Indicator section,
User ontrols).
8. To turn the Life hoice Activox PO
off, press the Power Button and
hold for 2 seconds until the green
Power LED goes off and the unit
shuts down. After the Life hoice
Activox PO is powered off, the
internal fans may continue to
operate for an additional period of
time until the temperature inside
the unit is sufficiently cool.

General
malfunction
of the
Life hoice
Activox PO
ontinuous
audible
alarm
Turn off
the unit.
hange to
another
source of
oxygen and
contact your
equipment
provider.
Alarm LED is solid Amber 15
A ARMS/ IGHT INDICATORS
NORMA OPERATION INDICATORS
14
Normal Audible ight Indication Action
Operation Alert
Life hoice
Activox PO
turned on
Brief
Audible
Tone
You may
begin to use
the Life hoice
Activox PO
unit in
30 seconds
Green LEDs of Power, Active-High and
Pulse Setting of 1 are illuminated.
Note: After the initial power on, the
Pulse Setting LED corresponding to the
last flow rate used will illuminate
A ARM TAB E
Life hoice
Activox PO
has detected
an inhalation/
breath and is
delivering a
pulse of oxygen
None Life hoice
Activox PO
is functioning
normally
Green LED of selected Pulse Setting
flashes with each breath
(( ))
Life hoice
Activox PO
powered by
internal battery
None Life hoice
Activox PO
is functioning
normally
Green LED light labeled INT
slowly flashes
(( ))
Life hoice
Activox PO
powered by
external battery
None Life hoice
Activox PO
is functioning
normally
Green LED light labeled EXT
slowly flashes
(( ))
Problem Audible ight Indication Action
Alert
No breath
detected by
the unit for
approximately
75 seconds
Intermittent
beep
heck the
cannula
connection.
Ensure that
you are
breathing
through your
nose. After the
system detects
a breath, the
alarm will
shut off.
If the alarm
persists,
contact your
equipment
provider.
Amber Alarm LED light flashes
Low Battery
Alarm
(Approximately
10 minutes of
attery power)
Single loud,
long beep
every 60
seconds
onnect the
Life hoice
Activox PO
to an external
power source,
or change to
another source
of oxygen.
Amber Battery Check LED labeled
INT slowly flashes
ritical
Low Battery
Alarm
(Approximately
30 seconds of
attery power)
Fast
audible
beep
(Unit shuts
down)
onnect the
Life hoice
Activox PO
to an external
power source,
or change to
another source
of oxygen.
Amber Battery Check LED labeled
INT begins flashing quickly
(((( ))))
(( ))
(( ))

17
ROUTINE MAINTENANCE
The Life hoice Activox PO does not have any end-user repairable parts.
Fan Inlet Vent Cleaning: There is one inlet vent located on the right
side (facing the unit) of the Life hoice Activox PO . The carrying case
for the unit has mesh material corresponding to the location of the fan
inlet vent. You should visually check the carrying case to ensure that
there is no build-up of lint, hair or other materials that could obstruct
the flow of air into the unit’s fan inlet vent. Remove any material build-up
using a soft brush.
At least once a month inspect the
grille of the intake vent and if it appears
dirty, use a dry cotton swab (Q-Tip®
or similar) to clean it. Never use
alcohol, solvents, or petroleum
distillates for cleaning.
Unit and Carrying Case Cleaning: If it is necessary to clean the unit
and/or carrying case, use only warm water and a mild liquid dish
detergent. Dampen a cloth in the solution and carefully wipe the
outside surfaces of the unit and/or carrying case.
WARNING: DO NOT use alcohol, solvents, polishes or any oily
substances since they may be flammable.
Nasal Cannula Inspection: At least once a week, visually inspect the
nasal cannula. Make sure there are no kinks or obstructions. Replace
as needed with a standard 7-foot (2-meter) single-lumen type cannula
from your local medical equipment supplier.
Reserve Oxygen Supply: Your equipment provider should provide
or suggest an alternative source of supplemental oxygen therapy in
case there is a mechanical failure or power outage.
Hour Meter: The hour meter function is for use by trained service
personnel.
SPECIFICATIONS
Maximum outlet pressure: 19 psi (131 kPa)
Device Electrical Classification: lass II, Type BF
Oxygen Concentration: 90% (+/- 3%)*
Oxygen Concentration Sensor Alarm: 85% or less
Flow Control Pulse Settings: 1, 2, & 3 LPM (Equivalent to
ontinuous Flow) ±20%
Mode of Operation: ontinuous equipment and mobility: portable
Weight: 4.83 lbs.
Dimensions: 9.05" Width x 7.875" Height x 4.38” Depth
Power Requirement: A 100 to 240 VA , 50 to 60 Hz;
D 12 to 19 VD or Battery 12-16.8 VD
ifeChoice Activox POC Input: 19VD , 6A
External AC Power Adapter/Charger, XYC103AC-Input:
100-240VA , 2.5A; Output: 19V, 6.3A
External DC Power Adapter/Charger, XYC104DC-Input:
11.5-16VD , 10A; Output: 19V, 6.3A
Approximate Internal Battery Duration: See Battery Life Table on
following page
Minimum Internal Battery Recharge Time: 4 Hours
Sound: At 10 Feet (3m) - 41 dB at 1 LPMeq, 44 dB at 2 LPMeq,
46 dB at 3 LPMeq
Environmental Conditions for Use, Storage and Transport:
Operational Temperature: 41°F to 104°F (5° to 40° )
Operational Humidity Range: 15% to 95%, Non-condensing
Operational Altitude Max: 10,000 Feet (3km) or 10.1 PSIA (522 mmHg)
Storage Temperature: 32°F to 140°F (0° to 60° )
Storage Humidity Range: Up to 95%, Non-condensing
* When powered by the external A adapter, external D adapter, or external or internal battery.
16
!

19
BATTERY IFE TAB E
18
Internal Battery 12 Hours 6 Hours 5 Hours
1 LPMeq 2 LPMeq 3 LPMeq
External Battery 3 Hours 3 Hours 2 Hours
Run Time (@15 BPM) Pulse Setting
EMC INFORMATION
Medical electrical equipment needs special precautions regarding
EM and needs to be installed and put into service according to the
EM information provided in the accompanying tables.
The Life hoice®Activox Portable Oxygen oncentrator Model XY 100B is
intended for use in the electromagnetic environment specified below. The
customer or the user of the Model XY 100B should assure that it is used in
such an environment.
NOTE: UTis the a.c. mains voltage prior to application of the test level
OXYGEN CONCENTRATION OVER A TITUDE AND F OW RATE
1 LPMeq 2 LPMeq 3 LPMeq
95%
90%
85%
80%
Flow Rate
Oxygen
Oxygen Concentration Over Altitude Range 0- 4000 m
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test IE 60601 ompliance Electromagnetic
Test Level Level Environment - Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
+/-6 kV contact
+/-8 kV air
+/-6 kV contact
+/-8 kV air
Electrostatic
Discharge (ESD)
IE 61000-4-2
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
3 A/m 3 A/m
Power Frequency
(50/60Hz)
Magnetic Field
IE 61000-4-8
Mains power quality
should be that of a typical
commercial or hospital
environment.
+/-2 kV
for power
supply lines
+/-1kV for
input/output
lines
+/-2 kV
for power
supply lines
Not applicable
Electrical Fast
Transient/Burst
IE 61000-4-4
Mains power quality
should be that of a
typical commercial
or hospital
environment.
+/-1 kV
line-to-line
+/-2 kV
line-to-earth
+/-1 kV
line-to-line
+/-2 kV
line-to-earth
Surge
IE 61000-4-5
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the Model XY 100B
requires continued
operation during power
mains interruptions, it is
recommended that the
Model XY 100B be
powered from an
uninterruptible power
supply or a battery.
< 5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
< 5% UT
(>95% dip in UT)
for 5 seconds
< 5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
< 5% UT
(>95% dip in UT)
for 5 seconds
Voltage dips,
short interruptions
and voltage variations
on power supply
input lines
IE 61000-4-11
SPECIFICATIONS
continued
SPECIFICATIONS
continued

2120
The Life hoice®Activox Portable Oxygen oncentrator Model XY 100B is
intended for use in the electromagnetic environment specified below. The
customer or the user of the Model XY 100B should assure that it is used in
such an environment.
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propa-
gation is affected by absorption and reflection from structures, objects, and people.
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and
TV broadcast cannot be predicted theoretically with accuracy. To assess the electro-
magnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the
Model XY 100B is to used exceeds the applicable RF compliance level above,
the Model XY 100B should be observed to verify normal operation. If the abnormal
performance is observed, additional measures may be necessary, such as
re-orienting or relocating the Model XY 100B.
bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test IEC 60601 Compliance Electromagnetic
Test evel evel Environment - Guidance
Portable and Mobile RF
communications equipment
should be used no closer
to any part of the Model
XY 100B, including cables,
than the recommended
separation distance
calculated from the equation
applicable to the frequency
of the transmitter.
Recommended
Separation Distance
d = 1.17 √P
d = 1.17 √P, 80MHz
to 800MHz
d = 2.33 √P, 800MHz to
2.5GHz
3 Vrms
150kHz-80MHz
3 V/m
80MHz-2.5GHz
3 V
3 V/m
onducted RF
IE 61000-4-6
Radiated RF
IE 61000-4-3
Where ‘P’ is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and 'd' is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey,ashould be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following
symbol:
The Life hoice®Activox Portable Oxygen oncentrator Model XY 100B is
intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Model XY 100B
can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters)
and the Model XY 100B as recommended below, according to the maximum
output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended
separation distance 'd' in meters can be estimated using the equation applicable
to the frequency of the transmitter, where ‘P’ is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80MHz and 800MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects, and people.
Recommended Separation Distances Between Portable
and Mobile RF Communications Equipment and The
ifeChoice®Activox Portable Oxygen Concentrator Model XYC100B
The Life hoice®Activox Portable Oxygen oncentrator Model XY 100B is
intended for use in the electromagnetic environment specified below. The
customer or the user of the Model XY 100B should assure that it is used in
such an environment.
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
SPECIFICATIONS
continued
SPECIFICATIONS
continued
Separation Distance According to the Frequency of Transmitter (m)
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.2 1.2 2.3
10 3.7 3.7 7.4
100 12 12 23
Rated Maximum
Output Power of
Transmitter (W)
150kHz to 80MHz 80MHz to 800MHz 800MHz to 2.5GHz
d=1.17 √P d=1.17 √P d=2.33 √P
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions, ISPR 11 Group 1
The Model XY 100B uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions, ISPR 11 lass B
Harmonic Emissions,
IE 61000-3-2 lass A
Voltage Fluctuations/Flicker
Emissions, IE 61000-3-3 omplies
The Model XY 100B is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.

23
ACCESSORIES (ADDITIONA /REP ACEMENT/SPARE)
Product Number: XY 105B
Description: 4-Way arry ase. omes complete with adjustable
strap system that allows case to be used as a backpack, shoulder strap,
waist strap or carried as a briefcase. Included with purchase of
LifeChoice Activox POC.
Product Number: XY 103A 10’ (3m)
Description: A Adapter/ harger. Used to charge and run the
Life hoice Activox PO simultaneously. Included with purchase of
LifeChoice Activox POC.*
Product Number: XY 104D 6’ (1.8m)
Description: D Adapter/ harger. Used to charge and run the
Life hoice Activox PO simultaneously. For use at home or in the car,
RV, motor boat, etc. Included with purchase of LifeChoice Activox POC.*
Product Number: XY 262
Description: 7’(2.1m) Standard Single Lumen annula. Included with
purchase of LifeChoice Activox POC.ϯ
Product Number: XY 340B
Description: Accessory Bag. Included with purchase of LifeChoice
Activox POC.
Product Number: XY 448A
Description: External Battery harger Module. Not included with
purchase of LifeChoice Activox POC Standard Package.
Product Number: XY 102B
Description: Extended Life External Battery. Not included with
purchase of LifeChoice Activox POC Standard Package.
*Accessories, adapters, and cables other than those specified, with the exception of
adapters and cables sold by the manufacturer of the medical electrical equipment as
replacement parts for internal components, may result in increased emissions or
decreased immunity of the Model XY 100B.
ϯannula may not be included in XY 100B packages in all countries based on local
international requirements.
TECHNICA SUPPORT
Please contact the local medical equipment company that provided or
sold you the Life hoice Activox PO for any technical or emergency
support.
If any additional information is needed, please contact Inova Labs’
ustomer are Team toll-free at 1.800.220.0977 or 1.512.617.1700.
WARRANTY
22
INOVA ABS, INC. IMITED WARRANTY AND DISC AIMER
(“ imited Warranty”)
NOTE: This Limited Warranty does not affect Purchaser’s statutory rights
under applicable national laws.
Inova Labs, Inc. (“Inova Labs”) warrants that each new Life hoice Activox™
Portable Oxygen oncentrator (“PO ”) shall be free from defects under normal
use, operation and service for three (3) years. Any Life hoice Activox™ PO
Accessories, the Sieve Bed and all Batteries are warranted for one (1) year.
See the table below for detailed coverage under this Limited Warranty.
The Limited Warranty extends only to the first Purchaser (it is not transferable)
and begins on the date of purchase as evidenced by the sales receipt directly
from Inova Labs or through one of its authorized distributors. This Limited
Warranty is only offered in connection with sales through Inova Labs’ authorized
distribution channels.
The Life hoice Activox™ PO produces enriched oxygen gas which accelerates
combustion. Do not allow smoking or open flames within 10 feet (3 meters) of this
device while in use. igarette/cigar/pipe smoking while using the Life hoice
Activox™ PO will void all warranties.
Extended Warranty is available — two additional years maximum. Extended
warranty covers the same components as the limited three-year warranty.
Batteries, sieve bed replacement and accessories remain at one-year warranty.
Extended warranties are offered to the original owner only and are
non-transferrable. Extended warranties must be purchased at the time of
original purchase of the Life hoice Activox™.
ITEM ENGTH OF COVERAGE
Outer Shell 3 years
ompressors 3 years
omputer ontroller & Power Boards 3 years
PSA Lung Assembly 3 years
Sieve Bed 1 year
Battery – Internal 1 year
Battery – External/Supplemental 1 year
Accessories 1 year

2524
The imited Warranty Excludes From Coverage
1. Defects caused by or related to:
a. Abuse, misuse, negligence or accident;
b. Failure to comply with instructions contained in the
Operator’s Manual;
c. Alteration or modification by Purchaser;
d. Unauthorized repairs or alterations;
e. Environmental conditions (including but not limited to water,
chemicals, fumes in the atmosphere, extreme heat, spillage of
food or liquid, sand, dirt or the like);
f. Lack of regular, preventive maintenance and cleaning;
g. Damage in shipment;
h. Other acts beyond the reasonable control of Inova Labs.
i. Any damage caused by improper packaging when returning
product to Inova Labs Inc.
Warranty Service and Coverage
Purchaser’s exclusive remedy and Inova Labs’ sole obligation hereunder shall be
limited to repair, replacement or refund of the purchase price, at Inova Labs’
option. However, this Limited Warranty shall not exclude or limit (1) Purchaser’s
statutory rights under the applicable national laws, or (2) Purchaser’s rights
against the authorized distributor or vendor.
All items claimed to be defective within the warranty period shall be properly
packaged and shipped on a prepaid basis to Inova Labs (USA sales) or its authorized
distributor at Purchaser’s expense. A prior obtained Return Authorization Number
issued by Inova Labs (USA) or its authorized distributor must properly identify
returned items together with proof of purchase date. Items returned without a
Return Authorization Number shall be refused and returned to Purchaser at
Purchaser’s expense. Inova Labs shall pay for regular shipment back to Purchaser
for repair or replacement of items under this Limited Warranty.
Purchaser shall notify Inova Labs or its authorized distributor of the alleged defect
within a reasonable time of discovery of the issue and in any event no later than
prior to the expiry of the Limited Warranty coverage.
When repairing or replacing the item, Inova Labs may use functionally equivalent
products or parts that are new, equivalent to new or re-conditioned.
ustomer is responsible for proper packaging of all returned items. Any product
damage caused by improper packaging will not be covered under warranty
repair. Warranty does not include shipping to Inova Labs for repair.
As far as national laws permit, the warranty coverage will not be extended or
renewed or otherwise affected due to Inova Labs’ authorized repair or
replacement. However, part(s) repaired or replacement product will be warranted
for the unexpired portion of the original Limited Warranty or for sixty (60) days
from the date of repair or replacement, whichever is longer.
NO REPRESENTATIVE OF INOVA ABS HAS AUTHORITY TO MAKE ANY
REPRESENTATIONS OR PROMISES EXCEPT AS STATED HEREIN. THIS
IMITED WARRANTY IS EXC USIVE AND GIVEN AND ACCEPTED IN
IEU OF ANY AND A OTHER WARRANTIES, EXPRESS OR IMP IED,
INC UDING WITHOUT IMITATION THE IMP IED WARRANTIES OF
MERCHANTABI ITY, FITNESS FOR A PARTICU AR PURPOSE AND
NON-INFRINGEMENT. INOVA ABS SHA HAVE NO FURTHER IABI ITY
FOR ANY DAMAGE, OSS, COST OR FEE OF WHATEVER KIND OR
NATURE, WHETHER FORSEEAB E OR NOT, INC UDING BUT NOT
IMITED TO ATTORNEYS’ FEES, CONSEQUENTIA , GENERA ,
SPECIA , INCIDENTA , INDIRECT, EXEMP ARY OR PUNITIVE
DAMAGES, REGARD ESS OF THE FORM OF THE C AIM, WHETHER IN
CONTRACT, TORT OR OTHERWISE, ARISING OUT OF OR RE ATING
TO INOVA ABS’ PRODUCTS, AND EVEN IF INOVA ABS HAS BEEN
ADVISED OF THE POSSIBI ITY OF SUCH DAMAGES, OSSES, COSTS OR
FEES. TO THE EXTENT PERMITTED BY APP ICAB E AW, INOVA
ABS’ IABI ITY HEREUNDER OR RE ATING HERETO SHA NOT
EXCEED THE PURCHASE PRICE PAID FOR THE PRODUCT, REGARD ESS
OF THE FORUM AND REGARD ESS OF WHETHER ANY ACTION OR
C AIM IS BASED ON CONTRACT, TORT OR OTHERWISE. PURCHASER
AGREES AND ACKNOW EDGES THAT THE PRODUCT IS OF A SIZE,
DESIGN AND CAPACITY SE ECTED BY ITS MEDICA PROVIDER. NO
AGREEMENT VARYING OR EXTENDING THIS IMITED WARRANTY
SHA BE BINDING UPON INOVA ABS UN ESS IN WRITING, SIGNED
BY A DU Y AUTHORIZED REPRESENTATIVE OF INOVA ABS.
Governing aw and Jurisdiction
The rights and obligations of the parties pursuant to this Limited Warranty and
any dispute arising hereunder shall be governed by and interpreted in accor-
dance with the internal (but not the conflicts) laws of the State of Texas, USA. The
1980 U.N. onvention on ontracts for the International Sale of Goods shall not
apply. If any provision of this Limited Warranty is inconsistent with applicable laws
relating to Purchaser, Purchaser agrees to waive any and all rights and remedies
it may have under such laws to the extent it may waive such rights and remedies.
This Limited Warranty is offered as an additional benefit to Purchaser’s statutory
rights and it does not affect such statutory rights in any way. The parties agree
and consent that the state or federal courts in Texas shall have exclusive jurisdic-
tion with respect to any dispute arising out of or relating to this Limited Warranty.
EA H PARTY EXPRESSLY WAIVES ALL RIGHTS TO A TRIAL BY JURY.
WARRANTY
continued
WARRANTY
continued

2726
DISPOSA
Inova Labs expects end users to dispose of Life hoice Activox PO in
an environmentally friendly way. Electrical and electronic equipment
is labeled with the crossed-out wheeled bin symbol indicating that the
equipment should be disposed of by the end user separate from other
types of waste. The Life hoice Activox PO device contains lithium-ion
batteries, and end users should contact Inova Labs or their local
distributor for disposal, collection and recycling options and terms
and conditions for their country. In 2002, the European Union
introduced the Directive on Waste Electrical and Electronic Equipment
(WEEE). The main aim of the Directive is to ensure that WEEE is
collected and treated separately. WEEE items may contain hazardous
substances that should not end up in the human environment and can
have adverse effects on it if they do.
OWNER’S NOTES

2928
OWNER’S NOTES OWNER’S NOTES
This manual suits for next models
1
Table of contents
Other Activox Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Eizo
Eizo CuratOR SP1-24 Instructions for use

Welch Allyn
Welch Allyn CP 50 Updating Software

Byond
Byond ResPlus B-30P Bi-Level PAP user manual

Whitehall Manufacturing
Whitehall Manufacturing Dignity Suite WHD-SHV10 Installation, operation & maintenance manual

Gima
Gima OXY-PED Use and maintenance book

Gima
Gima 27331 Use and maintenance book