Additel ADT685Ex User manual

685/685Ex
Advanced Digital Pressure Gauge
Manometry cyfrowe do pomiaru ciśnienia ADT685 i ADT685Ex (Additel) | http://acse.pl

ADT685/685EX Advanced Digital Pressure Gauges
-------User Manual
[Version:2103V01]
AdditelCorporation

STATEMENT
This user manual provides operatig and safety instructions for theADT685/685EXAdvanced
Digital Pressure Gauges. To ensure correct operation and safety, please follow the instructions in
this manual. Additel Corporation reserves the right to change the contents and other information
contained in this manual without notice.


I
CONTENT
1. INTRODUCTION.............................................................................................................................................................3
1.1 General Introduction...................................................................................................................................................3
1.2 Pressure Ranges and Accuracy .................................................................................................................................4
1.3 Technical Specifications.............................................................................................................................................9
1.4 Features...................................................................................................................................................................10
1.5 Basic Structure......................................................................................................................................................... 11
2. OPERATION.................................................................................................................................................................13
2.1 Main Display.............................................................................................................................................................13
2.2 Main Operation Interface..........................................................................................................................................14
2.3 Buttons.....................................................................................................................................................................16
2.4 Pressure Measurement............................................................................................................................................17
2.4.1 Pressure Percentage Indication.........................................................................................................................17
2.4.1 Pressure Zeroing...............................................................................................................................................18
2.4.2 Pressure Unit Switching ....................................................................................................................................19
2.4.3 PressureType Switching....................................................................................................................................19
2.4.4 Pressure Tare....................................................................................................................................................19
2.4.5 Alarm.................................................................................................................................................................19
2.5 Pressure Peak Recording.........................................................................................................................................19
2.6 Pressure Leak Test...................................................................................................................................................21
2.7 Data Logging............................................................................................................................................................24

II
2.8 Backlight ON/OFF....................................................................................................................................................26
2.9 Screen Lock and Unlock...........................................................................................................................................27
2.10 Date and Time........................................................................................................................................................27
2.11 Password Display and Editing.................................................................................................................................29
2.12 Number Editing.......................................................................................................................................................30
2.13 Cycle Selection Keyboard Display and Editing........................................................................................................31
3. SYSTEM SETTINGS.....................................................................................................................................................32
3.1 Communications ......................................................................................................................................................32
3.1.1 Bluetooth...........................................................................................................................................................32
3.1.2 RS232/RS485....................................................................................................................................................34
3.1.3 4-20mA..............................................................................................................................................................35
3.1.4 SWITCH Settings ..............................................................................................................................................36
3.2 Auto Power Off Settings............................................................................................................................................37
3.3 Filter Settings ...........................................................................................................................................................37
3.4 Pressure Reading Display Rate................................................................................................................................38
3.5 Calibration................................................................................................................................................................39
3.5.1 Calibration Interface ..........................................................................................................................................39
3.5.2 Pressure Calibration..........................................................................................................................................40
3.5.3 Barometric Calibration.......................................................................................................................................42
3.5.4 (4~20) mA Calibration (only available when using (4~20) mA module) .............................................................43
3.6 Restore to Factory Settings ......................................................................................................................................44
3.7 OPList Settings.........................................................................................................................................................44

III
3.7.1 Version Snformation..........................................................................................................................................44
3.7.2 Resolution Settings............................................................................................................................................44
3.7.3 Auto-backlight Time Settings.............................................................................................................................44
3.7.4 Barometric Pressure and Temperature Display.................................................................................................44
3.7.5 Overpressure Record Information .....................................................................................................................45
3.7.6 Touch Sound Settings........................................................................................................................................45
3.8 Malfunction Information............................................................................................................................................45
3.9 Operation Error Codes .............................................................................................................................................47
4. COPYRIGHT.................................................................................................................................................................48
APPENDIX A:RS232 MODULE DB9 PINS DESCRIPTION...........................................................................................48
APPENDIX B:RS485 MODULE DB9 PINS DESCRIPTION ..........................................................................................49
APPENDIX C:SWITCH MODULE PINS DESCRIPTION ...............................................................................................51
APPENDIX D:DUAL CABLE (4~20) MA MODULE PINS DESCRIPTION ....................................................................52
APPENDIX E:SPECIAL NOTICE FOR ADT685EX CABLE.............................................................................................53
APPENDIX F:ADT685EX EX-PROOF DESCRIPTION..................................................................................................54


1
Safety Instructions
Warning: The situation that poses a threat to user safety.
Attention: The condition that may cause damage to the gauges or affect the calibration result.
Warning:
To prevent the user from injury, please follow this user manual for use.
To prevent possible fire, electric shock or personal injury, please do as follows:
1. General:
The pressure medium should be confirmed before use. Please adhere to the following instructions:
◆Please read the user manual before using the product;
◆Before using the product, please check the appearance of the product to ensure there is no damage;
◆Please refer to the operation steps in the manual when using the product;
◆Please be sure to store, transport and use this product in the correct direction;
◆If the product is damaged or malfunctions, please do not use it and contact Additel;
◆Never use the non-Ex version in an explosive, steam or dust environment.
2. Electrical:
◆Before using the product, please make sure that the power supply is connected correctly, and the installation is
done properly.

2
Attention:
To prevent injury, please obey the instruction manual for use
To prevent possible damage, please do as follows:
◆Do not use the instrument in a high vibration environment;
◆If the gauge is abnormal, please stop using it and contact with Additel.

3
1. Introduction
1.1 General Introduction
ADT685 /ADT685EX Advanced Digital Pressure Gauge is the latest generation digital pressure gauges introduced by
Additel Corporation. It is mainly used to calibrate precision pressure gauges, general pressure gauges,
sphygmomanometers or other pressure instruments and it can also be widely used for precision pressure measurement.
The power consumption of ADT685/ADT685EX is very low, which is suitable for long-term continuous work (Up to
20,000 hours under ultra-low power consumption mode). TheADT685/ADT685EX also has excellent electromagnetic
compatibility characteristics, it has obtained the CE certification of the most authoritative organization and can be used in
various complex electromagnetic environments.
Contact us:
Additel Corporation
Tel: +1-714-998-6899
www.additel.com

4
1.2 Pressure Ranges and Accuracy
Table 1 Gauge Pressure[1](Sealed gauge pressure for above 1,000 psi)
P/N
Pressure Range
Media[2]
Accuracy
Burst Pressure
(psi)
(bar)
%FS
%RD
V15
-15
-1.0
G
0.02 (0.05)
N/A
3X
GP2
2
0.16
G,L
0.05
N/A
3X
GP5
5
0.35
G,L
0.05
0.1
3X
GP10
10
0.7
G,L[3]
0.02 (0.05)
0.1
3X
GP15
15
1.0
G,L[3]
0.02 (0.05)
0.1
3X
GP30
30
2.0
G,L[3]
0.02 (0.05)
0.1
3X
GP50
50
3.5
G,L
0.02 (0.05)
0.1
3X
GP100
100
7.0
G,L
0.02 (0.05)
0.1
3X
GP150
150
10
G,L
0.02 (0.05)
0.1
3X
GP300
300
20
G,L
0.02 (0.05)
0.1
3X
GP500
500
35
G,L
0.02 (0.05)
0.1
3X
GP600
600
40
G,L
0.02 (0.05)
0.1
3X
GP1K
1,000
70
G,L
0.02 (0.05)
0.1
3X
GP1.5K
1,500
100
G,L
0.02 (0.05)
0.1
3X

5
GP2K
2,000
140
G,L
0.02 (0.05)
0.1
3X
GP3K
3,000
200
G,L
0.02 (0.05)
0.1
3X
GP5K
5,000
350
G,L
0.02 (0.05)
0.1
3X
GP10K
10,000
700
G,L
0.02 (0.05)
0.1
3X
GP15K
15,000
1,000
G,L
0.05 (0.1)
0.1
1.5X
GP20K
20,000
1,400
G,L
0.05 (0.1)
N/A
1.5X
GP25K
25,000
1,600
G,L
0.05 (0.1)
N/A
1.5X
GP30K
30,000
2,000
G,L
0.05 (0.1)
N/A
1.5X
GP36K
36,000
2,500
G,L
0.05 (0.1)
N/A
1.5X
GP40K
40,000
2,800
G,L
0.05 (0.1)
N/A
1.35X
GP50K
50,000
3,500
G,L
0.1 (0.2)
N/A
1.2X
GP60K
60,000
4,200
G,L
0.1 (0.2)
N/A
11X
[1] Sealed gauge pressure for above 1,000 psi.
[2] G=Gas, L=Liquid.
[3] 0.02% FS for gas media only.

6
Table 2 Differential Pressure
P/N
Pressure Range
Media
Accuracy
(%FS)[1]
Burst Pressure
Static
Pressure
Range
(inH2O)
(mbar)
DP1
±1
±2.5
G
0.05[2]
100X
±10 psi
DP2
±2
±5.0
G
0.05[2]
100X
±10 psi
DP5
±5
±10
G
0.05[2]
50X
±10 psi
DP10
±10
±25
G
0.05[2]
20X
±10 psi
DP20
±20
±50
G
0.05
20X
±10 psi
DP30
±30
±75
G
0.05
20X
±10 psi
DP50
±50
±125
G
0.05
3X
±10 psi
DP100
±100
±250
G
0.02
3X
±15 psi
DP150
±150
±350
G
0.02
3X
50 psi
DP300
±300
±700
G
0.02
3X
50 psi
[1] FS specification applies to the span of the range. Accuracy includes one year stability.
[2] 0.05%FS accuracy (incl 6 months stability). One year accuracy is 0.05%FS calibration accuracy combined with 0.05%FS one year
stability.

7
Table 3 Compound Pressure
P/N
Pressure Range
Media[1]
Accuracy
Burst Pressure
(psi)
(bar)
%FS[2]
%RD
CP2
±2
±0.16
G
0.05
N/A
3X
CP5
±5
±0.35
G
0.02 (0.05)
0.1
3X
CP10
±10
±0.7
G
0.02 (0.05)
0.1
3X
CP15
±15
±1
G
0.02 (0.05)
0.1
3X
CP30
-15 to 30
-1 to 2
G
0.02 (0.05)
0.1
3X
CP100
-15 to 100
-1 to 7
G,L
0.02 (0.05)
0.1
3X
CP300
-15 to 300
-1 to 20
G,L
0.02 (0.05)
0.1
3X
CP500
-15 to 500
-1 to 35
G,L
0.02 (0.05)
0.1
3X
CP600
-15 to 600
-1 to 40
G,L
0.02 (0.05)
0.1
3X
CP1K
-15 to 1000
-1 to 70
G,L
0.02 (0.05)
0.1
3X
[1] G=Gas, L=Liquid.
[2] FS specification applies to the span of the range.

8
Table 4 Absolute Pressure
P/N
Pressure Range
Media[1]
Accuracy (%FS)
Burst Pressure
(psi)
(bar)
AP5
5
0.35
G,L
0.1
3X
AP10
10
0.7
G,L
0.1
3X
AP15
15
1.0
G,L
0.1
3X
AP30
30
2.0
G,L
0.1
3X
AP50
50
3.5
G,L
0.1
3X
AP100
100
7.0
G,L
0.05(0.1)
3X
AP300
300
20
G,L
0.05(0.1)
3X
AP500
500
35
G,L
0.05(0.1)
3X
AP1K
1000
70
G,L
0.05(0.1)
3X
AP3K
3000
200
G,L
0.05(0.1)
3X
AP5K
5000
350
G,L
0.05(0.1)
3X
[1] G=Gas, L=Liquid.
*Note
◆Temperature Compensation:(-10~50)℃, 1year accuracy;
◆Abs pressure, 0.05% accuracy, the time to guarantee this accuracy is 180days;
◆The medium requires non-corrosive.

9
1.3 Technical Specifications
Table 5 Technical Specifications
Model
ADT685
Digital Pressure Gauge
ADT685Ex
Intrinsically Safe Digital Pressure Gauge
Intrinsic Safety
&
European Compliance
CE
CE marked
ATEX certified intrinsically safe
II 1G EX ia IIC T4 Ga
Protection Level
IP67
Case Material
304 SS
Wetted Parts Material
316 SS[1]
Dimensions
4.65" (118mm) width X 1.77" (42mm) depth X 7.01" (178mm) height
Weight
1.50 lbs. (0.68 kg)
Display
Touch screen (3.4'' FSTN segment capacitive) - 5 1/2 full digits
Button
4 Function buttons, 1 Power ON/OFF button
Data Logging
(Included with -DL models)
Data storage: 10,000,000 records (Time+Pressure)
Communication
USB Type-C and Bluetooth included, RS-232 or RS-485[2] (optional)

10
Power
External power: 110/220 V adapter (5 VDC)
(optional, excl ADT685Ex)
Battery: Three AA alkaline batteries (included)
Battery life: 1500 hours typical
Environmental
Compensated Temperature: 14°F to 122°F (-10°C to 50°C)
Operating Temperature: 14°F to 122°F (-10°C to 50°C)
Storage Temperature: -4°F to 158°F (-20°C to 70°C)
Humidity: 0 - 95%, non-condensing
Barometric Accuracy
(Built-in barometer)
Range: (60~110) kPa.a, accuracy: 50Pa
[1] Wetted parts material types may vary by pressure range. Please refer to manual or contact Additel for more information.
[2] 4-20 mAoutput and RS-485 not available inADT685Ex models.
1.4 Features
Ultra-low power consumption design, long-term operationWide pressure ranges, up to 60,000 psi (4,200 bar)
Capacitive touch screen, interactive operation is convenient and easy to use Supports multiple type communication
modules to meet different needs, optional communication module including:
RS-232 or RS-485

11
1.5 Basic Structure
Figure 1 Basic Structure

12
Table 6 Basic Structure
No.
Part Name
Function
1
LCD Display
2
Back Light Button
Turn on/off the back light
3
Units Button
Switch the pressure engineering units
4
Zeroing/Clear Button
Readings zeroing or to clear the number in number editing interface
5
Power On/Off / Reset Button
Long press to power on/off the gauge,
short pressure to return to previous interface
6
Switchable Communication Module
Optional RS232, RS485, 2-wire (4~20) mA module and SWITCH
7
USB-Type C Port
Communication and power supply
8
Pressure module
9
Barometric pressure inlet port
Visible after removing the back cover (withATM logo), it is used for barometer
sensor calibration
This manual suits for next models
1
Table of contents
Other Additel Measuring Instrument manuals