AEMC SL261 User manual

AC/DC CURRENT
OSCILLOSCOPE PROBE
SL261
ENGLISH User Manual
ZERO
100 mV/A
10 mV/A
OFF
I
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
certifies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be
requested at the time of purchase, or obtained
by returning the instrument to our repair and
calibration facility, for a nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer.
Serial #: ________________________________
Catalog #: 1201.51
Model #: SL261
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

Table of Contents
1. INTRODUCTION............................................................................... 2
1.1 International Electrical Symbols................................................3
1.2 DenitionofMeasurementCategories .....................................3
1.3 ReceivingYourShipment..........................................................3
1.4 OrderingInformation.................................................................4
1.4.1 AccessoriesandReplacementParts............................4
2. PRODUCT FEATURES ...................................................................... 5
2.1 Description................................................................................5
2.2 Compatibility .............................................................................5
2.3 Features....................................................................................6
3. SPECIFICATIONS............................................................................. 7
3.1 ElectricalSpecications ............................................................7
3.2 MechanicalSpecications ........................................................8
3.3 EnvironmentalSpecications....................................................9
3.4 SafetySpecications ................................................................9
3.5 TypicalResponseCurves .......................................................10
4. OPERATION .................................................................................. 12
4.1 ZeroAdjustment......................................................................12
4.2 CurrentMeasurement.............................................................12
4.3 Battery Indication (Green LED)...............................................13
4.4 Peak Overload (OL) Indication (Red LED) .............................13
5. MAINTENANCE ............................................................................. 14
5.1 Maintenance ...........................................................................14
5.2 Cleaning..................................................................................14
5.3 ReplacingtheBatteries...........................................................15
Limited Warranty ...................................................................................17
WarrantyRepairs...................................................................................17
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

2
AC/DC Current Oscilloscope Probe Model SL261
CHAPTER 1
INTRODUCTION
WARNING
The safety warnings are provided to ensure the safety of
personnel and proper operation of the instrument. Read the
instruction completely.
• Usecautiononanycircuit:potentiallyhighvoltagesandcurrents
maybepresentandmayposeashockhazard.
• Donotusetheprobeifdamaged.Alwaysconnectthecurrent
probetothemeasuringdevicebeforeitisconnectedaroundthe
conductor
• Donotuseonnoninsulatedconductorwithapotentialtoground
greaterthan600VCATIIIpollution2.Useextremecautionwhen
clampingaroundbareconductorsorbusbars.
• Beforeeachuse,inspecttheprobe;lookforcracksinhousingor
outputcableinsulation.
• Donotuseclampinwetenvironmentorinlocationsthathazard-
ousgasesexist.
• Donotusetheprobeanywherebeyondthetactilebarrier.
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

AC/DC Current Oscilloscope Probe Model SL261
3
1.1 International Electrical Symbols
This symbol signifies that the instrument is protected by double or rein-
forced insulation.
This symbol on the instrument indicates a WARNING and that the operator
must refer to the user manual for instructions before operating the
instrument. In this manual, the symbol preceding instructions indicates that
if the instructions are not followed, bodily injury, installation/sample and
product damage may result.
Risk of electric shock. The voltage at the parts marked with this symbol may
be dangerous.
This symbol refers to a type A current sensor. This symbol signifies that
application around and removal from HAZARDOUS LIVE conductors is per-
mitted.
Probe fitted with an electronic output limiter providing protection against
voltage surges caused by the accidental opening of the probe secondary
circuit to 30V max. peak.
In conformity with WEEE 2002/96/EC.
1.2 Definition of Measurement Categories
CAT I: Formeasurementsoncircuits not directlyconnectedtotheAC
supply wall outlet such as protected secondaries, signal level,
andlimitedenergycircuits.
CAT II: For measurements performed on circuits directly connected to
the electrical distribution system. Examples are measurements
onhouseholdappliancesorportabletools.
CAT III: For measurements performed in the building installation at
the distribution level such as on hardwired equipment in xed
installationandcircuitbreakers.
CAT IV: For measurements performed at the primary electrical supply
(<1000V) such as on primary overcurrent protection devices,
ripplecontrolunits,ormeters.
1.3 Receiving Your Shipment
Uponreceivingyourshipment,makesurethatthecontentsareconsistent
with the packing list. Notify your distributor of any missing items. If the
equipmentappearstobedamaged,leaclaimimmediatelywiththecar-
rierandnotifyyourdistributoratonce,givingadetaileddescriptionofany
damage.Savethedamagedpackingcontainertosubstantiateyourclaim.
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

4
AC/DC Current Oscilloscope Probe Model SL261
1.4 Ordering Information
AC/DCCurrentProbeModelSL261
(10A-100mV/A,100A-10mV/A,BNC).................................. Cat. #1201.51
Includes a user manual and a product warranty and registration card.
1.4.1 Accessories and Replacement Parts
Adapter-BNCAdapterforusewithAC/DCCurrentProbe
ModelSL261&Meters8220&8335................................... Cat. #2140.40
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

AC/DC Current Oscilloscope Probe Model SL261
5
CHAPTER 2
PRODUCT FEATURES
2.1 Description
The SL261 AC/DC Current Oscilloscope Probe expands oscilloscope
applicationsinindustrial,automotiveorpowerenvironments,andisideal
for analysis and measurement of distorted current waveforms and har-
monics.Theprobepermitsaccuratedisplayandmeasurementofcurrents
from100mAto100Arms,DCto100kHzwithoutbreakingintothecircuit.
TheprobeusesHalleffecttechnologytomeasureACandDCsignals.The
probeconnectsdirectlytoanoscilloscopethrougha2metercoaxialcable
withaninsulatedBNC.
2.2 Compatibility
TheModelSL261iscompatiblewithanyanalogordigitaloscilloscopeor
othervoltage-measuringinstrumentwhichhasthefollowingfeatures:
• BNCinputconnector
• Rangecapableofdisplaying0.2to0.5Vperdivision
• Minimuminputimpedanceof1MΩ
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

6
AC/DC Current Oscilloscope Probe Model SL261
2.3 Features
ZERO
OFF
I
Battery
Compartment
Screw
Range Selection Switch
Power Indication LED
Zero Adjust Knob
Current Flow for DC Measurements
or Supply to Load for AC Measurements
Battery
Compartment Cover
Ø = 0.46"
(11.8 mm)
Peak Overload
Indication LED
Keep hands below
this point
100 mV/A
10 mV/A
WARNING: Always keep hand on the handle part of the probe.
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

AC/DC Current Oscilloscope Probe Model SL261
7
CHAPTER 3
SPECIFICATIONS
3.1 Electrical Specifications
*Referenceconditions:23°C±5°C,20to75%RH,DCto1kHz,probezeroed,1minute
warmup,batteriesat9V±0.1V,externalmagneticeld<40A/m,noDCcomponent,no
externalcurrentcarryingconductor,1MΩ/100pFload,conductorcenteredinjaw.
Current Range:
100mV/A:100mAto10Apeak
10mV/A:1to100Apeak
Output Signal:1000mVpeakmax
AC Current Accuracy:
Aftercalibrationandforoneyear
(zeroprobebeforemakingmeasurement)
Range Accuracy Phase Shift
100mV/A
(50mAto10Apeak) 3%ofreading±50mA <1.5°fromDCto65Hz
10mV/A
(500mAto40Apeak) 4%ofreading±50mA
<1°fromDCto65Hz
10mV/A
(40Ato100Apeak) 15%maxat100A
Frequency Range: DCto100kHz(-3dBwithcurrentderating)
Noise:
Range10mV/A:480µV
Range100mV/A:3mV
Slew Rate:
Range10mV/A:20mV/µs
Range100mV/A:0.3V/µs
Load Impedance: >1MΩ/100pF
Insertion Impedance (50/60 Hz): 100mV/A:3µs
10mV/A:0.01Ω
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

8
AC/DC Current Oscilloscope Probe Model SL261
Rise or Fall Time:Range100mV/A:3µs
Range10mV/A:4µs
Working Voltage: 600Vrmsmax
Common Mode Voltage: 600Vrmsmax
Inuence of Adjacent Conductor: < 0.2mA/A AC
Inuence of Conductor Position in Jaw: 0.5%ofreadingatkHz
Battery: 9Valkaline(NEDA1604A,IEC6LR61)
Low Battery: GreenLEDwhen≥6.5V
Overload Indication:
RedLEDindicatesinputgreaterthantheselectedrange
Typical Consumption: 8.6mA
Battery Life: 55hourstypical
3.2 Mechanical Specifications
Zero Adjustment: 20turnpotentiometer
Maximum Cable Diameter: 0.46" (11.8 mm)
Case Protection: IP20perIEC529
Drop Test:
1.0mon38mmofoakonconcrete;testaccordingtoIEC1010
Mechanical Shock: 100G;testperIEC68-2-27
Vibration:
TestperIEC68-2-6,frequencyrange10Hzto55Hz,amplitude0.15mm
Handle: Lexan®920A,UL94V2
Dimensions: 9.09x1.42x2.64"(231x36x67mm)
Weight: 11.6oz(330g)withbattery
Color: Lightgray
Output Lead: InsulatedcoaxialcablewithinsulatedBNCconnector
Cable Length: 6.5foot(2m)
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

AC/DC Current Oscilloscope Probe Model SL261
9
3.3 Environmental Specifications
Operating Temperature:0°to50°C(32°to122°F)
Storage Temperature:-30°to80°C(-22°to176°F)
Operating Relative Humidity:
10°to30°C:85±5%RH(withoutcondensation)
40°to50°C:45±5%RH(withoutcondensation)
Inuence of Temperature: <0.2%per°C
Altitude:Non-operating:0to12,000m
Operating:0to2000m
3.4 Safety Specifications
Electrical:Doubleinsulationorreinforcedinsulationbetweentheprimary
orsecondaryandtheoutercaseofthehandleperEN61010-2-032.
600V,CATIII,PollutionDegree:2
Electromagnetic Compatibility:
EN 50081-1 Class B
EN50082-2ElectrostaticdischargeIEC1000-4-2
Radiated Field IEC 1000-4-3
Fast Transients IEC 1000-4-4
MagneticFieldat50/60HzIEC1000-4-8
*Specications are subject to change without notice
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

10
AC/DC Current Oscilloscope Probe Model SL261
3.5 Typical Response Curves
0
-1
-2
-3
-4
-5
Frequency
Frequency in Hz
range 100 mV/A
range 10 mV/A
frequency response
measurement current: 1A sine
100 1000 10000 100000
gain in dB
0
-10
-50
-100
-150
Phase Shift
Frequency in Hz
range 100 mV/A
range 10 mV/A
phase graph
measurement current: 1A sine
100 10000 100000
phase in °
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

AC/DC Current Oscilloscope Probe Model SL261
11
0%
-2%
-4%
-6%
-8%
-10%
-12%
-14%
-16%
0A 20A 40A 60A 80A 100A
Accuracy
Uncertainty
Typical
Specification
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

12
AC/DC Current Oscilloscope Probe Model SL261
CHAPTER 4
OPERATION
4.1 Zero Adjustment
Theprobehasazeroadjustmentthatshouldbeadjustedbeforemeasure-
ment.Alternatively,youmay“zero”withtheoscilloscopeinstead.
4.2 Current Measurement
WARNING: Always connect the probe to the instrument before clamping onto
the circuit under test.
• ConnecttheModelSL261totheproperinputchannelontheoscil-
loscope.
•
Beginwiththeleastsensitiverangeonthecurrentprobe(10mV/A)
• Selectthe0.5V/Divisionrangeontheoscilloscope.
• Clamptheprobeontheconductortobemeasuredandreadthe
currentowingdirectlyonyouroscilloscope.
(Remember to unclamp the probe from the conductor before discon-
necting it from your meter or instrument.)
Youmayalsouseyouroscilloscopetoamplifythesignalwhileusingthe
100mV/A probe range (which offers the best accuracy and least phase
shift).
NOTE: It is possible to change the range on the current probe without remov-
ing the probe from the current carrying conductor, but it is important to
remember not to exceed the permissible peak ratings of 1000mV peak or
2000mV peak to peak maximum. The peak ratings by range are: 10A peak
on the 100mV/A range, 100A peak on the 10mV/A range.
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

AC/DC Current Oscilloscope Probe Model SL261
13
4.3 Battery Indication (Green LED)
TheprobehasabatteryconditionLED.Toensureproperreadingswith
yourcurrentprobe,besurethatthegreenLEDislitduringmeasurement.
Ifnot,replacethe9Vbattery.
4.4 Peak Overload (OL) Indication (Red LED)
TheSL261offersanoverloadindicator.IftheredLEDilluminatesduring
measurement,thisindicatesthatthepeakvalueexceedstheinstrument
responselevelandthattheoutputisdistorted.Switchtheprobetoahigher
rangeifpossible.
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

14
AC/DC Current Oscilloscope Probe Model SL261
CHAPTER 5
MAINTENANCE
5.1 Maintenance
WARNING
• Toensureoptimumperformance,itisimportanttokeeptheprobe
jawmatingsurfacescleanatalltimes.Failuretodosomayresult
inerrorinreadings.
• Formaintenanceuseonlyspeciedfactoryreplacementparts.
• Toavoidelectricalshock,donotattempttoperformanyservicing
unlessyouarequaliedtodoso.
• Toavoidelectricalshockand/ordamagetotheinstrument,donot
allowwaterorotherforeignsubstancesintothecase.
• Disconnecttheunitfromallcircuitsandtestcablesbeforeopen-
ingthecase.
5.2 Cleaning
• Tocleantheprobebody,useasoftclothdampenedinasolution
ofmilddetergentandwater.Tocleanthecore,openthejawand
cleantheexposed coresurfaceswithacottonswabdampened
with isopropyl alcohol (isopropanol) or ethyl alcohol (fotocol or
ethanol).Lubricatethejawsmatingsurfaceswithalightoil.
• Do not use chemicals containing benzine, benzene, toluene,
xylene,acetone,orsimilarsolvents.
• Donotimmersetheprobeinliquidsoruseabrasivecleaners.
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

AC/DC Current Oscilloscope Probe Model SL261
15
5.3 Replacing the Batteries
CAUTION: Risk of electric shock. Disconnect all input(s) from the unit or
remove the probe from any conductor before opening the battery cover to
change the batteries.
Ifthe greenbattery indication LEDdoes notlight upwhen the probe is
turnedon,thebatterieswillneedtobereplaced.
Donotreplacethebatterywhileprobeisinuse.
Toreplacethebattery:
• Disconnecttheprobefromthecircuitandtheoscilloscope.
• Turntheprobe“OFF”.
• Unscrewthebatterycompartmentscrewandpulloutthebatterycom-
partmentcover.
• Replacethebatteryandputthecoverbackon.
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

AC/DC Current Oscilloscope Probe Model SL261
17
Limited Warranty
TheAC/DCCurrentOscilloscopeProbeModelSL261iswarrantedtotheown-
erforaperiodofoneyearfromthedateoforiginalpurchaseagainstdefectsin
manufacture.ThislimitedwarrantyisgivenbyAEMC®Instruments,notbythe
distributorfromwhomitwaspurchased.Thiswarrantyisvoidiftheunithas
beentamperedwith,abusedorifthedefectisrelatedtoservicenotperformed
byAEMC®Instruments.
For full and detailed warranty coverage, please read the Warranty
Coverage Information, which is attached to the Warranty Registration
Card (if enclosed) or is available at www.aemc.com. Please keep the
Warranty Coverage Information with your records.
What AEMC®Instruments will do:
Ifamalfunctionoccurswithintheone-yearperiod,youmayreturntheinstrument
tousforrepair,providedwehaveyourwarrantyregistrationinformationonle
oraproofofpurchase.AEMC®Instrumentswill,atitsoption,repairorreplace
thefaultymaterial.
Warranty Repairs
What you must do to return an Instrument for Warranty Repair:
First, request a Customer ServiceAuthorization Number (CSA#) by phone
orbyfaxfromourServiceDepartment(seeaddressbelow),thenreturnthe
instrumentalongwith the signed CSAForm. Please write the CSA# onthe
outsideoftheshippingcontainer.Returntheinstrument,postageorshipment
pre-paidto:
Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments
15FaradayDrive•Dover,NH03820USA
Phone:(800)945-2362(Ext.360)
(603)749-6434(Ext.360)
Fax: (603)742-2346or(603)749-6309
E-mail:[email protected]
Caution:Toprotectyourselfagainstin-transitloss,werecommendyouinsure
yourreturnedmaterial.
NOTE: You must obtain a CSA# before returning any instrument.
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187

05/11
99-MAN100042v10
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
Shop for Data Logging products online at: www.DataLoggerStore.ca 1.800.561.8187
Other manuals for SL261
4
Table of contents
Popular Laboratory Equipment manuals by other brands

Thermo Scientific
Thermo Scientific Matrix 1029 user manual

Teledyne Cetac Technologies
Teledyne Cetac Technologies ASX-560 Alignment Guide

HURAKAN
HURAKAN HKN-UVA10 manual

Agilent Technologies
Agilent Technologies Cary 630 FTIR user guide

Thermo Scientific
Thermo Scientific TRACE 1600 Hardware manual

KNF
KNF LIQUIPORT NF 100.18 Series operating instructions