AEMC 2620 User manual

GROUND FAULT/LEAKAGE
DETECTOR 2620
with the Artifical Neutral AN-1
CHAUVIN ARNOUX GROUP
®
ENGLISH User Manual

Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
certifies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be
requested at the time of purchase, or obtained
by returning the instrument to our repair and
calibration facility, for a nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer. For recalibration, please
use our calibration services. Refer to our repair
and calibration section at www.aemc.com.
Serial #: ________________________________
Catalog #: 2125.52
Model #: 2620
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.
d.b.a AEMC®Instruments
www.aemc.com

Table of Contents
1. INTRODUCTION............................................................................... 3
1.1 International Electrical Symbols................................................3
1.2 DenitionofMeasurementCategories .....................................4
1.3 ReceivingYourShipment..........................................................4
1.4 OrderingInformation.................................................................4
1.4.1 AccessoriesandReplacementParts............................4
2. PRODUCT FEATURES ...................................................................... 5
2.1 Description................................................................................5
2.2 Connections..............................................................................6
3. SPECIFICATIONS............................................................................. 7
3.1 ReferenceConditions ...............................................................7
3.2 Electrical ...................................................................................7
3.3 Mechanical................................................................................8
3.4 Environmental...........................................................................9
3.5 Safety........................................................................................9
4. OPERATION .................................................................................. 10
4.1 PrincipleofOperation .............................................................10
4.1.1 MeasuringCurrentLeakage .......................................10
4.1.2 InterpretingtheMeasurement.....................................10
4.2 EquipmentTroubleshooting ....................................................11
4.3 TestingVariousCableCongurations.....................................12
4.3.1 MetalConduits............................................................12
4.3.2 GroundedNeutral .......................................................14
4.3.3 NeutralGroundedThroughanImpedance .................15
4.3.4 UngroundedSystem ...................................................15
4.4 CommonlyAskedQuestions...................................................15
5. ARTIFICIAL NEUTRAL.................................................................... 17
Warning...........................................................................................17
5.1 Description..............................................................................17

5.2 Specications..........................................................................19
5.2.1 Electrical .....................................................................19
5.2.2 Mechanical..................................................................19
5.2.3 Safety..........................................................................19
5.3 PrincipleofOperation .............................................................20
5.4 TestingUngroundedSystems.................................................21
6. MAINTENANCE ............................................................................. 22
6.1 Cleaning..................................................................................22
RepairandCalibration...........................................................................23
TechnicalandSalesAssistance ............................................................23
Limited Warranty ...................................................................................24
WarrantyRepairs...................................................................................24

GroundFault/LeakageDetectorModel2620
3
CHAPTER 1
INTRODUCTION
Warning
Thesesafetywarningsareprovidedtoensurethesafetyofperson-
nelandproperoperationoftheinstrument.
•Neverusethisclamponconductorswithavoltagepotentialabove
600Vrms.
•Neverclamparoundaconductorunlesstheclampisterminatedto
ameasuringinstrumentwithproperinputimpedance.
•Keep the jaw mating surfacesclean.Ifnecessary, use a slightly
oiledclothtoremoveoxidation.
•Ensurethatthecablesorbusbarareproperlycenteredwithinthe
clampjaws.
•Stayawayfromotherconductorswhichmaycreateinterference.
•Avoidleavingtheclampindampplacesorexposingittorunning
water.
•Duetoshockorrehazards,electricalconnectionsoftheinstrument
shouldbeperformedonlybyqualiedpersonnelandinaccordance
withlocal,state,andfederalelectricalrequirements.
1.1 International Electrical Symbols
Thissymbolsigniesthattheinstrumentisprotectedbydoubleor
reinforced insulation. Use only specied replacement parts when
servicingtheinstrument.
This symbol on the instrument indicates a WARNING and that
theoperatormustrefertotheusermanualforinstructionsbefore
operating the instrument. In this manual, the symbol preceding
instructionsindicatesthatiftheinstructionsarenotfollowed,bodily
injury,installation/sampleandproductdamagemayresult.
Risk of electric shock. The voltage at the parts marked with this
symbolmaybedangerous.

4
GroundFault/LeakageDetectorModel2620
1.2 Definition of Measurement Categories
CAT II: For measurements performed on circuits directly connected to
the electrical distribution system. Examples are measurements
onhouseholdappliancesorportabletools.
CAT III: For measurements performed in the building installation at
the distribution level such as on hardwired equipment in xed
installationandcircuitbreakers.
CAT IV: For measurements performed at the primary electrical supply
(<1000V) such as on primary overcurrent protection devices,
ripplecontrolunits,ormeters.
1.3 Receiving Your Shipment
Uponreceivingyourshipment,makesurethatthecontentsareconsistent
withthepackinglist.Notifyyourdistributorofanymissingitems.Iftheequip-
mentappearstobedamaged,leaclaimimmediatelywiththecarrierand
notifyyourdistributoratonce,givingadetaileddescriptionofanydamage.
Savethedamagedpackingcontainertosubstantiateyourclaim.
1.4 Ordering Information
Ground Fault/Leakage Detector Model 2620................... Cat. #2125.52
Includes user manual and warranty card.
1.4.1 Accessories and Replacement Parts
Adapter–4mmNon-insulatedforSafetyLeads................. Cat. #1017.45
(Converts male safety plugs to non-shielded male plugs)
ArticialNeutralModelAN-1................................................ Cat. #1971.01
Includes instrument with soft carrying case; eight 1.5 V “AA” batteries; four 5 ft (1.5 m) color-
coded leads (green, red, black, blue) and user manual.
Bag–LargeClassicToolBag .............................................Cat. #2133.71
Bag–SmallClassicToolBag............................................. Cat. #2133.72
Lead–Replacement ...........................................................Cat. #2950.29
Set of 4 Color-coded 5 ft Safety Leads (4mm straight banana plugs on both ends) for use with
the Articial Neutral Model AN-1. Rated 600V CAT IV

GroundFault/LeakageDetectorModel2620
5
CHAPTER 2
PRODUCT FEATURES
2.1 Description
TheModel2620measuresleakagecurrentshuntedtogroundcausedby
insulationbreakdownor leaks. Itenablestheoperator to locatefailures
whenorbeforetheyoccur,withoutshuttingdownequipmentorspending
hourstrouble-shooting.
It is designed specically for locating low current faults on high current
loads.ThedetectorisasensitiveACcurrenttransformercapableofmea-
suring dierential or leakage current from 500µA, and may be used to
measurecurrentupto400Acontinuousduty.
TheModel2620provides,on tworanges,1mV/mAACor1mV/AAC.The
outputleadsareterminatedwithstandard4mmbananaplugscapableof
interfacingwithanystandardmultimeter.
Useofadigitalmultimeterwithanalogbargraphisrecommended:
Digitaltoprovidethepropervoltageinputimpedance,andananalogbar
graphtotracktrends.
Whenthe2620isusedasacurrentleakagedetector,itmakesnodierence
ifthesystemissingle-phaseorpoly-phase,orifthecurrentsarein-phase
oroutofphase,balancedorunbalanced.
Thenetmagneticeldatanyinstantintimewillbezeroifalltheconductors
surroundedbythecurrentleakagedetectoraresupplyingallthecurrent
deliveredtoandreceivedfromtheload.
Ifany currentisdiverted throughanyalternate pathtoground, suchas
aninsulationbreakdown,thenetlosswillbedetectedproducingaoutput
proportionaltotheamplitudeofthefaultcurrent.
TheModel2620mayalsobeusedsimplyasahighlyaccurateclamp-on
currentprobe.Withits4"jawopeningandrangeof500µAto400A,the2620
providesaversatile way toanalyzeunbalancedcurrent measurements,
leakagevaluesongroundingconductorsandgroundloopcurrents.

6
GroundFault/LeakageDetectorModel2620
2.2 Connections
The Model 2620 is compatible with most DMMs, voltmeters, or other
voltagemeasuringinstrumentswiththefollowingfeatures:
• Rangeandresolutioncapableofdisplaying1mVACofinput.
• Voltmeteraccuracyof0.75%orbettertotakefulladvantageofthe
accuracyoftheprobe.
• Minimuminputimpedanceof1MΩ.
Connecttheprobetothemultimeterorotherinstrument.
SelecttheappropriateACvoltagerangeonyourmultimeter.
TheModel2620hastwoselectableoutputranges.The4Arangewillpro-
duceamV/mAwithanoutputof4VACat4A.The400Arangeproduces
1mV/Awith400mVAC at 400A.
Warning: User Safety
Always use a DMM, voltmeter or other displaying device, appropriately
ratedfortheworkingvoltageandforsafety.

GroundFault/LeakageDetectorModel2620
7
CHAPTER 3
SPECIFICATIONS
3.1 Reference Conditions
Ambient temperature: 23°C ± 3K
Relative humidity: 20to75%RH
Position of the conductor: Centeredinthejaws
Current frequency and form: Sinusoidal50/60Hz±0.2Hz,distortion<1%
Superimposed DC current: NoDCcurrent
Continuous magnetic eld: Eartheld<40A/m
Alternating magnetic eld: Noexternalalternatingmagneticeld
Proximity of external conductors: None
Measuring device’s impedance: ≥10MΩ/100pF
3.2 Electrical
4A Range
Nominal Current: 4AAC
Measurement Range: 0.5mA to 4AAC
Output Signal: 1mVAC/mAAC
4A Range 0.5 to 10mA 10 to 100mA 100mA to 4A
Accuracy 3%ofR±1mV 0.5%ofR±0.5mV 0.5%ofR±0.5mV
PhaseShift – < 15° < 10°
400A Range
Nominal Current: 400AAC
Measurement Range: 0.5A to 400AAC
Output Signal: 1mVAC/AAC
400A Range 0.5 to 10A 10 to 100A 100 to 400A
Accuracy 0.5%ofR±0.5mV 0.35%ofR±0.5mV 0.35%ofR±1mV
PhaseShift – < 1° < 0.6°

8
GroundFault/LeakageDetectorModel2620
Overloads
• Iplimitcurrent:permanent400ACRMS
• Peakcurrent:<1000A
• Permissibletransientdi/dt:≤30A/μs
• Conductortemperature:≤70°Cwithamaximumpeakof90°C
Load Impedance (DMM):1MΩmin
Frequency:48to1000Hz
Errors caused by external inuences:
Ambient temperature <0.1%per10K
Position of the gripped conductor
(maxwithnotcentredconductor)
0.1%typicofVOUT(non-dierential
current);0.2%max
Residual dierential
(maxwithnotcentredconductor)
0.1%typicofIP(dierentialcurrent);
0.2%max
External elds, 1V/A (1) <60mVofVs
External elds, 1mV/mA (1) <100μVofVs
Coupled DC current, 1V/A (2) <1mVforcontinuous1A
Coupled DC current, 1mVm/A (2) <0.1mVforcontinuous1A
Frequency, 1V/A (3) <1.5%from30Hzto1KHz
Frequency, 1mV/mA (3) <0.5%from30Hzto1KHz
(1): 400 A/m 50Hz eld perpendicular to the clamp opening
(2): DC current coupled onto an AC current
(3): Limited to 1KHz for 100A
3.3 Mechanical
Dimensions: 11.2x6.9x1.8"(285x175x45mm)
Weight:2.9lbs(1300g)approx
Jaw opening: 4.4"(112mm)
Maximum Jaw Gap:9.8"(250mm)
Maximum Clamping Capacity: Max4.5"(115mm)Øcable

GroundFault/LeakageDetectorModel2620
9
3.4 Environmental
Operating Temperature:
-14°to131°F(-10°to55°C);0to85%RH(non-condensing)
Storage Temperature:
-40°to158°F(-40°to70°C);0to85%RH(non-condensing)
Operating Altitude: ≤2000monuninsulatedconductors
Transport Altitude: ≤12000m
3.5 Safety
Electrical Safety
EN61010,600VCATIII
PollutionDegree:2
Electromagnetc Compatibility
IndustrialEnvironment:CriterionB
Emissivity(EN61326-1)
Susceptibility(EN61326-1)
Self-extinguishing Capability
JawsandCasing:VO(UL94)

10
GroundFault/LeakageDetectorModel2620
CHAPTER 4
OPERATION
4.1 Principle of Operation
4.1.1 Measuring Current Leakage
WhenageneratorsuppliesanACloadthroughapairofwiresinaninsulated
cable,thecurrentgoingoutononewireisequaltothecurrentreturning;
their vector sum equals zero.A ground fault changes this equality, and
theleakagecurrentdetectorpicksitup,measuresit,andprovidesanAC
voltageoutputproportionaltotheseverityoffaultcausingtheunbalance.
Thegroundconductormustnotpassthroughthedetector.
4.1.2 Interpreting the Measurement
Figure1isaschematicofasingle-phaseinstallationofamotorandan
oven,bothgrounded.Thenumbersindicatetheamplitudesofthecurrents
in dierent conductors. This installation has no faults. If you clamp the
detectoraroundthecablesatpointsA,B,CorD,itwilloutputzero.
115V/230V
15A
15A
10A
10A
AB
CD
5A 5A 10A10A
10A
5A
GND
MOTOR
ZERO
AMPERE
ZERO
AMPERE
OVEN
GND
Figure 1

GroundFault/LeakageDetectorModel2620
11
InFigure2,thereisafaultintheinstallation.AtpointA,theModel2620
detects 1A (the dierence between 16A and 15A), indicating a fault
downstream.AtpointC,itdetectszero;therefore,thefaultisnotinthe
oven.AtpointsBandD,thedetectorpicksup1A,showingthatthefault
isinthemotor.Ifitgetsworse,theleakagecurrentwillincreaseandthe
circuit breaker will trip. This motor should be disconnected immediately
andrepaired.
115V/230V 15A
15A
10A
11A
AB
CD
5A 5A 10A 11A
10A
5A
GND
MOTOR
1A
1A
OVEN
GND
1A
RESIDUAL CURRENT: 1A
Figure 2
In a three-phase installation the principle is the same. But here it is
necessarytopassallactiveconductorsthroughthedetector,includingthe
neutralifoneisused.Ifthisisdoneproperly,thedetectoroutputwillbe
zeroifthereisnocurrentleakingtoground.Ifitdetectssomethingother
thanzero,thereisagroundfaultthatshouldberepaired.
4.2 Equipment Troubleshooting
Tolocateafaultquickly,beginyourtroubleshootingattheincominglow-
voltagedistributionheader.Successivelycheckeachfeeder.IftheModel
2620 detects a ground fault, follow the feeder downstream, branch by
branch,totheequipmentwiththeproblem.
It is useful to take measurements periodically to create a history of the
qualityoftheinsulation.Thisisgoodpreventivemaintenancewhichcould
provevaluablelater.Ifyoudetectafaultcurrentthatisverylowinitially
butincreasesfromdaytoday,thismeansthatelectricalfailureisimmi-
nentand shouldbe takencare ofduring thenext plannedmaintenance
shutdown.

12
GroundFault/LeakageDetectorModel2620
Forbestresults,centertheconductorswithintheclampjaws.Checkthat
theedgesofthejawsareperfectlycleanandthattheyclosetightly.Tryto
avoidplacingtheconductorsnearthegaps,andgroupthemasmuchas
possible.
Ifyouarecareless,youcanaectthemagnitudeofthereading,depend-
ingonthecurrentbeingcarriedintheconductors.Incertain casesthis
requires an interpretation. Normally, however, the absolute value of the
groundcurrentisnotimportant.Allyouneedtoknowiswhethertheread-
ingiszero.
4.3 Testing Various Cable Configurations
Olderinstallationshave awidevariety ofelectricaldistribution systems.
Cablesareclampedtowallsandpassthroughmetalconduits,andthere
aremanycaseswhereashieldedcablewasinstalledduringrevampingof
anareaoradditiontoabuilding.Ifacableisclampedtowalls,detachit
andpullitawayfromthesurfaceatseveralpoints,sotheclampcansur-
roundalltheconductors.
4.3.1 Metal Conduits
Cablesenclosedinmetalconduitswhicharegroundedarenotaproblem.
Simplyapplytheclamparoundtheconduit;themetaltubewillnotblock
themeasurement.Cableswithametalshieldcanbehandledinthesame
wayascableinmetalconduit.
Agroundedshieldmaypresentaproblem.Ifitisasimplemetalshield,like
athinmetalwrapping,itcanbeconsideredinthesamewayasthemetal
conduit.Ifthearmorisgroundedbyaconductor,itmaybeaninconve-
nience,dependingontheterminalconnections,aswellasonthenumber
andtypeofgroundconnectionsalongthecable.
InFigure3thedetectorsensesthefaultbecause theshield isnotcon-
nected to the motor frame and the leakage current cannot ow back
throughthedetector.
InFigure4,however,theinstrumentcannot“see”thefaultbecausethe
shield is connected to the motor frame, letting the fault current return
through the detector and causing it to read zero. If possible, take your
measurementwhereyoucanavoidclosingtheshieldintheclamp,such
asthenearestcontrolpanelorinsidetheconnectingboxofthemotor.

GroundFault/LeakageDetectorModel2620
13
Figure 3 Figure 4
OnewaytocircumventthisshieldedcableproblemisshowninFigures
5and6.Heretheshieldisloopedthroughthedetectortwice,sothatthe
readingwillbethatofthefaultcurrent.
Figure 5 Figure 6

14
GroundFault/LeakageDetectorModel2620
4.3.2 Grounded Neutral
It is easy to take measurements on this type of network, where major
faultstripthecircuitbreaker. Here,thedetectorcansense minorfaults,
sorepairs canbemadebeforetheproblemgets worse.Theinspection
methodisthesameasforasingle-phaseinstallation.Rememberthaton
athree-phasenetwork,thecurrentsaddaccordingtothevectorsumlaw.
TheModel2620measuresthevectorsumoftheenclosedcurrents.Ifyou
have,forexample,a400mAfaultanda250mAfaultonthesamephase,
theModel2620 willdetect650mA. Butifthetwo faultsareon dierent
phases,theyaddvectoriallyandmayevensumuptozero.
Figure7showsanexampleinwhichtherearetwoseparatefaults:400mA
on a machine in one shop, 250mA on a machine in another shop. The
vectorsumisnot650mA,but350mA.Ifyouswitchothemachinewiththe
250mAfault,thereadingincreases,indicatingfaultsondierentphases.
Shop 2
Shop 1
Fault on Phase
1
400mA
Fault on Phase
2
250mA
Figure 7

GroundFault/LeakageDetectorModel2620
15
4.3.3 Neutral Grounded Through an Impedance
Followtheprocedurethesamewayasforagroundedneutralinstallation,
butwithcare.Whentheneutralisgroundedthroughanimpedance,afault
(evenaseriousone)maynottripthecircuitbreaker,andthecircuitwill
stay live.
4.3.4 Ungrounded System
In order to measure on an ungrounded system, the optional Articial
NeutralModelAN-1mustbeusedinconjunctionwiththeModel2620.
For measurement procedures, see the information in this manual in
Chapter5.
4.4 Commonly Asked Questions
Q. I have found an indication of 40 to 50mA on one of the main feed-
ers with 300A on the line. Is this value a fault?
A. With300Aontheline,itispossibletondanundesiredcurrentofthat
magnitude. Dierent cables in the installation may not have equivalent
capacitanceandresistance.Thesecablesprovidefor high current and,
ofcourse,weapplyahighcurrent.Therefore,wemayhaveacapacitive
inuencewhichcorrespondsto45mA.
Q. Must the conductors be centered within the jaws?
A. Youshouldavoidplacingtheconductorscloseagainstthemagnetic
circuit near the gaps, and should group the conductors as much as
possible. Best results are obtained while centering the conductor. You
mustalsoensurethatthejawmatingsurfaceareperfectlycleanandthat
nothingpreventsthemfromclosingtightly.
Q. Is it possible when checking cables which run parallel, yet supply
the same load, to clamp the detector over each cable separately?
A. Whenthereareparallelcables,themeasurementisvalidonlyifthe
detectorincludesthecablestogether.Infact,ifyoutakeeachonesepa-
rately, it is highly probable that the distribution of current in the parallel
cablesisnoteven.Thesedierencescanbecausedbythefactthatthe
cablesmaynothaveexactlythesameimpedance.

16
GroundFault/LeakageDetectorModel2620
Q. I have detected faults of 400mA and 250mA on branch circuits
being supplied by the feeder. But clamping onto the incoming feeder
resulted in an indication of 350mA. Can you explain why 400mA +
250mA = 350mA?
A. Thedetectormeasuresresidualcurrent;inotherwords,thevectorial
sumofthecurrents.Ifyouhavea400mAfaultanda250mAfaultonthe
samephase,youhaveeverychanceofnding650mA.But,ifyouhave
twofaultsondierentphases,theyaddasvectorsactually,inaverytheo-
reticalcasewecouldevenndzero.
Q. The Model 2620 detects faults caused by insulation failures. How
does detecting faults with the detector dier from information result-
ing from megohmmeter insulation tests?
A. TheModel2620willdetectelectricalfaultswhiletheequipmentisstill
inoperation.Thisoerstwodistinctadvantages.First,itisnotnecessary
toshutthe equipment down,eliminatingcostlydown time; andsecond,
you may detect faults that would not be evident under non-energized
conditions.
Q. Would it be possible to clamp the Model 2620 directly around the
grounding detector?
A. Yes,insomeinstancesitmaybepossibletodetectfaultcurrentona
specicmotorifthereisadedicatedgroundingpath.However,ifseveral
devicesaresharingacommonground,itshouldbenotedthatthefault
maybeoccurringonanyone(ormultiple)ofthedevices.

GroundFault/LeakageDetectorModel2620
17
CHAPTER 5
ARTIFICIAL NEUTRAL
Warning
• Themeasurementofleakagecurrentisdonespecicallywithaclamp
designedforthispurpose.
• TheArticial Neutral must be connected to a known low resistance
ground.
• Iffaultndingexceeds15minutes,removetheArticialNeutralfrom
itscarryingcase.
• TheArticialNeutralmustalwaysbeintheOFFpositionwhencon-
nectingtheleads.
• Always make connections from the instrument to the circuit under
test.
• Alwaysinspecttheinstrumentandaccessoryleadsforserviceability
priortouse.Replacedefectivepartsimmediately.
• Neverusetheinstrumentonelectricalconductorsratedabove600V.
5.1 Description
TheArticialNeutralisanaccessoryfortheAEMCModel2620Ground
Fault/LeakageDetector.Itgeneratesagroundedarticialneutraltoallow
themeasurementoffaultcurrentsonungrounded3phasesystems.
Inanungroundedsystem,theimpedancebetweenthetransformer,neu-
tral,andgroundisinnity.Therefore,independentoftheloadcurrent,no
groundfaultcurrentshouldow.TheArticialNeutralcreatesanintermittent
return path for the ground fault current enabling the troubleshooter to
distinguishitfromloadcurrent.
TheArticialNeutralisgroundedintermittentlythrougharelaydrivenby
aninternalclock.Thisperiodicgroundingtechniqueperformsthreefunc-
tions:
• Limitsthepossibilityofinsulationfaultswithinthecase.
• Allowsbetterdiscriminationofactualfaultcurrent.
• Limitsinstrumentoverheating

18
GroundFault/LeakageDetectorModel2620
Relayclosuretimeisselectedexternally(FastorSlow)sothattheground
faultclampcanbeusedoneitherananalogordigitaldisplay.Whenusing
ondigitalmultimeter,placetheswitchintheslowposition;whenusingan
analogmultimeter,settheswitchtothefastposition.
An internal buzzer, also driven by the relay, is activated each time the
neutralconnectionisgrounded.Threelights,oneforeachphase,show
thepresenceofeachofthethreephasesintheArticialNeutralpriorto
measurement.Afourthgreenlightindicatestheinstrumentisoperating.
WARNING
Connect and disconnect
only in the OFF position
ARTIFICIAL NEUTRAL
MODEL AN-1
ON
Slow
OFF
ON
Fast
SPECIFICATIONS
Working Voltage: 100 to 600V
Frequency: 45 to 65Hz
Batteries: 8 x 1.5V AA (LR6)
EN 61010, 600V, Cat. III, Pol. 2
600V max
ON
ON light
(green)
Male-male safety leads,
4 mm Ø, length 5 ft (1.5 m) double insulation
Carrying
Strap
Ventilation holes
3 position switch: •
ON, Slow test
•
OFF
•
ON, Fast test
[Carrying case not shown]
Safety alligator clips
C
V3
A
V1
B
V2
Red light, showing
voltage present and
phase fault
Figure 8
Table of contents
Other AEMC Security Sensor manuals