This quick start guide has been developed by the NHS and Cabinet Office and should be used in conjunction with the manufacturers Instructions for Use
CONNECT AND SET UP DEVICE
What controls does the ventilator have? (Image E)
3
SET THE CONTROLS ON THE VG70
AEONMED VG70 VENTILATOR: QUICK START GUIDE
CONNECT AND SET UP THE VG70
CLASS A ICU
•Check oxygen sensor is
connected under flap below
inspiratory port
•Attach HMEF to expiratory port (D)
•Connect patient circuit to
inspiratory port and to HMEF on
expiratory port.
•A standard humidification setup is
possible as shown
•The Y piece is then connected to a
test lung and then patient.
•The device is largely touch screen menu driven (image E)
with a control/confirm knob to the side of the screen
adjacent to an alarm mute button.
•The controls are accessed by pressing the main menu
button (highlighted yellow in the image)
•The standby button/start ventilation is in the bottom right
hand corner of the screen (highlighted blue in the image)
•These controls are used to input patient data and select the
standard modes of ICU ventilation and also to adjust the
parameters, e.g. Tidal volume, Respiratory rate, I:E, Fi02,
PEEP/CPAP etc, as well as to set limits.
Expiration valve
lock
Inspiration port
Expiration valve
Expiration port
Oxygen sensor
storage shell
Nebulizer
•Connect mains power adapter to the power plug
•Connect Oxygen from the wall or cylinder to the
oxygen input port.
•Air connection is not required as this ventilator is
turbine driven
How do I connect the ventilator?
(Image A)
1How do I set up the breathing system
and the HME filter? (Image B)
2
A
Power
Switch
Oxygen
input port
B
C
D
E
Latest Training Information can be found on:
https://portal.e-lfh.org.uk/Component/Details/606835
This quick start guide has been developed by the NHS and Cabinet Office and should be used in conjunction with the manufacturers Instructions for Use
The most important
alarms are:
Apnoea
Low Fi02
High or low airway
pressure
Disconnection
High/low VT
High PEEP
Circuit obstruction
Battery failure/mains
power failure
OPERATE THE VG70
AEONMED VG70 VENTILATOR: QUICK START GUIDE CLASS A ICU
V1.0
VP019
F
G
Select the mode by pressing the “mode” tab in the
left hand corner of the main menu screen.
How do I set the mode(s)?
(Image D)
4
•After selecting the mode of ventilation there will
be an option to change ventilatory parameters
in the settings tab
•The touch screen is used to adjust these values
or alternatively the dial can be used, pressing to
confirm entry
•Once the parameters are set. Press “Accept”
at the bottom left corner to start ventilation
•The ventilator in common with ICU ventilators
has a number of modes of ventilation including
weaning assist modes.
How do I set the parameters?
(Image F)
5
•The machine performs a Power On Self Test
•Before patient use, a test hose is connected between
the inspiratory and expiratory ports
•Follow on screen instructions to confirm the checks and
the device automatically performs a leak test, pressure
and flow sensor test and safety valve check
•If passed the machine goes into standby mode on
completion.
How do I check the device?
(Image G)
6
•High priority alarms flash red at the top of
the screen, Medium priority alarms flash
yellow at the top of the screen, Low priority
alarms show as a continual yellow light
•The alarms default to the manufacturers
settings, unless the values are altered by the
clinician on the “alarm setup” page (see
image). There are technical and clinical
alarms
The patient should be reviewed clinically and
action taken dependant on their status. The
components causing the alarm should be
reviewed. Alarm settings can be adjusted.
What do the alarms mean?
(Image I)
7I

Other manuals for vg70

1

Other Aeonmed Medical Equipment manuals

Aeonmed Aeon7400A User manual

Aeonmed

Aeonmed Aeon7400A User manual

Aeonmed Shangrila 590P User manual

Aeonmed

Aeonmed Shangrila 590P User manual

Aeonmed Shangrila510s User manual

Aeonmed

Aeonmed Shangrila510s User manual

Aeonmed vg70 User manual

Aeonmed

Aeonmed vg70 User manual

Popular Medical Equipment manuals by other brands

GI Supply RenovaRP operating manual

GI Supply

GI Supply RenovaRP operating manual

GE Venue 50 Basic user manual

GE

GE Venue 50 Basic user manual

Hausted ESC2EYEST operating manual

Hausted

Hausted ESC2EYEST operating manual

ST&G 1315S instructions

ST&G

ST&G 1315S instructions

Vaunn Medical M725N-DLWH-CHVM user manual

Vaunn Medical

Vaunn Medical M725N-DLWH-CHVM user manual

GE Medical Systems PRN 50 Service manual

GE Medical Systems

GE Medical Systems PRN 50 Service manual

Sizewise Evolution user manual

Sizewise

Sizewise Evolution user manual

PLUSOPTIX S16 user manual

PLUSOPTIX

PLUSOPTIX S16 user manual

Itamar Medical OM2196381 Operation manual

Itamar Medical

Itamar Medical OM2196381 Operation manual

AGFA DX-D user manual

AGFA

AGFA DX-D user manual

HoMedics MYB-S400 instruction manual

HoMedics

HoMedics MYB-S400 instruction manual

Zeiss Humphrey Matrix 800 user manual

Zeiss

Zeiss Humphrey Matrix 800 user manual

GENICON SINGLE-USE MONOPOLAR X-SURGE TIPS Instructions for use

GENICON

GENICON SINGLE-USE MONOPOLAR X-SURGE TIPS Instructions for use

Steelcase Sync Double-Sided Base Unit Assembly instructions

Steelcase

Steelcase Sync Double-Sided Base Unit Assembly instructions

Atos Medical PROVOX Liberty Electrolarynx Instructions for use

Atos Medical

Atos Medical PROVOX Liberty Electrolarynx Instructions for use

human care Kenmak BAMBINO Instruction for usage

human care

human care Kenmak BAMBINO Instruction for usage

Stryker SV2 Operation manual

Stryker

Stryker SV2 Operation manual

Arjo Evenda EV-ACC12 Accessories Instructions

Arjo

Arjo Evenda EV-ACC12 Accessories Instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.