
3
SYSTEM MAINTENANCE
Filter Housing/Bowls: Periodic cleaning of the polycarbonate bowls may become necessary. Remove the auto drains and
clean the bowls with a mild soapy solution. The auto drains may also be cleaned with a mild soapy solution at this time.
Dry and reinstall into the fi lter housing.
Filter Change: The fi ltration system consists of fi lter change indicators which will gradually change from green to orange
when fi lter life is spent.
Note: Air must be fl owing through the fi lters before the fi lter change indicators will function. With system pressurized, close
inlet ond outlet ball valves. Release the fi lter trio pressure by pressing the manual drain assembly to the side. Remove the
bowl assemblies by releasing the lock ring.
Calibration: Monitor calibration should be done monthly or whenever the reading may be questionable. A calibration date
sticker should be affi xed for future reference. To obtain an accurate calibration, we recommend the use of Air Systems’
calibration kits.
Part Number:
BBK-10 Canadian calibration kit for CO monitor; 10ppm CO, zero air, regulator and case - 17 liter size.
BBK-20 Calibration kit for CO monitor; 20ppm CO, zero air, regulator and case - 17 liter size.
BBK-20103 Calibration kit for CO monitor; 20ppm CO, zero air, regulator and case - 103 liter size.
BBG-O2 Oxygen sensor test gas, 20.9% oxygen - 17 liter size.
DECAL085CD Calibration decal, sold in sheet of 14.
To assure sensor accuracy, calibration of monitor is required. If you cannot obtain an accurate calibration, sensor replace-
ment may be necessary. Consult Repair Service Department before ordering.
Part Number:
CO-91NS Replacement CO sensor
CO2-O2 Replacement O2 sensor
CAUTION: Always depressurize the system before performing service.
The monitor will analyze the air sample and display the oxygen % and CO concentration in parts per million (ppm). The
system’s green “NORMAL” operation light will illuminate and the red “HIGH CO” light will fl icker approximately every sec-
ond when the CO level is below 10ppm (5ppm Canadian) and the oxygen level is above 19.5%. If the CO concentration
level exceeds the alarm set point, the green “NORMAL” light will turn off, the red “HIGH CO” light will illuminate, and the
audible alarm will sound. Once the CO concentration levels drop below the alarm set point, all alarm indicators will deac-
tivate and the unit will return to “NORMAL” operation. If oxygen levels drop below 19.5%, the green “NORMAL” light will
turn off, the red “LOW O2” light will illuminate, and the audible alarm will sound. Once the oxygen levels exceed 19.5%, all
alarm indicators will deactivate and the unit will return to “NORMAL” operation.
Carbon Monoxide/Oxygen Monitor System Components
Carbon Monoxide/Oxygen Monitor Specications
Maintenance Items
Size 2.75”H X 9.8”L X 5.1”W
Weight 3.2 lBS. (1.45kg.)
Case Extruded Aluminum
Anodized Black
Voltage
9 VDC
(9 Volt Batteries Only -
Intrinsically Safe)
Shielding Internal RFI/EMI fi lters
Operating
Temperature
4° to 113° Fahrenheit
(-15.5° to 45° Celcius)
Humidity
Range 10% to 90% relative humidity
Flow
Requirement 50 - 100 cc/min
Display 3 digit LCD - CO concentration
3 digit LCD - O2 percentage
Test Circuit Manually activated
Sensor Type Sealed electrochemical sensors for
Carbon Monoxide and Oxygen
Accuracy +/-1% full scale
Response 90% in 10-15 seconds
Detectable
Range
0-200 ppm CO
15-23% O2
Calibration Manual CO zero and span adjustments
Manual O2 adjustment
Alarm Setting 10 ppm CO (5 ppm - Canadian)
19.5% O2
Warning
Signals
Normal Operation - Green Light
High CO - Red Light
High CO/Low O2 - Audible Alarm
Low O2 - Red Light
Low Battery - Amber Light
Warranty 2 years from original date of purchase for
monitor and CO sensor, 1 year on O2 sensor