Parameters Contents Description of Setting
Detector size Connection size Size of detector connection (type)
KFactor KFactor Inherent constant of each detector (Deviation correction
factor).
Flow rate unit Flow rate unit Unit of flow rate at full scale.
Full scale Full scale Full scale flow rate value
Viscosity Function KinematicViscositysetting Set the kinematic viscosity of the service liquid. The output is
linearized automatically. The value as specified has been set
at delivery from factory.
KinematicViscosity KinematicViscositysettingvalue
Time constant Time constant
Response time. For example, “1s” means that measuring
value reaches to 63% of the step input signal after receiving
the signal in “1 second”.
Low cut off Low cut off
Low cut off can be set between 0.0% to 1.0%F.S. and
between 1 to 25%F.S. Each range has a different graduation
to set. Measuring value less than the set value is deemed as
0 automatically.
Burnout Burnout It determines analog output value when output is in fault.
Hi Hysteresis High hysteresis The gap between high alarm set point and reverting back to
normal. Setting value is fixed as 0%.
Lo Hysteresis Low hysteresis The gap between low alarm set point and reverting back to
normal. Setting value is fixed as 0%.
AL1, AL2 Alarm outputs Alarm setting points: high or low alarm, Setting points :in %
of F.S., Setting of make or break (NO or NC)
P. Out Contact output Frequencyoutput(0to1KHz),Settingoffaultoutputand
make or break contact.
Device setting Communication setting See Communication section on page 12.
Linearize Function Manual linearizer Linearizer can be set manually as an option. If set manually,
the factory settings are considered invalid.
Linearize Point Linearize points
Current output type Output type Set analog output type. 4 to 20 mA has been set at delivery.
0 to 10 V output is optional.
Parameter Setting
Contents of Parameters