Lovibond SD 50 pH User manual

SD 50
pH
Lock/Unlock Data!
C. Select pH/mV
D. Backlight ON/OFF
____________________________________________________________________________________________
Note: Icons/Characters in RED meaning "BLINK"
: Hit once / : Press and hold / : Release
3
F. Store Data
1. In lock mode: 2s
2. In measurement mode: 2s (or 2s )
3. When the storage is full, it will overwrite the oldest data.
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
G. System Setup
Menu selection:
2s (latest memory#)
(7.00 or 6.86) (dd-mm-yy/24hr or mm-dd-yy/12hr) (°C or °F)
(31-12 24 or 12-31 12ampm) .....
To enter:
!
F. pH Buffer 7.00 -> 4.00 -> 10.00
MEASUREMENT
pH CALIBRATION
Immerse the probe
into pH7 buffer
solution. Exit here for one
point calibration
Exit here for two
point calibration
or go for second point calibration as follows:
or go for third point calibration as follows:
wait
wait
wait
Clean the probe first,
then immerse into pH4
buffer solution.
Clean the probe first,
then immerse into
pH10 buffer solution.
(4) (7) (10)
on the Display
Remove protective cap from the probe
E. Store Data
Three point calibration
finished.
L M H
Low, Middle, High
Press
Press for
3 seconds
Insert the batteries
unlock
off
off
pH
unlock
off
Automatically turns off after 8 minutes of
inactivity.
Possible slight delay after pressing
before freeze of the data (!).
Mode „mV“ does not show ORP on the display,
only milivolts.
to switch, unfreeze data (see B).
After 3 sec. the file no. temporarily appears
on the display, e.g. 02.
Once the storage has reached max. capacity,
the data will overwrite the oldest files.
off
pH
lock
on
on
mV
• Re-Calibrate according to individual recommendations! Rinse the electrode with destilled water after each measurement.
• Choose Lovibond Standard Solutions or produce them with buffer tablets according to the kit manual (see backside).
Remove protective cap from the probe.
Power ON (see A)
GB
Characters in RED represent „Flashing symbols“.
4 5 6
+
+
+
1 2 3
+
-
--
-
www.ssb-berlin.de
V. Set Data/Time display type
VI. Electrode Replacement
SETUP
I. Recall Data
II. Select Buffer System
III. Set Date and Time
IV. Set Temperature system (°C/°F)
Recall Data
go to l.
Set Temperature
go to lV.
Select Buffer System
go to ll.
Set Data/Time display
type go to V.
Set Date and Time
go to lll.
To enter.
1x
4x
3x
2x
3 sec
3 sec
3 sec
3 sec
3 sec
3 sec
3 sec
3 sec
3 sec
Minutes
Minutes
24 h
AM
PM
Hours
Hours
Year
Year
Month
Day
Day
Month
pH Buffer
Order Codes
721250 pH buffer-set 4.00/7.00/10.00 (25°C)
721252 pH buffer 4.00 (25°C) 1 litre
721254 pH buffer 7.00 (25°C) 1 litre
721256 pH buffer 10.00 (25°C) 1 litre
515600 Buffer tablets pH 10 100 pc.
515601 Buffer tablets pH 10 250 pc.
515610 Buffer tablets pH 7 100 pc.
515611 Buffer tablets pH 7 250 pc.
515620 Buffer tablets pH 4 100 pc.
515621 Buffer tablets pH 4 250 pc.
Accessories
1950017 Battery CR 2032 (2 Batteries required)
194820 pH replacement electrode
384801 Beaker, 100 ml
Preparation of buffer solution
1. FilI a clean beaker with 20 ml deionised/
distilled water.
2. Add one buffer tablet of the required pH-
value and shake or crush with a clean
stirring rod until the tablet disintegrates.
3. Allow to stand for about two minutes
to ensure that the buffer is completely
dissolved and to allow any insoluble
particles to settle.
4. Immerse the electrode into the beaker
and calibrate the SD 50 in accordance
with this instruction manual.
5. Remove the electrode and rinse it tho-
roughly.
6. Use the buffer solution only once.
Note
It is important to rinse the
electrode before immerse in
a buffer solution of a different
value.
Accuracy
Fresh buffer solutions should
always be made up in dei-
onised/distilled water. The
maximum tolerance based
on the pH values 4, 7 and 10
is ±0.05 pH.
Warranty
The waterproof SD 50 is war-
ranted to be free from manu-
facturing defects for 2 years
and electrode (probe) modul
for 6 months.
G. System Setup
Unfreeze data first (see frontside B) before Menu selection.
Delete stored data.
Automatic Exit
Select V. `Set Data/ Time display type‘ first.
to increase (+);
new electrode
No.: 00386910
to decrease (-)
Adjustment:
1x
2x
3x
4x
Press for
further files. Exit
Exit
Exit
Exit
Exit
Exit
°C °C
select select
°F
123
4 5 6
Content
1 pH Meter SD 50 in a robust
plastic case with strap
2 Batteries
10 pH 4 Buffer tablets (1 strip)
10 pH 7 Buffer tablets (1 strip)
10 pH 10 Buffer tablets (1 strip)
1 Lenyard
1 Instruction Manual
Instruction manuals available in further languages, visit www.lovibond.com
SD 50 pH Specications/Features
am/pm
12 hour clock
or 24 hour
clock
am/pm 12 hour clock
24 hour clock
Characters in RED represent „Flashing symbols“.
www.ssb-berlin.de

Other Lovibond Measuring Instrument manuals

Lovibond Photometer-System MD100 User manual

Lovibond

Lovibond Photometer-System MD100 User manual

Lovibond Tintometer MD100 User manual

Lovibond

Lovibond Tintometer MD100 User manual

Lovibond SD 60 User manual

Lovibond

Lovibond SD 60 User manual

Lovibond SD 400 Oxi L User manual

Lovibond

Lovibond SD 400 Oxi L User manual

Lovibond TR 515 Assembly instructions

Lovibond

Lovibond TR 515 Assembly instructions

Lovibond TB300 IR User manual

Lovibond

Lovibond TB300 IR User manual

Lovibond Tintometer SensoDirect Salt110 User manual

Lovibond

Lovibond Tintometer SensoDirect Salt110 User manual

Lovibond Scuba User manual

Lovibond

Lovibond Scuba User manual

Lovibond Maxi Direct User manual

Lovibond

Lovibond Maxi Direct User manual

Lovibond MD 100 User manual

Lovibond

Lovibond MD 100 User manual

Lovibond Photometer-System MD100 User manual

Lovibond

Lovibond Photometer-System MD100 User manual

Lovibond RT 400 User manual

Lovibond

Lovibond RT 400 User manual

Lovibond EC 2000 Gardner Assembly instructions

Lovibond

Lovibond EC 2000 Gardner Assembly instructions

Lovibond Scuba II User manual

Lovibond

Lovibond Scuba II User manual

Lovibond Scuba+ User manual

Lovibond

Lovibond Scuba+ User manual

Lovibond SensoDirect Oxi200 User manual

Lovibond

Lovibond SensoDirect Oxi200 User manual

Lovibond CheckitDirect User manual

Lovibond

Lovibond CheckitDirect User manual

Lovibond Tintometer SensoDirect Salt110 User manual

Lovibond

Lovibond Tintometer SensoDirect Salt110 User manual

Lovibond SD 60 REDOX User manual

Lovibond

Lovibond SD 60 REDOX User manual

Lovibond Tintometer MD100 User manual

Lovibond

Lovibond Tintometer MD100 User manual

Lovibond PM 620 User manual

Lovibond

Lovibond PM 620 User manual

Lovibond LC 100 User manual

Lovibond

Lovibond LC 100 User manual

Lovibond SD 320 Con User manual

Lovibond

Lovibond SD 320 Con User manual

Lovibond PoolDirect Manual

Lovibond

Lovibond PoolDirect Manual

Popular Measuring Instrument manuals by other brands

GeoMax Zoom90 installation guide

GeoMax

GeoMax Zoom90 installation guide

Locus Solutions GO Real-Time 3G user manual

Locus Solutions

Locus Solutions GO Real-Time 3G user manual

OHAUS starter 3100 instruction manual

OHAUS

OHAUS starter 3100 instruction manual

AEMC instruments MN375 user manual

AEMC instruments

AEMC instruments MN375 user manual

WATANABE WPMZ-5 Quick instruction manual

WATANABE

WATANABE WPMZ-5 Quick instruction manual

Hioki RM3548 instruction manual

Hioki

Hioki RM3548 instruction manual

Magnet-physik FH 52 operating instructions

Magnet-physik

Magnet-physik FH 52 operating instructions

Bühler technologies GAS 222.20 Installation and operation instructions

Bühler technologies

Bühler technologies GAS 222.20 Installation and operation instructions

TESTO 606-1 instruction manual

TESTO

TESTO 606-1 instruction manual

Ludlum Measurements 18 manual

Ludlum Measurements

Ludlum Measurements 18 manual

Hitachi Relion 670 Series Applications manual

Hitachi

Hitachi Relion 670 Series Applications manual

Amprobe DM-5 instruction manual

Amprobe

Amprobe DM-5 instruction manual

CS Instruments VA 520 instruction manual

CS Instruments

CS Instruments VA 520 instruction manual

Lebow Products Inc. 7558 manual

Lebow Products Inc.

Lebow Products Inc. 7558 manual

LOVATO ELECTRIC DME D110T1 instruction manual

LOVATO ELECTRIC

LOVATO ELECTRIC DME D110T1 instruction manual

Fuji Electric FSJ instruction manual

Fuji Electric

Fuji Electric FSJ instruction manual

Rosemount 951C instruction manual

Rosemount

Rosemount 951C instruction manual

Panasonic BL-WV10A Service manual

Panasonic

Panasonic BL-WV10A Service manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.