Page
9Air Techniques, Inc.
AS10, AS12, AS21, AS22, AS30,
AS40, AS50, AS70, & AS100
SITE REQUIREMENTS
Site Electrical Requirements
AS10 AS12 AS21 AS22 AS30 AS40 AS50 AS70 AS100
Nominal Supply Voltage (VAC, see note) 120 220 120 220 220 220 220 220 220
Frequency (Hz) 60 60 60 60 60 60 60 60 60
Full Load Amps 8 4 15 8 8 12 16 24 20
Minimum Panel Breaker Rating (Amps) 20 10 30 20 20 20 30 40 30
Minimum. Wire Size (AWG) 12 16 10 12 12 12 10 8 10
Requirement
AirStar Model
Note: Install a buck or boost transformer if actual facility service is above or below the
supply voltage fluctuation of ±10% of nominal voltage ratings listed.
Figure 2. Overall Site Requirements
Remote 24 Volt Switch with
Pilot Light (sold separately)
3/8” FNPT End fitting
1/2” Copper Main Air Line
Control Cable 18 Gauge 4 Conductor
Connect to Pressure Switch.
See 24 Volt Circuit Connections.
13” max
Building Power
Supply Panel
Buck/Boost
Transformer
(optional)
Remote Air Intake Kit Manifold - Kit
includes 2 inch PVC Pipe and flexible
hose with 70 inches of clear tubing for
connection to the air intake of each
compressor. See Note 1.
Type Style
AS10 5-20R
NEMA*
AS21
AS12
6-20R
NEMA*
AS22
AS30
AS40
AS50 Hard
Wired**
See Figure 3
for Connection
Details.
AS70
AS100
Green
Dot
Green
Dot
OUTSIDE AIR PIPE
2-Inch Pipe for Air Intake.
Must be protected from rain and animals
Shroud &
Screen
Screen
Rear Clearance
12 inches minimum
from wall
Drip Leg with Drain Valve
See Note 2
36 Inch
Maximum
Height
Maximum 24
inches from
main air line
Membrane Dryer
Drain Valve
See Note 3
Notes:
1. Remote Air Intake Kit Manifold - Refer to the
Optional Accessories listing, page 18, for the kit part number
corresponding to the AirStar model.
2. Remote Air Intake Drip Leg & Valve - A drip leg
with drain valve must be installed at lower end of the
remote air intake pipe to collect condensation during
operation. Attach a drain tube to the drip leg valve to allow
drainage into oor drain/sink.
3. Membrane Dryer Drain Valve - See Key Parts
Identification, page 6, and install the Drain Tube found in
either accessory kit P/N 87133 or P/N 87134 used to drain
moisture collected in the Membrane Dryer.
Side
Clearance
12 inches
*Use Hospital Grade
Receptacle Only
** Only AS100 Models
use 3-Phase Power.
Disconnect Needed
when Servicing.