USAGE
Snap appropriate grip into the wrist dynamometer body to secure unit for test.
hold unit either by grasping the dynamometer body, or using a single or dual grip
handle. Set max indicator to zero. Have patient twist grip with maximum force,
note reading. Reset to zero for next test.
CALIBRATION
The BASELINE® wrist dynamometer is a sealed unit and calibrated at the facto-
ry. However, if indicator needle is out of “zero-range” it may be reset. Remove
clear cover by turning counter clockwise. Adjustment pin located by 220Kg mark-
ing. Turn pin to reset to zero. If unit is leaking hydraulic fluid it should be returned
to factory for repair and recalibration.
WARRANTY
The BASELINE® Hydraulic Wrist Dynamometer is warranteed for 1 (one) full year
(parts and labor) from date of purchase. If unit needs repair contact your local
dealer.
ALCO Sales & Service Co.
6851 High Grove Blvd
Burr Ridge, IL 60527
U.S.A.
TEL: 800-323-4282, 630-655-1900
FAX: 800-950-1167
HOW TO INSTALL HANDLES Dual Grip Handle
Single Grip Handle
Single Grip Handle:
1. Remove screws “D” and “J”
2. Position single grip handle and apply 2 thumb screws to connect the handle to the
openings at “D” and “H”
Dual Grip Handle:
1. Remove screws “A” “D” E” and “H”
2. Position dual grip handle and apply 4 thumb screws to connect the handle to the
openings at “A” “D” “E” and “H”
A B C D
E F G H
underside of wrist
dynamometer
800-323-4282 • 630-655-1900 • www.alcosales.com