All-sun EM6000 User manual

EM6000, TRUE RMS
DIGITAL MULTIMETER
OWNERS MANUAL
Read this owners manual thoroughly before use 6000
http://www.all-sun.com
OEM & ODM manufacturer of test & measurement instruments in China

This multimeter is compact 3 5/6-digit true rms digital
multimeter with the USB2.0 standard interface. It can be
used for measuring DC voltage, DC current, true-rms AC
voltage, true-rms AC current, resistance, capacitance,
frequency, diode, transistor, temperature, continuity and
duty cycle. It has the functions of polarity indication, data
hold, backlight, overrange indication, low battery indication,
relative value measurement, max/min value measurement
and automatic power-off. It can be operated easily and is
an ideal measurement tool.
1
GENERAL DESCRIPTION
This instrument is warranted to be free from defects in
material and workmanship for a period of one year. Any
instrument found defective within one year from the delivery
date and returned to the factory with transportation charges
prepaid, will be repaired, adjusted, or replaced at no charge
to the original purchaser. This warranty does not cover
expandable items such as batteries or fuses. If the defect
has been caused by a misuse or abnormal operating
conditions, the repair will be billed at a nominal cost.
WARRANTY
http://www.all-sun.com

2
FEATURES
1. Measures the true-rms values of ac voltage and ac
current.
2. Transfers the measurement readings of the meter to a
computer through standard USB port.
3. Provides analog bar graph, reading unit indication and
backlight.
4. The sampling rate is 3 times/sec. The bar graph updates
at 30 times/sec.
5. Provides automatic polarity indication, manual range
mode and autorange mode.
6. Provides low battery indication, automatic power-off,
manual power-off, relative value measurement,
and max/min value measurement.
7. Measures DC/AC voltage, DC/AC current, resistance,
capacitance, frequency, diode, transistor, temperature
(°C/°F), continuity and duty cycle.
8. Provides overload protection for all ranges.
9. Low power consumption.
http://www.all-sun.com

3
This meter has been designed according to IEC-61010
concerning electronic measuring instruments with a mea-
surement category (CAT II 600 V) and pollution degree 2.
To avoid possible electric shock or personal injury, follow
these guidelines:
Warning
Do not use the meter if it is damaged. Before you use
the meter, inspect the case. Pay particular attention
to the insulation surrounding the connectors.
Inspect the test leads for damaged insulation or exposed
metal. Check the test leads for continuity. Replace
damaged test leads before you use the meter.
Do not use the meter if it operates abnormally.
Protection may be impaired. When in doubt, have the
meter serviced.
Do not operate the meter around explosive gas, vapor,
or dust.
Do not apply more than the rated voltage, as marked
on the meter, between terminals or between any
terminal and earth ground.
Before use, verify the meter's operation by measuring
a known voltage.
When measuring current, turn off circuit power before
a.
b.
c.
d.
e.
f.
g.
SAFETY INFORMATION
http://www.all-sun.com

4
connecting the meter in the circuit. Remember to place
the meter in series with the circuit.
When servicing the meter, use only specified
replacement parts.
Use caution when working above 30Vac rms, 42V peak,
or 60Vdc. Such voltages pose a shock hazard.
When using the probes, keep your fingers behind the
finger guards on the probes.
When making connections, connect the common test
lead before you connect the live test lead. When you
disconnect test leads, disconnect the live test lead first.
Remove the test leads from the meter before you open
the battery cover or the case.
Do not operate the meter with the battery cover or
portions of the case removed or loosened.
To avoid false readings, which could lead to possible
electric shock or personal injury, replace the battery as
soon as the low battery indicator ( " " ) appears.
When in Relative Mode, the symbol " " is displayed.
Caution must be used because hazardous voltage
may be present.
When in MIN mode, the symbol "MIN"is displayed.
Caution must be used because hazardous voltage
may be present.
h.
i.
j.
k.
l.
p.
m.
n.
o.
http://www.all-sun.com

5
r.
s.
Do not use the meter in a manner not specified by this
manual or the safety features of the meter may be
impaired.
Remaining endangerment:
When an input terminal is connected to dangerous live
potential it is to be noted that this potential at all other
terminals can occur!
CATII - Measurement Category II is for measurements
performed on circuits directly connected to low
voltage
installation.(Examples are measurements on
household
appliances, portable tools and similar equipments .)
Do not use the meter for measurements within
Measurement Categories III and IV.
q.
http://www.all-sun.com

6
d.
e.
Disconnect circuit power and discharge all capacitors
before testing resistance, diode, capacitor, temperature
and continuity.
Use the proper terminals, function, and range for your
measurements.
Before measuring current, check the meter's fuses and
turn off the power to the circuit before connecting the
meter to the circuit.
Before rotating the range switch to change functions,
disconnect test leads from the circuit under test.
Remove test leads from the meter before opening
the meter case or the battery cover.
Because of anti-interference designs, the meter may
stop working in strong interference environment. Turning
on the meter again can solve this problem.
Caution
To avoid possible damage to the meter or to the equipment
under test, follow these guidelines:
a.
b.
c.
f.
http://www.all-sun.com

7
SYMBOLS
AC (Alternating Current)
DC (Direct Current)
DC or AC (alternating current or direct current)
Important safety information. Refer to the manual.
Dangerous voltage may be present. Be cautious.
Earth ground
Fuse
Conforms to European Union directives
Double insulated
Low battery
Diode
http://www.all-sun.com

8
INSTRUCTION
Figure 1
Meter Instruction
8 9 10
12
11
1
2
3
4
5
6
7
http://www.all-sun.com

9
1. Display
LCD display, with a max. reading of 9999
2. " " button
Press this button to set the meter to Relative Mode,
and "" will appear as an indicator. To exit Relative
Mode, press this button again, and "" will disappear.
Pressing and holding down this button for about 2
seconds set the meter to USB communication mode
(In this mode, you can transfer the readings of the meter
to a PC by the USB interface and the communication
application. ), and "USB" appears as an indicator.
To exit the USB communication mode, press and hold
down this button for about 2 seconds, and "USB"
disappears.
3. " " button
When measuring ac voltage ( or ac current ), you can
press this button to select frequency, duty cycle or ac
voltage ( or ac current ) measurement function, and the
display will show the relevant symbol.
In frequency or duty cycle measurement mode, you
can press this button to switch between frequency and
duty cycle measurements.
4. " " button
Press this button to hold the present reading on the
display, and ""will appear as an indicator. Press this
http://www.all-sun.com

10
button again to exit Data Hold mode, and ""will
disappear.
Press and hold down this button for about 2 seconds to
enable or disable the backlight. The backlight will turn
off automatically about 10 seconds later after it is turned
on.
5. " " button
*Press this button, the meter enters the MAX mode,
and the LCD shows "MAX"as an indicator, the present
reading on the LCD is the maximum reading of all
readings taken since the mode was activated. Press
this button again, the meter enters the MIN mode,
and the LCD shows "MIN" as an indicator, the present
reading on the LCD is the minimum reading of all
readings taken since the mode was activated.
Press and hold down this button for about 2 seconds
to exit MAX and MIN modes.
6. Function / Range switch
It can be used to select the desired function and range
as well as to turn on or off the meter.
To preserve battery life, set this switch to the "OFF "
position if you don't use the meter.
7. "A" jack
Plug-in connector for the red test lead for current
( 600mA*20A ) measurements.
http://www.all-sun.com

11
8. "*A/mA" jack
Plug-in connector for the red test lead for current
*<600mA *measurements.
9. "COM" Jack
Plug-in connector for the black test lead for all
measurements.
10. " V*Hz" jack
Plug-in connector for the red test lead for all
measurements except current measurements.
11."Range" button
In voltage, current or resistance function, pressing this
button causes the symbol "AUTO" on the display to
disappear, it means that the meter changes from
autorange mode to manual range mode. In manual
range mode, press this button to select desired range.
12. "Select" button
This button can be used to switch the meter :
a. between ac and dc measurement functions.
b. among resistance, diode and continuity measurement
functions.
c. between fahrenheit temperature measurement mode
and celsius temperature measurement mode.
Turning on the meter while holding down this button will
disable the automatic power-off feature, and the symbol
"APO"on the display will disappear.
http://www.all-sun.com

How to Connect the Adapter
Figure 2
Capacitance test socket
Transistor test socket
12
Figure 3
LCD Display
http://www.all-sun.com
10
12
15
16
6 5 4 3 12789
11
17
18
13
14

13
1. -----Duty Cycle is selected
2. -----Fahrenheit temperature test is selected
3. -----Celsius degree test is selected
4. --- Transistor hFE test is selected
5. ---- Continuity test is selected
6. --- Diode test is selected
7. ---- Relative mode is active
8. ---USB serial port communication is enabled
9. ---- Data Hold is enabled
10. ---Autorange mode is selected
11. --- Automatic power-off mode is selected
12. --- Maximum reading is being displayed
13. ----Minimum reading is being displayed
14. ----DC
15. --- Negative sign
16. --- AC
17. ---- Battery is low and should be replaced
immediately
18. Bar graph
http://www.all-sun.com

14
Using the Bar Graph
The bar graph at the bottom part of the LCD is a analog
display only for the measurements of ac/dc voltage, ac/dc
current and resistance. The length of its lit segments is
proportional to the present reading on the LCD. The bar
graph is unavailable for the other measurements. The bar
graph is like the needle on an analog meter. It has an
overload indicator ("") on its right and a negative polarity
indicator ("") on its left.
Because the bar graph updates about 30 times per second,
which is 10 times faster than the digital display, the bar
graph is useful for making peak and null adjustments and
for observing rapidly changing inputs.
The number of lit segments indicates the measured value
and is relative to the full-scale value of the selected range.
In the 600V range, for example, the major divisions on the
scale represent 0, 100, 200, 300, 400, 500 and 600V. An
input of
-
100V lights the negative sign and the segments
up to the "10"on the scale.
http://www.all-sun.com

15
Units on the LCD
mV,V
µ
A,mA,
A
Voltage unit
mV: Millivolt ; V: Volt;
1V=103mV
Current unit
µ
A: Microamp; mA: Milliamp; A: Ampere;
1A=103mA=106
µ
A
Ω, kΩ
MΩ
nF,
µ
F
°C
,
°F
Hz,
kHz,
MHz
Resistance unit
: Ohm; k : Kilohm; M : Megohm;
1M =103k=106
Capacitance unit
nF: Nanofarad;
µ
F: Microfarad;
1F=106
µ
F=109nF=1012pF
Temperature unit
°C
: Celsius degree;
°F
: Fahrenheit degree;
a (
°F
)= 32+1.8 x b (
°C
)
Frequency unit
Hz:Hertz; kHz:Kilohertz; MHz:Megahertz;
1MHz=103kHz=106Hz
http://www.all-sun.com

Maximum Voltage between
any Terminal and Earth Ground: 600Vrms
Fuse Protection for "µAmA" Jack Inputs:
500mA, 1000V, FAST, Min. Interrupt Rating 20000A
Fuse Protection for "A" Jack Inputs:
20A, 1000V, FAST, Min. Interrupt Rating 20000A
Display: LCD, with a max. reading 9999
Overrange Indication: "OL"shown on the LCD.
Negative Polarity Indication: "
-
"displayed automatically
Sampling Rate: 2~3 times/sec
Operating Temperature: 0°C,40°C,<75%RH
Storage Temperature: -30°C,60°C,<85%RH
Operating Altitude: 0 to 2000 meters
Battery: 9V, 6F22 or equivalent
Low Battery Indication: " " shown on the display
Size: 185X86X44mm
Weight: about 380g (including battery and the holster)
16
Accuracy is specified for a period of one year after
calibration and at 18°C to 28°C, with relative humidity
< 75%. Accuracy specifications take the form of:
*([% of Reading]+[number of Least Significant Digits])
GENERAL SPECIFICATION
SPECIFICATION

17
Input Impedance: 60mV and 600mV ranges: >100MΩ
the other ranges: 10MΩ
Overload Protection: AC600V; Crest Factor: 3.0
Reading: True rms
Frequency: 40Hz~400Hz
Note: When the input terminals are shorted, the display
may show a reading. It doesn't matter and will not
affect the measurement accuracy.
±(2.0%+10)
60mV
600mV
6V
60V
600V
10µV
0.1mV
1mV
10mV
0.1V ±(0.8%+5)
±(0.8%+5)
±(1.0%+7)
ResolutionRange Accuracy
AC Voltage
ResolutionRange Accuracy
60mV
600mV
6V
60V
600V
10µV
0.1mV
1mV
10mV
0.1V
±(1.5%+10)
DC Voltage
Input Impedance: 60mV and 600mV ranges: >100M
the other ranges: 10M
Overload Protection: AC600V
±(0.5%+5)
±(1.6%+10)

18
Overload Protection:
For "µA/mA"jack inputs: Fuse, 500mA/1000V, fast action,
For "A"jack inputs: Fuse, 20A/1000V, fast action,
Max. Input Current: 20A ( For inputs > 5A : measurement
duration<10 secs, interval >15 minutes )
DC Current
Overload Protection:
For "µA/mA"jack inputs: Fuse, 500mA/1000V, fast action,
For "A"jack inputs: Fuse, 20A/1000V, fast action,
Max. Input Current: 20A ( For inputs > 5A : measurement
duration<10 secs, interval >15 minutes )
ResolutionRange Accuracy
autorange
autorange
Remark
6000µA
60.00mA
600.0mA
20A
1.0µA
0.01mA
0.1mA
10mA ±(1.5%+7)
6.000A 1mA
600.0µA 0.1µA
±(1.0%+7)
autorange
ResolutionRange Accuracy
autorange
autorange
Remark
6000µA
60.00mA
600.0mA
20A
1.0µA
0.01mA
0.1mA
10mA ±(2.5%+10)
6.000A 1mA
600.0µA 0.1µA
±(2.0%+15)
autorange
AC Current

19
Overload Protection: AC600V
Open Circuit Voltage: < 0.7V
Resistance
600.0
6.000k
60.00k
600.0k
6.000M 1k
100
1
10
0.1 ±(1.0%+5)
±(0.8%+5)
60.00M 10k ±(3.0%+5)
ResolutionRange Accuracy
±(1.5%+5)
Frequency: 40Hz~400Hz
Crest Factor: 3.0
Reading: True rms
Note: When the input terminals are shorted, the display
may show a reading. It doesn't matter and will not
affect the measurement accuracy.
http://www.all-sun.com
Table of contents