manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Altronics
  6. •
  7. Measuring Instrument
  8. •
  9. Altronics Q0965 User manual

Altronics Q0965 User manual

Q0965 • High Current AC/DC Clamp Meter 600A
1. SUMMARY
This 3 5/6 Digit Highly Stable clamp meter uses two AAA (1.5V) batteries. It has a LCD with 18mm
digit display, which makes the reading clearer and the operation more convenient. It can test DCV, ACV,
DCA, ACA, resistance, capacitance, frequency, temperature, diode, continuity, as well as non-contact
AC voltage test. It is equipped with clamp jaw lighting, Backlight, unit symbol display, True RMS, data
hold function, Relative measurement, auto power off and warning functions. It works in both auto range
and manual range. To assure high accuracy and resolution, it adopts a micro-processor which drives the
LCD directly and a dual integral A/D converter IC. It is an ideal tool for labs, factories radio-technology
and household.
2. SAFETY NOTICE
The instrument is designed according to IEC1010-1 standard (safety standard issued by International
Electro technical Committee), CAT III 600V and UL3111-1. Please read the followings carefully before
any operation.
1. Introduction for safety symbols:
: Existing High Voltage. : Dual Insulation. : Operator must refer to manual. : Low
battery indication.
2. Notice:
(1) Please disconnect any signal and pluck out the test leads before opening the outside cover.
(2) To avoid electric shock and damage to the meter, water should not be leaked inside.
(3) DO NOT use this meter to do any test if the case hasn’t finished installation and its screws haven’t
been tightened.
(4) DO NOT input the value higher than the limit at every range.
(5) Please don’t input voltage in resistance terminal.
(6) Please turn the power switch to OFF after the test.
(7) To avoid the damage of battery leakage to the meter, please remove the battery for long-term storage.
(8) To avoid electric shock and damage to the meter, do not input voltage higher than DC 60V or AC
36V during the measurement.
(9) To avoid electric shock, please be seriously careful when nipping the bus bar or something insulated.
3. FEATURES
1. GENERAL FEATURES
(1) Display: LCD.
(2) Max. display: 6000 digits (3 5/6) , and auto polarity display .
(3) Clamp open: Max. Open: 28mm.
(4) Over-range display: “OL” displayed.
(5) Data hold.
(6) Relative Value measurement.
(7) Sampling rate: approx.3 times/sec.
(8) Low battery indicator: “ ”displayed.
(9) Continuity test: buzzer sounds when<50Ω.
(10) Range: Auto range or Manual range.
(11) Auto power off.
(12) Power consumption: about 3mA.
(13) Power: 2 ×1.5V batteries (AAA 7#battery).
(14) Working environment: (0~40)℃, relative humidity: <70%.
(15) Dimension: 185mm×65mm×32mm (length ×width ×height).
(16) Weight: approx. 200g (including battery).
(17) Accessories: user manual, test leads, temperature sensor TP01 banana probe, twoAAA batteries.
2. TECHNICAL FEATURES
(1) Accuracy:±(a% ×reading data + digits). To assure accuracy, the environment temperature should
be (23±5) ℃, relative humidity be <70%. One year accuracy guarantee since production date.
(2) Temperature coefficient: 0.1×specified accuracy/1℃(<18℃or>28℃).
(3) DC voltage (DCV)
Range
Accuracy
Resolution
600mV
±(1.0% reading+5)
0.1mV
6V
±(0.5% reading+5)
1mV
60V
10mV
600V
100mV
Input impedance: at 600mV range >40MΩ, at other ranges is 10MΩ.
Overload protection: 1000Vrms.
(4) AC voltage (ACV)
Range
Accuracy
Resolution
6V
±(0.8% reading+5)
1mV
60V
10mV
600V
100mV
Input impedance: 10MΩ.
Overload protection: 1000Vrms.
Frequency response: 600V at 40Hz~1kHz, other ranges 40Hz~2kHz.
Display: True Rms Response (calibration base on Sine Wave RMS).
(5) DV Current (DCA)
Range
Accuracy
Resolution
60A
±(1.5% reading+5)
0.01A
600A
0.1A
800A
1A
Overload protection: 1000A (input time <60 seconds).
NOTE: Clear to zero before measurement,The testing object shall be placed in the middle of the clamp
jaw.
(6) AC current (ACA)
Range
Accuracy
Resolution
60A
±(1.5% reading+5)
0.01A
600A
0.1A
800A
1A
Frequency response: Sine wave and triangle wave 40Hz- 1kHz; other waveforms 40Hz-200Hz.
Overload protection: 1000A (input time <60 seconds)
NOTE: The testing object shall be placed in the middle of the clamp jaw.
(7) Resistance
Range
Accuracy
Resolution
600Ω
±(0.8% +5)
0.1Ω
6kΩ
±(0.8% +1)
1Ω
60kΩ
10Ω
600kΩ
100Ω
6MΩ
1kΩ
60MΩ
±(1.0% +5)
10kΩ
Open circuit voltage: 500mV, Overload protection: 250Vrms.
NOTE: DO NOT input any voltage value at this range!
(8) Capacitance
Range
Accuracy
Resolution
6nF
±(5.0% +10)
1pF
60nF
±(2.5% +5)
10pF
600nF
100pF
6uF
1nF
60uF
10nF
600uF
±(5.0% +10)
100nF
6mF
1uF
60mF
10uF
Overload protection: 250Vrms. Warning: DO NOT input voltage value at this range!
(9) Frequency
Range
Accuracy
Resolution
10Hz
±(0.5% reading+4)
0.01Hz
100Hz
0.1Hz
1kHz
1Hz
10kHz
10Hz
100kHz
100Hz
1MHz
1kHz
10MHz
10kHz
Input sensitivity: >0.7V. Overload protection: 250Vrms.
(10) Temperature
Range
Accuracy
Resolution
(-40~1000 )℃
<400℃: ±(1.0%+5)
≥400℃: ±(1.5%+15)
1℃
(0~1832 )℉
<750℉: ±(1.0% + 5)
≥750℉: ±(1. 5% + 15)
1℉
Sensor: K type banana plug (Nickel-chromium-Nickel silicon).
Warning: DO NOT input voltage value at this range!
(11) Continuity Test
Range
Description
Test Conditions
Diode forward voltage drop
Forward DC current is approx0.8mA,
reverse voltage is approx2.2V.
When the resistance under test is less
than 50Ω, buzzer sounds continuously.
Open circuit voltage: 2V
Overload protection: 250V TRMS.
Warning: DO NOT input voltage value at this range!
4. OPERATION
4.1 PANEL DESCRIPTION
(1)Jaw: 0 to 800AC current and NCV detecting device.
(2)Clamp gunlock: Pressing the gunlock can turn on or off
the clamp.
(3)Hand protection: A safe design to protect users from
touching the dangerous area.
(4)Clamp light: Opening the clamp light can light up the
tested area in the dark to prevent the danger.
(5)NCV indicator: Detect the existence of the high voltage
around to prevent the electric shock.
(6)Function choosing switch: Used to select the function, range or turn on or turn off
the clamp meter.
Function
Description
NCV
Non-contact voltage test.
DCA, ACA measurement.
ACV measurement. Press “Hz/DUTY”to switch to Frequency/Duty cycle
measurement mode.
DCV measurement.
Press “SELECT”to switch to Diode, Continuity, Capacitance and Resistance
measurement mode.
Hz
Frequency measurement.
℃/℉
Temperature measurement. Press “SELECT” button to switch to ℃or ℉.
(7)Function key.
SELECT / KEY:
1) This is a key for Selecting function, Base on a working principle of Trigger action. Short press this
key to choose measurement modes: Choose DC or AC at “ ”, choose“Ω→→ → ” at “ ” ,
choose “℃” or “℉” when at Temperature range.
2) The Meter will be Auto-power off when it is not being using within 15 minutes, then it will be
entering into sleeping mode, the Buzzer will makes 5 times of alarm reminder within 1 minute before
Auto-power off; Please press this “SELECT” function or Power on/off button to turn on the Power if you
want to restart it.
3) Press this key and put it on Hold and then turn on the Power to awake it from sleeping mode, and then
Auto-power off function will be canceled.
4) Illuminating lamp control key, press it for more than 2 seconds to turn on Illuminating lamp, and then
press it for more than 2 seconds again to make it off.
5) Backlight control key, press it for more than 2 seconds to turn on Backlight (it will be shown on LCD)
then press it for more than 2 minutes again to make it off. Backlight function will be off automatically
when you release backlight control key after 30 minutes.
△/ RANGE key:
1.In and measurements, it is REL function. Press this key to clear “of readying” and enter
relative value measurement. REL symbol will appear on the LCD. Press this key again to exit REL
measurement. In and measurements, if the display dose not return to zero before measurement,
press this key to return to zero before measurement.
2.In voltage and resistance measurements, it is RANGE function, to choose from auto range or manual
range. The meter is default in auto range. The symbol “AUTO” appears on LCD. Press this key for one
time to switch to manual range. Press it again for one time to increase on step from low to high. Press it
for more than two seconds to return to AUTO range.
Hz/DUTY KEY: In ACA/ACV measurements, press this key to switch measurement between frequency,
duty ratio, voltage and current. In frequency measurement, press this
key to switch between measurement of frequency and duty ratio
(1-99%).
HOLD/ KEY:
1) It is a key for Data Hold, Base on a working principle of Trigger
action, short press this key to keep the current Data of measurement
on LCD, LCD will show symbol “ ”, press this key again to exit
Data Hold.
(8)LCD: Displaying measurement value and unit.
Sequence
Functions
Description
1
APO MAX-MIN
Auto power off function has been activated, High pressure warning.
MAX/MIN Null.
2
△
Relative measurement.
3
HOLD
Data hold
4
-
Indicates negative readings.
5
DC
DC measurement.
6
AC
AC measurement.
7
AUTO
Auto range.
8
Null.
9
Diode test.
10
Continuity beeper is on.
11
hFE
nF, μF
mV, V
uA,mA,A
%,℃,℉
NCV
MΩ,kΩ,Ω
Hz, kHz, MHz
Null.
Capacitance unit: nF, uF.
Non-contact voltage testing.
Current unit: uA, mA, A. Null: uA, mA.
Duty cycle, Degrees Celsius, Degrees Fahrenheit
Non-contact voltage testing.
Megohm, Kilohm, Ohm.
Hertz, Kilohertz, Megahertz
Low battery indication.
(9)V/Ω input terminal: Measurement inputs positive terminal (red test lead). COM
input terminal: Measurement inputs negative terminal (black test lead).
(10)Carrying belt.
4.2 DCAMEASUREMENT
(1) Press the clamp gunlock to open the clamp jaw to get
testing wire inserted in the middle and clipped
completely. To ensure measurement accuracy, please
check to ensure that the testing wire is placed in the
middle of clamp jaw. Measure one wire in one time. If
measure more than one wires in one time, the reading
will be inaccurate.
(2) Turn the rotary function switch to position. The meter is default at DCAmeasurement after
power on. Press SELECT to switch between DC measurement.
(3) When the meter is used in a strong magnetic field, the readings could be unstable or inaccurate. Press
REL to clear to zero before measurement.
NOTE:
1) The meter uses Hall component to detect current. Hall component is a sensitive component. It is
magnetic sensitive. It is also sensitive to temperature and mechanical stress. Any impact on it would
cause the changes on reading in a short time. This causes some remaining values on the DC ampere
reading when there is no current. To get accurate reading, press REL key to clear reading before
measurement, and the testing wire must be placed in the middle of the clamp jaw.
2) Press the clamp gunlock to open the clamp jaw to get testing wire inserted in the middle and clipped
completely. Ensure that the testing wire is placed in the middle of clamp jaw. Otherwise there will be
additional accuracy ±1.0% of reading. Read the measurement current value of the testing wire on the
LCD.
3) In the measurement of the DC current, if the reading is positive value, the current is top to down
(broad is top, bottom cover is down).
4) The max testing current is 800A. Exceeding the rated current for long time will damage the meter.
Follow below instructions to get more accurate DC current measurement:
①Switch off the testing wire.
②Press the clamp gunlock to open the clamp jaw to get testing wire inserted in the middle and clipped
completely. Ensure that the testing wire is placed in the middle of clamp jaw.
③When the meter reading is stabilized in the minimum value, press the RELkey to clear it.
④Switch on the testing wire. Read after clamp meter remains stable.
⑤When measuring current, to get more accurate readings, the ambient temperature must be 0 to 60℃.
4.3ACA MEASUREMENT
(1) Turn the rotary function switch to position. The meter is default at DCAmeasurement after
power on. Press SELECT to switch betweenAC measurement.
(2) Open the clamp jaw and grip one wire and get the reading directly.
To get the most precise reading, should user put the wire in the middle of the close jaw.
Note:
1. Firstly users should select the highest range, if users are not sure about the range of current under
test, and then select the proper range based on displaying value.
2. DO NOT input the current higher than the limit, or it would damage the meter.
4.4ACV MEASUREMENT
(1) Set the rotary switch to “ ” range.
(2) Turn the function swift to ACV range. Insert the black test lead into “COM”
terminal, and the red one into “V/Ω” terminal.
(3) Connect the test leads into the circuit and get the measurement value from the
LCD.
Note:
1. DO NOT input a voltage over the limit or It may cause damage to the circuit of the
meter.
2. Be careful while measuring a high voltage circuit. Do not touch the high voltage
circuit.
4.5 DCV MEASUREMENT
(1) Set the rotary switch to “ ” range.
(2) Turn the function swift to DCV range. Insert the black test lead into “COM” terminal, and the red
one into “V/Ω” terminal.
(3) Connect test leads to the test point. LCD will display polarity and voltage of the test point connected
by the red test lead.
Note:
1. DO NOT input a voltage over the limit or It may cause damage to the circuit of the meter.
2. Be careful while measuring a high voltage circuit. Do not touch the high voltage circuit.
4.6 Non-Contact Voltage Detecting
WARNING
This Function might be disturbing by different external Sources, then might be incurred a wrong alarm
reminder, test result is just for reference when using this function.
Turn Switch button to “NCV” position, the Target Circuit to be placed on the top of
Meter, Meter will shows strong and weak signals, meanwhile, Beeper will makes
“tick-tick-tick” of alarm reminder.
Note:
1. Even if there is no any indication, the voltage might still be there. Do not rely on
NCV detector as the only way to judge whether a Voltage is still existed on the Wire
lead or not.
2. Voltage detecting may be affected by power socket design, type of insulation and its
thickness and other factors.
3. Interference sources at the external environment such as flashing light, motor might cause a wrong
signal for a wrong judgement.
4.7 RESISTANCE MEASUREMENT
(1) Set the rotary switch to “ ” range and press “SELECT” to swift to “Ω” range.
(2) Insert the black lest lead into “COM” terminal and the red one into “V/Ω” terminal.
(3) Connect the test leads to the measured resistance to get the measurement value.
Note:
1. Before measuring in line resistor, make sure that the power is off and all capacitors
have discharged completely.
2. Do not apply any voltage at resistance range.
3. When input terminal is in open circuit, LCD will display “OL”.
4. If there is an “OL” displayed on the LCD, it means it is over range, switch to a higher range. When
the test leads are shorted, there would be about 1Ω stray resistance.
4.8 DIODE AND CONTINUITY TEST
(1) Set the rotary switch to “ ” range. The original state is diode measurement mode.
(2) Forward measurement: Connect red test lead to the positive polarity and the black test lead to the
cathode polarity of the diode. LCD will display the approx. value of forward voltage
drop.
(3) Backward measurement: connect red test lead to the cathode polarity and the black
test lead to positive polarity of the diode. LCD will display “OL”.
(4) The complete diode testing includes forward and backward measurement. If the
result doesn’t meet the descriptions above, it means the diode is broken.
(5) Press “SELECT” key to select the continuity measurement mode.
(6) Insert the black test lead into “COM” terminal, and the red one into “V/Ω” terminal.
(7) Connect test leads to two points of tested circuit, if the resistance is less than 50 Ω, the buzzer
sounds.
Note:
DO NOT input any voltage value at this range!
4.9 CAPACITANCE TEST
(1) Set the rotary switch to “ ” range and press “SELECT” to swift to “ ” range.
(2) Insert the black test lead into “COM” terminal, and the red one into “V/Ω” terminal.
(3) Press “REL” key to clear the reading to make the LCD displayed zero.
(4) Connect the capacitor to “COM” and “V/Ω” terminal. (Note: the red test leads is for
positive pole +). LCD displays capacitance value.
Note:
1. Don’t input voltage or current at Capacitance range.
2. In order to assure the accuracy, please press “REL” to clear the reading before testing.
3. There is only the auto range mode under the capacitance range.
4. The capacitor must be completely discharged before testing.
5. The reading of 200uF range will take more than 15 seconds to be stable.
4.10 FREQUENCY MEASUREMENT
(1) Turn the function swift to Hz range. Insert the black lest lead in “COM” terminal and
the red one in “V/Ω” terminal.
(2) Connect the test leads to the measured circuit and get the measurement value.
Note:
1. In noisy environment, it's better to use shield cable to measure a low signal.
2. When measuring high voltage circuit, do not touch the high voltage circuit.
3. Don’t input voltage higher than 250Vrms or it may damage the meter.
4.11 TEMPERATURE TEST
(1) Turn Switch button to “℃/℉”, press “SELECT/ ”key to choose ℃ or ℉
measurement mode.
(2) Insert the black plug of the thermocouple sensor into “COM” and red one into
“V/Ω” jack, and put the working terminal (temperature measuring end) of
thermocouple on the surface or inside the object to be tested. Then you can read
temperature from the screen, and the data is in Centigrade.
Note:
1. The meter should be far away high temperature. Use the thermocouple probe within
the specified temperature range.
2. Don’t change the temperature probe randomly, or the accuracy will not be guaranteed.
3. Don’t input voltage at temperature range.
4.12AUTO POWER OFF
(1) The meter will auto power off if there is not any operation in 15 minutes and come into dormant
mode. The beeper will sound five times within 1 minute before power-off.
To restart the meter, please press any function key or turn to the function swift.
(2) Press “SELECT” key and turn on the meter or press “SELECT” when the meter is in dormant mode
to wake up the meter and the “AUTO POWER OFF” function will be cancelled
5. MAINTENANCE
The meter is a precise instrument. Random changes to the circuit should be avoided.
Note:
1. DO NOT put voltage higher than 600V rms.
2. DO NOT apply voltage in current, resistance, diode and continuity range.
3. DO NOT make any measurements when the battery isn’t installed or the back cover isn’t fixed.
4. Before replacing battery or fuse, please remove the test leads from the measuring point and turn off
the power.
5. Keep the meter away from water, dust and shock.
6. DO NOT use the meter under high temperature, high humidity, combustible, explosive and strong
magnetic environments.
7. Clean the case with a damp cloth and mild detergent only. DO NOT use abrasives or alcohol or any
other strong solvents to clean the meter.
8. To avoid leakage damage, please take out the battery if the meter will not be used for a long time.
9. When “ ” symbol is displayed, you should replace the battery according to the following steps:
a) Unscrew the screw on the battery door and remove the cover.
b) Replace the old battery with a new one. (For long life, it's better to use alkaline battery.)
c) Replace the battery door and tighten the screw.
6. TROUBLE SHOOTING
If the meter does not work properly, please check the meter as following steps:
(If the problems still cannot be solved, please refer to repairing center or contact the local dealers.)
Fault
Solution
No reading on LCD
Turn on the power
Replace battery
Release the HOLD key
signal appears
Replace battery
No current or temperature input
Replace fuse
Big error Value
Replace battery
■The specifications are subject to changes without prior notice.
■The content of this manual is regarded as correct. If users find out any mistakes or omissions,
please kindly contact the manufacturer.
■The manufacturer will not be responsible for accidents and damage caused by improper
operations.
■The functions described in this User Manual shall not be considered as the reason for any
special usages.
EN-2016_V0.1

Other Altronics Measuring Instrument manuals

Altronics PerformAIRE PRO User manual

Altronics

Altronics PerformAIRE PRO User manual

Altronics Redback A 2067 User manual

Altronics

Altronics Redback A 2067 User manual

Altronics Air Quality Computer User manual

Altronics

Altronics Air Quality Computer User manual

Altronics PerformAire Eclipse User manual

Altronics

Altronics PerformAire Eclipse User manual

Altronics Redback Q 2001 User manual

Altronics

Altronics Redback Q 2001 User manual

Altronics PH-01 User manual

Altronics

Altronics PH-01 User manual

Altronics Q1270B User manual

Altronics

Altronics Q1270B User manual

Popular Measuring Instrument manuals by other brands

Hioki CT6500 instruction manual

Hioki

Hioki CT6500 instruction manual

TSI Incorporated Q-TRAK 7575 Operation and service manual

TSI Incorporated

TSI Incorporated Q-TRAK 7575 Operation and service manual

Omega RH820 instruction sheet

Omega

Omega RH820 instruction sheet

PRESIDIUM PEGG User handbook

PRESIDIUM

PRESIDIUM PEGG User handbook

Signal SOLAR 4 Series operating manual

Signal

Signal SOLAR 4 Series operating manual

General Homeline THE-1 instruction manual

General Homeline

General Homeline THE-1 instruction manual

David White AL8-22 instruction manual

David White

David White AL8-22 instruction manual

THORLABS upLED Operation manual

THORLABS

THORLABS upLED Operation manual

NT-MDT NTEGRA Spectra instruction manual

NT-MDT

NT-MDT NTEGRA Spectra instruction manual

PCB Piezotronics 356A71 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 356A71 Installation and operating manual

Elcontrol STAR 3 user manual

Elcontrol

Elcontrol STAR 3 user manual

IME Conto D4-Pt MID manual

IME

IME Conto D4-Pt MID manual

Schäfer RMS 442 quick start guide

Schäfer

Schäfer RMS 442 quick start guide

Namiba Terra Biostat 1000 instructions

Namiba Terra

Namiba Terra Biostat 1000 instructions

Teledyne Analytical Instruments GFC 7002T Operation manual

Teledyne Analytical Instruments

Teledyne Analytical Instruments GFC 7002T Operation manual

Endress+Hauser Proline Promass I 300 PROFINET operating instructions

Endress+Hauser

Endress+Hauser Proline Promass I 300 PROFINET operating instructions

SEFRAM 3515F user manual

SEFRAM

SEFRAM 3515F user manual

Rosemount 8732 Series Quick installation guide

Rosemount

Rosemount 8732 Series Quick installation guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.