amc VIBRO AV CONDITIONER 1000 User manual

USER MANUAL
2018
AV CONDITIONER 1000
amc VIBRO Sp. z o.o.
Pilotow 2e
31-462 Krakow, Poland
T: +48 (12) 362 97 60
S: + 48 (12) 362 97 63
info@amcvibro.pl
KRS No.: 0000618618
REGON No.: 364497010
VAT No.: 6772403385
www.amcvibro.pl
SINGLE CHANNEL CONDITIONER
ICP/IEPE

2
AV CONDITIONER 1000 / ©AMC VIBRO Sp. z o.o.
Table of contents
1. Introduction ...........................................................................................................................................................................3
2. Technical data .......................................................................................................................................................................3
3. Description of the AVC 1000...........................................................................................................................................4
4. Power supply .........................................................................................................................................................................4
4.1. Installing / replacing batteries..............................................................................................................................5
4.2. Power supply...............................................................................................................................................................5
5. Using the conditioner ........................................................................................................................................................6
6. Troubleshooting...................................................................................................................................................................6
7. Recycle .....................................................................................................................................................................................7
7.1. Hazardous Materials.................................................................................................................................................7
7.2. Recycling Facilities.....................................................................................................................................................7

3
AV CONDITIONER 1000 / ©AMC VIBRO Sp. z o.o.
1. Introduction
AV CONDITIONER 1000 (AVC 1000) is a universal conditioner designed to interface with sensors made
in the ICP/IEPE standard.
Thanks to its small size and weight, robust design, very good operational parameters and ease of use, it
is ideally suited for carrying out measurements both in the laboratory and in the field.
The conditioner is equipped with input circuits enabling powering up the ICP/IEPE sensor, while
maintaining wide frequency response, very good linearity and low noise. Is also equipped with an LED
indicating open or short circuit in the sensor loop.
The device can be powered up either by an adapter or an internal battery pack 4xAAA.
2. Technical data
Parameters
number of channels
1
input type
ICP/IEPE 24V DC / 2,4 mA
input / output connector
BNC / BNC
input impedance
100 kΩ
output impedance
120 Ω
gain
1x (0 dB)
gain error
< 0,1%
gain drift
< 5 ppm /°C
output range
±10V
offset
< 10 mV
SNR
> 90 dB
frequency range (- 3 dB)
0,16 Hz –100 kHz *
power supply
9V DC/100 mA
batteries
4 x AAA
dimensions
60 x 100 x 30 mm
weight
225 g
operating temperature
0°C - 60°C
storage temperature
-10°C – 70 °C (without batteries)
*) optional 500 kHz

4
AV CONDITIONER 1000 / ©AMC VIBRO Sp. z o.o.
3. Description of the AVC 1000
Fig. 3.1 shows a plan view of the AVC 1000 conditioner with marked important elements. They are
described in table below.
Item
Description
1
BNC connectors used to connect the sensor (IN) and measuring equipment (OUT)
2
Power switch, switching to the ON position activates the device.
3
Red LED Sensor Error, continuous light indicates open or short-circuit
4
Red LED Power; continuous light indicates operation
5
Red LED Low Batt; continuous light indicates the need to replace batteries
6
Connector of AC adapter 9V DC/100mA
4. Power supply
The AVC 1000 conditioner can be powered by AC adapter supplying 9V DC / 100 mA or internal AAA
batteries. When the external power supply is used, the internal battery is automatically disconnected to
prevent the discharge.
Figure 3.1 View of the AVC 1000 conditioner

5
AV CONDITIONER 1000 / ©AMC VIBRO Sp. z o.o.
4.1. Installing / replacing batteries
In order to install or replace the batteries, remove the front panel of the casing after removing two
screws marked with the number 1 on Fig. 4.1, so that container for four AAA cells can be reached (Fig.
4.2). After putting new AAA cells into the container, insert it into the housing, and mount the front panel
back on its place with the screws.
The need to replace the cells is indicated by the ‘Low Batt’ LED.
WARNING!
Do not leave discharged batteries in the device, as it may cause a leak, resulting in damaging of
the conditioner.
WARNING!
If the conditioner is not used for a long time, remove the batteries to avoid leakage and damage.
4.2. Power supply
The unit can be powered with external power supply with 9V DC output voltage and minimal current
load of 100 mA. AC adapter is available as an option. It is recommended to use the AC transformer with
linear stabilization. Using a power supply with pulse stabilization can drastically increase the noise level
at the output of the conditioner.
WARNING!
Pay attention to the polarity of the adapter plug –it should be in line with the outlet polarity
marked on the top of the cabinet. Connecting the power supply incorrectly may damage the
conditioner.
Figure 4.1 View of the screws allowing access to
the battery compartment.
Figure 4.2 View of the removed battery container

6
AV CONDITIONER 1000 / ©AMC VIBRO Sp. z o.o.
5. Using the conditioner
Operation is reduced to few steps:
»Connect an ICP/IEPE standard sensor with the IN connector using a shielded cable, terminated
with BNC plug
»Connect the conditioner output to measuring equipment (oscilloscope, data acquisition card,
etc.)
»Connect an external power supply - necessary if the batteries haven’t been installed in the device
»Turn on the conditioner - switch should be placed to the ON position
When the conditioner is turned on, the red LED labeled Power will lit, which means that the device is
ready for use.
6. Troubleshooting
The table contains the solution to some problems that the user may encounter during the operation of
the conditioner.
Problem
Solution
The unit will not turn on
Check if the switch is set to the ON position.
If the conditioner is battery powered, make sure that the batteries are installed in
the device and that they are in working order.
If an external power supply is used, make sure that the power adapter is connected
correctly (with compatible polarity), in case of a regulated power supply make sure
that it is set to supply 9V, check that the power supply is turned on.
No signal on
conditioner’s output
Check if the ‘Sensor Error’ LED shines, if yes - check connection of the sensor to the
conditioner and make sure that the sensor cable is OK - any breaks, cuts and
abrasions qualify it to be replaced. Make sure that the sensor is operational.
Output signal
conditioner is noisy
and/or disturbed
When an external power supply is used, make sure that it is a transformer with
linear stabilization. Using a power supply with pulse stabilization (inverter) may
increase the noise level.
Make sure the signal cable used to connect the sensor and the measuring
equipment are shielded cables.
Make sure that nearby there are no electrical devices turned on, that may interfere
with the operation of the conditioner.
Disconnect the AC adapter from the conditioner and power it up with batteries - in
some cases, excessive noise can penetrate from the network through the AC
adapter and affect the operation of the conditioner
If the above table does not include the problem or given solution has been unsuccessful, contact AMC
VIBRO service.

7
AV CONDITIONER 1000 / ©AMC VIBRO Sp. z o.o.
7. Recycle
7.1. Hazardous Materials
AVC 1000 system does not use any hazardous materials outlined by RoHS. These
regulations confirm that lead, mercury, cadmium, hexavalent chromium,
polybrominated biphenyls, polybrominated diphenyl ether, or other battery related
materials are limited to no more than trace amounts.
7.2. Recycling Facilities
When decommissioning out of use devices, minimize the impact of the waste created. Refer to local
waste removal administration for current information on proper material collection and recycling.
This manual suits for next models
1
Table of contents
Other amc VIBRO Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Flybox
Flybox Omnia57 MAP Installation and user manual

Konica Minolta
Konica Minolta T-10A instruction manual

Metrohm
Metrohm 940 Professional IC Vario ONE/ChS/LPG Manual - Short Instructions

Nikon
Nikon COOLSHOT LITE instruction manual

Xylem
Xylem FLYGT Concertor APP 411 Installation, operation and maintenance manual

RadiantLife
RadiantLife LTs user guide