AMEC MR410 User manual

AC/DC CURRENT PROBE MR410
MR520
ENGLISH User Manual
ON
OFF
1500A
600V CAT III - 1500A
!
OUTPUT:
1500 A: 1 mV/A
AC/DC CURRENT PROBE
Model MR 520
®
INSTRUMENTS
ON
OFF
600A
600V CAT III - 600A
!
OUTPUT:
600A: 1 mV/A
AC/DC CURRENT PROBE
Model MR 410
®
INSTRUMENTS
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
certifies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be
requested at the time of purchase, or obtained
by returning the instrument to our repair and
calibration facility, for a nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer.
Serial #: ______________________________
Catalog #: 1200.70/1200.71
Model #: MR410 / MR520
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.
d.b.a AEMC®Instruments
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents
1. INTRODUCTION ........................................................................ 3
1.1 International Electrical Symbols.....................................................3
1.2 DenitionofMeasurementCategories...........................................4
1.3 ReceivingYourShipment...............................................................4
1.4 OrderingInformation ......................................................................4
1.4.1 AccessoriesandReplacementParts ................................4
2. PRODUCT FEATURES ............................................................... 5
2.1 Description .....................................................................................5
2.2 MR410Features ............................................................................6
2.3 MR520Features ............................................................................7
2.4 DMM/DisplayCompatibility ..........................................................8
3. SPECIFICATIONS ...................................................................... 9
3.1 MR410Specications ....................................................................9
3.1.1 Electrical............................................................................9
3.1.2 Mechanical ........................................................................9
3.2 MR520Specications ....................................................................9
3.2.1 Electrical............................................................................9
3.2.2 Mechanical ......................................................................10
3.3 CommonSpecications(MR410&MR520) ................................10
3.3.1 Electrical..........................................................................10
3.3.2 Mechanical ...................................................................... 11
3.3.3 Safety .............................................................................. 11
4. OPERATION............................................................................. 12
4.1 MakingMeasurements.................................................................12
4.1.1 DCMeasurement ............................................................12
4.1.2 ACMeasurement ............................................................13
4.1.3 IndicatorLights:GreenLED ............................................13
4.2 TipsForGettingtheBestAccuracy..............................................13
4.3 OperationExamples ....................................................................14
4.4 TypicalResponseCurves ............................................................17
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

5. MAINTENANCE ....................................................................... 18
5.1 Maintenance ................................................................................18
5.1.1 BatteryReplacement.......................................................18
5.1.2 Cleaning ..........................................................................18
Limited Warranty .........................................................................................20
WarrantyRepairs.........................................................................................20
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

AC/DCCurrentProbeModelsMR410andMR520
3
CHAPTER 1
INTRODUCTION
Warning
Thesesafetywarningsareprovidedtoensurethesafetyofpersonnel
andproperoperationoftheinstrument.
• Readthisinstructionmanualcompletelyandfollowallthesafetyinfor-
mationbeforeattemptingtouseorservicethisinstrument.
• Usecautiononanycircuit:Potentiallyhighvoltagesandcurrentsmay
bepresentandmayposeashockhazard.
• ReadthesafetyspecicationssectionbeforeusingtheModelMR410
or MR520. Never exceed the maximum working voltage ratings
given.
• Safetyistheresponsibilityoftheoperator!
• Neveropenthebackoftheinstrumentwhileconnectedtoanycircuit
orinput.
• ALWAYSconnecttheModelMR410orMR520tothedisplaydevice
beforeclampingtheprobeontothesampleundertest.
• Always inspect the instrument and lead prior to use. Replace any
defectivepartsimmediately.
• NEVER use the Model MR410 or MR520 on electrical conductors
ratedabove600V.
1.1 International Electrical Symbols
Thissymbolsigniesthattheinstrumentisprotectedbydoubleorrein-
forcedinsulation.Useonlyspeciedreplacementpartswhenservicing
theinstrument.
This symbol on the instrument indicates a WARNING and that the
operatormustrefertotheusermanualforinstructionsbeforeoperat-
ingtheinstrument.Inthismanual,thesymbolprecedinginstructions
indicatesthatiftheinstructionsarenotfollowed,bodilyinjury,installa-
tion/sampleandproductdamagemayresult.
Riskofelectricshock.Thevoltageatthepartsmarkedwiththissymbol
maybedangerous.
ThisisatypeAcurrentsensor.Thissymbolsigniesthatapplication
aroundandremovalfromHAZARDOUSLIVEconductorsispermitted.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

4
AC/DCCurrentProbeModelsMR410andMR520
1.2 Definition of Measurement Categories
CAT II: For measurements performed on circuits directly connected to
the electrical distribution system. Examples are measurements
onhouseholdappliancesorportabletools.
CAT III: For measurements performed in the building installation at
the distribution level such as on hardwired equipment in xed
installationandcircuitbreakers.
CAT IV: For measurements performed at the primary electrical supply
(<1000V) such as on primary overcurrent protection devices,
ripplecontrolunits,ormeters.
1.3 Receiving Your Shipment
Uponreceivingyourshipment,makesurethatthecontentsareconsistent
withthepackinglist.Notifyyourdistributorofanymissingitems.Iftheequip-
mentappearstobedamaged,leaclaimimmediatelywiththecarrierand
notifyyourdistributoratonce,givingadetaileddescriptionofanydamage.
Savethedamagedpackingcontainertosubstantiateyourclaim.
1.4 Ordering Information
AC/DCCurrentProbeModelMR410
(Lead-1mV/A-600Amax) ................................................. Cat. #1200.70
AC/DCCurrentProbeModelMR520
(Lead-1mV/A-1500Amax) ............................................... Cat. #1200.71
Both models include a 9V battery and a user manual.
1.4.1 Accessories and Replacement Parts
Bananaplugadapter(tonon-recessedplug) ...................... Cat. #1017.45
Banana(female)-BNC(male)adapter............................... Cat. #2118.46
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

AC/DCCurrentProbeModelsMR410andMR520
5
CHAPTER 2
PRODUCT FEATURES
2.1 Description
The MR probes are designed to the meet the latest safety and perfor-
mancestandards.Twodierenthook-shapedjawsareoered,bothper-
mitting the user to “pry” into or “hook” onto cables (will accept 2 x 500
MCM)orsmallbusbars.
The MR series uses Hall eect technology. The electronics and batter-
iesareself-containedinthehandles.TheoutputoftheAC/DCprobesis
1mV/A.
WhentakingDCmeasurements,amechanicalDCzeroallowstheuserto
adjustforanyDCoset.Thereisnooutputltering-TrueRMSwithDC
componentsispossible.
Thephaseshiftisexcellent,makingtheMRserieswellsuitedforpower
andpowerqualityapplications.
ModelMR410measures400AAC(600Apeak),600ADC.
ModelMR520measures1000AAC(1500Apeak),1500ADC.
BoththeMR410andMR520haveproportionalmVoutputfordirectread-
ingsonmultimeters,recorders,loggersandotherinstrumentsaccepting
bananaplugs.
AnoptionalbananaplugtoBNCadaptorisavailableasanaccessory.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

6
AC/DCCurrentProbeModelsMR410andMR520
2.2 MR410 Features
ON
OFF
600A
600V CAT III - 600A
!
OUTPUT:
600A: 1 mV/A
AC/DC CURRENT PROBE
Model MR 410
®
INSTRUMENTS
1
2
3
4
5
6
7
8
Ø = 1.18" (30mm)
Cable max.
Ø = 0.95" (24mm)
Dual Cable max.
Figure 1
1. Jaws
2. Conductor
3. Protectivenon-slipguard
4. Greenlight(onwhenbattery
voltage≥6.5V)
5. Two-positionrangeselection
switch
6. Zeroadjustmentknob
7. Lead,5ft(1.5m)
8. SafetybananaplugØ4mm
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

AC/DCCurrentProbeModelsMR410andMR520
7
2.3 MR520 Features
ON
OFF
1500A
600V CAT III - 1500A
!
OUTPUT:
1500 A: 1 mV/A
AC/DC CURRENT PROBE
Model MR 520
®
INSTRUMENTS
1
2
3
4
5
6
7
8
Ø = 1.5" (39mm)
Cable max.
1.96 x 0.19" (50 x 5mm)
Dual Bus Bar max.
Ø = 0.98" (25mm)
Dual Cable max.
Figure 2
1. Jaws
2. Conductor
3. Protectivenon-slipguard
4. Greenlight(onwhenbattery
voltage≥6.5V)
5. Two-positionrangeselection
switch
6. Zeroadjustmentknob
7. Lead,5ft(1.5m)
8. SafetybananaplugØ4mm
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

8
AC/DCCurrentProbeModelsMR410andMR520
2.4 DMM / Display Compatibility
TheModelsMR410andMR520currentprobesarecompatiblewithany
multimetervoltmeterorothervoltagemeasuringinstrumentwhichhasthe
followingfeatures:
• Inputjackthataccepts4mmsafetybananaplug
• Rangeandresolutioncapableofdisplaying1mVofoutputper
ampofmeasuredcurrent
• Voltmeteraccuracyof0.3%orbettertotakefulladvantageofthe
probeaccuracy
• Inputimpedanceof1MΩ/100pForgreater
Whentheprobeismakingameasurement,thecurrent-carryingconductor
isnotbrokenandremainselectricallyisolatedfromtheprobeoutput.
Asaresult,theprobeoutputcommonmaybeeitheroated(isolated)or
grounded.
Warning: User Safety
Always use a DMM, voltmeter or other displaying device, appropriately
ratedforsafety.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

AC/DCCurrentProbeModelsMR410andMR520
9
CHAPTER 3
SPECIFICATIONS
3.1 MR410 Specifications
3.1.1 Electrical
Measurement Range:1to400AAC(600APeak);1to600ADC
Accuracy*: 1to100A:1.5%reading±1A;100to400A:2.0%reading;
400to600ADConly:2.5%reading
Phase Shift: 45to65Hz;10to200A:≤2.5°
200to400A:≤2°
Overload: 1500ADCand1000AACcontinuousupto1kHz
3.1.2 Mechanical
Jaw Opening: 1.2"(31mm)
Maximum Cable Diameter: One1.18"(30mm)ortwo0.95”(24mm)or
twobusbars1.2x0.4"(31.5x10mm)
Dimensions (without zero knob):8.8x3.82x1.73"(224x97x44mm)
Weight: 15oz(440g)
3.2 MR520 Specifications
3.2.1 Electrical
Measurement Range:1to1000AAC(1500APeak);1to1500ADC
Accuracy*:1to100A:1.5%reading±1A
100to800A:2.5%reading
800to1000A:4%reading
1000to1400ADConly:4%reading
Phase Shift: 45to65Hz
10to200A:≤2.5°;200to1000A:≤1.5°
Overload: 1500AACand1000ADCcontinuousupto1kHz
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

10
AC/DCCurrentProbeModelsMR410andMR520
3.2.2 Mechanical
Jaw Opening: 1.55"(40mm)
Maximum Cable Diameter:
One1.5"(39mm)ortwo.98"(25.4mm)or
twobusbars1.96x.19”(50x5mm)oronebusbar1.96x.49"(50x12.5mm)
Dimensions (without zero knob):9.31x3.82x1.73"(236.5x97x44mm)
Weight:1.06lbs(480g)
*Reference conditions: 18° to 28°C , 20 to 75% RH, external magnetic eld <40 A/m, no DC component,
no external current carrying conductor, test sample centered, 1MΩ≤100pF load, zero adjustment prior to
measurement [DC only] DC to 65Hz, battery voltage 9V ± 0.1V.
3.3 Common Specifications (MR410 & MR520)
3.3.1 Electrical
Output Signal: 1mV/A
Frequency Range:DCto10kHz@-3dB
Noise: DCto1kHz:≤1mVpeaktopeak
DCto5kHz:≤1.5mVpeaktopeak
0.1Hzto5kHz:≤500µVpeaktopeak
Rise and Fall Time: Rise:≤100µsfrom10to90%Vout
Fall:≤120µsfrom10to90%Vout
Load Impedance:>100kΩ/100pF
Insertion Impedance: 0.39mΩ@50Hz,58mΩ@1000Hz
Working Voltage: 600Vrms
Common Mode Voltage: 600Vrms
Inuence of Adjacent Conductor:
<10mA/Aat50Hzat23mmfromtheprobe
Inuence of Conductor in Jaw Opening:0.5%reading(DCto440Hz)
Power Source: 9Valkaline(NEDA1604A,IEC6LR61)
Battery Life: Approximately100Htypical
Low Battery: GreenLEDonwhenbatteryvoltage≤6.5V
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

AC/DCCurrentProbeModelsMR410andMR520
11
3.3.2 Mechanical
Output: Doubleinsulated5ft(1.5m)leadwithsafetybananaplugs
Zero Adjustment: ±10Aapprox.by10turnknobatbaseofcase
Case Protection: IP30perIEC529
Drop Test: 1.0mon38mmofoakonconcrete;testaccordingtoEN61010
Mechanical Shock: 100G,testperIEC68-2-27
Vibration: TestperIEC68-2-6
Frequency Range: 5to15Hz,Amplitude:1.5mm
15to25Hz,Amplitude:1mm
25to55Hz,Amplitude:0.25mm
Handle: UL94V0
Jaws: UL94V0
3.3.3 Environmental
Operating Temperature Range: 14°to131°F(-10°to+55°C)
Storage Temperature Range: -40°to176°F(-40°to+80°C)
Temperature Inuence: ≤300°ppm/°Kor1%/10°K
≤0.2A/°KonZero
Operating Relative Humidity:
10°to35°C:90±5%RH(withoutcondensation)
40°to55°C:70±5%RH(withoutcondensation)
Humidity Inuence: 10to90%RH@referencetemperature≤0.1%R
Altitude: Operating:0to2000m
Non-operating:0to12,000m
3.3.3 Safety
Electrical:
EN61010-2-32,600VCATIII,Pollution:2
Electromagnetic Compatibility:
EN50081-1ClassB
EN50082-2ElectrostaticdischargeIEC1000-4-2
RadiatedeldIEC1000-4-3
FasttransientsIEC1000-4-4
Magneticeldat50/60HzIEC1000-4-8
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

12
AC/DCCurrentProbeModelsMR410andMR520
CHAPTER 4
OPERATION
4.1 Making Measurements
• Plugtheprobeintothedisplaydevice(e.g.,DMM,logger).
Notethepolarityoftheprobeoutputbananaplugs(red=positive[+],
black=negative[-]).
• Selecttheappropriaterangeonthedisplaydevice.
Notethattheprobe’soutputis1mV/A.
• Turndisplaydevicepoweron.
• Turn on the instrument.The Green LED should be on. If the Green
LEDdoesnotcomeonorgoesoafterafewminutes,itisnecessary
toreplacethebattery(see“BatteryReplacement”§5.1.1).
4.1.1 DC Measurement
• SelecttheappropriateDCmVorVrangeontheDMMormeasuring
instrument.
• “Zero”theprobe-withtheprobedisconnectedfromtestsamples(no
conductorinprobejawwindow),pushinandturnthezeroadjustment
knobtoobtainareadingbetterthan0.5mVontheDMM.
• Forbestaccuracy,especiallyonlow-levelmeasurement,itisrecom-
mendedthatyouzerotheprobebeforeeachmeasurement.
• Clamptheprobearoundtheconductortobetested.TheDMMordis-
play device should now display the measured conductor current. A
positivereadingindicatescurrentowinginthedirectionofthearrow
locatedonthesideofthejaw.Anegativereadingindicatescurrentow
intheoppositedirectionofthearrow.
• Apply the conversion ratio (1mV/A) to get the value of the current.
(e.g.If themultimeter reading is 250mVDC, the measured current is
250ADC).
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

AC/DCCurrentProbeModelsMR410andMR520
13
4.1.2 AC Measurement
• SelecttheappropriateACmVorVrangeontheDMMormeasuring
instrument.
• TheDCzeroadjustmentisnotrequiredwhenmeasuringACcurrent
withavoltmeterthatisACcoupled.TheDCzeroadjustmentis
requiredifyouareusingavoltmeterthatisDCcoupled.
• SelecttheappropriatemVAC or VACrangeontheDMMormeasuring
instrument.
• Clamptheprobearoundtheconductortobetested.Thedisplaydevice
shouldnowdisplaythemeasuredconductorcurrent.
• Applytheconversionratio(1mV/A)togetthevalueofthecurrent.
4.1.3 Indicator Lights: Green LED
• TheGreenLEDindicatesthattheprobeisonandthatthebatteryis
good.
• Itwillnotlightunderlowbatteryconditions.
• Replacethe9Vbatteryifitisnotlit.
4.2 Tips For Getting the Best Accuracy
TheModelsMR410 and MR520 arecapableof measuring DCandfre-
quencycurrentsoverawiderange.Herearesomekeyconsiderationsfor
gettingthemostaccuracyfromyourdisplayinstrument:
• WhenusingtheModelsMR410andMR520withaDMMorother
meter,itisimportanttoselecttherangethatprovidesthebest
resolution.
• Makesurethatprobejawmatingsurfacesarefreeofdustand
contamination.
• Bewareofshort-circuitcurrents.Largein-rushDCcurrents
(whichcanoccurwhenpowerisrstappliedinacircuit)and
largehigh-currenttransientsmaycausevaryingdegreesof
residualreadings.
• Ifindoubtofaparticularreading,removetheprobefromthe
conductorundertestandchecktoseethatthedisplaydevice
returnstozero.Ifnot,itwillbenecessarytorezerotheprobe.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

14
AC/DCCurrentProbeModelsMR410andMR520
4.3 Operation Examples
DC Current Measurement Example for MR410
• Conductorcarrying25ADCinthedirectionofthearrow
• DMMplacedinDCvoltsmode
• DMMdisplays25.0mVDC=25ADC
ACV DCV
25.0 mV
COM +
25 A
DC
ON
OFF
600A
600V CAT III - 600A
!
OUTPUT:
600A: 1 mV/A
AC/DC CURRENT PROBE
Model MR 410
®
INSTRUMENTS
Figure 3
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

AC/DCCurrentProbeModelsMR410andMR520
15
Measuring the DC Component of an (AC+DC) Waveform for MR410
• Conductorcarrying300ADC+25.0AAC
• DMMplacedinDCvoltsmode
• DMMdisplays300.0mVDC=300ADC
ACV DCV
300.0
mV
COM +
300 A
DC
ON
OFF
600A
600V CAT III - 600A
!
OUTPUT:
600A: 1 mV/A
AC/DC CURRENT PROBE
Model MR 410
®
INSTRUMENTS
Figure 4
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

16
AC/DCCurrentProbeModelsMR410andMR520
Measuring the AC Component of an (AC+DC) Waveform for MR520
• Conductorcarrying20ADC+250AAC
• VoltmeterplacedinACvoltsmode
• Voltmeterdisplays250.0mVAC=250AAC
ACV DCV
250.0
mV
COM +
250 A
DC
ON
OFF
1500A
600V CAT III - 1500A
!
OUTPUT:
1500A: 1 mV/A
AC/DC CURRENT PROBE
Model MR 520
®
INSTRUMENTS
Figure 5
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

AC/DCCurrentProbeModelsMR410andMR520
17
4.4 Typical Response Curves
Model MR410
110100 1000
2
0
-2
-4
-6
-8
-10
-12
-14
-16
-18
Primary current in amperes
Typical Accuracy in DC
Error in %
Model MR520
Typical Accuracy in DC
1
2
0
-2
-4
-6
-8
10 1001000 10,000
Primary current in amperes
Accuracy in %
-10
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

18
AC/DCCurrentProbeModelsMR410andMR520
CHAPTER 5
MAINTENANCE
5.1 Maintenance
WARNING:
• Formaintenanceuseonlyspeciedreplacementparts.
• Toavoidelectricalshock,donotattempttoperformanyservicing
unlessyouarequaliedtodoso.
• Toavoidelectricalshockand/ordamagetotheinstrument,donot
getwaterorotherforeignagentsintothecase.Turnthecurrent
probe OFF and disconnect the unit from all the circuits before
openingthecase.
5.1.1 Battery Replacement
Whentheprobeisturnedon,thegreenLEDshouldlightup.Ifitdoesnot,
replacethe9Vbattery.
• Completelydisconnecttheprobefromthecircuitundertest,and
fromtheDMMormeasuringinstrument.
• TurntheprobeOFF.
• Unscrewthebatterycompartmentscrewandremovecover.
• Replacethebatteryandputthecoverbackon.
Donotreplacethebatterywhiletheprobeisinuse.
5.1.2 Cleaning
• Cleanthe bodyof theclamp witha clothlightly moistenedwith
soapywater.
• Wipecleanwithaclothmoistenedwithcleanwateranddry.
• Donotusesolvent.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
This manual suits for next models
1
Table of contents