pH as measured through the probe with the slide switch in the “read” position. The
pH LED is off when the display is indicating either the high or low controller
setpoints.
CENTER (Controller Setpoint Block)
There are two potentiometers that are used to adjust the controller
setpoints. They control the RANGE (Fig. F) and CENTER VALUE (Fig. G) of the
controller, respectively. Detailed information regarding their adjustment is given
later in the ADJUSTING THE CONTROLLER section.
LOWER RIGHT (Display Mode)
The position of the SLIDE SWITCH (Fig. H) determines what the LED
display will show. At the center position, the display indicates the pH that the
probe is measuring, and the red pH LED (Fig. E) at the lower left of the instrument
will be illuminated. Sliding the switch up or down will cause the display to show
the high and low controller setpoints, respectively.
UPPER RIGHT (Controller Status LED’s)
There are a total of 4 LED’s here. The upper row (Fig. I) shows the status
of the 110 VAC outlets, which you will find, on the back of the controller beneath
the LCD pH display. The lower LED lights (Fig. J) are activated when the slide
switch is in either the HIGH or LOW position.
110 VAC OUTLETS
There are two outlets, which will be energized when the pH is above
(outlet #1) or below (outlet #2) the controller setpoints. A common application for
the controller will be to control the addition of CO2 to the system. Since the
addition of carbon dioxide lowers the pH, a NORMALLY CLOSED solenoid valve
controlling the flow of CO2 would be plugged into the HIGH control outlet (Outlet
#1). If a basic substance is being dispensed into the system (i.e. a lime water
reactor for a reef aquarium) the device(s) controlling the flow would be attached to
the LOW control outlet # 2.
NORMALLY CLOSED means that the valve will be closed when NOT energized.
A pH electrode is supplied with the controller. A plastic fluid-filled cap
will protect the end of the electrode. The fluid inside the cap is a mixture of #4.0
calibration fluid + 3M KCL. This is commonly known as storage fluid. There is a
suction cup attached to the electrode, which may be used to secure it, or a user-
devised scheme of holding the electrode may be employed.
IV CALIBRATING THE pH METER SUBSYSTEM
This is a two-point calibration instrument therefore TWO different
calibrations must always be used. A pH 7.00 fluid must be used to set the CALIB
and a pH 10.00 or 4.00 fluid is used to set the slope.
COMPLETE INSTRUCTIONS:
Before you begin the calibration, you should disconnect the pumps or
valves that you are controlling, because the pH electrode will be put into solutions
with a pH that is probably far outside the controller limits you have established.
Before you begin, make sure that you have the pH calibration fluids, and
some room temperature tap water to rinse the electrode between calibration
solutions. You must have a #7.00 calibration solution, and another solution with
either be higher or lower pH (typically 4 or 10). If you are doing this for the first
time, read through these instructions once to determine what is required before
you begin.
1.Set the display mode switch (far right) to the central position (read). The
pH LED (lower left) should be illuminated.
2.Remove the plastic cap protecting the end of the electrode if you are
doing this for the first time. If the electrode has already been in service you should
note the condition of the electrode and clean it if required.
3.Rinse the electrode with room temperature tap water taking cares to
remove any accumulated salt. Rinsing the electrode prevents you from carrying
over contaminating substances into the calibration solutions. Gently shake the
electrode to remove any clinging drops of water.
4.Immerse the tip (bottom 1 inch) of the electrode into the pH 7 calibration
fluid. It is important to use the pH 7 solution first. Be sure that the tip of the probe
is fully immersed in the calibration fluid for a stable and reliable result.
5. Keeping the glass bulb immersed, gently swirl the ends if the electrode
in the calibration fluid until the pH reading stabilizes. If the electrode is in good
condition, the reading should stabilize in a few seconds. If the electrode does not