5. Function Mode
The function mode stores parameters to control the weighing module.
The parameters are stored even without power supplied.
5.1.
Summary
............ Calibration function
............ Linearity adjustment function
............ Basics function
............ Hold function
............ Flow rate function
............ Control I/O function
............ Current loop output function
............ Analog 4-20mA output function
※
When only a numerical value is displayed, the decimal point flashes to distinguish
that from a weighing value.
5.2.
Operations and Types
5.2.1. Select Modes Under Function Mode
+
.....Proceeds to function mode from the weighing mode.
........... Selects a type of select mode (3 upper figures).
....... Enters a selected mode.
........... Selects an item under the selected mode (2 lower figures).
....... Enters the item.
....... Stores parameters and returns to weighing mode.
5.2.2. Change Value
........... Moves selected digit.
........... Changes numerical value.
....... Activates (store) value and returns to select mode.
....... Deactivates value and returns to select mode.
5.3.
Function Table
※
1Decimal point depends on .
※
2When "span calibration using mass ( )" is performed by the input voltage from the
load cell in the calibration mode, "input voltage at zero ( )", "span calibration input
voltage ( )" and "weight against span calibration input voltage( )" change.
※
3If pressing the key while displaying or , the current weighing value
can be monitored. Press the key again to return function mode.
5.3.1. Calibration Function ( )
Item & Function Descriptions, Range & Factory settings
Unit
0 : No used 1: g 2 : kg 3: t 4: N 5: kN
Decimal point position
Decimal point (D.P.) position.
0 : 0 1 : 0.0 2 : 0.00 3 : 0.000 4 : 0.0000
Minimum division
Minimum division of value. (A scale / digit)
1 : 1 2 : 2 3 : 5 4 : 10 5 : 20 6 : 50
Weighing capacity
Measurement can be displayed up to +8 digits (8 scales) from
capacity.
※
11 to 70000 to 99999
Zero range
The range that the key (zero key) works. Deviation from the
calibrated zero point [%]. 0 to 2 to 100
Zero tracking time
Used with for zero tracking.
Scale: 0.1 sec. 0.0 to 5.0
Zero tracking band
Used with for zero tracking.
Scale: 0.1 digit 0.0 to 9.9
Stability detection time
Used with for stability detection.
Scale: 0.1 sec. 0.0 to 1.0 to 9.9
Stability detection width
Used with for stability detection.
Scale: 1digit 0 to 2 to 100
Tare and zero at unstable
weight value
Tare and zero operation when unstable
0 : Disables both functions. 1 : Enables both functions.
Tare when the gross
weight is negative
Tare when the gross weight is negative.
0 : Disables tare. 1 : Enables tare.
Output when out of
range and unstable
0 : Disables output. 1 : Enables output.
Exceeding negative
gross weight
To judge when the negative gross weight is exceeded.
1 : Gross weight < -99999 3 : Gross weight < -19 digits
2 : Gross weight < -Capacity
Exceeding negative net
weight
To judge when the negative net weight is exceeded.
1 : Net weight < -99999 2 : Net weight < -Capacity
Clear the zero value
Select whether or not to clear the zero value.
0 : Disable 1 : Enable
Zero setting when power
is turned on
Select whether or not to perform zero setting when power is
turned on.
0 : Disable 1 : Enable
Input voltage at zero
Input voltage from a load cell at zero.
※
2
Scale: 0.0001 mV/V -7.0000 to 0.0000 to 7.0000
Span input voltage
Input voltage from a load cell at span.
※
2
Scale: 0.0001 mV/V 0.0100 to 3.2000 to 9.9999
Weight against span
input voltage
The calibration weight value corresponding to the input voltage at
.
※
1
※
21 to 32000 to 99999
Gravity acceleration of
the calibration place
Gravity acceleration of the place where the scale is calibrated.
Scale: 0.0001 m/s29.7500 to 9.8000 to 9.8500
Gravity acceleration of use
place
Gravity acceleration of the place where the scale is being used.
Scale: 0.0001 m/s29.7500 to 9.8000 to 9.8500
Suppression of the hold
function
0 : Permission. 1 : Prohibition.
~
Reserved internally
5.3.2. Linearity Adjustment Function ( )
Item & Function Descriptions, Range & Factory settings
Number of input points
Number of adjustment point. If 0, 1 or 2 is selected, this linearization
is not used. 0 to 5
Linear-zero
Input voltage for linear-zero input.
Scale: 0.0001 mV/V -7.0000 to 0.0000 to 7.0000
Setting value for linear 1
The setting value of weights for linear 1 input.
※
1
0 to 99999
Span at linear 1
The span voltage between linear-zero and linear 1 input.
Scale: 0.0001 mV/V 0.0000 to 9.9999
Setting value for linear 2
The setting value of weights for linear 2 input.
※
1
0 to 99999
Span at linear 2
The span voltage between linear-zero and linear 2 input.
Scale: 0.0001 mV/V 0.0000 to 9.9999
Setting value for linear 3
The setting value of weights for linear 3 input.
※
1
0 to 99999
Span at linear 3
The span voltage between linear-zero and linear 3 input.
Scale: 0.0001 mV/V 0.0000 to 9.9999
Setting value for linear 4
The setting value of weights for linear 4 input.
※
1
0 to 99999
Span at linear 4
The span voltage between linear-zero and linear 4 input.
Scale: 0.0001 mV/V 0.0000 to 9.9999
5.3.3. Basics Function ( )
Item & Function Descriptions, Range & Factory settings
Key switch disable
Each digit corresponds to to a key switch. Only available in weighing
mode.
4 fig. 3 fig. 2 fig. 1 fig. 0 : Permission 1 : Prohibition
0 0 0 0 to 1 1 1 1
key function
0 : None 7: Zero clear
1 : Manual print command 8: Reserved internally
2 : Hold 9: Reserved internally
3 : Operation switch 1 10: Reserved internally
4 : Operation switch 2 11: Reserved internally
5 : Display exchange 12: mV/V monitor
6 : Tare clear 13: Digital filter 2
14: Display output data selected in
15: Display output data selected in
Display refresh rate
1 : 20 times/sec. 2 : 10 times/sec. 3 : 5 times/sec.
display
0 : None
1 : Zero tracking in progress
2 : Alarm
3 : Display operation switch status as on or off
4 : Near-zero
5 : HI output
6 : OK output
7 : LO output
Digital filter 1
Selects a cutoff frequency.
※
3
0 : None 6 : 20.0 Hz 12 : 2.8 Hz
1 :100.0 Hz 7 : 14.0 Hz 13 : 2.0 Hz
2 : 70.0 Hz 8 : 10.0 Hz 14 : 1.4 Hz
3 : 56.0 Hz 9 : 7.0 Hz 15 : 1.0 Hz
4 : 40.0 Hz 10 : 5.6 Hz 16 : 0.7 Hz
5 : 28.0 Hz 11 : 4.0 Hz
Digital Filter 2
Selects a cutoff frequency.
0 : None 6 : 20.0 Hz 12 : 2.8 Hz 18 : 0.40 Hz
1 : 100.0 Hz 7 : 14.0 Hz 13 : 2.0 Hz 19 : 0.28 Hz
2 : 70.0 Hz 8 : 10.0 Hz 14 : 1.4 Hz 20 : 0.20 Hz
3 : 56.0 Hz 9 : 7.0 Hz 15 : 1.0 Hz 21 : 0.14 Hz
4 : 40.0 Hz 10 : 5.6 Hz 16 : 0.7 Hz 22 : 0.10 Hz
5 : 28.0 Hz 11 : 4.0 Hz 17 : 0.56 Hz 23 : 0.07 Hz
Hold function
1 : Normal hold 2 : Peak hold 3 : Averaging hold
5.3.7. Current loop Output Function ( )
Item & Function Descriptions, Range & Factory settings
Serial data
1 : Displayed weighing value 3 : Net 5 : Gross / Net / Tare
2 : Gross 4 : Ta r e
Communication mode
1 : Stream 3 : Manual print
2 : Automatic print
Baud rate
1 : 600 bps 2 : 2400 bps
5.3.8. Analog 4-20mA Output Function ( )
Item & Function Descriptions, Range & Factory settings
Output data
1 : Displayed weighing value (Digital filter 1)
2 : Gross (Digital filter 1)
3 : Net (Digital filter 1)
4 : Displayed weighing value (Digital filter 2)
5 : Gross (Digital filter 2)
6 : Net (Digital filter 2)
7 : Flow rate 1
8 : Flow rate 2
9 : Flow rate 1 or Flow rate 2 (Select in control input)
Mass/flow rate
at 4mA output
Select mass/flow rate by setting output data ( )
Decimal point position linkage:
Mass
Flow rate (setting magnification of flow rate) +
-99999 to 0 to 99999
Mass/flow rate
at 20mA output
Select mass/flow rate by set output data ( )
Decimal point position linkage:
Mass
Flow rate (setting magnification of flow rate) +
-99999 to 70000 to 99999
Flow rate unit
1 : Seconds
2 : Minutes
3 : Hours
IOUT1
Flow rate setting
magnification
(times)
1 : 1
2 : 10
3 : 100
4 : 1000
5 : 10000
Output data
1 : Displayed weighing value (Digital filter 1)
2 : Gross (Digital filter 1)
3 : Net (Digital filter 1)
4 : Displayed weighing value (Digital filter 2)
5 : Gross (Digital filter 2)
6 : Net (Digital filter 2)
7 : Flow rate 1
8 : Flow rate 2
9 : Flow rate 1 or Flow rate 2 (assign to input terminals)
Mass/flow rate
at 4mA output
Select mass/flow rate by set output data ( )
Decimal point position linkage:
Mass
Flow rate (setting magnification of flow rate) +
-99999 to 0 to 99999
Mass/flow rate
at 20mA output
Select mass/flow rate by set output data ( )
Decimal point position linkage:
Mass
Flow rate (setting magnification of flow rate) +
-99999 to 70000 to 99999
Flow rate unit
1 : Seconds
2 : Minutes
3 : Hours
IOUT2
Flow rate setting
magnification
(times)
1 : 1
2 : 10
3 : 100
4 : 1000
5 : 10000
Near
-
zero
※
1
-99999 to 10 to 99999
Comparison mass at
near-zero
1 : Gross weight 2 : Net weight
Upper limit value
※
1
-99999 to 10 to 99999
Lower limit value
※
1
-99999 to -10 to 99999
Comparison mass of
upper and lower limit
1 : Gross weight 2 : Net weight
5.3.4. Hold Function ()
Item & Function Descriptions, Range & Factory settings
Average time
Time to calculate the average. 0.00 is not averaged.
Scale: 0.01 sec. 0.00 to 9.99
Start wait time
Waiting time to commence a holding or averaging.
Scale: 0.01 sec. 0.00 to 9.99
Condition of
automatic start
The condition to commence a holding or averaging.
0 : Not used 2 : Above the near-zero
1 : Above the near-zero, and stable
Release using control
input
Release when control input of the hold terminal is falling.
0 : Do not release 1 : Release
Release time
Release after a set amount of time has passed. 0.00 is not averaged.
Scale: 0.01 sec. 0.00 to 9.99
Release using
fluctuation range
Release when fluctuation from the holding value exceeds a set value.
※
1
0 : Continue 0 to 99999
Release at near-zero
Release when the weighing value is in the near-zero range.
0 : Do not release 1 : Release
5.3.5. Flow Rate Function ( )
Item & Function Descriptions, Range & Factory settings
Filter of flow rate 1
Filter of flow rate 2
1 :Digital filter 1
12 :Digital filter 2
Damping time of flow rate 1
Damping time of flow rate 2
Suppress shaking of flow rate.
The higher value setting, the less shaking.
Scale: 1 sec. 1 to 5 to 1000
5.3.6. Control I/O Function ()
Item & Function Descriptions, Range & Factory settings
IN 1
0 to 7 to 30
IN 2
0 to 8 to 30
IN 3
0 t o 30
IN 4
0 to
30
IN 5
0 to
30
IN
IN 6
0 : Not used 22 : Zero clear
1 to 6 : Reserved internally 23 : Tare clear
7 : Zero 24 :
Operation same as a F key
8 : Ta r e 25 : Prohibit update of flow rate 1
9 : Hold 26 : Prohibit update of flow rate 2
10 : Gross / Net exchange 27 : Initialize flow rate 1
11 : Diagnose 28 : Initialize flow rate 2
12 : Print command 29 : Specify flow rate in
※
13 to 21 : Reserved internally 30 : Specify flow rate in
※
※
Select flow rate 1 or 2 0 t o 30
OUT 1
0 to 18 to 36
OUT 2
0 to 9 to 36
OUT 3
0 t o 36
OUT 4
0 t o 36
OUT 5
0 t o 36
OUT 6
0 t o 36
OUT 7
0 t o 36
OUT 8
0 : Not used
1 to 8 : Reserved internally
9 : Stability
10 : Over capacity 30 : In weighing (
ON
)
11 : Net display 31 : In weighing (
1 Hz
)
12 : During tare 32 : In weighing (
50 Hz
)
13 : Hold 33 : Alarm
14 : Hold busy 34 : Output operation switch is
15 : HI output on or off
16 : OK output 35 : Approximate flow rate value
17 : LO output of flow rate 1
18 : Near-zero 36 : Approximate flow rate value
19 to 29 : Reserved internally of flow rate 2 0 t o 36
OUT 1
OUT 2
OUT 3
OUT 4
OUT 5
OUT 6
OUT 7
OUT
OUT 8
1 : Inverting output 2 : Non inverting output
-