
Page 8 of 45
Neerlandweg 21
www.ankersmidsampling.com
© Reproduction of this document or its
allowed without permission of Ankersmid.
Technical Details Subject to Change
1.6 Mounting
The 19”-rack mount device should only be used in stationary conditions, the portable
device can be used on site, both with correct selection of the installation point and safe
installation. When used and properly installed in the prescript area ANKERSMID
SAMPLING guarantee a long-time of maintenance-free and satisfaction use.
Before switching ON the analyzer always check the gas inlet and outlet tubing are
correctly connected to the respective ports of the analyzer.
Sample gas pressure must be within the specification range in order to avoid gas leak
due to excessive pressure or wrong operation of the gas analyzer. A leakage of toxic or
explosive gas can lead to serious accident.
Sampling gas path should be based on the specific circumstances and make gas
conditioning device or system, otherwise it will cause the instrument does not work
properly.
Do not use the piping, pressure reducing valve or other sampling devices with grease
which may block the gas path or cause a fire.
Please use the power supply that complies with the power ratings of the analyzer, to
avoid fire and or abnormal operation.
Ensure the power supply is switched OFF when connecting the gas analyzer to the
process to avoid any accident by electric discharge.
Ground connection of the analyzer must be carried out according to the local and or
international regulations to prevent from any damage of injuries.
Ensure the power supply is switched OFF before connecting a data transmission line to
the serial RS232 port.
Insulation of all electrical connections and cables must be controlled before switching ON
the analyzer to avoid any dangerous and accidental electrical discharge.
NOTE
Test the gas analyzer response exclusively on target gases with certified gas
canisters from known concentration.
Span gas concentration need to be within 90% to 100% of the designed full scale range,
otherwise the accuracy and linearity of the analyzer can be affected. Use pure nitrogen
N2 5.0 quality for zero calibration.
Never let the moisture enter the analyzer to avoid electric discharge and risk of short
circuit.
Always prevent dust and oil traces for entering the analyzer to avoid damages to the
measuring cells.
Never switch OFF & ON successively the analyzer power supply without reasons,
otherwise the analyzer lifetime might be shortened, or the analyzer might be damaged.
Do not touch the input and output terminal site by metal and fingers etc..
Otherwise, it may result in electric shock
Smoking nearby to the analyzer is not allowed and could lead to serious fire hazards
Do not allow moisture intrusion analyzer. Otherwise, it may result in electric shock or
analyzer internal fire.