  9. Hanna Instruments HI 701 User manual

Hanna Instruments HI 701 User manual

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments Pool Line HI95654 User manual

Hanna Instruments

Hanna Instruments Pool Line HI95654 User manual

Hanna Instruments HI 2550 User manual

Hanna Instruments

Hanna Instruments HI 2550 User manual

Hanna Instruments HI 93102 User manual

Hanna Instruments

Hanna Instruments HI 93102 User manual

Hanna Instruments HI 5421 User manual

Hanna Instruments

Hanna Instruments HI 5421 User manual

Hanna Instruments HI 758 User manual

Hanna Instruments

Hanna Instruments HI 758 User manual

Hanna Instruments HI 9210N User manual

Hanna Instruments

Hanna Instruments HI 9210N User manual

Hanna Instruments HI 98402 User manual

Hanna Instruments

Hanna Instruments HI 98402 User manual

Hanna Instruments HI 96771C User manual

Hanna Instruments

Hanna Instruments HI 96771C User manual

Hanna Instruments pH 21 User manual

Hanna Instruments

Hanna Instruments pH 21 User manual

Hanna Instruments HI 9142 User manual

Hanna Instruments

Hanna Instruments HI 9142 User manual

Hanna Instruments HI 254 User manual

Hanna Instruments

Hanna Instruments HI 254 User manual

Hanna Instruments HI99151 User manual

Hanna Instruments

Hanna Instruments HI99151 User manual

Hanna Instruments HI 96729C User manual

Hanna Instruments

Hanna Instruments HI 96729C User manual

Hanna Instruments HI 9828 User manual

Hanna Instruments

Hanna Instruments HI 9828 User manual

Hanna Instruments HI 83540 User manual

Hanna Instruments

Hanna Instruments HI 83540 User manual

Hanna Instruments HI98164 User manual

Hanna Instruments

Hanna Instruments HI98164 User manual

Hanna Instruments HI 96762 User manual

Hanna Instruments

Hanna Instruments HI 96762 User manual

Hanna Instruments HI 9033 User manual

Hanna Instruments

Hanna Instruments HI 9033 User manual

Hanna Instruments HI 84432 User manual

Hanna Instruments

Hanna Instruments HI 84432 User manual

Hanna Instruments HI 758 User manual

Hanna Instruments

Hanna Instruments HI 758 User manual

Hanna Instruments Color of Water HI 727 User manual

Hanna Instruments

Hanna Instruments Color of Water HI 727 User manual

Hanna Instruments HI 83748 User manual

Hanna Instruments

Hanna Instruments HI 83748 User manual

Hanna Instruments HI 993310 User manual

Hanna Instruments

Hanna Instruments HI 993310 User manual

Hanna Instruments pHep HI 98107 User manual

Hanna Instruments

Hanna Instruments pHep HI 98107 User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

IRIMO 984-6D-1 owner's manual

IRIMO

IRIMO 984-6D-1 owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Free Chlorine Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 701 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• HI 701-25 Reagents for 25 tests
• 1 x 1.5V AAA Battery
• Instruction Manual and Quick Reference
HI 701
Technical specifications:
Range 0.00 to 2.50 ppm
Resolution 0.01 ppm
Accuracy ±0.03 ppm ±3% of reading @ 25 °C / 77 °F
Light Source Light Emitting Diode @ 525 nm
Light Detector Silicon Photocell
Method Adaptation of USEPA method 330.5. The reaction
between free chlorine and the DPD reagent causes a
pink tint in the sample.
Environment 0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type 1 x 1.5V AAA
Auto-Shut off After 2 minutes of non-use
Dimensions 86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)
Weight 64 g (2.3 oz.)
Functional description:
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.00” indicates that the sample absorbs
less light than the zero reference. Check the procedure and make sure
you use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.
1. Dust cover.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button.
For more details about spare parts and accessories
see “Accessories”.
www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
additional technical information, do not hesitate to e-mail us at [email protected]
or view our worldwide contact list at www.hannainst.com.
[email protected]
[email protected]
www.ponpe.com
•Turn the meter on by pressing the button.
All segments are displayed. When the display
shows “Add”, “C.1” with “Press” blinking,
the meter is ready.
•Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.
•Press the button. When the display shows
“Add”, “C.2” with “Press” blinking the
meter is zeroed.
•Remove the cuvette from the meter and
unscrew the cap. Add the content of one
packet of HI 701-25 reagent. Replace the
cap and shake gently for 20 seconds.
Replace the cuvette into the meter.
•Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement)
or, alternatively, wait for 1 minute and press
the button.
•The instrument directly displays the
concentration of free chlorine in ppm. The
meter automatically turns off after 2 minutes.
10 mL
•It is important that the sample does not contain any debris.
•Whenever the cuvette is placed into the measurement cell, it must be dry
outside, and completely free of fingerprints, oil and dirt. Wipe it thoroughly with
HI 731318 or a lint-free cloth prior to insertion.
•Shaking the cuvette can generate bubbles, causing higher readings. To obtain
accurate measurements, remove such bubbles by swirling or by gently tapping
the cuvette.
•Do not let the reacted sample stand for too long after reagent is added, as
accuracy will be affected.
•After the reading it is important to immediately discard the sample, otherwise
the glass might become permanently stained.
REAGENT SETS
HI 701-25 Reagent for 25 Free Chlorine tests
OTHER ACCESSORIES
HI 701-11 Free Chlorine Certified Standard Kit
HI 731318 Cloth for wiping cuvettes (4 pcs.)
HI 731321 Glass cuvettes (4 pcs.)
HI 731225 Black Cuvette Cap for Checker HC (4 pcs.)
HI 740028 1.5V AAA batteries (4 pcs.)
HI 93703-50 Cuvette cleaning solution (230 mL)
Recommendations for Users
Before using this product, make sure that it is entirely suitable for your specific application and for the
environment in which it is used.
Operation of this instrument may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damage or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find a Hanna Office in your area,
visit our web site:
www.hannainst.com
Hanna Instruments reserves the right to modify the design, construction or
appearance of its products without advance notice.
Battery management
To save the battery, the instrument shuts down after 2 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
•Turn the instrument off by holding the button until the meter shuts off.
•Turn the instrument upside down and remove the battery cover with a screwdriver.
•Remove the battery from its location and replace it with a new one, inserting the
negative end first.
•Insert the battery cover and replace the screw with a screwdriver.
IST701 06/15