
AQC Dust Collecting Systems
Follow these steps to install the drum back into
position:
1. Ensure the flange gasket seal is in place
and intact. Replace if damaged.
2. Slide the drum into the ramp slides.
3. Check to make sure the drum flange is
pushed at the end of the slide.
4. Lift drum up.
5. Attach the drum to the quick clip
system.
6. Check the seal assembly. Adjust quick
clip system if there is a leak.
Platform and ladder
Platforms and ladders are available for 8 to 80
cartridges Maxiflo-VC units. They may be
stationary or attached to the dust collector
structure. Contact an A.Q.C Inc. Representative
to know which platform design is best suited to
your needs.
Instructions for platform and ladder assembly
are included on delivery.
Warning!
Ensure platform and ladder are correctly
installed before climbing on the platform.
Explosion relief vent
Note on explosion venting panels : a minimum
clearance of 25’ (8 meters) free of obstacles,
pedestrian walkway, building walls, trees or
bushes is required to allow dispersion of
possible blast. Contact factory for details.
A.Q.C. Inc. will not be held liable nor
responsible for any injury or damage caused by
fire or explosion as per the agreement issued
upon construction of the dust collector.
However, the explosion relief vent is designed
to meet NFPA-68 criteria, related to explosion
venting systems. The entire venting surface is
calculated using such criteria. The explosion
venting door is held shut using handles
especially designed to open under a specific
pressure. Handles were factory adjusted as per
the manufacturer’s requirements considering
the explosion venting surface, the type of
handles and the type of attachments.
Warning!
The explosion vent must point in a
direction away from workers area,
offices, pedestrian walkways or any
other area usually accessible, vegetation
or any matter that would sustain
damages caused by an explosion.
Magnehelic gauge
The Magnehelic is a differential pressure gauge
used to measure the difference between the
clean and dirty air. This allows a visual
reference on the filters status and indicates
when it is time to replace.
This gauge is generally factory installed unless
under specific request to A.Q.C.. If the
Magnehelic gauge is not part of the collector,
connect the (HIGH-PRESSURE) tube on the
‘’dirty’’ side of the collector. Connect the other
(LOW-PRESSURE) tube on the ‘’clean’’ side
of the collector.
This gauge is not available when the collector is
supplied with a DCT-1000 electronic board and
a DCP pressure module since it is this one that
will read the pressure differential.
18