Arizona Instrument JEROME J605 User manual

AZI P/N 700-0108-E
Last update June 2013
USER MANUAL
JEROME®J605
HYDROGEN SULFIDE ANALYZER
June 2013
Firmware v1.2X
ARIZONA INSTRUMENT LLC
3375 N Delaware Street | Chandler, AZ 85225 USA
800.528.7411 | 602.470.1414 | f 602.281.1745
www.azic.com
Email:
General – [email protected]
International – intl@azic.com

AZI Customer Service 800-528-7411 or 602-470-1414 Page 2 of 57
JEROME®J605
Hydrogen Sulfide Analyzer
Operation Manual
PROPRIETARY RIGHTS NOTICE
This manual contains valuable information and material developed by Arizona Instrument LLC
for use with the Jerome®J605 Hydrogen Sulfide Analyzer. No part of this manual can be
reproduced or transmitted in any form or by any means, electronic, mechanical or otherwise. This
includes photocopying and recording or in connection with any information storage or retrieval
system without the express written permission of Arizona Instrument LLC.
ALL RIGHTS RESERVED
Copyright 2010-2013 Arizona Instrument LLC
Resisorb®is a registered trademark of Avantor Performance Materials.
Tygon®is a registered trademark of Saint-Gobain Performance Plastics Corporation.
Acrodisc®is a registered trademark of Pall Corporation.
Windows®is a registered trademark of Microsoft Corporation in the United States and other
countries.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 3 of 57
Table of Contents
1. UNPACKING THE INSTRUMENT.....................................................................................................4
2. INTRODUCTION ...................................................................................................................................6
3. JEROME®J605 TECHNICAL SPECIFICATIONS ...........................................................................8
4. PRINCIPLE OF OPERATION..............................................................................................................9
Zero Air Filter (AZI P/N Z2600 3905).......................................................................................................11
5. INSTRUMENT OPERATION...............................................................................................................12
J605 Main Screen Display..........................................................................................................................12
J605 Back Panel Connections.....................................................................................................................13
J605 User Interface Main Menu Structure..................................................................................................14
SAMPLE menu ..........................................................................................................................................15
REGEN menu.............................................................................................................................................17
DATA menu...............................................................................................................................................18
SYSTEM menu ..........................................................................................................................................20
Daily Operations.........................................................................................................................................22
Sensor Regeneration...................................................................................................................................23
Sample Mode..............................................................................................................................................24
Survey Mode ..............................................................................................................................................25
Operating on AC Power or Generator........................................................................................................26
Operating on Internal Battery Power..........................................................................................................26
Battery Management ..................................................................................................................................26
External Battery Power...............................................................................................................................27
Charging Internal Battery...........................................................................................................................28
Retrieving Data...........................................................................................................................................29
Instrument I/O Interface.............................................................................................................................30
4-20mA Output...........................................................................................................................................30
6. MAINTENANCE.....................................................................................................................................32
Preventive Maintenance Calendar..............................................................................................................32
Flow System...............................................................................................................................................33
.25 inch Fritware Filter........................................................................................................................33
Internal Filters .....................................................................................................................................34
Replacing the Battery Pack ........................................................................................................................35
7. CALIBRATION.......................................................................................................................................36
Factory Calibration Service........................................................................................................................36
Verification of Functionality and Quality Control .....................................................................................36
8. J605 TROUBLESHOOTING.................................................................................................................37
Potential Interferences................................................................................................................................39
9. ACCESSORIES & MAINTENANCE PARTS.....................................................................................40
Spare Parts & Accessories..........................................................................................................................42
10. APPENDIX A – USB/HYPERTERMINAL SETUP ............................................................................45
11. APPENDIX B - J605 FUNCTIONAL TEST MODULE......................................................................55
Simple Operation........................................................................................................................................55
12. WARRANTY...........................................................................................................................................56
An up-to-date electronic copy of this manual can be found on the AZI website at
http://www.azic.com/
NOTE: For levels less than 10 ppb, it is necessary to run a warm-up routine before
sampling. To initiate the automatic five minute warm-up, install a Zero Air Filter (AZI
P/N Z2600 3905) in the intake and select Warmup from the REGEN menu. Refer to the
Daily Operations section for more information.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 4 of 57
1. UNPACKING THE INSTRUMENT
This manual contains details that will optimize the results and the life of your instrument. Read
and refer to the manual for complete details on operation, maintenance and troubleshooting,
special voltage inputs and data output.
The Jerome®J605 is easy to operate and ready for use upon receipt from the factory.
•Remove the instrument from the packing material.
Retain all packaging materials for any future shipment of the
instrument.
If the instrument is returned to AZI for any reason, it must be
placed in the original packaging materials that have been tested and
proven to be effective protection during shipment.
•Call AZI Customer Service at 800-528-7411 or 602-470-1414 or go to the AZI website
www.azic.com for Return Material Authorization (RMA) information prior to returning an
instrument.
•Boxes and packing materials for all shipments are available from AZI.
•Pack the Jerome®instrument only in a Jerome®shipping container.
AZI WILL NOT BE RESPONSIBLE FOR SHIPPING DAMAGE.
IF YOU RETURN THE INSTRUMENT IMPROPERLY PACKAGED OR
SHIPPED,
YOU SHOULD INSURE IT FOR FULL VALUE.
•Check for any damage and confirm receipt of all parts
on your packing list. Contact Arizona Instrument
Customer Service at (800) 528-7411 or
(602) 470-1414 if you have any questions.
•Press the I/O button. The display will light up and
show instrument serial number and software revision.
If necessary, press ESC to clear any calibration
reminders. Call AZI Customer Service or e-mail
[email protected], to schedule instrument
calibration.
•The digital meter displays 0.000 ppm (or 0.00 ppb).
•Look at the battery icon in the top center of the display
to determine the current battery level.
If the battery icon is empty and flashing, recharge
the battery. See page 28.
•The included AC power supply/charger can utilize 110V or 220V AC power, and it is not
necessary to manually select the input voltage.
•The J605 contains an auto-resetting fuse that does not require care, maintenance or
replacement by the user.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 5 of 57
•Perform a sensor regeneration by following these steps:
Connect the AC power supply/charger between the matching (DIN) connector on the back
of the J605 and an AC power outlet or connect the external battery pack to the back of the
J605.
If the J605 is turned off, press the I/O power button to turn the instrument on.
Press the RIGHT arrow button (►) to enter the main menu.
Press the DOWN arrow button (▼) to move the cursor to Regen.
Press the RIGHT arrow button (►) to select Regen from the menu.
Press the ENTER/START button on the keypad to select Regen Now from the Regen
menu.
The instrument will now begin a 45-minute regeneration cycle, indicated by
“Regeneration in Progress” on the display. Do not interrupt this cycle. For a
complete description of this process, see page 23.
If any error message appears on the display, see the J605 TROUBLESHOOTING
section beginning on page 37.
•Ensure the instrument has been powered on for at least five (5) minutes prior to sampling.
•The instrument is now ready to sample.
•The instrument is designed for ambient air monitoring.
•The instrument is not explosion
proof.
•Press the SAMPLE BUTTON at the
end of the handle of the J605 to start a
sampling cycle.
•When the instrument measures
hydrogen sulfide, the 0.000 ppm (or
0.00 ppb) display will be replaced
with a value.
•To ensure the input to the instrument contains no hydrogen sulfide or mercaptans, install a
Zero Air Filter (AZI P/N Z2600 3905) in the intake of the instrument. The Zero Air Filter
cleans the air sample and should produce sample readings of 0.000 ppm (or 0.00 ppb).
Therefore, use the filter to:
Equilibrate the instrument to temperatures that are higher or lower than the instrument.
Sample with the filter installed until the reading is 0.000 ppm (or 0.00 ppb).
Identify contamination within the instrument.
Confirm the presence of hydrogen sulfide when readings are elevated. Install the Zero Air
Filter and verify that the readings go down with the filter installed.
•Perform sensor regeneration before each day’s use.
•Perform another sensor regeneration after each day’s use.
•During periods of storage or inactivity, perform sensor regeneration every 30 days.
Call AZI Customer Service, at (800) 528-7411 or (602) 470-1414 if you have any questions.
WARNING
DO NOT allow the probe or the instrument’s intake to be exposed to any liquid.
DO NOT obstruct the intake or exhaust ports of the J605, as this could cause errors
in readings and damage to the flow control system.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 6 of 57
2. INTRODUCTION
The Jerome®J605 Hydrogen Sulfide Analyzer is an ambient air analyzer with a range of 3 parts
per billion (ppb) to 10 parts per million hydrogen sulfide (ppm H2S). Readings of 100 ppb or less
are displayed in units of ppb, and readings above 100 ppb (0.100 ppm) are displayed in ppm.
Readings of zero are displayed as 0.000 ppm or 0.00 ppb depending on the Range in use.
CAUTION:
The Jerome®J605 is for gaseous vapor use only.
DO NOT allow the probe or the instrument's intake
to be exposed to any liquid, dust
or other foreign material.
The J605 is designed to be easy to operate for quick and accurate analysis of hydrogen sulfide
levels. It has few maintenance requirements. However, please take a moment to read this manual
before attempting operation. If you have any questions about your application or operation,
please call AZI Customer Service at (800) 528-7411 or (602) 470-1414 or e-mail
J605 Features
•Accurate analysis of hydrogen sulfide concentrations in seconds.
•3 ½ inch (9 cm) backlit display, showing everything you need to be confident in the
instrument, including 5/8 inch (1.6 cm) tall character readout, battery charge indicator,
and sensor saturation percentage.
•Microprocessor ensures a linear response throughout the entire range of the sensor.
•Inherently stable gold film sensor.
•Wide detection range allows for multiple applications.
•Survey mode for rapid source detection of hydrogen sulfide concentrations.
•Rechargeable internal battery pack for portability.
•Easy to understand diagnostic and error detection.
•Auto zero bridge adjustment.
•Updated electronics including 24-bit a/d conversion, and on-board storage for 20,000
samples.
•Controlled temperature film heat ensures the sensor removes all hydrogen sulfide and
returns to its original state.
•The Jerome®J605 can be operated for 18 hours from the internal 12V battery or it can
be operated using the external AC power supply/charger.
Accessories and Maintenance Parts
The Accessories and optional items available to support the J605 are listed and pictured beginning
on page 40.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 7 of 57
Applications
•Regulatory compliance
•Ambient air analysis
•Odor nuisance monitoring
•Control room corrosion monitoring
•Waste water treatment facilities
•Quality control
•Scrubber efficiency testing
•Accuracy check for other hydrogen sulfide monitors and control systems
•Hydrogen sulfide source detection
•Leak detection
•Portable hydrogen sulfide detection. (The Jerome®J605 is operated from the internal
12V battery. For charging the battery, an external AC power supply/charger is
included.)

AZI Customer Service 800-528-7411 or 602-470-1414 Page 8 of 57
3. JEROME®J605 TECHNICAL SPECIFICATIONS
Range*
3 ppb (0.003 ppm) to 10 ppm H2S in three graduated ranges
Resolution
0.02 ppb (20 ppt) H2S (range dependent)
Response time – sample mode
1.0 to 10.0 ppm (Range 2)
0.10 to 1.0 ppm (Range 1)
3 to 100 ppb (Range 0)
3 ppb to 10.0 ppm (full range)
12 seconds (manual range selection)
18 seconds (manual range selection)
12 to 27 seconds (manual range selection)
12 to 52 seconds (auto range mode)
Response time – survey mode
Manual range selection:
Range 2
Range 1
Range 0
Auto range mode
12 seconds for each reading
18 seconds for each reading
12 to 27 seconds for each reading
12 to 52 seconds for each reading
Flow rate
150ml/min (0.15 liters/min)
Power requirements
12 VDC for the instrument
(provided by the internal battery, AC power supply/charger,
external battery pack or car accessory cable)
100-240VAC, 47-63Hz, 3.2A for the AC power supply/charger
Fuse
Auto-resetting fuse
Internal battery pack
Rechargeable nickel metal hydride (NiMH)
Operating environment
0 °C to 40 °C, non-condensing, non-explosive
Case construction
Aluminum, powder coated
Dimensions
11 in L x 6 in W x 6.5 in H (28 cm L x 16 cm W x 17 cm H)
Weight
5.4 pounds (2.5 kilograms)
Digital meter display
3 ½ inch (9 cm) backlit liquid crystal display (LCD)
Data storage capacity
20,000 samples
Output Digital: USB serial data to PC, printer, or USB flash drive
Analog: 4-20mA current loop (requires external power source)
accurate to 0.3% of output
Certifications
TUV 61010, CE
Accuracy and Precision:
Range
Accuracy
Precision (RSD)
Range 0:
5 ppb H2S
±
1 ppb
10%
50 ppb H2S
±
3 ppb
5%
Range 1:
0.50 ppm H2S
±
0.03 ppm
5%
Range 2:
5.0 ppm H2S
±
0.3 ppm
5%
*Sample values below 3 ppb (0.003 ppm) are recorded and displayed as 0.00 ppb by the J605. Sample
values above 10 ppm (or above the upper limit of the chosen range) are recorded and displayed as “High
Concentration.” “High Concentration” may occur in Auto-range if the gas concentration increases while
the instrument is switching down ranges. In this situation, sample again and use the following result or
lock the J605 in the desired range.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 9 of 57
4. PRINCIPLE OF OPERATION
The heart of the J605 is the well-proven gold film sensing technology. A thin gold film, in the
presence of hydrogen sulfide, undergoes an increase in electrical resistance proportional to the
mass of hydrogen sulfide in the sample.
When the SAMPLE button is pressed, an internal pump pulls ambient air over the gold film
sensor for a precise period of time. The gold film sensor adsorbs the hydrogen sulfide, and the
instrument determines the amount adsorbed and displays the measured concentration of hydrogen
sulfide in parts per million (ppm) or parts per billion (ppb) depending on the Range in use.
During normal sampling in Range 1 or 2, the ambient air sample is diluted in the flow system at a
ratio of 100:1.
Initial Pump, Range 1 or 2
Sample Period, Range 1 or 2

AZI Customer Service 800-528-7411 or 602-470-1414 Page 10 of 57
When sampling in Range 0, (where low levels of hydrogen sulfide are expected) undiluted air
samples are drawn across the gold film sensor.
Initial Pump, Range 0
Sample Period, Range 0
The instrument’s microprocessor automatically re-zeroes the digital meter at the start of each
sample cycle and holds the meter reading until the next sample cycle is activated.
During the sample mode cycle, a sensor saturation meter on the LCD represents the percentage of
sensor saturation, or adsorbed hydrogen sulfide collected on the gold film. With use, the sensor
becomes saturated and needs to be cleaned. This is accomplished by a manually activated 45-
minute sensor regeneration cycle, which removes the hydrogen sulfide from the sensor. The
bypass solenoid closes during the sensor regeneration cycle, causing the air to pass through the
scrubber filter, providing clean air for the regeneration process. The closed loop film heat cycle
ensures the sensor is cleaned to its original state. The hydrogen sulfide released during
regeneration is absorbed by the flow system’s final scrubber filter (AZI P/N: Z2600 3930) to
prevent any external contamination from the desorbed hydrogen sulfide. The 45-minute
regeneration process includes a cool-down phase, so the instrument is ready for use as soon as the
regeneration process finishes.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 11 of 57
Zero Air Filter (AZI P/N Z2600 3905)
The Zero Air Filter (AZI P/N Z2600 3905) removes hydrogen sulfide, mercury vapor and
mercaptans from the air sample. Readings with the filter installed should be near 0.000 ppm (or
0.00 ppb). For maximum accuracy, wait 15 seconds between samples to allow the sensor to
restabilize.
Because air that is cooler than the instrument will cause low readings and warmer air will cause
higher readings, the Zero Air Filter should be used to equilibrate the instrument to ambient air.
Repeated sampling with clean air will not cause saturation of the gold film sensor but will
equalize temperatures faster to allow accurate analysis to begin sooner. For maximum accuracy,
wait 15 seconds between samples to allow the sensor to restabilize.
The Zero Air Filter can also be used to identify contamination within the instrument. If the
readings do not reduce to near 0.000 ppm with the filter installed, contamination should be
suspected. If the readings do drop to near 0.000 ppm with the filter installed but elevate with the
filter removed, the presence of hydrogen sulfide at the sampled location is confirmed. Again, for
maximum accuracy, wait 15 seconds between samples to allow the sensor to restabilize.
The Zero Air Filter should be inserted into the intake of the instrument when running the
instrument Warm-up.
For more information on the use of the Zero Air Filter, contact customer service at 1-800-528-
7411 or 1-602-470-1414.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 12 of 57
5. INSTRUMENT OPERATION
J605 Main Screen Display
The J605 Main Screen is depicted below, followed by explanations of the labeled items.
SYSTEM TIME – As set in the SYSTEM menu. (See SYSTEM menu on page 20)
BATTERY – Indicates current charge level and charging status (See Charging Internal Battery
on page 28)
SOUND ICON – Indicates if the audible alarm is turned on or muted, as set in the SYSTEM
menu. (See SYSTEM menu on page 20).
SYSTEM DATE – As set in the SYSTEM menu. (See SYSTEM menu on page 20)
SENSOR SATURATION METER – Graphically indicates the hydrogen sulfide saturation level
of the J605 sensor. (See REGEN menu on page 17).
USB ICON – Indicates proper functioning of USB ports, and flashes during USB transmission of
data. (See Retrieving Data on page 29). If the USB icon is not present, this option is not
activated. Contact your AZI Sales Representative for details on the USB option.
RANGE – Indicates in which Range the J605 is set to sample: 0, 1, 2 or A (Auto-range), as set in
the SAMPLE menu. (See Range in SAMPLE menu on page 15).
AUTO SAMPLING INDICATOR – Indicates if the J605 is set to sample automatically at
regular intervals, as set in Sample Mode in the SAMPLE menu. (See SAMPLE menu on
page 15).
HYDROGEN SULFIDE READING – The hydrogen sulfide concentration detected by the
J605, in the units indicated.
UNITS – Units for the hydrogen sulfide reading: parts per million (ppm) or parts per billion
(ppb), depending on the Range in use and the sample concentration detected. Readings of
zero are displayed as 0.000 ppm or 0.00 ppb, depending on the Range being used.
AVAILABLE KEYPAD COMMANDS – The bottom line of the screen indicates the currently
available keypad commands. From the main screen, as shown here, the Main Menu can
be accessed using the ► key.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 13 of 57
J605 Back Panel Connections
The above-labeled connections are available from the back panel of the J605. If the USB option
has not been activated, the USB ports will not be present. Contact your AZI Sales Representative
for information on the USB option.
POWER CONNECTOR – Connect the supplied AC power supply/charger here to power the
instrument or recharge the internal battery, or connect the optional external battery pack (AZI
P/N: 990-0214) or the optional car accessory cable (AZI P/N: 200-0170) here to provide power to
the instrument.
USB TYPE “A” PORTS – The USB Type “A” ports are used to connect a USB printer or USB
memory drive for printing or saving sample data or to connect a USB keyboard for menu
navigation or data entry. The two ports are identical, and either can be used. For more
information, see DATA menu on page 18.
USB TYPE “B” PORT – The USB Type “B” port is used to connect the J605 directly to the
USB port of a computer, using the included USB A to B cable (AZI P/N: 200-0165). For more
information see DATA menu on page 18.
DB25 CONNECTOR – There are multiple specialized uses for the DB25 connector. See
Instrument I/O Interface on page 30 for more information. NOTE: This is not a printer port.
EXHAUST PORT – After sampling, the sampled air is cleaned of any hydrogen sulfide and
expelled from the exhaust port on the back of the J605.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 14 of 57
J605 User Interface Main Menu Structure
NOTE: The main menu also contains a selection labeled “Factory.” This option is for AZI
factory use only and is not accessible by the end user.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 15 of 57
A USB keyboard (AZI P/N: 990-0230) can be used to navigate the menu system and for data
entry for items such as Location Names. Connect the USB keyboard to one of the two USB Type
“A” ports on the back of the instrument. Either upper or lower case characters may be entered
when using a USB keyboard, while only upper case letters are available from the J605 keypad.
The keyboard arrow keys can be used to navigate the menu system.
SAMPLE menu
SAMPLE MODE
The Sample Mode setting provides the choice between auto-sampling and manual sampling. For
normal use, use manual sampling, which is the instrument default setting. If auto-sampling is
selected, the instrument will automatically sample at the interval set in SET SAMPLE
INTERVAL (below) without monitoring or attendance by the user. Use the ▲and ▼arrows to
toggle between Manual and Auto, and use ENTER/START to save the selection. Use the ESC
key to exit without saving changes. If auto-sampling is selected, the word Auto will appear below
the system date on the instrument’s main screen.
RANGE
Three different ranges and an Auto-ranging function are available in the J605. If the approximate
value of the expected H2S concentration is known, the instrument can be set into that specific
range prior to sampling (0, 1, or 2). If the expected H2S concentration is not known, Auto-
ranging can be used instead, and the J605 will test each sample in each range before displaying
the concentration. Using Auto-ranging results in longer test times, however accuracy is not
affected. The selected range also determines what units the J605 will use to display the results.
The three ranges and their default display units (ppb or ppm) are shown below:
RANGE
CONCENTRATION
0
3 to 100 ppb
1
0.10 to 1.0 ppm
2
1.0 to 10.0 ppm
The range can be set or Auto-ranging can be toggled on or off by selecting Range from the
SAMPLE menu. Use ▲ and ▼ to scroll through the ranges 0, 1, 2 and Auto, and press
ENTER/START to make a selection. Press ESC to exit without saving changes. The selected
range will be displayed on the main screen as a number (0, 1, 2) or as an Afor Auto-ranging. See
J605 Main Screen Display on page 12 for the location of the Range indicator.
SET SAMPLE INTERVAL
If auto-sampling is being used (as set in SAMPLE MODE above), the sample interval
determines how often the J605 will automatically take samples. Use the ▲and ▼arrow keys to
scroll through the available sample intervals of 1, 2, 5, 10, 15, 30, 45, 60, 90 and 120 minutes, and
use ENTER/START to save the selected interval. Use the ESC key to exit without saving
changes.
The sample intervals refer to specific times on the instrument clock as measured in whole
multiples of the selected interval. For example, selecting a 10-minute sample interval will result

AZI Customer Service 800-528-7411 or 602-470-1414 Page 16 of 57
in samples being taken on the hour, 10 minutes after the hour, 20 minutes after the hour, etc., not
every ten minutes from when auto-sampling is initiated. Refer to the chart below for examples of
how these intervals work. The Main Screen must be displayed for autosampling takes place, and
autosampling will start when the clock reads :00 seconds.
Interval
Samples on the:
Examples:
1 minute 1-minute marks 12:00, 12:01, 12:02, 12:03, 12:04, 12:05, 12:06, etc.
2 minutes 2-minute marks 12:00, 12:02, 12:04, 12:06, 12:08, 12:10, 12:12, etc.
5 minutes 5-minute marks 12:00, 12:05, 12:10, 12:15, 12:20, 12:25, 12:30, etc.
10 minutes 10-minute marks 12:00, 12:10, 12:20, 12:30, 12:40, 12:50, 1:00, etc.
15 minutes Quarter hour 12:00, 12:15, 12:30, 12:45, 1:00, 1:15, 1:30, etc.
30 minutes Half hour 12:00, 12:30, 1:00, 1:30, 2:00, 2:30, 3:00, etc.
45 minutes
Times that are an even
45 minutes from
midnight or noon
12:00, 12:45, 1:30, 2:15, 3:00, 3:45, 4:30, 5:15, 6:00,
6:45, 7:30, 8:15, 9:00, 9:45, 10:30, 11:15, etc.
60 minutes Hour 12:00, 1:00, 2:00, 3:00, 4:00, 5:00, 6:00, etc.
90 minutes
Times that are an even
90 minutes from
midnight or noon
12:00, 1:30, 3:00, 4:30, 6:00, 7:30, 9:00, 10:30, etc.
120 minutes
Times that are an even
120 minutes from
midnight or noon
12:00, 2:00, 4:00, 6:00, 8:00, 10:00, etc.
SET LOCATION NAME
Up to 80 unique location presets are available using the J605. This feature is useful if sampling is
done in multiple locations and there is a need to record which readings came from which
locations. Use the ▲and ▼arrows to scroll through the list of 80 presets until the location to be
edited is highlighted. Press ENTER/START to begin editing the name of the selected location.
Use the ▲and ▼arrows to scroll through the available alphanumeric characters and punctuation.
When the desired character is displayed, use the ► arrow to move to the next character field. If a
mistake is made, the ◄ arrow can be used to return to the previous character. When the location
has been named as desired, press ENTER/START to save the location name, and return to the
list of locations. To select an existing location from the list for the samples to be taken, use the
▲and ▼arrows to scroll to the desired location, and press ENTER/START twice. Note the
asterisk (*) located to the left of the currently selected location. The selected location name will
be stored with the sample data for all samples taken until a different location is selected instead.
Use ESC to return to the sample menu. See Retrieving Data on page 29 for more info.
WARNING
The 605 is not intended for continuous sampling 24 hours a day / 7 days a week.
Continuous sampling may reduce sensor life and may affect your warranty.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 17 of 57
SURVEY MODE
Highlight SURVEY MODE in the SAMPLE menu and press ENTER/STARTto begin
immediate sampling in survey mode. In survey mode, after an initial four-second pause to purge
the flow system, the J605 will sample continuously. For complete information on survey mode,
see Survey Mode on page 25.
REGEN menu
The sensor saturation meter on the main screen (to the right of s. sat) indicates the current level of
saturation of the sensor with hydrogen sulfide.
When the sensor becomes saturated with hydrogen sulfide, as indicated by the sensor saturation
meter, a regeneration (regen) must be performed on the sensor to release the accumulated
hydrogen sulfide, before additional samples can be taken. See Sensor Regeneration on page 23
for more information on the regeneration procedure.
The AC power supply/charger or the external battery pack must be plugged into the back of the
instrument before a regeneration can be performed.
The following options are available from the REGEN menu:
REGEN NOW
Connect external power to the instrument before performing a regeneration.
Highlighting REGEN NOW and pressing ENTER/START will immediately start the sensor
regeneration process. Do not disconnect external power during the regeneration.
WARMUP
To ensure maximum accuracy when sampling at levels less than 10 ppb, the J605 must be
warmed up with the Zero Air Filter in the intake, at the location to be sampled. To run the warm-
up, insert the Zero Air Filter in the intake, highlight WARMUP and press ENTER/START.
While warming up, the pump will run and the J605 will display a 5-minute countdown. When the
warm-up finishes, remove the Zero Air Filter and begin normal sampling. If the J605 is idle for
more than 20 minutes, another warm-up is required to maintain maximum accuracy at these low
levels.
Clean/Regenerated Sensor
Partially Saturated Sensor
Saturated Sensor –
Regeneration Required

AZI Customer Service 800-528-7411 or 602-470-1414 Page 18 of 57
SET INTERVAL
If auto-sampling is being used (as set in SAMPLE MODE above), the J605 can be configured to
automatically perform sensor regenerations at a predetermined interval. Use the ▲and ▼arrows
to scroll through the available intervals of 6, 12, 24, and 48 hours, and press ENTER/START to
save the selected interval. The J605 will now automatically perform a sample followed by a
regeneration every 6, 12, 24, or 48 hours as selected, starting at the time set in SET HOUR
below. Use ESC to exit without saving.
SET HOUR
Use the ▲and ▼arrow keys to select the hour of the day (using a 24-hr clock) to start
regeneration. Press ENTER/STARTto save the setting. At the time set, if auto-sampling is
being used, the J605 will automatically take one sample and then perform a regeneration. This
process will repeat at the interval indicated in SET INTERVAL above. Use ESC to exit without
saving.
NOTE: If the sensor is saturated and a regeneration cycle is performed and completed before the
set hour, the instrument will perform another regeneration at the set hour. However, if the
regeneration cycle does not complete prior to the set hour, the regeneration at the set hour will be
skipped.
DATA menu
Sample data can be stored within the instrument, up to a maximum of twenty thousand (20,000)
data points. On instruments that have the USB Communications option, in addition to viewing
the sample data on the instrument, data can be retrieved from the instrument in one of three ways:
1. SEND TO PC
2. SAVE TO USB MEMORY DRIVE
3. PRINT DIRECTLY TO AN AZI APPROVED PRINTER (AZI P/N: Y990-0098)
The presence of the USB icon on the main screen indicates that the Communications option has
been purchased and that the USB ports are functioning properly. The USB icon will flash during
data transmission. See J605 Main Screen Display on page 12 for the location of the USB icon.
The data is stored in non-volatile memory and will not be affected if the battery is changed or
disconnected.
VIEW STORED DATA
Select VIEW STORED DATA to view previous sample and regeneration history on the screen
of the J605. The Location Name (if specified) corresponding to the data is displayed at the top of
the screen and three sample values at a time are displayed at the bottom of the screen, along with
the time and date the sample was taken. The most recent sample is displayed at the top of the list,
and regenerations are indicated by the word “Regen” instead of a sample value. Use the ▲ and
▼ arrow keys to scroll through the data. If Location Names are used, the Location Name
displayed will automatically update during scrolling to reflect the sampling location of the data
being displayed. The J605 will beep-beep upon reaching the end of the list. After viewing the
data, press ESC to exit and return to the DATA menu.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 19 of 57
On instruments that have the USB Communications option, all of the data is accessible from the
VIEW STORED DATA menu selection. On instruments that do not have the USB
Communications option, only the previous two samples are stored.
SAVE TO USB DRIVE
Before selecting SAVE TO USB DRIVE, connect the target USB memory drive to either USB
TYPE “A” PORT on the back panel of the J605. When ENTER/START is pressed, all of the
stored data on the J605 will be transmitted to the target drive as a comma-delimited text file. See
Retrieving Data on page 29 for more information.
NOTE: The J605 is NOT compatible with USB memory drives that have the U3 program
pre-installed. Do not use USB memory drives that have the U3 program with the J605.
SEND TO PC
Before using SEND TO PC, the USB driver must be installed on the target PC, and
HyperTerminal must be configured on the target PC. The USB driver is on the CD with this
operations manual and on the AZI website at www.azic.com. See APPENDIX A –
USB/HYPERTERMINAL SETUP on page 45 for detailed instructions on installing the USB
driver, configuring HyperTerminal, and downloading data from the J605. Once the target PC has
been configured, connect the included USB A to B cable (AZI P/N: 200-0165) between the USB
TYPE “B” PORT on the back panel of the J605 and the USB PORT of the target computer, and
launch HyperTerminal. On the J605, with SEND TO PC highlighted, press ENTER/START,
and all of the stored data on the J605 will be transmitted to the target computer as a comma-
delimited text file. See Retrieving Data on page 29 for more information.
PRINT DATA
Before selecting PRINT DATA, connect an AZI printer (AZI P/N: Y990-0098) to either USB
TYPE “A” PORT on the back panel of the J605 using the cable supplied with the printer. When
ENTER/START is pressed, all of the stored data on the J605 will be printed on the attached
printer. See Retrieving Data on page 29 for more information.
VIEW MEMORY USAGE
The J605 can store up to 20,000 data points. Selecting VIEW MEMORY USAGE will display
what percentage of the data storage area is in use. Once the data memory is full, no additional
data can be stored until the data on the instrument is deleted. If the memory is full, the J605 will
continue to perform as expected, but it will indicate that the memory is full and no new sample
data will be retained. The storage area can be cleared using DELETE DATA as described
below.
DELETE DATA
Selecting DELETE DATA will delete all sample data stored on the J605. Prior to deletion, the
J605 will prompt the user for confirmation as a precaution.

AZI Customer Service 800-528-7411 or 602-470-1414 Page 20 of 57
SYSTEM menu
The SYSTEM menu is password protected to prevent unwanted or accidental changes to
operating parameters. The default initial password is AZI, but should be changed to something
unique if restricted access to the SYSTEM menu is desired. The password can be up to 10
characters long.
When the SYSTEM menu is selected from the Main Menu, the J605 will prompt for the
password. Input the password using ▲ and ▼ to change the alphanumeric character, and ► and
◄ to move between characters. Press ENTER/START to proceed after entering the password.
The SYSTEM menu will remain unlocked until the instrument is powered off and back on.
After the correct password is entered, the password can be changed using the CHANGE
PASSWORD selection at the bottom of the SYSTEM menu.
SET ALARM LEVELS
When sampling, if the hydrogen sulfide level exceeds the selected alarm level, the word
“ALARM” will flash on the J605 display. If the AUDIBLE ALARM is set ON, the J605 will
also beep three times. Two different alarm levels can be set using this menu option. Use the
▲and ▼arrow keys to change the value of the blinking field, and the ► arrow to move to the
next character in the alarm level setting. Typically, the high alarm is used for an industrial level,
while the low alarm is set to a residential value. Use the Aand Bbuttons to switch between the
high and low alarm settings by pressing A to select the high alarm or Bto select the low alarm.
Once the alarm level has been set, press ENTER/START to select the highlighted alarm level as
the current alarm level to be used. Upon returning to the SET ALARM LEVELS screen, there
will be asterisk next to the currently selected alarm level (either high or low). It is recommended
to set the alarm level higher than .003 ppm. The alarm can be muted using AUDIBLE ALARM
ON/OFF in the SYSTEM menu.
AUDIBLE ALARM ON/OFF
Use the ▲and ▼arrow keys to toggle the audible alarm on or off, and use ENTER/START to
save the desired setting. Press ESC to exit without saving changes. The main screen display has
a sound icon that indicates if the alarm is currently turned on or muted. The action lines in front
of the sound icon on the main screen will disappear when the audible alarm is muted. See J605
Main Screen Display on page 12 to locate the sound icon.
DISPLAY MODE
From the DISPLAY MODE screen, use the ▲and ▼arrow keys to toggle between Threshold
and Scientific display modes, and press ENTER/START to save the setting. In Scientific display
mode, the J605 displays the sample reading for every sample taken. In Threshold display mode,
the sample reading must be within 85% of the set alarm level to be displayed. Results below 85%
of the alarm level value will be displayed as 0.000 ppm or 0.00 ppb depending on the current
Range selected. Use ESC to exit without saving. See SET ALARM LEVELSabove to set
Threshold limits.
Table of contents
Other Arizona Instrument Measuring Instrument manuals

Arizona Instrument
Arizona Instrument COMPUTRAC VAPOR PRO User manual

Arizona Instrument
Arizona Instrument COMPUTRAC MAX 4000 User manual

Arizona Instrument
Arizona Instrument Computrac MAX-5000XL User manual

Arizona Instrument
Arizona Instrument Jerome J505 User manual

Arizona Instrument
Arizona Instrument JEROME J405 User manual

Arizona Instrument
Arizona Instrument JEROME 431-X User manual

Arizona Instrument
Arizona Instrument Jerome J431-x Installation and operating instructions