ASCOM NUBM3-HE User manual

INSTALLATION GUIDE
Bedside Module (NUBM3-HE)

Symbols
INSTALLATION GUIDE
Bedside Module (NUBM3-HE)
1 Symbols
Important Safety Information
The Bedside Module (NUBM3–HE) Installation Guide contains important instructions when installing and
maintaining the NUBM3–HE. To ensure a safe working environment during the installation and operation of
the NUBM3–HE, the following safety symbols appear throughout this document to indicate dangerous
conditions and important safety instructions.
Caution: Use extreme caution and follow instructions carefully.
Pay particular attention to statements referred to by a note.
1TD 93505EN / 28 February 2022 / Ver. A

INSTALLATION GUIDE
Bedside Module (NUBM3-HE) Description
2 Description
The Bedside module is a wall-mounted device that is suitable for general room applications. The module
connects to an active room bus and is typically mounted near a bed.
Packing list:
• 1 x Bedside module NUBM3–HE
Available Accessories
Accessories Bedside Extension Module, EU (P/N NUBM3X-HE)
Installation Frame 1U, EU (P/N NUF1U-HE)
Installation Frame 2U, EU (P/N NUF2U-HE)
Installation Frame 3U, EU (P/N NUF3U-HE)
Button Insert Kit EU (P/N NUCBK-HE)
Button Insert Kit UK (P/N NUCBK-HK)
Surface Mounting Spacer, EU (P/N NUSP1-HE)
NIAT TV Audio Interface (P/N 660382)
Accessories are not included and must be ordered separately.
TD 93505EN / 28 February 2022 / Ver. A 2

Mounting
INSTALLATION GUIDE
Bedside Module (NUBM3-HE)
3 Mo nting
The base of the module mounts directly onto flat walls or ceiling surfaces.
Caution: To prevent a fire hazard because of dust buildup inside the module, regular
inspection and/or cleaning of the module is required.
Take proper measures to avoid spreading dust and other particles when applying
maintenance to a module that is mounted inside a clinical area.
3.1 Mo nt the NUBM3–HE onto a backbox
To attach an adapter to a backbox:
1. For EU-style or UK-style backboxes, loosen the screws from the backbox so that approximately 5mm
extend out from the backbox and the heads of the screws can pass through the keyhole slots on the
adapter plate.
The screw distance for the backboxes must be 60mm (2.36 in.).
2. Place the adapter (with the arrow facing up, see below) over the backbox and ensure that it is level,
and then tighten the screws.
3.2 Mo nt the NUBM3–HE onto a spacer
To attach an adapter to a spacer (NUSP1–HE):
1. Place the spacer against a flat wall and orient it so that “TOP” (printed on the inside of the spacer) is
pointing up and that the spacer is level. Use a leveling device to check, if necessary.
3TD 93505EN / 28 February 2022 / Ver. A

INSTALLATION GUIDE
Bedside Module (NUBM3-HE) Mounting
2. Using the spacer as a template, mark two holes for the screws using the outer fitting holes located in
the corners of the spacer.
3. Remove the spacer and drill screw holes for the screws that will be used, such as wood, concrete, or
drywall screws, or screws with anchors.
4. Remove the knockouts in the spacer for the cable wires.
5. Place the spacer over the holes, insert the screws, and then tighten.
6. Place the adapter over the spacer ensuring that it fits inside the spacer’s housing.
7. Insert the screws though the adapter plate and into the spacer, and then tighten the screws.
8. Pull the connection cables through the knockout holes.
TD 93505EN / 28 February 2022 / Ver. A 4

Module Connectors
INSTALLATION GUIDE
Bedside Module (NUBM3-HE)
4 Mod le Connectors
The module includes two RJ45 jacks for room bus connections, a connector for light relays, a connector for
the NUBM3X-HE extension module, and DIP switches for room bus addressing. Figure 1 shows the
connectors and DIP switches on the back of the module.
Figure 1. Module connections
Legend
1. DIP switches for room bus addressing
2. 4–pole connector for light relays
3. 2 x RJ45 connections to room bus
4. 6–pole connector to NUBM3X–HE extension module
4.1 Setting the Room B s Address
The room bus address is a unique number assigned to a room module when modules of the same type are
attached to the room bus.
If you do not use DIP switches to set the room bus address, be sure all the DIP switches are in the OFF
position on the back of the module.
To set the DIP switches:
1. Set the DIP switches for the correct room bus address.
2. Using a small screwdriver, gently slide the switch up for “ON” or slide it down for “OFF”.
5TD 93505EN / 28 February 2022 / Ver. A

INSTALLATION GUIDE
Bedside Module (NUBM3-HE) Connecting to light relays
5 Connecting to light relays
The NUBM3–HE offers two light connections: two relay circuits for switching external light sources that are
operated via the handset connected to the module.
Each relay circuit is suitable for switching an external bi-stable 24 volt DC relay. The maximum switching
current for each relay must not exceed 0.2A at max. 30VDC.
Caution: A diode (1N4004 or equivalent) must be connected across the coil of the bed light
switching relay to prevent surges caused by the relay coil from interfering with the NUBM3–
HE’s internal circuitry.
The diode(s), power supply and external relay(s) are not included and must be provided by the installer.
The following diagram shows a typical example of where to connect the diodes (5).
Figure 2. NUBM3–HE Electrical connections for two light relays
Legend
1. Module
2. Bed light
3. Room light
4. DC power for relays, max. 30Vdc
5. Protection (surge damping) diodes
TD 93505EN / 28 February 2022 / Ver. A 6

Mount the NUBM3–HE
INSTALLATION GUIDE
Bedside Module (NUBM3-HE)
6 Mo nt the NUBM3–HE
1. Pull the room buses through the adapter plate and the frame.
See 8 Terminating the Room Bus RJ45 Connectors, page 9.
2. Insert the room buses into the RJ45 connectors on the rear side of the NUBM3–HE.
3. Place the frame with the module over the adapter and insert the top of the module first into the adapter
(item 1) and then click the bottom of the module into the adapter (item 2).
4. Perform a test to verify that the LEDs function correctly.
Only in a correctly configured system will a call or presence generate an alert on display
devices such as corridor lamps.
7TD 93505EN / 28 February 2022 / Ver. A

INSTALLATION GUIDE
Bedside Module (NUBM3-HE) Removing the module
7 Removing the mod le
Removal of the NUBM3–HE from wall or ceiling.
1. Remove the module:
• Insert a small flat tipped screwdriver into the hole on the bottom of the frame (item 1).
• Push the screwdriver up until the module pops out (item 2).
Caution: Do not insert the screwdriver into the bottom corner of the frame as this may
damage the frame or module.
It may be necessary to wiggle the screwdriver a bit for the module to come forward.
2. Pull the module gently away from the base. The module is still connected to the room bus.
3. Disconnect the room bus connectors and all other optional connectors from the rear of the module.
TD 93505EN / 28 February 2022 / Ver. A 8

Terminating the Room Bus RJ45 Connectors
INSTALLATION GUIDE
Bedside Module (NUBM3-HE)
8 Terminating the Room B s RJ45 Connectors
The NUBM3–HE includes two active room bus RJ45 sockets. Crimp the RJ45 connector(s) to the room bus
cable(s) using the Ethernet T-568B termination color scheme. The following figure shows the correct pinout
for terminating the active room bus cables.
RJ45 connector for active room b s cable
For easy mounting of the RJ45 connector, use Easy RJ45 connectors that allow feeding excess wire length
through the connector. When crimping, the excess wire length will be cut automatically.
(Easy) RJ45 connectors are not included and must be ordered separately.
Easy RJ45 connectors require a dedicated crimping tool (not included) that includes a cutting
mechanism to cut the excess wire when crimping the connector on the cable.
RJ45 Connector on Cat 6/7 Cables
Caution: Cat 6/7 cables may be too stiff to handle using the normal terminating procedure,
which can result in putting to much stress onto the switch module when it is mounted onto
the backbox.
Therefore remove an appropriate portion (1) of up to 15cm (6in.) of the outer jacket before
crimping the RJ45 connector onto the Cat 6/7 cable.
Note that the individual twisting of the wire pairs should remain intact up to the RJ45
connector.
9TD 93505EN / 28 February 2022 / Ver. A

INSTALLATION GUIDE
Bedside Module (NUBM3-HE) Replacing the button insert
9 Replacing the b tton insert
1. Remove current button insert by gently pulling the small tab at the lower end of the insert with a small
pair of pliers.
2. Insert a new button insert from the lower end of the module. Make sure the insert fits completely in the
membrane.
The button insert shall be water resistant.
3. Check that all of the button LEDs are visible.
Perform a test to verify that all LEDs work properly.
TD 93505EN / 28 February 2022 / Ver. A 10

Specifications
INSTALLATION GUIDE
Bedside Module (NUBM3-HE)
10 Specifications
Wire/terminations Cat 5/5e/6/7, U/UTP
Cat 6/7 cable will work electrically but may be too stiff for
some back boxes.
Compatible electrical boxes (metal
or plastic) • EU box (or equivalent) Single backbox with mounting holes
60mm (2.36in.)
• MK Honeywell models (or equivalent) 1-gang: 861ZIC, 866ZIC,
877ZIC Note: UK backboxes require single (NUMP86-HK) or
dual (NUMPD86-HK) mounting plates.
11 TD 93505EN / 28 February 2022 / Ver. A

Ascom (Sweden) AB
Grimbodalen 2
SE–417 49 Göteborg
Sweden
Phone +46 31 55 93 00
www.ascom.com
TD 93505EN / 28 February 2022 / Ver. A
last page
Table of contents
Other ASCOM Control Unit manuals
Popular Control Unit manuals by other brands

Kitz
Kitz Y Series Operation manual

TP-Link
TP-Link kasa smart KE100 V1.20 quick start guide

STW
STW TC3G Quick Start Guide for Starting Up

Georg Fischer
Georg Fischer Signet 9900 instruction sheet

ElMod
ElMod Famo Instructions for Assembly and Handling

Cisco
Cisco WS-SVC-WISM2-1-K9 Installation and Verification Note