PLEASE READ THE CRITICAL INFORMATION BELOW:
• Ensure the user, carer and/or equipment is positioned to avoid being trapped
or crushed during bed adjustment.
• Ensure the handset cord is not stretched excessively.
• Ensure the handset and electrical leads cannot become entangled with moving
parts of the bed.
• DISCONNECT the power lead from the mains and stow before transporting the bed.
• If the power lead shows any sign of damage, replace immediately.
UNDERSTANDING THE RELEVANT HANDSET
ComfiMotion Beds are powered by five-to-six electric actuators, which are controlled by the Handset.
Adjustment occurs only while a button is pressed. Adjustment will stop when the button is released,
or the selection reaches the end of its range.
6.1 SETTING UP YOUR HANDSETS
NOTE: Handset cable is attached to control box out of factory; this guide is to assist if it becomes
knocked or removed during assembly or if a replacement is needed.
1. Disconnect power to the bed by unplugging from wall socket
2. Disconnect the battery backup from the bed by disconnecting the below plug from the control box
3. Wait 10 seconds for all power to deplete from the control box
4. Disconnect the existing handset(s) from the control box (if replacing) and attach your handset(s)
to the ports shown overleaf
NOTE: only connect handsets in the configurations shown overleaf to avoid electronic system malfunction
5. Route the handset cable(s) carefully to the side of the bed that best suits the user/usage environment
6. Reconnect the battery backup
7. Reconnect power plug to wall socket
8. Wait 10 seconds
9. Conduct the Electronic System Reset procedure (see function 13 in section 6.2 Handset Operation)
to ensure software is calibrated to the new handset
6. SETTING UP AND OPERATING THE HANDSET
WARNING
LONG SINGLE AND KING SINGLE BEDS
EITHER
STANDARD HANDSET
GSP640220
SIMPLIFIED HANDSET
SOLD SEPARATELY
BEA830214
LONG DOUBLE AND LONG QUEEN BEDS
Handset 1 Handset 2
STANDARD HANDSET
GSP640220
PARTNER HANDSET
BEA830213
SIMPLIFIED HANDSET
SOLD SEPARATELY
BEA830214
STANDARD x1 x1
OPTION 1 x1 x1
OPTION 2 x1 x1
OPTION 3 x2

comfimotion.com.au 43
STANDARD AND PARTNER HANDSETS
Press & Hold
to initiate function
Release
button to stop
USING YOUR HANDSET
Press & Hold
to initiate function
Release
button to stop
USING YOUR HANDSET
SIMPLIFIED HANDSET
(Sold separately)
1
2
4
6
8
10
12
3
5
7
9
Battery Indicator
Headrest Tilt
Legrest
(Up & Down)
Trendelenburg
(& Reverse)
Exit Height
(Raise & Lower)
Memory
Night Assist
(Underbed Lighting)
Backrest
(Forwards & Back)
Hi/Lo
(Rise & Lower)
Seating/Dining
(Seated Position)
Zero Gravity
Massage ON/OFF 11
LED Indicator 1
Backrest
(Forwards & Back) 3
Hi/Lo
(Raise & Lower) 5
Zero Gravity 7
Massage ON/OFF 9
Headrest Tilt
(Forwards & Back)
2
Legrest
(Up & Down)
4
Exit Height
(Raise & Lower)
6
Night Assist
(Underbed Lighting)
8
WARNING
Extra care should be taken
whilst operating the bed
functions to ensure hands
and fingers are free from
potential squeeze hazards.
WARNING
Extra care should be taken
whilst operating the bed
functions to ensure hands
and fingers are free from
potential squeeze hazards.
6.2 HANDSET OPERATION
1Battery Indicator (standard handset)
LED Indicator (simplified handset)
2
Headrest Tilt: Adjusts the headrest
between 65 – 85. This can help alleviate
neck muscle strain, providing optimal
support while reading or watching TV.
3
Backrest: Adjust the backrest (up to
60) providing postural support, easing
symptoms of reflux, respiratory conditions,
snoring and sleep apnoea.
4
Legrest: Adjust the legrest to promote
better circulation, relieving tired, swollen
and/or painful limbs.
5
Hi / Lo: Raise and lowers the height of the
bed. Consider lowering the bed to assist
with bed entry and exit.
NOTE: The bed will intermittently stop
during its descent to minimum height.
Keep pressing down button to continue
lowering bed.
6
Trendelenburg: (-14º) and Reverse (14º):
This position helps facilitate blood pressure
localisation as well as promote low cardiac
output. This medically prescribed position
is not recommended for those with heart
issues or those susceptible to strokes.
7
Seating/Dining Mode: Simultaneously
raises the backrest (30) and legrest (40)
to maximise comfort and relaxation.
8
Exit Height: Lowers mattress deck to flat
position and bed to a safe exit height. To
set custom exit height, ensure mattress
deck is flat and then raise/lower bed to
desired height. Hold Exit Height and
Night Assist buttons for 5 seconds to set
the current height.
9
Zero Gravity: Simultaneously raises the
backrest to 20 and legrest (also known
as knee-break) to a 35 angle. This position
helps to relax the circulatory system and
muscles of the lower back, creating a sense
of weightlessness.
10
Memory Preset: Not available on simplified
handset. Nominate and adjust to your
preferred position (x1 PRESET ONLY). To set
up your memory function, first move the
bed to desired positions. Then, hold ‘M’
and Night Assist button for 5 seconds.
11
Massage On / Off: First press turns ON
massage at low speed, second press
increases to medium and third press to
high. Fourth press turns massage OFF.
Massage will automatically turn off after
30 minutes.
12
Night Assist: Luminous underbed halo
(lighting) for safe bed entry and exit
throughout the night. Under bed light will
turn off automatically after 12 hours.
13
Electronic System Reset: Resets
electronic system in case of malfunction.
Simultaneously press backrest up and
backrest down buttons until mattress
deck goes to flat position, bed lowers to
minimum height and battery/LED indicator
flashes three times.
14
Electronic System Factory Reset: Resets
bed to default settings (including exit and
memory functions). Press Night Assist
button for 5 seconds until LED
starts flashing. While LED light flashes,
simultaneously press Zero Gravity
and legrest up buttons until mattress
deck goes to flat position, bed lowers to
minimum height and battery/LED indicator
flashes three times.