Assicontrol MACH-I User manual

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 1 of 37
MACH-I
OPERATING MANUAL
STRUMENTI &SERVIZI
per il sistema qualità
V
ia S. Silvestro, 92 21100 Varese
Tel + 39 0332 - 213045
212639- 220185
Fax +39 0332 - 822553
www.assicontrol.com
C.F. e P.I. 02436670125
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 2 of 37
TABLE OF CONTENTS
1.0 Mach-I Introduction…………….…………………………………………………….1
2.0 Ultrasonic Flaw Detector Basics ………..……………………………………………4
3.0 Mach-I Controls ……….. ………….…………………………………………………7
4.0 Operating Detail ………………………………………………………………………11
5.0 Time/Date setting, Factory Defaults ……………………………………………….. 28
6.0 Printer Interface ……………………………………………………………………...29
7.0 Power Supply …………………………………………………………………………30
8.0 Precautions ……………………………………………………………………………31
9.0 Technical Specifications ……….…….………………………………………………..32
10.0 Menu Tree …………………………………..................................................................36

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 3 of 37
1.0 Mach-I Introduction
The Mach-I is a portable, compact, lightweight and user friendly Digital Ultrasonic Flaw Detector. The
Mach-I has a COLOR LCD display which provides a wide viewing angle and allows for fast scanning
speeds. The Mach-I’s fully charged battery gives continuous operation of eight hours and charging
time is just 2 to 3 hours. A-scan trace pattern can be directly printed out on a conventional PC printer
having serial port interface or can be transferred to a PC via RS232 Serial port for storage of data.
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 4 of 37
2.0 Ultrasonic Flaw Detector Basics
The Mach-I is a single channel ultrasonic testing instrument used for the inspection of homogeneous
materials for the presence of inclusions, porosity and other discontinuities that can affect the
performance of material and components. It can also be used for thickness gauging of homogeneous
materials, requiring access from only one side of the test piece. High frequency sound (ultrasonic
sound) waves are introduced into the test material or part from a transducer/probe that is usually
coupled to the test piece by water or other suitable coupling liquid. The transducer converts electric
signals to ultrasound and vice versa. A short burst of ultrasound is introduced into test material so
some or all of the energy is reflected by discontinuities. The reflection of the ultrasound
energy is a function of the ratio between the
acoustic impedance of the discontinuity and the
base material. The greater the impedance ratio
the more sound energy will be reflected. The
principle of ultrasonic testing is shown in figure1.
It shows the ultrasonic energy generated in the
test piece and resulting instrument display.
Thickness gauging with the Mach-I
operates on the principle of the time-of-flight

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 5 of 37
measurement. This principle utilizes the precise timing of the transit time of a short burst of ultrasound
energy through a material under test. The ultrasound waves travel to the far side of the test piece and
reflect back to the transducer/probe and a measurement is obtained.
Transducers
Basically three types of transducers/probes are available for different types of applications:
1. Straight beam Probe (Normal probe)
2. TR probe (Dual crystal probe)
3. Angle beam Probe
1. Straight beam Probe: A straight beam probe introduces ultrasound normal to the test piece
surface utilizing longitudinal or compression waves. Straight beam probes are used primarily for
flaw detection and thickness gauging (ref. Figure 1.).
2. Dual probe: A dual probe contains separate transmitting
and receiving elements (as shown in Figure 2)
usually mounted on delay lines. This design
improves near surface resolution by separating
the initial pulse from the received echoes. Dual
probes are suitable for thickness gauging of pitting
and corrosion and are also better for near surface resolution.
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 6 of 37
3. Angle Beam Probe: An angle beam probe introduces ultrasound at an angle to the surface of
the test piece. In most angle beam probes, the
wave energy is mode converted from a
longitudinal wave to a shear wave by the
refraction principle . These probes are suitable
for inspection of welds. The reason
for this is its ability to position the
transducer/probe away from the weld bead
yet propagate energy into the weld zone.
Another reason to use angle beam testing on
welds is to position the sound beam more
normal to the expected discontinuities since
the flaws in welds are usually perpendicular to
the test piece (except porosity). Figure 3 shows the principle of angled beam weld testing.

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 7 of 37
3.0 Mach-I Controls
1
2
3
4
5 6 78
9
10
11
12
13
14
15
20
1716 19
18
21
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 8 of 37
1. POWER Key The Mach-I can be switched ON and OFF using this key.
2. ZOOM Key By pressing the Zoom key, A-Scan expands to full screen. Menu parameters
disappear from the LCD and only A-Scan is displayed. By pressing any other
key, A-Scan is restored to normal size and menu parameters appear again
3. FREEZE Key By pressing the Freeze key once, A-Scan is frozen. Pressing the key a second
time selects Peak Freeze mode. In the Peak Freeze mode, the Mach I holds
an echo dynamic pattern which is helpful for an angle beam inspection (in
order to locate peak signals).
4. LOCK Key By pressing the Lock key, all function keys are locked which prevents
accidental pressing of any key. To unlock the function keys, simultaneously
press the Hot Key (9) and the Lock Key (4).
5. LCD Screen Displays A-Scan Trace pattern and Menu Parameters.
6. Battery Status The display of three full cells indicates that the battery is fully charged.
As battery strength decreases, the Battery Status indicator will show a
diminishing number of cells.

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 9 of 37
7. Gain Step Display The Gain Step Display displays Gain Step value. This value can be changed
by using the dB STEP Key (10). Pressing the Db STEP Key changes the
dB step sequence in increments of 0.5, 1, 2, 6, 12 and 20.
8. Menu Keys By using these keys the Menu can be changed.
9. Hot Key When the Hot Key is pressed simultaneously with any other parameter
key, the change in value takes place in larger steps.
10. dB STEP Key Any gain step of 0.5, 1, 2, 6, 12 or 20 dB can be selected using this key.
11. GAIN UP/DN Gain can be increased or decreased using these keys in the step selected
by dB STEP key.
12. UP/DN key Respective parameters can be increased or decreased using these keys.
13. REJECT Reject function activation is indicated by this LED.
LED
14. Gate LED The Gate LED’s provide visual indication of Gate alarm preset conditions.
(LED 1 and LED 2 represent Gate 1 and Gate 2 respectively).
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 10 of 37
15. Parameter All function (parameter) values are displayed in this area of LCD.
Values
16. RX Connector Probes are connected to the Mach I receiver through this connector.
(Choice of BNC or LEMO)
17. TX Connector Probes are connected to the Mach I transmitter through this connector.
(Choice of BNC or LEMO)
18. Charge Connector Connect the charger here.
19.RS-232 Connector Serial interface connector.
20. Battery The Mach I battery is fitted in this compartment.
Compartment
21. Mode Status Indicator Mode selection SINGLE or DUAL is displayed here.

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 11 of 37
4.0 Operating Detail
The Mach-I is a user friendly, menu driven Ultrasonic Flaw Detector. To access any parameter,
various menus are available. Details of all menus and parameters are below.
First menu is the Mach I’s basic menu. In this menu, basic functions of an Ultrasonic Flaw
Detector like Range, Zero Offset, Delay and Gain functions are covered.
ZERO Calibrates artificial zero of the probe.
RANGE The required test range can be set directly in terms of inches (mm).
The range available for steel is from 0.5 in to 200 inches (10 to
5000mm). It steps through 0.5,1,2,5, 10, 15, 20, 25, 30, 50, 100, 150,
and 200 inches by pressing the Hot key or in steps of 0.2 inches (10,
25, 50, 100, 125 ,200, 300, 500, 1000, 2000, 3000, 4000, and 5000 mm
or in steps of 1 mm).
VELOCITY The required velocity can be set directly in terms of inches/us (m/sec).
By pressing the Hot Key it steps through preset values. In fine mode
mode 0.0001 inches/us (1 meter/second).
ZERO
2200
RANGE
005.00
VEL I/S
0.2330
DELAY
000.0
REF 25.0
GAIN 1
25.0 dB
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 12 of 37
DELAY Partial Range can be set using these keys. Exact shift/delay in inches
(mm) is displayed at the bottom.
GAIN STEP Gain step value can be set by using the GAIN STEP key. It adjusts dB in
steps of 0.5, 1, 2, 6, 12 or 20 dB.
GAIN Gain can be increased or decreased as per value of step by using
GAIN UP/DN keys. For setting of REF (Reference) gain refer DAC
Setting.

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 13 of 37
●Mach-I calibration procedure for 4 inch test range using single element (Normal) probe
1. Set the required range to 4 inches using RANGE UP/DN keys.
2. Couple the probe on 1 inch thickness of V1 reference block and set first echo at 1 inch
using ZERO UP/DN keys.
3. Set 4th echo at 4 inches using VELOCITY UP/DN keys, if required.
4. Repeat steps 2 and 3 until both the echoes are at their respective positions.
5. Amplitudes of the echoes can be varied using GAIN keys.
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 14 of 37
Second Menu covers GATE 1 functions.
GATE 1 Gate can be changed to PLOGIC, NLOGIC, DAC, DAC –6,
DAC-14, trigger, EXPAND or OFF.
PLOGIC When PLOGIC is selected, the Gate is triggered if an echo is present
within the gate and its height is equal to or crosses the set threshold level.
NLOGIC When NLOGIC is selected, the MONITOR is triggered if there is no
echo within the gate or the height is lower than the threshold level.
DAC If DAC is turned ON and DAC is selected and if an echo signal
equals or crosses the DAC curve then the gates trigger.
DAC-6 If DAC is turned ON 6/14 and DAC-6 is selected then if any echo signal
Crosses the DAC-6 dB curve then the gates trigger.
DAC-14 If DAC is turned ON 6/14 and DAC-14 is selected then if the echo signal
equals or crosses the DAC-14 dB curve then the gates trigger.
EXPAND When EXPAND is selected, the Scale expansion is activated and the
GATE 1
OFF
START
1.500
END
5.000
LEVEL
040 %
REF 25.0
GAIN 1
25.0 dB

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 15 of 37
GATE width is expanded to Full-screen width for a magnified view.
START The Gate starting position can be set using START UP/DN keys. Using
these keys, Gate start can be directly set it terms of inch/mm.
END Gate End position can be set using END UP/DN keys. Using these keys
Gate end can be directly set in terms of inch/mm.
LEVEL Gate Threshold Level can be directly set in terms of the percentage of
Full Scale Height (FSH).
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 16 of 37
●Set up gate to trigger when any echo crosses the 50% FSH between 1.0 in. and 4.0 inches.
1. In the Gate setting, select PLOGIC using GATE UP/DN keys.
2. Set Gate START to 1.0in. using START UP/DN keys.
3. Set Gate END to 4.0in. using END UP/DN keys.
4. Set Gate LEVEL to 50% using LEVEL UP/DN keys.

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 17 of 37
Third Menu covers GATE 2 functions.
GATE 2 Gate can be changed to PLOGIC, NLOGIC, or OFF.
PLOGIC When PLOGIC is selected, the Gate is triggered if an echo is
present within the gate and its height crosses the set threshold level.
NLOGIC When NLOGIC is selected, the Gate is triggered if there is no echo
within the gate or the height is lower than the threshold level.
START Gate starting position can be set using START UP/DN keys.
Using these keys, Gate start can be directly set in terms of in/mm.
END Gate End position can be set using END UP/DN keys. Using these keys
the Gate end can be directly set in terms of in/mm.
LEVEL Gate Threshold Level can be directly set in terms of the percentage of
the Full Scale Height (FSH).
GATE 2
OFF
START
2.000
END
5.000
LEVEL
050 %
REF 25.0
GAIN 1
25.0 dB
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 18 of 37
Fourth Menu covers Memory storage/recall and Trace printout functions.
MEMORY Memory function can be selected A SCAN or SET-UP
using MEMORY UP/DN keys.
MEM NO Desired memory location can be selected using MEM NO UP/DN keys
When changing the Memory No, If Data is stored and valid, then an
asterisk (“*”) will be displayed next to MEM NO.
ACTION Required functions like SAVE, RECALL, SEND, PRINT or DELETE,
DETAIL can be selected using ACTION UP/DN keys. When DETAIL
is selected it shows NOTE data of Stored A Scan. It can be scrolled by
pressing MEM NO UP/DN keys.
ENTER By pressing the Right side key of the ENTER, the ACTION selected is
performed. By pressing the Left side key, the NOTE Menu will be
Menu will be displayed and the Side text menu will be changed for text
editing.
MEMORY
A SCAN
MEM NO *
001
ACTION
RECALL
NOTE
ENTER
REF 25.0
GAIN 1
25.0 dB

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 19 of 37
●To SAVE current Calibration and A scan in Memory number 3.
1. Get required pattern to be stored. Press FREEZE key.
2. Select A SCAN using Memory UP/DN keys.
3. Set MEM NO to 3 using MEM NO UP/DN keys.
4. Set ACTION to SAVE using ACTION UP/DN keys.
5. Then press NOTE/ENTER Left key so Note menu will be displayed. If required, then edit it and
exit from note menu.
6. Then press NOTE/ENTER Right key to store A Scan. By pressing Right key Current
Calibration, A Scan pattern and Note DATA will be stored at memory number 3.
Note: If data is already stored at this location, then the message will be displayed on
the LCD screen as “OVER WRITE DATA” “YES - PRESS HOT KEY”,
“NO – PRESS ANY KEY”.
OPERATOR:
ABC
JOB DETAIL:
NIL
DEFECT DETAIL:
NIL
REMARK:
NIL
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 20 of 37
Fifth Menu covers Reject of unwanted signals, Set Reference, Video setting and Testing mode
selection function.
REJECT For the reject of unwanted signals. Using REJECT UP/DN keys,
the rejection level can be set as required. It can be set directly in terms of
percentage of FSH. Activation of this function is indicated by REJ LED
(13).
POWER Hi or Low Pulser power can be selected using POWER UP/DN keys.
DAMPING Probe Damping Hi/Low can be selected using DAMPING UP/DN keys.
MODE Test Mode for the use of Single or Dual Crystal probes can be selected
MODE UP/DN keys.
REJECT
000 %
POWER
LOW
DAMPING
LOW
MODE
SINGLE
REF 25.0
GAIN 1
25.0 dB

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 21 of 37
Sixth menu covers the required functions for DAC curve.
DAC DAC setting can be switched to DRAW, ON or OFF using DAC
UP/DN keys. DRAW mode, peak points of echo signals are selected.
REF Gain When DAC DRAW is selected and if gain is changed, then that gain
value is stored as a Reference gain. ON mode, the selected peak points
are digitally connected to form a DAC curve. ON 6/14 mode, -6dB DAC
and –14dB DAC will be also displayed. OFF mode, DAC curve will not
be displayed.
CURSOR The position of the Cursor can be set using CURSOR UP/DN keys.
To select any peak of the echo signal, move the cursor on the echo
signal and then press the SET UP/DN key. POINT value is automatically
automatically incremented to indicate the next point.
POINT Indicates how many points are selected to draw the DAC curve.
PRESS Press SET UP/DN keys to accept the peak points selected by the
CURSOR.
DAC
OFF
CURSOR
001
POINT
001
PRESS
ENTER
REF 25.0
GAIN 1
25.0 dB
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 22 of 37
●To digitally draw DAC curve to join four peaks of the echo signals using 1.0 inch thick
reference block and a range of 5.0 inches.
1. Set Gain so that all four peaks are visible on the LCD screen above 5% FSH.
2. Set DAC to DRAW using DAC UP/DN keys.
3. Move cursor using CURSOR UP/DN keys to cover the first echo.
4. Press SET UP/DN key to accept it and Point No. is incremented automatically.
5. Repeat above two steps for second, third and forth echo.
> Set DAC to ON using DAC UP/DN key. Now DAC Curve will be displayed on
LCD.
Note: Cursor does not need to touch the echo for selection and acceptance of the peak point.

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 23 of 37
Seventh menu covers the required functions for Thickness/Distance measurements.
MEASURE MEASURE setting can be switched to ON or OFF using the
MEASURE UP/DN keys. In ON mode it will display thickness
/distance of the echo signal which is inside the gate. In OFF mode
it will not display thickness/distance.
X-OFF Set X-OFF value to same as Angle probe index point to Front face
distance so Surface Distance displayed on LCD will be the distance
from the angle probe front face to the defect position.
ANGLE Value to be set as per Angle of the probe. If angle value is greater than
zero and Measure is set to ON then Surface and Depth of the defect are
calculated and displayed on the LCD.
THICK Thickness value should be set as per the thickness of the test piece, so
as to enable in calculation of depth/distance accounting for multiple
skips.
MEASURE
OFF
X-OFF
0.000
ANGLE
000 DEG
THICK
4.000
REF 25.0
GAIN 1
25.0 dB
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 24 of 37
●To measure Sound path, Surface and Depth distance when 45 deg. probe is used.
1. Go to GATE 1 menu and enable Gate to PLOGIC/NLOGIC.
2. Set Gate START and Gate END value so defect signal is within the GATE.
3. Set Gain or Threshold so height of the echo/defect is beyond the GATE.
4. Go to measure menu and set MEASURE to “ON”.
5. Set angle value to 45 DEG. using angle UP/DN keys.
6. Set appropriate THICK value so it can account multiple skips for DEPTH distance
calculation.

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 25 of 37
Eighth menu covers the required functions for Thickness/Distance measurements.
TRIGGER Select Measurement point to trigger by FLANK or by PEAK.
HUD When Heads Up Display is ON then Thickness/Distance will be
displayed in a large font at the top of the LCD display.
SET REF To set any A-scan pattern as a reference, achieve reference pattern,
then set SET REF ON, so reference pattern is displayed in the
background for easy comparison. A-scan Pattern From Memory
location can be recalled and set as a reference.
UNIT To set Measurement/Calibration Units. Inches or Metric can be
selected by pressing UNIT UP/DN keys.
Note: When changing UNIT, Calibration will be changed.
TRIGGER
FLANK
HUD
OFF
SET REF
OFF
UNIT
INCHES
REF 25.0
GAIN 1
25.0 dB
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 26 of 37
Ninth Menu covers miscellaneous functions, which are described below.
CLOCK Clock can be switched ON or OFF using CLOCK UP/DN keys.
If CLOCK is set to ON then date and time will be displayed at the
bottom of A-Scan.
COLOR LCD Color combination can be selected using COLOR UP/DN keys.
VIDEO Echo signals can be selected to appear as line (ENVELOPE) or solid
(FILLED) using VIDEO UP/DN keys. In ENVELOPE mode, echo
signals appear similar to an Analog Ultrasonic Flaw Detector. In
FILLED mode the echo signals look solid for better visibility.
BRIGHT Brightness of LCD can be adjusted for better visibility using the
BRIGHT UP/DN keys. Set to Low brightness indoors for longer battery
life.
CLOCK
OFF
COLOR
005
VIDEO
ENVELOP
BRIGHT
050 %
REF 25.0
GAIN 1
25.0 dB

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 27 of 37
Tenth menu covers the functions for HORN and Key Beep Setting.
HORN Horn can be switched ON or OFF using HORN UP/DN keys. When the
horn is switched ON, it produces an audible alarm each time the gate is
triggered.
BEEP Key press will be acknowledged by sound if BEEP is
selected to ON. Key press will be silent when set to OFF.
HORN
OFF
BEEP
ON
REF 25.0
GAIN 1
25.0 dB
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 28 of 37
5.0 Time/Date, Factory Defaults
5.1 Time/Date. Keep key 1 pressed and switch ON Mach-I.
After display of menu parameter, release Key 1 and again press
key 1 so Time setting menu will be displayed. Next, set the
correct time by pressing the UP/DN parameter key. Then press
Key 2 to store/set correct time to value set as per display.
Press key 3 to exit/cancel time setup.
5.2 Factory Default Calibration set-up
In the Mach-I, the last calibration set-up remains stored.
The Mach-I will utilize the last calibration set-up at start up.
However, if the stored calibration becomes corrupted, the unit
will start with the factory default calibration set up. To recall the
factory default calibration set-up, press and hold Key 4. Next
turn on the Mach-I. After the unit displays menu parameters,
release key 4.
1
2
3

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 29 of 37
6.0 Printer Interface
The Mach-I can print a copy of A-Scan patterns with full calibration set-up values. This requires a
conventional PC printer having a Serial Communication connector. The printer should be an IBM-
PC compatible with the serial port – Serial printer EPSON FX/LQ type is recommended.
Printer settings are below:
Baud Rate 9600 Parity bit None
Data Bits 8 Stop Bits 1
To print out trace pattern, follow these steps.
●Connect Serial cable (supplied with Mach-I) between connector (20) of Mach-I and serial
printer.
●Switch ON Serial PC printer.
●Switch ON Mach-I.
●Get the required Trace pattern and freeze using Freeze key (3).
●Using Menu key go to memory menu (6).
●Using ACTION UP/DN key select “PRINT”.
●Press ENTER UP/DN key to initiate printing.
For Interface with PC Using Serial port refer to Help.txt file supplied in Software CD.
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 30 of 37
7.0 Power Supply
The Mach-I is powered by a Lithium-Ion battery pack (Mach001) which is fitted in Mach-I using
two screws. It electrically connects to Mach-I via two internal terminal contact points. With fully
charged battery pack, Mach-I can operate continuously for eight hours under typical operating
conditions. The battery pack can be charged when inside Mach-I by connecting Battery charger
Mach002 to the connector (19). Alternatively, the battery pack can be charged outside Mach-I
which allows uninterrupted use of Mach-I using multiple battery packs.
Battery Charging
The battery Charger (Mach002) supplied with Mach-I is a special quick-charge charger. The
charging time is 2.5 hours. The Mach002 battery charger is suitable for use with mains supply from
100 to 240 volts AC/50/60 Hz. The LED on charger illuminates to indicate power on. When the
battery voltage is low (discharged condition) the LED glows RED. When the battery is almost
charged, the LED turns to YELLOW. When the battery is fully charged, the LED glows GREEN.
The output voltage from the charger is 16.4 Volts DC. The charging current is constant at 2.0
Amps.

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 31 of 37
8.0 Precautions
●Avoid shock to the Mach-I and probes.
●Avoid High-Energy electrical static sources (welding equipment) which can cause erratic
operational behavior in sensitive digital equipment.
●Wipe couplant off probes and Mach-I after each use.
●When battery pack is out-of Mach-I, keep battery contacts away from metal objects.
●Never heat the battery pack (Mach001) nor throw into a fire.
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 32 of 37
9.0 Technical Specifications
Test Range : 0.500 in to 200 in (10 mm to 5 meter) (in steel). In Hot Key mode
it has 13 preset values. In fine mode it is adjustable in steps of
0.2 in (1mm).
Velocity : 0.0400 in/µs to 0.4000 in/µs (1000 m/sec to 9999 m/sec). In Hot
Key mode it has 7 preset values. In Fine Mode it is adjustable in
steps of 0.0001 in/µs (1 met/sec)
Delay : Variable from 0 to 100 in ( 0 to 3000 mm). In Hot Key mode 0.2
in (5 mm). In Fine mode 0.002 in (0.5mm).
Gain : 100dB calibrated gain adjustable in 0.5, 1, 2, 6, 12 or 20 dB steps.
Rejection : 0 to 80%FSH with LED indicator.
Rectification : Full-wave rectified with filtering.
Frequency : Broad Band amplifier 0.5 MHz. to 10MHz.
Test Modes : Pulse echo and Transmit/Receiver.
Pulser : 350 Volt Negative Spike with Hi/Low power selection.
Damping : Probe Damping Hi/Low selection.
Connectors : BNC and LEMO (Size 1) both are provided.

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 33 of 37
Freeze/Peak Freeze : A Scan freeze and Peak Freeze available. In Peak freeze it holds
echo dynamic pattern which is useful for angle beam probe to
locate peak signal.
Gates : Dual gates adjustable in 1% of Screen width with
Positive/Negative Logic, Gate Expand modes.
Gate Expand : Expands Range to width of the gate.
A-Scan memory : 200 Trace Patterns can be stored, recalled, printed or transferred to
PC via RS-232 serial port.
Calibration Set-up : 50 different calibration set-ups can be Stored and Recalled.
Software : Using “NewSoft TFT” software A-Scan with calibration set-ups
can be transferred to PC via RS-232 serial link for storage in PC
or printing.
Printer Attachment : IBM compatible Printer having serial port can be directly
attached to main unit for printing of stored A-Scan waveform
with calibration data.
Display : High brightness Color TFT LCD Display. Display area 320 x 240
pixel (122 x 92 mm). Eleven different color scheme options.
Full Screen : By pressing Zoom key A-Scan can be displayed in Full Screen
area.
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 34 of 37
Reference A-Scan : Reference A-Scan pattern of standard test object can be
saved and recalled in the background for easy comparison during
testing.
DAC : DAC curve can be entered using minimum 2 to maximum 10
points and then digitally drawn with –6dB and –14dB DAC
curves. DAC curve can be set as flaw monitor gate.
Digital Readout : Thickness/Depth can be displayed in digital readout when
using a normal probe. Beam path, Surface Distance and
Depth are directly displayed when angle probe is in use.
Update Rate : 50 Hz.
Power : Lithium-Ion Battery pack 14.4 VDC, 4AH, gives 8 hours
continuous operation from fully charged battery.
Charger : Input voltage 100 to 240 VAC. Charge status indicators provided.
Temperature : 32º 131º F (0 to 55º C).
Size : 5.98 in H x 10.04 in W x 2.76 in D (152 x 255 x 70 mm).
Weight : 5.07 lbs (2.3 kg.) with Battery.
--27--

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 35 of 37
Menu Tree
First Menu ZERO RANGE VEL DELAY
Second Menu GATE 1 START END LEVEL
OFF
PLOGIC
NLOGIC
DAC
DAC –6
DAC –14
EXPAND
Third Menu GATE 2 START END LEVEL
OFF
PLOGIC
NLOGIC
Forth Menu MEMORY MEM NO ACTION NOTE
SET-UP RECALL
SAVE
DELETE
A SCAN RECALL
SAVE
Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 36 of 37
SEND
PRINT
DELETE
DETAIL
Fifth Menu REJECT POWER DAMPING MODE
LOW LOW SINGLE
HIGH HIGH DUAL
Sixth Menu DAC CURSOR POINT PRESS
OFF ENTER
DRAW
ON
ON 6/14
Seventh Menu MEASURE X-OFF ANGLE THICK
OFF
ON
Eighth Menu TRIGGER HUD SET REF UNIT
FLANK SND PTH OFF INCHES
PEAK DEPTH ON METRIC
SURFD
ALL
Ninth Menu CLOCK COLOR VIDEO BRIGHT

Mach-I OPERATING MANUAL Rev. 2.0 1/12/06 PAGE 37 of 37
OFF ENVELOP
ON FILLED
Tenth Menu HORN BEEP
OFF OFF
ON ON
Table of contents
Other Assicontrol Measuring Instrument manuals