AST ADL 280 RFID User manual

Subject to technical changes 11/2016 Seite 1 von 12
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
Data Logger
ADL 280 RFID
Original
Operating Manual
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
Marschnerstraße 26 01307 Dresden
Telefon (03 51) 44 55 30 Telefax (03 51) 44 55 555
www.ast.de

Manual ADL 280 RFID
Seite 2 von 12 Subject to technical changes 11/2016
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
List of Contents
1. Safety instructions ......................................................................................................3
1.1. Intended use..................................................................................................................3
1.2. Maintenance instructions...............................................................................................3
2. Method of functioning.................................................................................................3
3. Specifications ..............................................................................................................4
3.1. Technical data ...............................................................................................................4
3.2. Dimensions/ connections...............................................................................................4
4. Operation oft the appliance ADL 280 RFID................................................................4
4.1. Power supply.................................................................................................................4
4.2. Operating controls .........................................................................................................5
4.3. Implementation of measurement procedure ..................................................................7
4.4. Display of measurement results ....................................................................................7
4.5. Switching off the appliance............................................................................................7
5. Operation of software „xks280“ .................................................................................7
5.1. Installation of software „xks280“ ....................................................................................7
5.2. Reading out data from Data Logger „ADL 280 RFID“....................................................7
5.3. Configuration of appliance.............................................................................................9
5.4. Additional functions of Data Logger.............................................................................11
5.5. Functions of transponder.............................................................................................11
6. Troubleshooting assistance .....................................................................................12

Manual ADL 280 RFID
Subject to technical changes 11/2016 Seite 3 von 12
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
1. Safety instructions
This operating manual describes how to put the Data Logger “ADL 280 RFID” into service and
how to operate the appliance.
The appliance contains a lithium polymer rechargeable battery, for which reason it must not be
exposed to temperatures exceeding 60°C or directly to fire.
CAUTION! Failure to observe safety instructions may lead to material damage or
personal injury occurring.
1.1. Intended use
The Data Logger “ADL 280 RFID” is intended to be used for implementing a quasi-static force
measurement process by means of force sensors equipped with an A.S.T. RFID transponder.
Any other use above and beyond this is deemed to be not for the intended purpose. The
manufacturer accepts no liability for resulting damage due to improper use. All such risk is
borne by the user alone.
1.2. Maintenance instructions
The appliance does not contain any parts requiring maintenance. The lithium polymer
rechargeable battery used on the appliance is a durable wearing part designed to give long
service. It may only be replaced by the manufacturer. The same applies to all other repairs
which may only be carried out by the manufacturer.
2. Method of functioning
The Data Logger “ADL 280 RFID” is a reading device designed for use in conjunction with an
A.S.T. RFID transponder. Power is supplied to the transponder by means of an alternating
magnetic field with a frequency of 125 kHz. The transmission of commands to the transponder
takes place via amplitude modulation of the alternating magnetic field, response being
accomplished by load modulation. Each transponder possesses a unique 48-bit identity
number. The following are also stored in the transponder: Name of the measuring point (11
characters), measurement parameters, nominal value and measurement unit of the force
sensor.
After power has been supplied to the transponder, transmission takes place of all parameters
along with the raw measurement value. From this information the Data Logger calculates the
current measured value and stores the result in an EEPROM. A dataset comprises the
following:
•Sequential number
•Name of measuring point
•Measured value (sign, numerical value and unit of measurement)
•Identification number of the transponder
•Date and time of the measurement
All recorded datasets can be displayed on the appliance but not manipulated (i.e. altered or
deleted). Reading out of all datasets on a PC is possible via the USB connector in conjunction
with the supplied software “xks280”. In addition, this software package facilitates deletion of the
entire list on the appliance as well as the setting of dates and times

Manual ADL 280 RFID
Seite 4 von 12 Subject to technical changes 11/2016
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
3. Specifications
3.1. Technical data
Number of measurements in memory
storage
500
Clock rate deviation
< 1 minute/month
Reading range
up to 20 mm
Power supply
- integral lithium polymer rechargeable battery 3.7 V, 2Ah
- USB connector
Battery life
- sufficient for 2,500 measurements
Max. current consumption
- at USB port 105 mA
- on battery charger 500 mA
Max. charging time
- at USB port only partial charging within 6 hours
- on battery charger 6 hours
Ambient conditions
-20° C to + 60° C
3.2. Dimensions/ connections
Housing material: Polyamide
Dimensions:
•Width: 115 mm
•Length: 206 mm (display 85 mm)
•Height: 85 mm
Connection: Micro USB 2.0 Type B (socket)
4. Operation oft the appliance ADL 280 RFID
4.1. Power supply
Power is supplied to the Data Logger by means of an integral rechargeable battery. If
connected up to the USB port of a PC or a USB charger, power is supplied via the USB. This
ensures immediate operating availability even if the battery is flat while also facilitating
charging of the integral battery.
The charging level of the integral battery can be gauged via the symbol appearing on the
bottom left-hand side of the display. As soon as the battery is almost flat, an alert message will
be displayed.
A.S.T. GmbH
ADL 280 RFID
-- Batterie leer --
28.01.15 17:32:58
Figure 1 – Alert message on start screen indicating that the battery is flat
Any further discharging of the battery will result in the appliance automatically switching itself
off although the integral clock will continue to function for the time being.

Manual ADL 280 RFID
Subject to technical changes 11/2016 Seite 5 von 12
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
4.2. Operating controls
Three pushbutton controls are provided for operating the Data Logger along with a display and
an audible signal tone.
Figure 2: Operating controls and display functions on the Data Logger
The pushbutton controls trigger the following functions:
Pushbutton
Action
Function
Press briefly
(< 5 sec.)
Previous dataset
Press and hold
(> 5 sec.)
First dataset
(beginning of list)
Press briefly
(< 5 sec.)
•Switches on the appliance
•Performs measurement
(pressing a second time)
Press and hold
(> 5 sec.)
Switches off the appliance
Press briefly
(< 5 sec.)
Next dataset
Press and hold
(> 5 sec.)
Last dataset
(end of list)
LEDs
Display
Pushbutton
controls

Manual ADL 280 RFID
Seite 6 von 12 Subject to technical changes 11/2016
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
These various statuses are signalled by the LEDs as follows:
LED
Meaning
Integral battery being charged
Magnetic field switched on
Measurement successful
(simultaneously a brief audible signal tone will sound)
After the Data Logger has been switched on, the following messages will appear in the display:
Figure 3: Display after appliance has been switched on
If measurement has proved successful, the display will contain the following information:
Figure 4: Anzeige eines Messdatensatzes
A.S.T. GmbH
ADL 280 RFID
frei: 495 / 500
21.01.15 17:32:58
Manufacturer
Designation of appliance
Enough storage space remaining for dataset
Clock time format
“Hours : minutes : seconds”
“
Date format: “Day ,month , year”
Battery display with symbols:
Total storage space available
Power supply via USB
Battery topped up … Battery flat
4/005: KAU 100t
+50.1 t
Id: 920000000102
21.01.15 17:32:58
Number of current dataset
Total quantity of available datasets
Name of measuring point (11 characters)
Result of measurement (sign, value, unit of measurement)
Identification number of measuring point
Time of measurement
Date of measurement
Current battery display

Manual ADL 280 RFID
Subject to technical changes 11/2016 Seite 7 von 12
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
4.3. Implementation of measurement procedure
Pressing the button briefly will switch on the Data Logger. Pressing the same button a
second time will switch on the magnetic field for 10 seconds. Within this period the Data
Logger must be held against the transponder. A successful measurement is indicated by the
green LED and a short audible signal tone. In addition, the appliance will switch over to display
the measured value. In the event of no measurement proving successful within the 10 second
period, the start screen will be displayed and the magnetic field will be switched off. A new
measurement is possible by briefly pressing the button once more.
4.4. Display of measurement results
Following successful measurement, the appliance will automatically switch over to the
displayed list of the measurement results. Manual changeover to this display is possible
following switch-on by briefly pressing or holding the or buttons. (Figure 2) The
display will then reveal a single measurement result (Figure 4).
4.5. Switching off the appliance
The appliance is switched off by pressing and holding down the key for longer than 5
seconds or automatically after approximately 2 minutes inactivity.
5. Operation of software „xks280“
5.1. Installation of software „xks280“
The software package “xks280” calls for Microsoft “NET Framework 4” (Client Profile). If this is
available on the PC, the software can be run directly. For German-user menu texts, a
German-language Windows version must be available in addition to the file
“xks280.resources.dll” in the subdirectory “DE”. If these facilities are not available, all menus
will appear in English by default.
If the “NET Framework” is not installed on the PC, use can be made of the
“Setup_xks280.msi” installer included in the scope of delivery. As a first step this will initiate
installation of the Framework, the actual program installation then taking place in a second run
as follows:
•Copying of the software into a user-defined subdirectory.
•Creation of a link on the desktop.
•Entry in “Start“ -> all programs” -> “AST ”
For complete removal of the installed software, the installer needs to be run again.
5.2. Reading out data from Data Logger „ADL 280 RFID“
Data from the Data Logger can be transferred to a PC by performing the following steps:
1. Starting up of software package “xks280“.

Manual ADL 280 RFID
Seite 8 von 12 Subject to technical changes 11/2016
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
Figure 5: Software “xks280“: Continuation of an existing list
2. Connection of the Data Logger “ADL 280 RFID” to a USB port on the PC. Data readout
takes place automatically.
Figure 6: Software “xks280“: Display of list showing all datasets
The internal list can be sorted according to the content of the column by clicking onto
the column header. The columns are modifiable according to their order by clicking onto
them with the mouse and dragging them as required.
3. Saving the data:
a. Export to “Excel”. Save in “Excel” program
b. as “comma-separated values (CSV)” file:
Number of datasets stored
on the appliance
List of datasets

Manual ADL 280 RFID
Subject to technical changes 11/2016 Seite 9 von 12
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
Figure 7: Software “xks280“: Options for saving the list as a file
A dialog box appears, in which the drive, path and file name can be defined. The preset default
is a file name in the format
“Report_Date(year)(month)(day)_Time(hour)(minute)(second).csv”
Afterwards the list of datasets can be completely emptied from the Data Logger ADL 280 RFID
by actuating the command button “Delete from appliance”.
CAUTION!
The data should first be saved before pressing the command button
“Delete from appliance”. Data deleted from the ADL 280 RFID are irretrievably lost!
5.3. Configuration of appliance
The appliance is configured by performing the following steps:
1. Starting up of software “xks280”.
2. Connection of Data Logger “ADL 280 RFID” up to USB port of PC.
3. Setting of date and time:
Figure 8: Software “xks280“: Setting the clock on the Data Logger
1. With / without header
2.
Specifying the separator
3. Saving the data as a file

Manual ADL 280 RFID
Seite 10 von 12 Subject to technical changes 11/2016
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
a. automatically by activating the command “Set clock on PC” or
b. by performing manual entry of the date and time and activating the command
“Write data in device”.
Note:
After the date and time has been successfully altered on the appliance, a
corresponding message will appear briefly within the status bar of the program
window..
4. Switchover to “Tools” window“.
Figure 9: Software “xks280“: Calling up the tools
5. Selection of language on the Data Logger:
Figure 10: Software “xks280“: Selection of language for Data Logger

Manual ADL 280 RFID
Subject to technical changes 11/2016 Seite 11 von 12
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
5.4. Additional functions of Data Logger
The following are additional functions of the Data Logger provided by the software:
•Restarting of appliance (refer to Figure 8) on bottom right-hand side of the displayed
image.
•Display of the following in the “Tools” window (refer to Figure 9):
oUSB vendor ID
oUSB product ID
oSerial number Serien-Nummer
oA.S.T.-internal hardware ID
oVersion of firmware
oMenu language of Data Logger
5.5. Functions of transponder
By means of the software with the Data Logger connected up to the PC, use can be also be
made of functions on the A.S.T. RFID transponders. Applicable preconditions for this are as
follows:
•The Data Logger must be connected up to the PC.
•The transponder must be located within the reading range of the Data Logger.
•The field on the Data Logger must be switched on (“RF Field” button).
Due to limited power consumption at the USB port, charging of the integral battery is
interrupted whenever the magnetic field is switched on.
Figure 11: Software “xks280“ Tools:
The following functions will then be available via the various pushbutton controls:
•Reading of transponder ID (tag ID).
•Reading and writing of measuring-point name (max. 11 characters).
•Performance of a measurement which is not saved as a dataset on the Data Logger.
•Setting of zero value to the current measured value.
•Restoring of factory settings on transponder.
•
A brief message will appear within the status bar of the program window denoting the success
of the last two functions.
Note: Zeroing is possible up to a maximum of 20% of the final value by actuating the
> 0 < pushbutton control.

Manual ADL 280 RFID
Seite 12 von 12 Subject to technical changes 11/2016
A.S.T. - Angewandte System Technik GmbH, Mess- und Regeltechnik
6. Troubleshooting assistance
This section deals with possible causes of problems and provides information on suggested
corrective measures:
Phenomenon Cause Remedy
Appliance cannot
be switched on Battery flat
Connect up to a USB port/battery charger. This will
ensure instant power supply while also charging the
integral battery.
Sensor not
detected or cannot
be readout
Shielding of
magnetic field
Remove metallic objects away from the magnetic
field.
Interference
affecting
magnetic field
Eliminate source of interference (e.g. additional RFID
reader) or move away from surrounding area.
Insufficient
magnetic
coupling
Vary the distance between the sensor and Data
Logger or alter the position of the Data Logger.
Appliance not
detected by the
software
Contact
problems
Check all connectors of USB cable for correct fit.
If necessary disconnect and afterwards reconnect
the cable.
Problems with
the USB port
Disconnect and afterwards reconnect the cable.
Use a different USB on the PC.
“Microsoft Excel”
fails to start up
Excel not
installed Install “Microsoft Excel” or “Office” package.
Table of contents