ATAGO PAL-83S User manual

Measurement Value
The instrument measures the KMW (or Babo) degree of grape must.
The scale indicates the ripeness of grape must for wine-making.
Automatic Temperature Compensation
The readings are corrected based on the temperature of the prism within the automatic temperature compensation range.
【Caution】
◇ Measurements may fluctuate with hot or cold samples. Wait for approximately 20 seconds to press the START button.
Measurements will stabilize once the instrument acclimates to the sample temperature.
Storage and Maintenance
Store the instrument in
a dry place away from
direct sunlight.
Exposure to humidity
and heat may damage
the instrument.
Clean and dry the sample
stage thoroughly following
the “Cleaning” instructions.
Store the unit away from
direct sunlight at a stable
temperature with as little
fluctuation as possible.
Repair and Warranty
The instrument is warranted for one year from the date of purchase. This warranty is void if the instrument shows evidence of the
following. Send the included batteries as well if they are still in use.
•Having been disassembled by unauthorized personnel
•Damages to the prism and/or sample stage
•Water damage or having been dropped
•Having been misused and/or operated outside the environmental specifications
•Leakage from batteries other than those included with the unit
Repair services are available for a fee after the warranty expires.
Contact an ATAGO authorized service center for service and support.
Please have the serial number information ready when contacting a service center.
Specifications
Measurement range KMW (or Babo) 0.0 to 40.0° Maximum number 100
10.0 to 40.0℃ of data history
esolution 0.1°/ 0.1℃ Output NFC Forum Type 4 Tag
Accuracy ±0.2°/ ±1℃ ISO/IEC 14443 Type A
Automatic temperature 10 to 40℃ Output category
compensation range Date Time, KMW(Babo) [deg], Brix [%],Temp [degC]
Ambient temperature 10 to 40℃ (e.g.) 2017/08/17 09:30:45, 16.9, 12.3, 20.4
range Power supply Two (2) AAA alkaline batteries
Sample volume At least 0.3mL Battery life Approx. 11,000 measurements
Measurement time Approx. 3 seconds (when using alkaline batteries)
Backlight The backlight stays on for International IP65
30 seconds after Protection class
any button is pressed. Dimensions and weight 55 (W) x 31 (D) x 109 (H)mm, 100g (main unit only)
The product is in conformity with the requirements of the EMC Directive 2004/108/EC.
Patent Granted in Japan United States Germany China and Taiwan.
Headquarters: The Front Tower Shiba Koen,
23rd Floor 2-6-3 Shiba-koen, Minato-ku,
Tokyo 105-0011, Japan
http://www.atago.net/
TEL: 1-425-637-2107
TEL: 91-22-2854-4915 / 4071-3232
TEL: 66-21948727-9 ,66-21171549
TEL: 55 16 3913-8400
TEL: 39 02 36557267
1709K Printed in Japan
Digital Hand-held “Pocket” Wine Refractometer
PAL-83S Cat.No.4483
4483-E05
Instruction Manual
Parts
Contents
Main unit.........1 Instruction Manual (this book) .....1 Calibration Re ort .. 1 AAA batteries ...........2
AAA alkaline batteries are included. Remove the tape from the battery compartment before inserting the batteries.
ATAGO instruments are rigorously inspected to ensure each unit meets the highest standards of quality assurance.
Introduction
Thank you for purchasing the instrument. Carefully read and follow all instructions. Keep this manual for future reference.
Safety Instructions
ead and follow all safety instructions before operating the instrument. Failure to comply with the following instructions may result in personal injury or
property damage.
◇ Ensure safety when handling hazardous materials. Observe precautionary measures and use protective equipment, Be aware of the hazards of such
chemicals and emergency response guidelines.
◇ ATAGO may not be held liable for any injury or damage arising in connection with handling of hazardous materials during the use of the instrument.
◇ Do not drop the instrument or subject it to strong physical shock.
◇ Do not attempt to repair, modify, or disassemble the instrument.
◇ Carefully read this manual to have basic knowledge of the function of each component.
◇ ATAGO is not liable for any loss and damage caused by the measurement and use of this instrument.
◇ Some acids may corrode the glass prism and/or metal sample stage, which may cause erroneous measurements.
◇ Do not use metal tools, such as a spoon, as they may scratch the prism, resulting in erroneous measurements.
◇ Do not use water above 30℃ to rinse the instrument.
◇ Only use the specified battery type. Observe proper polarities, properly aligning the anodes and cathodes.
◇ Store the instrument away from direct sunlight/heat sources and excessive amounts of dust/debris.
◇ Do not expose the instrument to a rapid change in ambient temperature.
◇ Do not subject the instrument to strong vibration.
◇ Do not subject the instrument to extreme cold temperature.
◇ Do not place the instrument under anything heavy.
◇ Loosen the battery compartment cover for air transportation.
<International Protection Classification IP6 >
◇ The instrument is water-resistant, not waterproof, and should not be submerged.
<Chemical Resistance of Body Case>
◇ The body case is made of ABS resin. Do not expose it to water vapor Some solvents may compromise the structural integrity of the instrument.
Press to take measurements and
hold down to turn off the display.
START button
(Power button)
Place and remove batteries
from here.
Battery com artment
Sam le stage
Apply water and samples on the glass prism
located in the center of the sample stage.
ZERO button
Press to perform zero-setting.
Lanyard hole
Measurement results prism
temperature remaining battery
charge etc. are displayed.
The displayed value is an example.
LCD
START button & ZERO button
Press to set date time and delete data
history.
Image is for explanation purposes only. It may
be different than the actual product purchased.

Zero-setting and Measurement
Zero-setting
【Caution】
◇ Zero-set the instrument at the beginning of each day before use as well as after replacing the batteries.
◇ Let water on the prism acclimate to the temperature of the instrument before zero-setting.
◇ When “AAA” is displayed wipe the prism clean apply water and press the ZERO button again.
3.
Measurement : not 0.0
Press the ZERO
button while the
water is on the
prism.
4. Zero-setting is
complete when “000”blinks
twice and then stops.
Measurement : 0.0
1. Wipe the prism clean and
apply about 0.3mL of
tap/distilled water.
2. Press the START
button and after “---”
appears the
measurement and the
prism temperature are
displayed.
The instrument is zero-set
correctly.
Proceed to measurement.
Wipe the prism clean with
tissue paper.
Measurement
【Caution】
◇ Do not use metal tools to apply samples on the prism as they may scratch the prism.
◇ Initial measurements may fluctuate with hot or cold samples. Wait for the instrument to acclimate to the sample temperature
approximately 20 seconds to press the START button. Alternatively press the START button multiple times until
measurements become stable.
◇ Do not splash water above 30℃. The plastic may warp which may compromise the water resistance.
•When measuring hot samples place only the necessary amount and do not let it overflow from the sample stage well.
•When hot water is necessary to clean off hardened samples use water-soaked gauze around the prism area and keep hot
water away from the body case.
◇ The displayed temperature is that of the prism and may not necessarily match the temperature of the sample.
1. Wipe the prism clean and
apply about 0.3mL of the sample.
2. Press the START
button
3. After “---”appears the
measurement and the prism
temperature are displayed.
<LCD Auto Shut-off>
The instrument will turn itself off after 2 minutes of inactivity. To manually turn it off hold down the START button for more than
2 seconds.
<For oily/fatty samples>
Try stirring the sample on the prism while measuring to improve the
repeatability of oily/fatty samples.
Cleaning
【Caution】
◇Do not scratch the prism.
◇The instrument is water-resistant not waterproof and should not be submerged.
1. Wipe off the sample.
2. Clean the prism and sample stage using a mild soap and thoroughly rinse with water.
3. Dry the area with tissues thoroughly.
Error Messages
The following messages alert the user when an operation has failed.
The battery is low. The prism temperature is below the temperature
range.
The ZERO button was pressed with nothing o
r
something other than water on the prism.
The prism temperature is above the temperature
range.
The START button was pressed with nothing or an
insufficient amount of sample on the prism.
The sample measured outside the measurement
range.
The instrument is faulty.
(Replace the batteries. Contact ATAGO if this
error persists.)
Too much light is entering the prism and the
instrument cannot measure accurately.
(Shade the sample stage with your hand and take a
measurement again.)
Replacing the Batteries
【Caution】
◇ Fasten the battery compartment cover tightly to prevent water ingress or poor connection which will
cause erroneous measurements. Push the cover in firmly and turn.
◇ When the O-ring on the cover is dirty or damaged the water resistance may be compromised. (Figure
1)
◇ When the battery icon indicates the low power level ( ) replace both batteries with a brand new
set of AAA alkaline batteries (1.5V).
◇ Static images may occasionally appear on LCD. Such retained pixel charges do not indicate a faulty
display consume the battery power or affect the instrument’s performance in any way.
◇ Check the expiration dates on batteries before purchase.
◇ Zero-set the instrument after the batteries are replaced.
1. Insert a coin in the
groove on the battery
compartment cover. Turn
the coin counterclockwise
to remove the cover.
2. Insert batteries
observing the correct
polarity.
3. Align the cover and push
it down.
4. Close the battery
compartment cover by
pushing the cover in with a
coin in the groove and turning
it clockwise until it stops.
Groove
Approx.
2mm
Approx. 0.3mL of sample
Approx. 0.3mL of water
Approx.
2mm
Figure1
O-ring
Mild
soap
This manual suits for next models
1
Other ATAGO Measuring Instrument manuals

ATAGO
ATAGO PR-50HO User manual

ATAGO
ATAGO 4440 User manual

ATAGO
ATAGO PAL-14S User manual

ATAGO
ATAGO RX-7000a Manual

ATAGO
ATAGO PAL-54S User manual

ATAGO
ATAGO PAL-Easy ACID4 User manual

ATAGO
ATAGO PAL-OWE2 User manual

ATAGO
ATAGO PAL-Easy ACID1 User manual

ATAGO
ATAGO PALETTE Series User manual

ATAGO
ATAGO PAL-mOsm. BLT/i User manual

ATAGO
ATAGO PAL-63S User manual

ATAGO
ATAGO AP-300 User manual

ATAGO
ATAGO PAL-24S User manual

ATAGO
ATAGO PAL-HIKARi 2 User manual

ATAGO
ATAGO PAL-102S User manual

ATAGO
ATAGO PAL-67S User manual

ATAGO
ATAGO PAL-3 User manual

ATAGO
ATAGO PAL-89S User manual

ATAGO
ATAGO PAL-80S User manual

ATAGO
ATAGO PAL-Easy ACID101 User manual