manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. ATAGO
  6. •
  7. Measuring Instrument
  8. •
  9. ATAGO PAL-89S User manual

ATAGO PAL-89S User manual

Measurement Value
The main scale indicates the concentration of Propylene glycol and the secondary scale indicates the freezing point of Propylene
glycol.
Freezing point changes depending on the concentration of the Propylene glycol in the solution.
Automatic Temperature Compensation
The readings are corrected, based on the temperature of the prism, within the automatic temperature compensation range.
【Caution】
◇ Measurements may fluctuate with hot or cold samples. Wait for approximately 20 seconds to press the START button.
Measurements will stabilize once the instrument acclimates to the sample temperature.
Storage and Maintenance
Store the instrument in a dry place away from
direct sunlight. Exposure to humidity and heat
may damage the instrument.
Clean and dry the sample stage thoroughly,
following the “Cleaning” instructions.
Store the unit away from direct sunlight at a
stable temperature with as little fluctuation as
possible.
Repair and Warranty
The instrument is warranted for one year from the date of purchase. This warranty is void if the instrument shows evidence of the
following. Send the included batteries as well if they are still in use.
•Having been disassembled by unauthorized personnel
•Damages to the prism and/or sample stage
•Water damage or having been dropped
•Having been misused and/or operated outside the environmental specifications
•Leakage from batteries other than those included with the unit
Repair services are available for a fee after the warranty expires.s
Contact an ATAGO authorized service center for service and support.
Please have the serial number information ready when contacting a service center.
Specifications
Measurement range Main scale : Propylene glycol 0.0 to 90.0% (V/V) Maximum number 100
Secondary scale : Freezing point 32 to -60oF of data history
Resolution Main scale : Propylene glycol 0.2% (V/V) Output NFC Forum Type 4 Tag
Secondary scale : Freezing point 1 oF ISO/IEC 14443 Type A
Output category :
Accuracy Main scale : Propylene glycol ±0.4% Date Time, Propylene glycol [mL/100mL],
Freezing point [degF], Temp [degC]
Secondary scale : Freezing point ±1 oF
Automatic temperature 50 to 104 oF (10 to 40℃) (e.g.) 2017/08/17 09:30:45, 12.4, -6, 20.4
compensation range Power supply Two (2) AAA alkaline batteries
Ambient temperature 50 to 104 oF (10 to 40℃) Battery life Approx. 11,000 measurements
range (when using alkaline batteries)
Sample volume At least 0.3mL International IP65
Measurement time Approx. 3 seconds Protection class
Backlight The backlight stays on for 30 Dimensions 55 (W) x 31 (D) x 109 (H)mm,
seconds after any button is pressed. and weight 100g (main unit only)
Patent Granted in Japan, United States, Germany, China and Taiwan.
Headquarters: The Front Tower Shiba Koen,
23rd Floor 2-6-3 Shiba-koen, Minato-ku,
Tokyo 105-0011, Japan
TEL: 81-3-3431-1943 [email protected]
http://www.atago.net/
TEL: 1-425-637-2107
[email protected]
TEL: 91-22-2854-4915 / 4071-3232
[email protected]
TEL: 66-21948727-9 ,66-21171549
[email protected]
TEL: 55 16 3913-8400
[email protected]
TEL: 39 02 36557267
[email protected]
TEL: 86-20-38108256
[email protected]
TEL: 7-812-777-96-96
[email protected]
TEL: 234-707-558-1552
[email protected]
TEL: 7-727-257-08-95
[email protected]
2108K Printed in Japan
4489-E09
Instruction Manua
“Pocket” Coolant Refractometer
PAL-89S Cat.No.4489
Parts
Contents
Main unit ......... 1 Instruction Manual (this book) ..... 1 Calibration Report .. 1 AAA batteries ........... 2
AAA alkaline batteries are included. Remove the tape from the battery compartment before inserting the batteries.
ATAGO instruments are rigorously inspected to ensure each unit meets the highest standards of quality assurance.
Introduction
Thank you for purchasing the instrument. Carefully read and follow all instructions. Keep this manual for future reference.
Safety Instructions
Read and follow all safety instructions before operating the instrument. Failure to comply with the following instructions may result in personal injury or
property damage.
◇ Ensure safety when handling hazardous materials. Observe precautionary measures and use protective equipment, Be aware of the hazards of such
chemicals and emergency response guidelines.
◇ ATAGO may not be held liable for any injury or damage arising in connection with handling of hazardous materials during the use of the instrument.
◇ Do not drop the instrument or subject it to strong physical shock.
◇ Do not attempt to repair, modify, or disassemble the instrument.
◇ Carefully read this manual to have basic knowledge of the function of each component.
◇ ATAGO is not liable for any loss and damage caused by the measurement and use of this instrument.
◇ Some acids may corrode the glass prism and/or metal sample stage, which may cause erroneous measurements.
◇ Do not use metal tools, such as a spoon, as they may scratch the prism, resulting in erroneous measurements.
◇ Do not use water above 50℃ to rinse the instrument.
◇ Only use the specified battery type. Observe proper polarities, properly aligning the anodes and cathodes.
◇ Store the instrument away from direct sunlight/heat sources and excessive amounts of dust/debris.
◇ Do not expose the instrument to a rapid change in ambient temperature.
◇ Do not subject the instrument to strong vibration.
◇ Do not subject the instrument to extreme cold temperature.
◇ Do not place the instrument under anything heavy.
◇ Loosen the battery compartment cover for air transportation.
<International Protection Classification IP65>
◇ The instrument is water-resistant, not waterproof, and should not be submerged.
<Chemical Resistance of Body Case>
◇ The body case is made of resin. Do not expose it to water vapor. Some solvents may compromise the structural integrity of the instrument.
Press to take
measurements and
hold down to turn off
the display.
START button
(Power botton)
Place and remove
batteries from here.
Battery
compartment
ZERO button
Press to perform
zero-setting.
START button &
ZERO button
Press to set date, time,
and delete data history.
Lanyard hole
Sample stage
Apply water and
samples on the glass
prism located in the
center of the sample
stage.
Image is for explanation purposes only. It may be
different than the actual product purchased.
LCD
Measurement results,
prism temperature,
remaining battery
charge, etc., are
displayed.
The displayed value
is an example.
The warranty period will
be extended from 1 year
to 2 years when you
register customer
information.
ATAGO Logger NFC
can also be downloaded at the same
time.
Trouble scanning the code? Access
this link
https://www.atago.net/ur/
index.php?l=en
QR code
The warranty period
extension method
1 year 2 years
Zero-setting and Measurement
Zero-setting
【Caution】
◇ Zero-set the instrument at the beginning of each day before use as well as after replacing the batteries.
◇ Let water on the prism acclimate to the temperature of the instrument before zero-setting.
◇ hen “AAA” is displayed, wipe the prism clean, apply water, and press the ZERO button again.
1. ipe the prism clean and
apply about 0.3mL of tap water.
2. Press the START
button, and after “---”
appears, the
measurement and the
prism temperature are
displayed.
3. 4. Zero-setting is
complete when “000” blinks
twice and then stops.
Measurement : not Brix 0.0%
Press the ZERO
button while the
water is on the
prism.
Measurement : Brix 0.0% Proceed to measurement.
Wipe the prism clean with
tissue paper.
The instrument is zero-set
correctly.
Measurement
【Caution】
◇ Do not use metal tools to apply samples on the prism as they may scratch the prism.
◇ Initial measurements may fluctuate with hot or cold samples. Wait for the instrument to acclimate to the sample temperature,
approximately 20 seconds, to press the START button. Alternatively, press the START button multiple times until
measurements become stable.
◇ Do not splash water above 50℃. The plastic may warp, which may compromise the water resistance.
•When measuring hot samples, place only the necessary amount and do not let it overflow from the sample stage well.
•When hot water is necessary to clean off hardened samples, use water-soaked gauze around the prism area and keep hot
water away from the body case.
◇ The displayed temperature is that of the prism and may not necessarily match the temperature of the sample.
1. Wipe the prism clean and
apply about 0.3mL of the sample.
2. Press the START
button
3. After “---” appears, the measurement
and the prism temperature are displayed.
<LCD Auto Shut-off>
The instrument will turn itself off after 2 minutes of inactivity. To manually turn it off, hold down the START
button for more than 2 seconds.
<For oily/fatty samples>
Try stirring the sample on the prism while measuring to improve the repeatability of oily/fatty samples.
Cleaning
【Caution】
◇ Do not scratch the prism.
◇ The instrument is water-resistant, not waterproof, and should not be submerged.
1. Wipe off the sample.
2. Clean the prism and sample stage using a mild soap and thoroughly rinse with water.
3. Dry the area with tissues thoroughly.
Error Messages
The following messages alert the user when an operation has failed.
The battery is low. The prism temperature is below the temperature
range.
The ZERO button was pressed with nothing or
something other than water on the prism.
The prism temperature is above the temperature
range.
The START button was pressed with nothing or an
insufficient amount of sample on the prism.
The instrument is faulty.
(Replace the batteries. Contact ATAGO if this
error persists.)
The sample measured outside the measurement
range.
Too much light is entering the prism, and the
instrument cannot measure accurately.
(Shade the sample stage with your hand and take a
measurement again.)
Replacing the Batteries
【Caution】
◇ Fasten the battery compartment cover tightly to prevent water ingress or poor connection, which will
cause erroneous measurements. Push the cover in firmly and turn.
◇ When the O-ring on the cover is dirty or damaged, the water resistance may be compromised.
(Figure 1)
◇ When the battery icon indicates the low power level ( ), replace both batteries with a brand new
set of AAA alkaline batteries (1.5V).
◇ Static images may occasionally appear on LCD. Such retained pixel charges do not indicate a faulty
display, consume the battery power, or affect the instrument’s performance in any way.
◇ Check the expiration dates on batteries before purchase.
◇ Zero-set the instrument after the batteries are replaced.
1. Insert a coin in the
groove on the battery
compartment cover. Turn
the coin counterclockwise
to remove the cover.
2. Insert batteries,
observing the correct
polarity.
3. Align the cover and push
it down.
4. Close the battery
compartment cover by
pushing the cover in with a
coin in the groove and turning
it clockwise until it stops.
*Turning excessively may cause
malfunction.
Approx.
2mm
Approx. 0.3mL of sample
Groove
Approx. 0.3mL of water
Approx.
2mm
Figure1
O-ring
Mild
soap

This manual suits for next models

1

Other ATAGO Measuring Instrument manuals

ATAGO PAL-3 User manual

ATAGO

ATAGO PAL-3 User manual

ATAGO PAL-Fish Fillets User manual

ATAGO

ATAGO PAL-Fish Fillets User manual

ATAGO PAL-84S User manual

ATAGO

ATAGO PAL-84S User manual

ATAGO PAL-17S User manual

ATAGO

ATAGO PAL-17S User manual

ATAGO PAL-22S User manual

ATAGO

ATAGO PAL-22S User manual

ATAGO PAL-79S User manual

ATAGO

ATAGO PAL-79S User manual

ATAGO PAL-64S User manual

ATAGO

ATAGO PAL-64S User manual

ATAGO PAL-29S User manual

ATAGO

ATAGO PAL-29S User manual

ATAGO PAL-HIKARi 5 User manual

ATAGO

ATAGO PAL-HIKARi 5 User manual

ATAGO PRM-TANK100 (FER) User manual

ATAGO

ATAGO PRM-TANK100 (FER) User manual

ATAGO PAL-Easy ACID3 User manual

ATAGO

ATAGO PAL-Easy ACID3 User manual

ATAGO PAL-Easy ACID91 User manual

ATAGO

ATAGO PAL-Easy ACID91 User manual

ATAGO PAL-06S User manual

ATAGO

ATAGO PAL-06S User manual

ATAGO PAL-55S User manual

ATAGO

ATAGO PAL-55S User manual

ATAGO PAL-14S User manual

ATAGO

ATAGO PAL-14S User manual

ATAGO DPH-2 User manual

ATAGO

ATAGO DPH-2 User manual

ATAGO PAL-30S User manual

ATAGO

ATAGO PAL-30S User manual

ATAGO PAL-03S User manual

ATAGO

ATAGO PAL-03S User manual

ATAGO PAL-pH User manual

ATAGO

ATAGO PAL-pH User manual

ATAGO PAL-83S User manual

ATAGO

ATAGO PAL-83S User manual

ATAGO PAL-50S User manual

ATAGO

ATAGO PAL-50S User manual

ATAGO PAL-39S User manual

ATAGO

ATAGO PAL-39S User manual

ATAGO PAL-60S User manual

ATAGO

ATAGO PAL-60S User manual

ATAGO PR-50HO User manual

ATAGO

ATAGO PR-50HO User manual

Popular Measuring Instrument manuals by other brands

Emerson Rosemount 8800D Series Reference manual

Emerson

Emerson Rosemount 8800D Series Reference manual

Fluidwell F014-A user manual

Fluidwell

Fluidwell F014-A user manual

Panametrics DewPro MMR31 Installation & operation manual

Panametrics

Panametrics DewPro MMR31 Installation & operation manual

Roche CoaguChek XS Pro Operator's manual

Roche

Roche CoaguChek XS Pro Operator's manual

TESTO 325-I instruction manual

TESTO

TESTO 325-I instruction manual

Teledyne 100E instruction manual

Teledyne

Teledyne 100E instruction manual

IDEAL Networks LanXPLORER Quick reference guide

IDEAL Networks

IDEAL Networks LanXPLORER Quick reference guide

Siemens Sitrans F Series operating instructions

Siemens

Siemens Sitrans F Series operating instructions

Keysight N6854A Configuration guide

Keysight

Keysight N6854A Configuration guide

MARTINDALE PSI4000 instruction manual

MARTINDALE

MARTINDALE PSI4000 instruction manual

KROHNE OPTIWAVE 6300 C Handbook

KROHNE

KROHNE OPTIWAVE 6300 C Handbook

Ankom A2000 Service Procedure

Ankom

Ankom A2000 Service Procedure

techem ultra S3 operating instructions

techem

techem ultra S3 operating instructions

Netatmo Smart Anemometer manual

Netatmo

Netatmo Smart Anemometer manual

Senquip ORB-X1-G Faq

Senquip

Senquip ORB-X1-G Faq

DRUCK & TEMPERATUR Leitenberger LR-Cal LTC 100 manual

DRUCK & TEMPERATUR Leitenberger

DRUCK & TEMPERATUR Leitenberger LR-Cal LTC 100 manual

Ambient Weather WS-3000 user manual

Ambient Weather

Ambient Weather WS-3000 user manual

Victron energy SmartShunt Quick installation guide

Victron energy

Victron energy SmartShunt Quick installation guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.