ATP Instrumentation 8820 User manual

INSTRUCTION MANUAL
MODEL 8820
Multi-Function Environment Meter
4 IN 1
SOUND LEVEL
LIGHT
HUMIDITY
TEMPERATURE
TABLET OF CONTENTS
TITLE
.
INTRODUCTION
2.
FEATURES
3.
SPECIFICATIONS
4.
PANEL DESCRIPTION
5.
OPERATING INSTRUCTION
6.
MAINTENANCE

2 1
1. INTRODUCTION
The 4 in 1 digital multi- Multi-Function Environment Meter has been
designed to combine the functions of Sound Level Meter, Light Meter,
Humidit Meter, and Temperature Meter. It is an ideal Multi-Function
Environment Meter Instrument with scores of practical applications
for professional and home use.
The Sound Level function can be used to measure noise in factories,
schools, offices, airports, home, etc., checking acoustics of studios,
auditoriums and hi-fi installations.
The Light function is used to measure illuminance in the field. It is
full cosine corrected for the angular incidence of light. The light
sensitive component used in the meter is a ver Stable, long life
silicon diode.
The Humidit /Temperature is for use a humidit /semiconductor
sensor and K t pe thermocouple. This operations manual contains
general information and specification
2. FEATURES
4 functions measure Sound level, Light, Humidit and
Temperature
3 1/2 large LCD displa with units of Lux, ℃℃℃RH and C & dB,
A & dB indication.
Eas to use
Light measuring levers ranging from 0.01 lux to 20,000 lux.
Sound level range:
A LO (low) – Weighting: 35-100 dB
A HI (High)- Weighting: 65-130 dB
C LO (low) – Weighting: 35-100 dB
C HI (High)- Weighting: 65-130 dB
Resolution: 0.1 dB
Humidit measurement from 25℃RH to 95℃RH with 0.1℃RH
resolution and fast time response.
Temperature measuring levers ranging from – 20.0℃℃℃750℃/-4
℃℃℃1400℃
3. SPECIFICATIONS
Display Large 1999 counts LCD displa with function of Lux , x10
Lux, ℃, ℃, %RH and dB, A & dB ,C & dB, Lo & dB, Hi & dB,
MAX HOLD, DATA HOLD indication.
Polarity Automatic, (-) negative polarit indication.
Over-range "OL" mark indication.
Low battery indication The "BAT" is displa ed when the batter
voltage drops below the operating level.
Measurement rate 1.5 times per second, nominal.
Storage temperature -10℃ to 60℃(14℃ to 140℃) at℃ 80℃ relative
humidit
Auto Power Off Meter automaticall shuts down after approx.10
minutes of inactivit .
Power One standard 9V, NEDA1604 or 6F22 batter .
Dimensions/Wt. 251.0 (H) x 63.8 (W) x 40 (D) mm/250g
Photo Detector Dimensions 115 X 60 X 27 mm

4 3
Sound Level
Measurement range:
A LO (low) – Weighting: 35-100 dB
A HI (High)- Weighting: 65-130 dB
C LO (low) – Weighting: 35-100 dB
C HI (High)- Weighting: 65-130 dB
Resolution: 0.1 dB
Typical instrument frequency range 30Hz-10KHz
Frequency Weighting A, C –weighting
Time Weighting Fast
Maximum Hold Deca ℃1.5dB/3 min
Accuracy ±3.5 dB at 94 dB sound level, 1KHZ sine wave.
Microphone Electric condenser microphone.
Light
Measuring Range 20, 200, 2000, 20,000lux
(20,000lux range reading x10)
Overrate Display Highest digit of "1" is displa ed.
Accuracy ±5℃ rdg ℃10 dgts (calibrated to standard incandescent
lamp at color temperature 2856k).
Repeatability ±2℃.
Temperature Characteristic ±0.1℃/℃
Photo detector One silicon photo diode with filter.
Humidity/Temperatur
e
Measurement Range
Humidit 25℃℃95℃RH
Temperature -20.0℃℃℃50.0℃ -4℃℃℃122℃
(K-type) -20.0℃℃℃200.0℃ -20℃℃℃750℃;
-4.0℃℃℃200℃, -4℃℃℃1400℃.
Resolution 0.1℃RH, 0.1℃, 1℃/ 0.1℃, 1℃.
Accuracy (after calibration)
Humidity ±5℃RH (at 25℃ , 35℃℃95℃RH)
Response time of the humidity sensor approx. 6min.
Temperature
±3%rdg±2℃(at-20.0℃℃+200.0℃)
±3.5%rdg±2℃(at-20.0℃℃+750℃)
±3%rdg±2℃(at-4.0℃℃ +200.0℃)
±3.5%rdg±2℃(at-4℃℃+1400℃)
Input Protection: 60V dc or 24V ac rms.
4. PANEL DESCRIPTION

6 5
1. LCD displa : 3 1/2 digits LCD displa with units of Lux, x10 Lux,
℃,℃,%RH, dB, A, C, Lo, Hi and low batter "BAT" MAX HOLD,
DATA HOLD indication.
2. Power Button: Selects meter's power ON or power OFF.
3. Selection Button: Selects meter's Functions and ranges.
4. MAX HOLD: If ou press the MAX button, the maximum reading
will be held. Press once again the button, will release the hold and
allow a further measurement.
5. DATA HOLD: The reading will be held when Data Hold button
Switch is pressed. If the button Switch is pressed once again, will
release the hold and allow a further measurement.
6. Function Switch: Selects measurement functions of Lux,
Temperature, Humidit and Sound Level.
7. Microphone: Electric condenser microphone inside.
8. Photo Detector: Long life silicon photo diode inside.
9. Humidit at Temperature: Humidit Sensor and Semiconductor
Sensor inside.
10.
Temperature Terminal: Insert the temperature probe in this terminal.
5. OPERATING INSTRUCTION
Measuring Sound Level
1. Turn the function Switch to "dB" position.
2. Remove the meter and face the microphone to sound source in a
horizontal position.
3. Press Select Button: Selects A & dB, C & dB, Lo & dB and Hi & dB.
4. The A, C-weighting curve is nearl uniform over the frequenc
range form 30 to 10 KHz, thus giving an indication of overall Sound
level.
5. The Fast response is suitable to measure shout bursts and peak
values form sound source.
6. The sound level will be displa ed.
7. Note: Strong wind (over 10m/sec.) striking the microphone can
cause misreading for measurement in wind locations, a windscreen
should be used in front of microphone.
Measuring Light
1. Turn the function Switch to select the "Lux"
2. Remove the detector and face the photo detector to light source in a
horizontal position.
3. Press Select Button: Selects 20, 200, 2000, 20,000 LUX ranges.
4. Read the illuminance nominal from the LCD displa .
5. Over-range: If the instrument onl displa one "1" in the M.S.D. the
input signal is too strong, and a higher range should be selected.
6. When the measurement is completed. Replace the photo detector
from the light source.
7. Spectral sensitivit characteristic: To the detector, the applied photo
diode with filters makes the spectral sensitivit characteristic almost
meet C.I.E. (International Commission on Illumination) photopia
curve V (λ) as the following chart described.

8 7
100%(Relative Sensitivit Spectral Sensitivit
80% V(λ)
60%
40%
20%
0
400 500 600 700
Wavelength (nm)
8. Recommended Illumination:
Locations
*Office
Conference, Reception room.
Clerical work
T ping drafting
*Factor
Packing work, Entrance passage
Visual work at production line
Inspection work
Electronic parts assembl line
*Hotel
Public room, Cloakroom
Reception, Cashier
*Store
Indoors Stairs Corridor
Show window, Packing table
Forefront of show window
Lux
200 ~ 750
700 ~ 1,500
1000 ~ 2,000
150 ~ 300
300 ~ 750
750 ~ 1,500
1500 ~ 3,000
100 ~ 200
200 ~ 1,000
150 ~ 200
750 ~ 1,500
1500 ~ 3,000
*Hospital
Sickroom, Warehouse
Medical Examination room
Operation room
Emergenc Treatment
*School
Auditorium, Indoor G mnasium
Class room
Laborator Librar Drafting room,
100 ~ 200
300 ~ 750
750 ~ 1,500
100 ~ 300
200 ~ 750
500 ~ 1,500
Measuring Humidity/Temperature
1.Humidit Measurement:
℃ Set the function Switch to "%RH" position.
℃ Then the displa will show the humidit reading value
(%RH) directl .
℃ When the tested environment humidit value changed. It
need to a few minutes to get the stable "%RH" reading.
Warning:
Don't expose the humidit sensor to direct sunlight.
Don't touch or manipulate the humidit sensor.
2.Temperature Measurement:
℃ Set the function Switch to "TEMP"
℃ Press Select Button: Selects "0.1℃ or 1℃ and 0.1℃ or 1℃"
range.
℃ Then the displa will show the environment temperature
reading value (℃/℃) directl .
℃ Insert the temperature probe into the K-t pe thermocouple
socket.

9
℃ Touch the end of the temperature sensor to the area or
surface of the object to be measured. The displa will show
the temperature reading value (℃/℃) directl .
Warning:
When function switch on temperature "0.1℃ or 1℃ and 0.1℃ or 1℃
"range, Never attempt a voltage measurement with the test leads
inserted into the K-t pe thermocouple socket.
You might be injured or damage the meter.
6. MAINTENANCE
Battery Replacement
If the sign "BAT" appears on the LCD displa , it indicates that the
batter should be replaced. Open the batter case and replace the
exhausted batter with new batter . (1 x 9V batter NEDA 1604, 6F22
or equivalent)
Patent Pending
Table of contents
Other ATP Instrumentation Measuring Instrument manuals

ATP Instrumentation
ATP Instrumentation LX-1309 User manual

ATP Instrumentation
ATP Instrumentation IEC61672-1 CLASS2 User manual

ATP Instrumentation
ATP Instrumentation ET-951W User manual

ATP Instrumentation
ATP Instrumentation MT-913 User manual

ATP Instrumentation
ATP Instrumentation AET-951W User manual

ATP Instrumentation
ATP Instrumentation SL-8851 User manual

ATP Instrumentation
ATP Instrumentation MT-901A User manual

ATP Instrumentation
ATP Instrumentation 1309 User manual

ATP Instrumentation
ATP Instrumentation 8928 User manual