www.madur.com Electrochemical cells calibration MaMoS gas analyser
I. GAS CALIBRATI N F ELECTR CHEMICAL CELLS
1. Turn the analyser ON and connect it usin proper connection interface to the PC computer.
2. Start the service pro ram MaMoS on the connected PC computer and communicate with the analyser.
3. Press F6 and enter password: 010939
4. Make sure that:
-the analyser is in the measurements phase (user's manual chapter 7.3.8. Work modes)
-time of the measurement cycle is lon enou h to prevent switchin the analyser durin calibration to
ventilation mode.
-analyser is suckin fresh air
5. Open Service->CPU from the top menu.