Description
The MiOXSYS System uses the Static Oxidation-Reduction Potential (sORP) Test to measure
oxidative stress in a biologic sample.
Intended Use
The MiOXSYS Test performed on the MiOXSYS Analyzer is a rapid in vitro diagnostic semen analysis
test utilizing an electrochemical technology for the qualitative measurement (millivolts mV) of static
Oxidation Reduction Potential (sORP) in human semen.
The MiOXSYS System is intended for professional use in conjunction with standard semen analysis
parameters (ejaculate volume, total sperm, sperm concentration, total motility, progressive motility,
and morphology) as an aid to assess semen quality. For in vitro diagnostic use only.
The MiOXSYS System is intended for use in hospital, reference or state laboratory settings. The
device is not intended for point-of-care use.
The MiOXSYS System is indicated for use in adult males aged 21-45 yrs old who are undergoing
semen analysis.
Warnings and Precautions
It is essential that you read the following warnings and precautions in order to avoid risks to persons
and damage to the analyzer and other equipment. Aytu BioScience does not accept responsibility
for damage or injury resulting from a lack of observance of the instructions in this manual.
Follow the User Guide!
Each time the MiOXSYS Analyzer is used,
knowledge of and attention to these
operating instructions is required. Use the
Aytu BioScience MiOXSYS Analyzer only for
the purpose it is intended.
Check that the AC voltage and frequency
printed on the AC power adapter label
match your electrical socket and whether
the shape and configuration of the plug
contacts are compatible.
Keep the Analyzer Away From Liquids!
The Aytu BioScience MiOXSYS Analyzer
is not waterproof. Fluids entering the
instrument could damage the electrical
components in the analyzer. Cleaning and
disinfecting methods commonly used on
laboratory instruments may be used. Please
see the notes on care in the MiOXSYS
Maintenance section of these instructions.
Allow the Instrument to
Reach Room Temperature!
Particularly on moving from a cold into
a warm environment (ie, after storage or
transport), condensation can form inside
and on the outside of the instrument. Wait
an appropriate time (approximately 1 hour)
before connecting to the main power supply
or switching on the instrument.
Use Only the Original Equipment!
Do not attach equipment that is not
expressly approved for use with the
Aytu BioScience MiOXSYS Analyzer.
Aytu BioScience does not recommend or
guarantee the function of the instrument
with other equipment.
Never Open the Analyzer!
There are no serviceable parts inside.
Service to any component of this device is
to be performed only by Aytu BioScience.
Unauthorized repairs or modifications
will void the warranty and may violate
conformity of the MiOXSYS Analyzer
with the regulatory requirements of the
Regulation (EU) 2017/745 of the European
Parliament and of the Council of 5 April
2017 on Medical Devices. Further notes on
maintenance can be found in the MiOXSYS
Maintenance section of these instructions.
Be Aware of the Danger of
Fatal Electric Shock!
Under no circumstances should you open
the AC power adapter. There are no
elements inside that require servicing
or maintenance.
Never use a mechanically damaged AC
power adapter — live connections might
be exposed.
The AC power adapter is not waterproof.
Therefore, never let liquids come into
contact with it. A lightly dampened cloth,
however, can be used to clean it when
disconnected. Please see the notes on care
in the MiOXSYS Maintenance section of
these instructions.
Use the AC power adapter only in a socket
that has been properly installed.
Do Not Use in Areas Where There is
A Risk Of Explosion!
This instrument is not approved for use in
areas where there is a risk of explosion.