Bante Instruments PHscan10 User manual

Bante is committed to the reduction and eventual elimination of all hazardous
substances in both the manufacturing process and finished products we supply.
We have an active manufacturing and procurement program to minimize and
eliminate the use of harmful heavy metals such as cadmium, lead, mercury and
the like.
New technologies and design parameters are also promoting these efforts and
we expect to have little or no such materials in our product in the coming years.
We welcome our customers suggestions on how to speed up these efforts.
Hazardous Substance Statement
Warranty
The warranty period for PHscan series tester is 12 months from the date of
shipment. Above warranty does not cover pH electrode module and calibration
solutions. For more information, please contact nearest authorized distributor.
BANTE INSTRUMENTS LIMITED
E-mail: [email protected]
Phone: 0086-21-6404-1598
Website: www.bante-china.com
The information in this document is subject to change without notice.
Copyright © Bante Instruments Limited, 2013. All rights reserved.
PHscan10 Pocket pH Tester
Instruction Manual
Technology to create the future...
BANTE INSTRUMENTS LIMITED
02
Introduction
Thank you for selecting the PHscan10 pocket pH tester. This manual provides
a step-by-step guide to help you operate the meter, please carefully read the
following instructions before use.
Inserting the Batteries
insert batteries into the battery compartment.
— Remove the battery compartment lid.
— Insert the batteries into the battery compartment, note polarity.
— Replace the battery compartment lid and turn clockwise until it is tight.
Keypad
PHscan10 tester includes 3 multifunction keys, names and symbols describe
the function key controls.
Take out the three G13A button cell batteries from the packaging. Follow the
steps below to
— Power the tester ON/OFF.
— Freezes the measured values on the display, press the
key again to resume measuring.
— In calibration mode, exits calibration and returns to
measurement mode.
— Press and hold the key to enter calibration mode.
— Press the key again to select the pH buffer option.
— Confirms the calibration or option.
KEY FUNCTION
MEAS/HOLD
CAL
ENTER
Meas
Hold
PHscan10
01
PHscan10
4.01 7.00 10.01
Electrode Bulb
PHscan
POCKET PH TESTER
— Remove the electrode protective cap from tester.
— If the electrode bulb dries out, soak the electrode in storage solution
or tap water for at least 15 minutes. Do not use ,
—
—
Prior to Use
module
distilled or deionized water
it will shorten the life of sensor.
Preparation of pH Standard Buffer Solutions
The tester is packaged with three pH buffer packets required for calibration.
Open the pH7.00 buffer packet, place the powder into a 250ml volumetric
flask. Pour distilled water 250ml to scale line, mix the solution until reagent
is completely dissolved.
Preparation of pH4.01 and 10.01 standard buffer solutions are the same as
above. Prepared standard buffer solutions should be stored in hermetically
sealed glass containers.
03
PHscan10 PHscan10
04
Cal Cal
Enter
Hold
Meas
— Press MEAS key to turn on the tester, the meter shows measured values.
—
— If
—
—
Press and hold the CAL key the display shows “CAL 7-4”. If
necessary, press CAL key again, the display shows “CAL 10-7”.
Choose the pH buffer option you want to use, press ENTER key to confirm.
The tester shows first calibration point, pH7.00.
4. Dip the electrode into the pH7.00 standard buffer solution, the end of the
electrode must be completely immersed into the calibration solution. Stir
the tester gently.
Power On/Off
Press and hold the MEAS key for 5 seconds, the tester will turn off.
you do not press any key for 8 minutes, the tester will automatically turn
off to conserve energy.
Calibration
The tester must be calibrated prior to first use or new electrode replaced.
To ensure accuracy, regular calibration is recommended.
Do not reuse calibration solution after calibration, contaminants in solution
will affect the calibration and eventually the accuracy of the measurement.
1. Rinse the pH electrode with distilled water, press MEAS key to turn on the
tester.
2. for 3 seconds,
3.
Enter
or
Enter
Enter
CAL
PHscan
POCKET PH TESTER
5. Press ENTER key to confirm. W
Dip the electrode into the pH4.01
or 10.01 standard buffer solution. Stir the tester gently.
7. Press ENTER key to confirm. W
ait for the measured value to stabilize, the
reading will flash 3 times indicating that first calibration point is completed.
The tester automatically shows “CAL 4” or “CAL 10” depending on the option
selected in the step 2.
6. Rinse the electrode with distilled water.
ait for the measured value to stabilize, the
reading will flash 3 times, “CAL” indicator disappears from display, the tester
returns to measurement mode. Calibration is completed.
EXIT THE CALIBRATION:
During the calibration, if you want to exit calibration mode, press MEAS key,
the tester returns to measurement mode immediately.
05
PHscan10 PHscan10
06
Hold
Meas
Slot
pH Measurement
Rinse the electrode thoroughly with distilled water,
ait for the to stabilize, record
Electrode Replacement
When the tester fails to calibrate or gives fluctuating readings for calibration
standards, you need to replace the electrode module.
1. Twist the electrode collar counter clockwise, pull the old electrode module
away from the tester.
2. Align the slot on the new electrode module, gently push the module into the
tester.
3. Twist the electrode collar clockwise until it is tight.
1. press MEAS key to turn
on the tester.
2. Dip the electrode into the sample solution, stir the tester gently.
3. W reading the measured value on the display.
Hold Function
1. During the measurement, press HOLD key,
, “HOLD” indicator appears on the display.
2. Press the key again to release the measured reading, you can continue to
take measurement.
Electrode Care and Maintenance
— For best results, please always keep the sensor’s glass bulb wet.
— Ensure that the sensor is thoroughly washed with distilled water after each
use.
— If you do not use the tester for long periods, please store the sensor with
electrode storage solution. DO NOT use de-ionized or distilled water.
Installation is completed.
the tester immediately freezes
current displayed value
Troubleshooting
— Electrode dried out.
Soak the electrode in tap water for 10 minutes.
— Measured value is out of range.
Check the electrode whether clogged, dirty or broken.
module
— Incorrect buffer used or contaminated buffer solution.
Use fresh pH buffer solutions for calibration.
— Electrode is broken.
Replace the electrode .module
LCD DISPLAY CAUSE
Optional Electrode Modules
ORDER CODE DESCRIPTION
E-PHSCAN-S
Standard pH Electrode
Used for routine pH measurements,
E.g. drinking water, rainwater, etc.
E-PHSCAN-F
Flat Surface pH Electrode
Used for measuring the skin, fruits,
meats etc.
E-PHSCAN-L
Long-Reach pH Electrode
Used for measuring the samples in
the test tube.
Specifications
Model
pH Range
Accuracy
Calibration Points
Calibration Solutions
Operating Temperature
PHscan10
0.0~14.0pH
±0.1pH
2 points
USA (pH4.01, 7.00, 10.01)
o o
0~60 C, 32~140 F
Power Requirement
Dimensions
Weight
3 1.5V “ Batteries×G13A”
185(L) 40(Dia.)mm×
100 grams

Other Bante Instruments Measuring Instrument manuals

Bante Instruments 9 Series User manual

Bante Instruments

Bante Instruments 9 Series User manual

Bante Instruments POL-200 User manual

Bante Instruments

Bante Instruments POL-200 User manual

Bante Instruments TB100 User manual

Bante Instruments

Bante Instruments TB100 User manual

Bante Instruments Bante520 User manual

Bante Instruments

Bante Instruments Bante520 User manual

Bante Instruments 320 User manual

Bante Instruments

Bante Instruments 320 User manual

Bante Instruments Bante221 User manual

Bante Instruments

Bante Instruments Bante221 User manual

Bante Instruments PHscan10 User manual

Bante Instruments

Bante Instruments PHscan10 User manual

Bante Instruments TB100 User manual

Bante Instruments

Bante Instruments TB100 User manual

Bante Instruments TB100 User manual

Bante Instruments

Bante Instruments TB100 User manual

Bante Instruments Bante810 User manual

Bante Instruments

Bante Instruments Bante810 User manual

Bante Instruments 210 User manual

Bante Instruments

Bante Instruments 210 User manual

Bante Instruments TDSscan20 User manual

Bante Instruments

Bante Instruments TDSscan20 User manual

Bante Instruments 520 User manual

Bante Instruments

Bante Instruments 520 User manual

Bante Instruments 810 User manual

Bante Instruments

Bante Instruments 810 User manual

Bante Instruments TB200 User manual

Bante Instruments

Bante Instruments TB200 User manual

Bante Instruments DDS-12DW User manual

Bante Instruments

Bante Instruments DDS-12DW User manual

Bante Instruments A Series User manual

Bante Instruments

Bante Instruments A Series User manual

Bante Instruments 904P User manual

Bante Instruments

Bante Instruments 904P User manual

Bante Instruments Bante221 User manual

Bante Instruments

Bante Instruments Bante221 User manual

Bante Instruments DDS-12DW User manual

Bante Instruments

Bante Instruments DDS-12DW User manual

Bante Instruments 820 User manual

Bante Instruments

Bante Instruments 820 User manual

Bante Instruments PHscan20 User manual

Bante Instruments

Bante Instruments PHscan20 User manual

Bante Instruments 220 User manual

Bante Instruments

Bante Instruments 220 User manual

Bante Instruments Bante900P User manual

Bante Instruments

Bante Instruments Bante900P User manual

Popular Measuring Instrument manuals by other brands

Yamaha MZ-N10 manual

Yamaha

Yamaha MZ-N10 manual

Geokon GK-604D quick start guide

Geokon

Geokon GK-604D quick start guide

SCHUNK VERO-S WDM-5X Assembly and operating manual

SCHUNK

SCHUNK VERO-S WDM-5X Assembly and operating manual

Emerson Rosemount X-Stream Series manual

Emerson

Emerson Rosemount X-Stream Series manual

TSI Instruments Alnor Velometer 6000 Series Operation and service manual

TSI Instruments

TSI Instruments Alnor Velometer 6000 Series Operation and service manual

LaserLiner SuperSquare-Laser 4 quick start guide

LaserLiner

LaserLiner SuperSquare-Laser 4 quick start guide

X-Rite RapidMatch XI user guide

X-Rite

X-Rite RapidMatch XI user guide

Bushnell 20-0002 instruction manual

Bushnell

Bushnell 20-0002 instruction manual

Amway ARTISTRY ASA-100 manual

Amway

Amway ARTISTRY ASA-100 manual

Eastron SDM120 Series manual

Eastron

Eastron SDM120 Series manual

Tektronix KEITHLEY 3765 user manual

Tektronix

Tektronix KEITHLEY 3765 user manual

Bacharach 3010 Product data

Bacharach

Bacharach 3010 Product data

Velleman DVM6013 manual

Velleman

Velleman DVM6013 manual

GMC Camille Bauer SINEAX AM2000 Device handbook

GMC

GMC Camille Bauer SINEAX AM2000 Device handbook

Extech Instruments AN340 user guide

Extech Instruments

Extech Instruments AN340 user guide

VOLTCRAFT SEM6000 operating instructions

VOLTCRAFT

VOLTCRAFT SEM6000 operating instructions

BENDIX SmartTire Quick setup guide

BENDIX

BENDIX SmartTire Quick setup guide

Kusam-meco KM 071 Operation manual

Kusam-meco

Kusam-meco KM 071 Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.