BANTE 530 User manual

Bante530 Portable Conductivity/TDS Meter
Instruction Manual
BANTE INSTRUMENTS LIMITED

Bante530 Portable Conductivity/TDS Meter
1
Introduction
Thank you for selecting the Bante530 portable conductivity/TDS meter. This manual provides a step-by-step guide to help you operate the meter, please
carefully read the following instructions before use.
UNPACKING THE METER:
Before unpacking, ensure that the current work environment meet following conditions.
●Relative humidity is less than 80%.
●Ambient temperature is greater than 0oC and less than 60oC.
●No potential electromagnetic interference.
The following list describes the standard accessories of the meter. After the unpacking, please check all accessories are complete. If any are damaged or
missing, please contact nearest distributor.
ACCESSORIES:
●Conductivity Electrode
●Temperature Probe
●Conductivity Calibration Solutions (84µS/cm, 1413µS/cm, 12.88mS/cm)
●AA Batteries
Conductivity Electrode
Temperature Probe
Conductivity Calibration Solutions

Bante530 Portable Conductivity/TDS Meter
2
Display
Bante530 portable conductivity/TDS meter is equipped with a clear and bright LCD display that used to show measured values, mode indicators and help
message. The following table describes the meaning of each indicator.
INDEX:
Measurement mode indicator:
Indicates meter is in the measurement mode
Low Battery Alarm:
When battery is depleted, the indicator will disappear
Calibration mode indicator:
Indicates meter is in the calibration mode
Stable indicator:
Indicates the measured value has stabilized
Setup mode indicator:
Indicates meter is in SETUP mode
Hold indicator:
Indicates the displayed value has been frozen
Memory mode indicator:
Indicates data is stored into memory
Calibration Due Reminder:
Prompts user to calibrate the meter regularly

Bante530 Portable Conductivity/TDS Meter
3
Keypad
The meter has a succinct membrane keypad, names and symbols describe the each function key controls.
INDEX:
KEY
DESCRIPTION
MEAS|
●Power the meter ON/OFF.
●Freezes the measured value on the display, press the key again to resume measuring.
●In the calibration or setting mode, exits current mode and returns to measurement.
MODE|oC
●Toggles between available measurement modes (Conductivity, TDS).
●Press and hold the key to enter temperature setting mode.
CAL|
●Press the key to enter the calibration mode.
●Press and hold the key to enter the setup menu.
MI|
●Press the key to store current measured value.
●Press key in setup mode to scroll up through menu.
●Press key in temperature setting mode to increase the setting value.
MR|
●Press the key to view calibration report or stored data.
●Press key in setup mode to scroll down through menu.
●Press key in temperature setting mode to decrease the setting value.
ENTER|
●Confirms the calibration, setting value or displayed option.
●Press and hold the key to turn On/Off the backlight.

Bante530 Portable Conductivity/TDS Meter
4
Connectors
Bante530 portable meter provides 2 connectors for connecting the various types of sensors. Listed in the below table are the details of these connectors.
INDEX:
NO.
CONNECTOR
FUNCTION
1
6-pin Connector
For connecting the conductivity electrode
2
Phone Jack
For connecting the temperature probe
Inserting the Batteries
Before using the meter, insert three 1.5V “AA” alkaline batteries into battery compartment, follow the procedure outlined below.
1. Remove the battery cover from meter’s backside.
2. Insert the batteries into the battery compartment, note polarity.
3. Replace the battery cover into its original position. Installation is completed.
When batteries are depleted, the meter allows you to use the USB cable connected to computer as a temporarily power supply.
2
1

Bante530 Portable Conductivity/TDS Meter
5
Connecting the Conductivity Electrode
Take out the conductivity electrode from the carrying case. Insert the 6-pin connector into corresponding connector socket. Ensure the connector is fully
seated. After connection is completed, DO NOT pull on the sensor cord. Always make sure that the connector is clean and dry.
Prior to Use
Soak the conductivity electrode for a few minutes in tap water to remove dirt and oil stains on the electrode.
Power On/Off
●Press MEAS key to turn on the meter, the display shows measured values, mode indicators and help messages.
●Press and hold the MEAS key for 3 seconds, the meter will turn off.
●If you do not press any key within the specified time period, the meter will turn off automatically.
To disable the auto-off function, please read the SETUP MENU section.

Bante530 Portable Conductivity/TDS Meter
6
Setup Menu
Bante530 portable conductivity/TDS meter contains an integrated setup menu that allows you to customize each displayed option to meet measurement
requirements.
PARAMETER
DESCRIPTION
OPTIONS
DESCRIPTION
DEFAULT
Cell Constant:
Select the cell constant of conductivity electrode.
K=0.1
K=1
●
K=10
User-Defined
Temperature Coefficient:
Select the temperature coefficient of sample.
Setting Range: 0.0 to 10.0%/oC
2.10
Calibration Points:
Select the number of calibration points.
1 point
●
2 points
3 points
4 points
5 points
Pure Water Coefficient:
Pure water coefficient will be calculated and applied
automatically for ultra pure water measurement if
enabled.
Enable
Disable
●
Normalization Temperature:
Sets the normalization temperature for conductivity
measurement and calibration.
25oC
●
20oC
TDS Factor:
Sets the default TDS conversion factor.
Setting Range: 0.40 to 1.00
0.5
Measurement Unit:
Sets the default temperature unit.
Degrees Celsius
●
Degrees Fahrenheit
Stability Criteria:
Sets the stability criteria for measurement.
When the “LO” option is enabled, measuring value
will stabilize quickly, but reading is less accurate.
When the “HI” option is enabled, measuring value
will stabilize slowly, but guarantees high accuracy.
Low
●
High
Auto-Hold:
When the auto-hold function is enabled, the meter
will automatically sense a stable end-point reading
and freeze it.
Enable
Disable
●

Bante530 Portable Conductivity/TDS Meter
7
Auto-Power Off:
When the auto-off power is enabled, if you do not
press any key within a specified time period, the
meter will automatically turn off.
10 minutes
20 minutes
30 minutes
Disable
●
Calibration Due:
When calibration due reminder is enabled, if you do
not recalibrate meter within a specified time period,
the meter will automatically show indicator.
…
1 to 31 days
Disable
●
Date and Time:
Sets the date and time of the meter.
---
---
Clear stored data:
Clear all stored data.
Enable
Disable
●
Reset:
Reset function allows user to restore the meter
back to factory default parameters. When this
function is enabled, all calibration values and
selected parameters will be lost or reset.
Enable
Disable
●
Setting the Default Parameters
1. Press and hold the key for 3 seconds, the meter enters setup menu, the display shows selectable parameter and page number.
2. Press or key to scroll through menu, select the parameter you want to set (Refer to Setup Menu section).
…………Parameter
…………Page number

Bante530 Portable Conductivity/TDS Meter
8
3. Press ENTER key to confirm, the display shows an option in the submenu.
4. Press or key to select the desired option.
5. Press ENTER key to confirm, the meter returns to measurement mode. Setting is completed.
EXIT THE SETUP MENU:
During the setup mode, press MEAS key, the meter will return to measurement mode immediately.
MI
Enter
…………Parameter
…………Option

Bante530 Portable Conductivity/TDS Meter
9
Setting the Date and Time
Bante530 portable meter has a real time clock that is used to time-stamp stored measured value and calibration data. Follow the steps below to set the
date and time during the first use.
1. Press and hold the key for 3 seconds to enter the setup menu.
2. Press or key until the display shows “DATE”option.
3. Press ENTER key to confirm, the meter shows current year.
4. Press or key to set the year.
5. Press ENTER key to confirm, the meter shows current date and time.
MI
MI
Enter
…………Year
…………Format: Month - Day
…………Format: Hour - Minutes

Bante530 Portable Conductivity/TDS Meter
10
6. Press or key to set the date and time.
7. Press ENTER key to confirm, the meter returns to measurement mode. Setting is completed.
Temperature Compensation
In order to get accurate measuring results, you need to enable the manual or automatic temperature compensation before measurement or calibration.
AUTOMATIC TEMPERATURE COMPENSATION:
●Insert the connector of temperature probe into the meter’s phone jack.
●The “ATC” indicator will show on the display, the meter is now switched to automatic temperature compensation mode.
MANUAL TEMPERATURE COMPENSATION:
1. DO NOT connect the temperature probe to meter.
2. Press and hold the oC key for 3 seconds to enter temperature setting mode.
3. Press or key to set the temperature value of sample.
4. Press ENTER key to confirm, the meter returns to measurement mode. Setting is completed.
In the temperature setting mode, press or key once, the setting value will increase or decrease by 0.1. Press and hold the or key, the
setting value will increase or decrease by 1.

Bante530 Portable Conductivity/TDS Meter
11
Conductivity Calibration
Bante530 portable conductivity/TDS meter is capable of using the three types of conductivity electrodes. Before calibration or measurement, you need to
select the different sensors as per your sample concentration. The table below shows available sensors and its effective measuring ranges.
CONDUCTIVITY ELECTRODE
MEASURING RANGES
CELL CONSTANT
CON-0.1
0.1~100µS/cm
K=0.1
CON-1
0.01~10mS/cm
K=1
CON-10
0.1~200mS/cm
K=10
Make sure that you selected the cell constant between 0.1, 1 or 10 in the setup menu. If the current option is “USER”, the automatic calibration function
will be disabled (Refer to Setup Menu section).
In the automatic calibration mode, the meter allows up to 5 points calibration. To ensure higher accuracy, we recommend that you perform 3 points
calibration or select a standard value close to the sample value you are measuring. The meter will automatically detect these conductivity standard solutions
and prompt the user to calibrate the meter. When the calibration is completed, all new calibration values will automatically override existing data.
The following table shows acceptable conductivity range of calibration solution for each measuring range.
MEASURING RANGE
CALIBRATION SOLUTION RANGE
DEFAULT
0~20µS/cm
7~17µS/cm
10µS/cm
20~200µS/cm
70~170µS/cm
84µS/cm
200~2000µS/cm
700~1700µS/cm
1413µS/cm
2~20mS/cm
7~17mS/cm
12.88mS/cm
20~200mS/cm
70~170mS/cm
111.8mS/cm
SINGLE POINT CALIBRATION:
1.1 Press MODE key until the meter shows indicator.
1.2 Rinse the conductivity electrode with distilled water, then rinse with a small amount of calibration solution.
1.3 Press CAL key, the meter enters the calibration mode.

Bante530 Portable Conductivity/TDS Meter
12
1.4 Dip the conductivity electrode into the calibration solution, the meter automatically shows current calibration standard (e.g., 1413µS/cm).
1.5 Press ENTER key to confirm, “Calibration” indicator begins flashing.
1.6 Wait for the measured value to stabilize, the display shows “END”. The meter returns to measurement mode automatically. Calibration is completed.
Enter

Bante530 Portable Conductivity/TDS Meter
13
MULTI-POINT CALIBRATION:
2.1 Make sure that you have selected 2 to 5 points calibration in setup menu.
2.2 Repeat steps 1.2 to 1.5 above, when the first calibration point is completed, the display will show “CAL2”. The meter prompts you to continue with
second point calibration.
2.3 Rinse the conductivity electrode with distilled water. Dip the electrode into the next calibration solution, the meter automatically shows current
calibration standard (e.g., 12.88mS/cm).
2.4 Press ENTER key to confirm, “Calibration” indicator begins flashing.
2.5 Wait for the measured value to stabilize, the display shows “CAL3”. The meter prompts you to continue with third point calibration.
2.6 Repeat steps 2.3 to 2.4 above until the display shows “END”. The meter returns to measurement mode automatically. Calibration is completed.
MANUAL CALIBRATION:
Bante530 portable meter provides a quick manual calibration mode that allows user to easily calibrate the meter.
3.1 Record the cell constant value on the electrode (e.g., K=1.08).
Enter

Bante530 Portable Conductivity/TDS Meter
14
3.2 Press and hold the key for 3 seconds to enter setup menu, the display shows “CELL” option.
3.3 Press ENTER key to confirm, the meter enters the cell constant setting mode.
3.4 Press or key to select “USER” option.
Enter
MI

Bante530 Portable Conductivity/TDS Meter
15
3.5 Press ENTER key to confirm, the meter enters User-Defined mode.
3.6 Press or key to set each digit according to previous record.
3.7 Press ENTER key to confirm until the setting values stop flashing. The meter returns to measurement mode automatically. Setting is completed.
●During the calibration process, if you want to exit calibration, press MEAS key, the meter will return to measurement mode immediately.
●Performing the conductivity calibration will simultaneously calibrate the corresponding TDS value.
Enter

Bante530 Portable Conductivity/TDS Meter
16
Conductivity Calibration Report
This program lets you check the calibration factor of the conductivity electrode.
1. Press MR key in the conductivity measurement mode, the display shows “LOC/P-01”.
2. Press or key until the meter shows “ELE/P-02” (Electrode Diagnosis).
3. Press ENTER key to confirm, the meter shows the last calibration date (Format: mm-dd).
4. Press key, the meter shows calibration point and calibration factor (e.g., K=0.998).
5. After the browsing, press MEAS key to exit the current mode.
MR
MR
MR
……... Calibration Point
...…... Calibration Factor

Bante530 Portable Conductivity/TDS Meter
17
Temperature Calibration
During the measurement, when automatic temperature compensation is enabled, if the temperature reading displayed differs from that of an accurate
thermometer, you need to calibrate the meter.
1. Press and hold the oC key for 3 seconds to enter temperature calibration mode, the display shows current temperature reading.
2. Press or key to set the temperature value.
3. Press ENTER key to confirm. Calibration is completed.

Bante530 Portable Conductivity/TDS Meter
18
Switching the Measurement Mode
1. In the conductivity mode, press MODE key until the display shows indicator. The meter is now enters to TDS measurement mode.
2. Press MODE key again, the meter returns to conductivity mode.
Measurement
1. Rinse the conductivity electrode thoroughly with distilled water.
2. Dip the electrode into the sample solution, stir the sensor gently.
3. Wait for the reading to stabilize, record the measured value on the display.
Hold Function
The meter contains two data hold modes. When the Auto-Hold function is enabled, the meter will automatically sense a stable endpoint reading and freeze it,
“HOLD” indicator appears on the display. If the Auto-Hold function is disabled, press key, the meter will immediately freeze currently displayed value.
Press the key again to resume measuring.

Bante530 Portable Conductivity/TDS Meter
19
Storing and Recalling Data from Memory
Bante530 portable conductivity/TDS meter allows up to 500 data sets to be stored and recalled.
MEMORY INPUT:
During the measurement process, press MI key to input measured value into the memory, “Memory” indicator appears on the display.
MEMORY RECALL:
1. Press MR key in the measurement mode, the meter shows “LOC/P-01” (Data Log).
2. Press ENTER key to confirm, the meter shows page number of the stored data.
MI
MR
Enter
..…..….Page number
Table of contents