BANTE Bante321 User manual

Bante321 Portable Ion Meter
Instruction Manual
BANTE INSTRUMENTS CO., LTD

Bante321 Portable Ion Meter
1
Introduction
Thank you for selecting the Bante321 portable ion meter. This manual provides a step-by-step guide to help you operate the meter, please carefully read the
following instructions before use.
Unpacking
Before unpacking, ensure that the current work environment meets following conditions.
•Relative humidity is less than 80%.
•Ambient temperature is greater than 0°Cand less than 60°C.
•No potential electromagnetic interference.
The following list describes the standard components of the meter. After the unpacking, please check all components are complete. If any are damaged or
missing, please contact nearest distributor.
Bante321 Ion Meter
Ion Selective Electrode
TP-10K Temperature Probe
Electrode Clip
Carrying Case

Bante321 Portable Ion Meter
2
Display
The Bante321 portable ion meter is equipped with an easy-read LCD display that used to show the measured values and mode icons. The following table
describes the function of each icon.
INDEX:
Measurement mode icon:
Indicates the meter is in the measurement mode.
Stable icon:
Indicates the measuring value has stabilized.
Calibration mode icon:
Indicates the meter is in the calibration mode.
Hold icon:
Indicates the measuring value has been locked.
Setup mode icon:
Indicates the meter is in the setting mode.
Calibration Due Alarm:
Prompts the user to calibrate the meter.
Memory icon:
Indicates the data is stored into memory.
Automatic Temperature Compensation:
Indicates the temperature compensation is enabled.
Low battery alarm:
When the battery is depleted, the icon will disappear.

Bante321 Portable Ion Meter
3
Keypad
KEY
FUNCTION
•Switches the meter ON/OFF.
•Locks the measured value, press the key again to resume measuring.
•Exits the calibration or setting and returns to measurement.
•Toggles between available measurement modes (Ion concentration and mV).
•Sets the temperature (Press and hold the key for 3 seconds).
•Starts calibration.
•Enters the setup menu (Press and hold the key for 3 seconds).
•Stores current reading to memory.
•Increase value or scroll up through the menu item.
•Views the calibration report or data logs.
•Decrease value or scroll down through the menu item.
•Confirms the calibration, settings or displayed options.
•Turn on/off the backlight (Press and hold the key for 3 seconds).
Connectors
INDEX:
NO.
CONNECTOR
DESCRIPTION
1
BNC Connector
Used for connecting the ion selective electrode
2
Phone Jack
Used for connecting the temperature probe
3
USB
Used for connecting the USB cable

Bante321 Portable Ion Meter
4
Installing the Batteries
•Remove the battery cover from backside of the meter.
•Insert three AA batteries into the battery compartment, note polarity.
•Replace the battery cover into its original position. Installation is completed.
When the batteries are depleted, the meter allows using the DC5V power adapter with USB cable for power supply. NOTE, take out the batteries.
Connecting the Sensors
•Take out the ion selective electrode from the packaging. Insert the connector into the BNC connector socket on meter, rotate and push the connector
clockwise until it locks. After the connection is completed, DO NOT pull on the cable. Always make sure that the connector is clean and dry.
•Insert the connector of temperature probe into to the corresponding connector socket. Ensure the connector is fully seated.

Bante321 Portable Ion Meter
5
Prior to Use
•Remove the protective cap from the bottom of the ion selective electrode.
•Soak the electrode in the 100ppm standard solution for at least 20 minutes.
Switching the Meter On and Off
•Press the Meas key to switch on the meter, the display shows the measured value.
•Press and hold the Meas key for 5 seconds, the meter will switch off.
To enable the Auto-Power Off feature, please refer to chapter SETUP MENU.
Setup Menu
The Bante321 portable ion meter contains an integrated setup menu that is used to customize the displayed option to meet measurement requirements. The
following table describes the functions of the menu items.
INDEX:
MENU
DESCRIPTION
OPTIONS
DESCRIPTION
DEFAULT
Measurement Unit:
Set the ion concentration and temperature units.
ppm
Parts per million
ppm, °C
mg/L
Milligrams per liter
mol/L
Moles per liter
Degrees Celsius
Degrees Fahrenheit
Calibration Points:
Set the number of calibration points.
2 points
2 points
3 points
4 points
5 points
Ionic Valency:
Set the ion valence of sample.
Monovalent
Monovalent
Divalent
Stability Criteria:
When the LO option is enabled, the Stable icon will
quickly appear on the display.
When the HI option is enabled, the icon will take
longer to appear, but guarantees high accuracy of
the measurement.
Low
Low
High

Bante321 Portable Ion Meter
6
Auto-Hold:
When the option is enabled, the meter will
automatically sense a stable reading and lock the
measurements.
Enable
Disable
Disable
Auto-Power Off:
When the option is enabled, the meter will
automatically turn off if no key is pressed within a
specified time period.
10 minutes
Disable
20 minutes
30 minutes
Disable
Calibration Due:
When the option is enabled, if the meter does not
calibrated within a specified time period, the meter
will automatically show the icon.
…
1 to 31 days
Disable
Disable
Date and Time:
Set the current date and time.
---
Year-month-day, hour-minutes
Clear Stored Data:
Delete all stored readings in the memory.
Enable
Disable
Disable
Factory Reset:
If enabled, all of the calibration data and selected
parameters will back to factory default settings,
the meter must be recalibrated.
Enable
Disable
Disable
Setting the default option
1.1 Press and hold the key for 3 seconds to enter the setup menu and the or key to select the menu item (e.g., CAL/P-02).
1.2 Press the Enter key, the display shows an option.
1.3 Press the or key to select the desired option.
1.4 Press the Enter key to confirm, the meter returns to the measurement mode. Setting is completed.
If you want to exit the setting, press the Meas key.
Menu
Page
Option
Menu

Bante321 Portable Ion Meter
7
Setting the date and time
2.1 Press and hold the key for 3 seconds to enter the setup menu and the or key until the display shows DATE/P-08.
2.2 Press the Enter key, the meter shows current year (e.g., 2018).
2.3 Press the or key to set year and the Enter key to confirm, the display shows current date and time (Format: month-day, hour-minutes).
2.4 Press the or key to set the date and time, press the Enter key to confirm until the meter returns to the measurement mode. Setting is
completed.
Year
Date
Time

Bante321 Portable Ion Meter
8
Temperature Compensation
In order to get accurate measuring results, we recommend that the standards and samples should be measured at the same temperature. If you need to
enable the temperature compensation, follow the steps below.
Automatic Temperature Compensation
Connect the temperature probe to the meter (Refer to page 4 “Connecting the Sensors”). The ATC icon immediately appears on the display, the meter is now
switched to the automatic temperature compensation mode.
Manual Temperature Compensation
If the meter does not detect a temperature probe, the °C icon will show on the display indicating that the meter is switched to the manual temperature
compensation mode. To set the temperature value of sample, follow the steps below.
1. Press and hold the oCkey for 3 seconds to enter the temperature setting mode.
2. Press the or key to modify the temperature value.
3. Press the Enter key to confirm.
Press the or key once, the setting value will increase or decrease by 0.1. Press and hold the or key, the setting value will increase or
decrease by 1.
ATC
oC

Bante321 Portable Ion Meter
9
Ion Concentration Calibration
The Bante321 portable ion meter is capable of 2 to 5 points ion calibration with standard solutions, available calibration points include the following options.
MEASUREMENT UNITS
CALIBRATION POINTS
ppm
0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000
mg/L
0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000
mol/L
0.001, 0.01, 0.1, 1, 10
mmol/L
0.001, 0.01, 0.1
In order to get accurate measuring results, we recommend that adding the ionic strength adjuster to all standards and samples. A typical addition would be
2ml ISA to 100ml of standards and samples. If the meter does not calibrated or calibration is not successfully, the display will always show 0.000.
During the calibration, ensure that the selected calibration points cover the anticipated range of the samples.
1.1 Press the Mode key until the meter shows icon.
1.2 If necessary, select the concentration unit and ion valence in the setup menu (Refer to page 5
SETUP MENU).
The meter provides three measurement units, including the ppm, mg/L and mol/L, the factory
default is ppm. If the measurement unit has converted, the display will always show “CAL”and
wait for calibrating the meter.
1.3 Press the Cal key, the meter enters the calibration mode. The display shows 0.001ppm (or
mg/L, mol/L, mmol/L).
1.4 Press the or key to select the calibration point (e.g., 100ppm). The meter will
automatically perform the calibration from the low to high concentrations.

Bante321 Portable Ion Meter
10
1.5 Rinse the ion selective electrode with distilled water, then rinse with asmall amount of
standard solution. Place the electrode into corresponding standard solution. Stir the electrode
gently to create a homogeneous solution. Press the Enter key, the Calibration icon begins
flashing.
1.6 Wait for the reading to stabilize, the display shows 1000ppm/CAL2. The meter prompts you to
continue with second point calibration.
1.7 Rinse the ion selective electrode with distilled water, then rinse with a small amount of
standard solution. Place the electrode into the next standard solution. Stir the electrode
gently. Press the Enter key, the Calibration icon begins flashing.
1.8 Wait for the reading to stabilize, the meter automatically show END and return to the
measurement mode. Calibration is completed.
•If you have selected the multi-point calibration in the setup menu, the display will show CAL3.
The meter prompts you to continue with third point calibration. Repeat the step 1.7 above
until the display shows END. The meter will automatically return to the measurement mode.
•If you want to exit the calibration, press the Meas key.

Bante321 Portable Ion Meter
11
Viewing the ion calibration report
2.1 Press the MR key in the ion measurement mode, the meter shows LOC/P-01.
2.2 Press the or key until the meter shows ELE/P-02.
2.3 Press the Enter key, the meter shows the last calibration date (Format: month-day).
2.4 Press the key, the meter shows the first calibartion point and mV value (e.g., 100ppm,
59mV).
2.5 Press the key, the meter shows the second calibartion point and mV value (e.g., 1000ppm,
118mV).
2.6 To exit the calibration report, press the Meas key.

Bante321 Portable Ion Meter
12
Temperature Calibration
During the measurement process, if the temperature reading displayed differs from that of an accurate thermometer, the meter needs to be calibrated.
1. Connect the temperature probe to the meter and place into a solution with a known accurate temperature.
2. Press and hold the oCkey for 3 seconds to enter the temperature setting mode.
3. Press the or key to set the temperature value.
4. Press the Enter key to confirm. Calibrating is completed.
Ion Concentration Measurement
Before measuring, ensure that the temperature of samples are the same as the standard solutions, the maximum error should be controlled within the 1°C.
For low level measurements or samples contain the interference ions, adding the ionic strength adjuster to sample and using the plastic beaker are
necessary.
1. Press the Mode key until the display shows icon.
2. Rinse the ion selective electrode thoroughly with distilled water. Place the electrode into the sample solution, stir the electrode gently. Record the
measured value when the reading is stable.
mV Measurement
1. Press the Mode key until the meter shows measurement unit “mV”.
2. Rinse the electrode thoroughly with distilled water. Place the electrode into the sample solution. Record the measured value when the reading is
stable.
Auto-Hold
The Bante321 portable ion meter contains an Auto-Hold function. If enabled, the meter will automatically sense a stable reading and lock the measurements,
the HOLD icon appears on the display. If disabled, press the key, the meter will immediately lock the displayed value. Press the Meas key to resume
measuring.

Bante321 Portable Ion Meter
13
Storing and Recalling Data
The meter is capable of storing and recalling up to 500 data sets.
Storing readings into memory
During the measurement process, press the MI key to store the reading into the memory, the Memory
icon appears on the display.
Viewing stored readings
1. Press the MR key in the measurement mode, the meter shows LOC/P-01 (Data Log).
2. Press the Enter key, the meter shows the serial number of the stored data.
3. Press the key, the meter shows the date and time of the stored data (Format: month-day,
hour-minutes).
4. Press the key, the meter shows the stored data.
5. Press the key again, the meter shows next data set.
6. Press the Meas key, the meter returns to the measurement mode.
Clearing the memory
Please refer to page 5 SETUP MENU.

Bante321 Portable Ion Meter
14
Communication
Bante Instruments provides a Data Acquisition System that can be used to transfer data, receive the measuring values or import the data to Excel. You are
able to download this software from our official website at www.bante-china.com. Before installation, ensure that Windows 7/8/10 operating system has
been installed on your computer.
Receiving data
•Connect the USB cable to meter and computer. Click the DAS icon, the system will automatically scan an available communication port and show the
message box “Found a port on your computer”.
•Click the OK button, the application starts.
•Click the Connect button, the screen shows “Port is connected” indicate that the communication between the meter and the computer has been
established.
•Click the OK button to confirm.
•Click the Receive button, the stored data automatically transfer to computer.
Interval recording
This function is used for recording the measuring value within the specify time period.
•Click the Interval Recording button and select a time option.
•Click the Receive button, the measured value will automatically send to data sheet.
•The first data need 1 minute to be shown on screen.
•Do not press any key on meter during the Interval Recording mode, it will cause communication interruption.
Graph mode
This function is used for viewing the variations of the measured parameter continuously.
Click the Graph button, the screen immediately shows the curve graph. To quit current mode, click the Back button.
Create the excel file
When the transfer is completed, click the Save as Excel button, the measured values in the data sheet will automatically convert to Excel file.
Once the software is closed, all received data will be lost and can not be recovered.

Bante321 Portable Ion Meter
15
Electrode Care and Maintenance
•Ensure that the ion selective electrode is thoroughly washed with distilled water after use.
•DO NOT scratch the sensitive membrane on electrode.
•If you do not use the electrode for long periods, store the electrode in a dry, cool and well-ventilated area.
Troubleshooting
LCD DISPLAY
CAUSE
CORRECTIVE ACTION
Electrode has dried out
Soak the ion selective electrode in 100ppm standard solution for 2 hours.
Measured value is out of range
Check the electrode whether clogged, dirty or broken.
Incorrect calibration solutions
Using the fresh calibration solutions for calibration.
Electrode has expired
Replace the electrode.
Keypad is not working properly
Replace the batteries.
Specifications
Ion
Model
Bante321
Range
0.001~19999ppm, mg/L, mol/L (Depending on the range of ISE)
Accuracy
± 0.5% F.S (Monovalent), ± 1% F.S (Divalent)
Resolution
0.001/0.01/0.1/1
Calibration Points
2 to 5 points
Calibration Solutions
0.001/0.01/0.1/1/10/100/1000/10000ppm, mol/L, mg/L
mV
Range
-1999.9~1999.9mV
Accuracy
± 0.2mV
Resolution
0.1/1mV, Selectable
Calibration Points
1 point (Only for relative mV mode)
Temperature
Range
0~105°C, 32~221°F
Accuracy
± 0.5°C
Resolution
0.1°C
Calibration Points
1 point
General
Temperature Compensation
0~100°C, 32~212°F, Manual or Automatic
Memory
Stores up to 500 data sets
Output
USB communication interface
Connector
BNC
Power Requirements
3 × 1.5V “AA” batteries
Dimensions
170 (L) × 85 (W) × 30 (H)mm
Weight
300g

Bante321 Portable Ion Meter
16
Addendum 1: Ion Selective Electrode Selection Guide
ORDER CODE
ION TYPE
RANGE
F-US
Fluoride (F-)
0.02ppm~Saturation
CL-US
Chloride (Cl-)
1.8~35500ppm
Br-US
Bromide (Br)
0.4~79900ppm
Cn-US
Cyanide (Cn-)
0.2~260ppm
Na-US
Sodium (Na+)
0.1~23000ppm
NO3-US
Nitrate (NO3-)
0.4~62000ppm
Ca-US
Calcium (Ca2+)
0.02~40000ppm
NH4-US
Ammonium (NH4+)
0.1~18000ppm
Cd-US
Cadmium (Cd2+)
0.01~11200ppm
Cu-US
Cupric (Cu2+)
0.006~6400ppm
I-US
Iodide (I-)
0.06~127000ppm
Pb-US
Lead (Pb2+)
0.2~20700ppm
K-US
Potassium (K+)
0.04~39000ppm
Ag-US
Silver (Ag+)
0.01~107900ppm
S-US
Sulphide (S2-)
0.003~32100ppm
NH3-US
Ammonia (NH3)
0.02~17000ppm
Addendum 2: Preparation of Ion Standard Solutions (1000ppm)
To prepare these solutions, half fill a 1 liter volumetric flask with distilled water and add the analytical grade reagent below.
ION TYPE
REAGENT
WEIGHT
Fluoride (F-)
Sodium Fluoride
2.21g
Chloride (Cl-)
Sodium Chloride
1.65g
Bromide (Br)
Sodium Bromide
1.29g
Cyanide (Cn-)
Sodium Cyanide
1.88g
Sodium (Na+)
Sodium Chloride
2.542g
Nitrate (NO3-)
Sodium Nitrate
1.37g
Calcium (Ca2+)
Calcium Chloride
3.67g
Ammonium (NH4+)
Ammonium Chloride
2.97g
Cadmium (Cd2+)
Cadmium Nitrate
2.74g
Cupric (Cu2+)
Copper Nitrate
3.80g
Iodide (I-)
Sodium Iodide
1.18g
Lead (Pb2+)
Lead Perchlorate
2.22g

Bante321 Portable Ion Meter
17
Potassium (K+)
Potassium Chloride
1.91g
Silver (Ag+)
Silver Nitrate
1.57g
Sulphide (S2-)
Sodium Sulfide
7.49g
Ammonia (NH3)
Ammonium Chloride
3.82g
Swirl the flask gently to dissolve the reagent and fill to the mark with distilled water. Cap the flask and upend several times to mix the solution.

Hazardous Substance Statement
Bante Instruments is committed to the reduction and eventual elimination of all hazardous substances in both the manufacturing process and finished
products we supply. We have an active manufacturing and procurement program to minimize and eliminate the use of harmful heavy metals such as
cadmium, lead, mercury and the like. New technologies and design parameters are also promoting these efforts and we expect to have little or no such
materials in our product in the coming years. We welcome our customer suggestions on how to speed up these efforts.
Warranty
The warranty period for meter is one year from the date of shipment. Above warranty does not cover the sensor and calibration solutions. Out of warranty
products will be repaired on a charged basis. The warranty on your meter shall not apply to defects resulting from:
•Improper or inadequate maintenance by customer.
•Unauthorized modification or misuse.
•Operation outside of the environment specifications of the products.
For more information, please contact the nearest authorized distributor.
BANTE INSTRUMENTS CO., LTD
E-mail: banteinstruments@yahoo.com
Phone: +8621-6404-1598
Website: www.bante-china.com
The information in this document is subject to change without notice. Copyright © Bante Instruments, 2018. All rights reserved.
Table of contents
Other BANTE Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

PCB Piezotronics
PCB Piezotronics 356A14 Installation and operating manual

Campbell
Campbell CS140 product manual

BC Group International
BC Group International BC Biomedical SA-2600 user manual

Flow vision
Flow vision FC01-Ex User handbook

KPS
KPS KPS-AN10 instruction manual

PCB Piezotronics
PCB Piezotronics IMI SENSORS M603M170 Installation and operating manual