E-Line 90 Refractometer
Instruction Manual
Correcting the reading for temperature:
The refractometer is calibrated to read correctly at 20˚C. The refractive index of the measured sample will vary with temperature and the
reading must be corrected to take account of this. Please use the Table 2 below to adjust your reading accordingly.
Temperature Correction Table (Table 2)
This table gives mass fraction corrections to refractometric tables for sucrose solutions at 589nm for temperature different from 20˚C. Taken
from ICUMSA, Appendix 2, SPS-3 (1998) page 8.
Temperature (°C) Measured Sucrose ˚Brix (Mass Fraction)
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85
15 -0.29 -0.30 -0.32 -0.33 -0.34 -0.35 -0.36 -0.37 -0.37 -0.38 -0.38 -0.38 -0.38 -0.38 -0.38 -0.38 -0.37 -0.37
16 -0.24 -0.25 -0.26 -0.27 -0.28 -0.28 -0.29 -0.30 -0.30 -0.30 -0.31 -0.31 -0.31 -0.31 -0.31 -0.30 -0.30 -0.30
17 -0.18 -0.19 -0.20 -0.20 -0.21 -0.21 -0.22 -0.22 -0.23 -0.23 -0.23 -0.23 -0.23 -0.23 -0.23 -0.23 -0.23 -0.22
18 -0.12 -0.13 -0.13 -0.14 -0.14 -0.14 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15 -0.15
19 -0.06 -0.06 -0.07 -0.07 -0.07 -0.07 -0.07 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.08 -0.07
20 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00
21 0.06 0.07 0.07 0.07 0.07 0.07 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.07
22 0.13 0.14 0.14 0.14 0.15 0.15 0.15 0.15 0.16 0.16 0.16 0.16 0.16 0.16 0.15 0.15 0.15 0.15
23 0.20 0.21 0.21 0.22 0.22 0.23 0.23 0.23 0.23 0.24 0.24 0.24 0.24 0.23 0.23 0.23 0.23 0.22
24 0.27 0.28 0.29 0.29 0.30 0.30 0.31 0.31 0.31 0.32 0.32 0.32 0.32 0.31 0.31 0.31 0.30 0.30
25 0.34 0.35 0.36 0.37 0.38 0.38 0.39 0.39 0.40 0.40 0.40 0.40 0.40 0.39 0.39 0.38 0.38 0.37
26 0.42 0.43 0.44 0.45 0.46 0.46 0.47 0.47 0.48 0.48 0.48 0.48 0.48 0.47 0.47 0.46 0.46 0.45
27 0.50 0.51 0.52 0.53 0.54 0.55 0.55 0.56 0.56 0.56 0.56 0.56 0.56 0.55 0.55 0.54 0.53 0.52
28 0.58 0.59 0.60 0.61 0.62 0.63 0.64 0.64 0.64 0.65 0.65 0.64 0.64 0.63 0.63 0.62 0.61 0.60
29 0.66 0.67 0.68 0.70 0.71 0.71 0.72 0.73 0.73 0.73 0.73 0.73 0.72 0.72 0.71 0.70 0.69 0.67
30 0.74 0.76 0.77 0.78 0.79 0.80 0.81 0.81 0.82 0.82 0.81 0.81 0.80 0.80 0.79 0.78 0.76 0.75
31 0.83 0.84 0.85 0.87 0.88 0.89 0.89 0.90 0.90 0.90 0.90 0.89 0.89 0.88 0.87 0.86 0.84 0.82
32 0.92 0.93 0.94 0.96 0.97 0.98 0.98 0.99 0.99 0.99 0.99 0.98 0.97 0.96 0.95 0.93 0.92 0.90
33 1.01 1.02 1.03 1.05 1.06 1.07 1.07 1.08 1.08 1.08 1.07 1.07 1.06 1.04 1.03 1.01 1.00 0.98
34 1.10 1.11 1.13 1.14 1.15 1.16 1.16 1.17 1.17 1.16 1.16 1.15 1.14 1.13 1.11 1.09 1.07 1.05
35 1.19 1.21 1.22 1.23 1.24 1.25 1.25 1.26 1.26 1.25 1.25 1.24 1.23 1.21 1.19 1.17 1.15 1.13
36 1.29 1.30 1.31 1.33 1.34 1.34 1.35 1.35 1.35 1.34 1.34 1.33 1.31 1.29 1.28 1.25 1.23 1.20
37 1.39 1.40 1.41 1.42 1.43 1.44 1.44 1.44 1.44 1.43 1.43 1.41 1.40 1.38 1.36 1.33 1.31 1.28
38 1.49 1.50 1.51 1.52 1.53 1.53 1.54 1.54 1.53 1.53 1.52 1.50 1.48 1.46 1.44 1.42 1.39 1.36
39 1.59 1.60 1.61 1.62 1.63 1.63 1.63 1.63 1.63 1.62 1.61 1.59 1.57 1.55 1.52 1.50 1.47 1.43
40 1.69 1.70 1.71 1.72 1.73 1.73 1.73 1.73 1.72 1.71 1.70 1.68 1.66 1.63 1.61 1.58 1.54 1.51
Cleaning the prism:
Thoroughly clean the prism after use with water or other suitable solvent and dry with clean tissue.
The prism surface could be damaged by strong alkalis or acids if left in contact for long periods of time. Clean samples from the
prism as soon as practicable.
Wiping the prism surface occasionally with alcohol will remove any build-up of oils left from the samples.
44-892 Iss 01 Rev