Besantek BST-MG06 User manual

BST-MG06 Multi-gas Detecting Alarm
Manual Instruction

1
This BST-MG06 multi gas detecting alarm is applied to underground coal mine, petrochemical industry,
municipal, environmental protection and other spaces where multi gas may exist.This device can be
used for detecting methane (CH4) or combustible gas (LEL), oxygen (O2), carbon monoxide (CO) and
hydrogen sulfide (H2S) gases simultaneously. When the concentration of any gas in the environment
is higher or lower than that of set alarm point, this device with send out both sound and light alarms to
warn users who wears it to take protective actions. This device is portable and can be also used as a
fixed device. It’s anti-explosion and intrinsically safe, the anti-explosion sign is ExibΙ
1. Application
2. Working principle
In this device, the methane/combustible gas sensor adopts hot catalytic working principle and it
consists of black and white elements inside. The black element is the catalytic carrier of
methane/combustible gas, and will occur thermal catalytic reaction with the methane/LEL gas to
change its own resistance value; the white element is a pure carrier and its resistance value is very
stable. According to this sensor’s characteristics, the signal is enlarged by the wheatstone bridge and
displays the value, thus concentration detection methane/ combustible gas is achieved.
3. Charging this device
Notice: To charge the battery, ensure you are in a safe, non-hazardous area.
①Input voltage: 110-240V, 50-60Hz Output voltage: 5V, 500mA.
②It can be charged from 5V USB socket through connecting USB cord with this device.
USB socket could from computer, phone adapter, power station, etc.
③After connecting USB cord with this device, green LED flashing indicates regularly charging.
④When green LED stops flashing and remains light, it indicates charging is basically completed,
to enlarge battery life, please don’t unplug charging device right away, instead remain it
on charge for another hour.
< >
The CO sensor, O2sensor and H2S sensor all adopt electrochemical principle. The inside electrode
will react with the being detected gases in the role of catalyst to achieve the directional movement
of electron between electrodes, and make the electrical signal amplified and displayed by amplifying
circuit technology and other existing technologies so that to achieve the detection of concentration.

4. The drawing sketches for all components of this device.
1
2
3
4
5
6
7
8
9
10
11
12
13
Item Description
face cover
alarm light
front cover
lcd
usb port
buttons
buzzer
knob
screw
back cover
lable
calibration mask
5. Technical specification: ( 1ppm=1X 10 )
-6
Technical specification Parameter
Detecting gas
Measurement range
Measurement error
Respond time
CH4/LEL CO O2H2S
CH4
( 0.00~1.00 ) %CH4: ±0.10%CH4( 1.00~3.00 ) %CH4: true value ±10%
( 3.00~5.00 ) %CH4: ±0.30%CH4
≤20S( T90 )
(0.00~5.00)%Vol
Resolution
Alarm point
0.01%vol
1.00%vol(can be adjusted in whole range)
clip
(optional)
(optional)
2

Technical specification Parameter
Measurement error
Measurement error
Measurement error
Measurement error
Respond time
Respond time
Respond time
Respond time
LEL
Resolution
Resolution
Resolution
CO
Alarm point
Alarm point
Measurement range
Measurement range
Measurement range
Measurement range
O2
Low Alarm point
( 0~1000 )ppm
( 0~20 ) ppm: ±2ppm ( >20~100 )ppm: ±4ppm
(> 100~500 )ppm: true value ±5% ( >500~1000) %ppm: true value ±6%
≤45S( T90 )
1ppm
24ppm(can be adjusted in whole range)
(0.0 ~ 30.0)
( 0.0~5.0 ) %vol:±0.5%vol ( >5.0~30.0 )%vol: 0.9%vol
≤35S( T90 )
0.1%vol
18.0%vol(can be adjusted in whole range)
25.0%vol(can be adjusted in whole range)
( 0 ~ 100 )ppm
( 0~50 ) ppm: ±3ppm ( >50~100 )ppm: True value ±10%
≤45S( T90 )
( 0 ~ 100 ) %LEL
( 1 ~ 20 )LEL: ±2%LEL
( 60 ~ 100 ) %LEL: ±6%LEL
( 20 ~ 60 ) %LEL: true value ±2%
1%LEL
≤20S( T90 )
20%LEL(can be adjusted in whole range)
High alarm point
H2S
3
%vol

Technical specification Parameter
Resolution
Alarm point
Alarm sound intensity
Alarm way
Alarm indicator light
Sensor life
Working current
Battery item
Battery short circuit
current
Charging time
Protection level
Size
Working temperature
Sound, light alarm
Can be visible in the distance of not less 20m
1ppm
10ppm(can be adjusted in whole range)
≥85dB at 1m audible
95mA
≥13hours
3.7V/1500mA
2A
4-5hours
IP65
130×66×30 (mm)
-25℃~55℃; short time for -40℃~55℃
≥2year( CO, O2, H2S ) ≥1year( LEL,CH4)
Continuous operating
hours
6. Detecting operation
① Button introduction
Button Description
On/ off button
yes button press OK
(change between detecting interface and alarm point interface)
start this device or shut it down, please long press
up button set the data value
down button set the data value
4

Detection
interface
Funcion
interface
Funcion
interface
Zero setting
interface
long press
OK button
short press
OK button
short press
OK button
short
press
long press
OK button
Step Display
②On/ off operation
Step
Long press on/off button to shut down this device in the state of running
long press
OK button
long press
OK button
In the state of closing, long press the
on/off button to open this device,
display our company’s logo “ ”,
then change into the alarm. point
display interface,and at last, go into
the normal detecting interface.
Display
③The interface switch operation
CH
4
H
2
S
CO
O
2
0.05
20.9
0
0
2013.02.18
[ ]
%VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
09 30
:
CH
4
H
2
S
CO
O
2
0.05
20.9
0
0
2013.02.18
[ ]
%VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
09 30
:
short
press
short
press
Zero Calibr-
-ation
Alarm
Time
Setting
Zero
Zero
Zero
Zero
Zero
LEL
H
2
S
CO
O
2
20.9
0
0
0
2013.02.18
[ ]
%LEL
[ ]
%VOL
[
]
PPM
[
]
PPM
09 30
:
LEL
H
2
S
CO
O
2
20.9
0
0
0
2013.02.18
[ ]
%LEL
[ ]
%VOL
[
]
PPM
[
]
PPM
[
]
PPM
[
]
PPM
09 30
:
Calibr-
-ation
Alarm
Time
Setting
CH
4
H
2
S
CO
O
2
0.05
20.9
0
0
[ ]
% VOL
[ ]
%VOL
Zero
Calibr-
-ation
Alarm
Time
Setting
Calibr-
-ation
Alarm
Time
Setting
LEL
H
2
S
CO
O
2
20.9
0
0
0
[ ]
% LEL
[ ]
%VOL
[
]
PPM
[
]
PPM
Zero
Calibr-
-ation
Alarm
Time
Setting
Calibr-
-ation
Alarm
Time
Setting
short
press
(Zero)
CH4
H2S
CO
O2
0.00
20.9
0
0
2013.02.18
[ ]
%VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
09 30
:
CH
4
H
2
S
CO
O
2
1.00
18.0
24
10
[ ]
%VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
Alarm
LEL
H2S
CO
O2
0
20.9
0
0
2013.02.18
[ ]
%LEL
[ ]
%VOL
[
]
PPM
[
]
PPM
09 30
:
LEL
H2S
CO
O
2
20
18.0
24
0
[ ]
%LEL
[ ]
%VOL
[
]
PPM
[
]
PPM
Alarm
[
L
]
[
L
]
5
BESANTEK
BESANTEK
BESANTEK

Funcion
interface
Funcion
interface
Funcion
interface
Alarm point
setting interface
Time setting
interface
Calibration
setting
interface
short press
OK button
short press
OK button
short press
OK button
short press
OK button
short press
OK button
short press
OK button
Step Display
short
press
short
press
short
press
short
press
short
press
short
press
Funcion
interface
Setting
interface
CH
4
H
2
S
CO
O
2
1.00
18.0
24
0
[ ]
% VOL
[ ]
%VOL
O
2
[ ]
%VOL
[
]
PPM
[
]
PPM
H
2
S
CO
[
]
PPM
[
]
PPM
Alarm
Zero Calibr-
-ation
Time
Setting
Alarm
Alarm Alarm
Zero Calibr-
-ation
Time
Setting
Alarm
Zero Calibr-
-ation
Time
Setting
LEL
20
18.0
24
0
[ ]
% LEL
Alarm
Zero Calibr-
-ation
Time
Setting
(Alarm)
Time
Zero Calibr-
-ation
Alarm
Time Setting
Zero Calibr-
-ation
Alarm
Time Setting
Year: 2013
02
18
09
30
Month:
Day:
Hour:
Min:
(Time)
CH
4
H
2
S
CO
O
2
0.00
20.9
0
0
[ ]
% VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
Zero Calibr-
-ation
Calibration
Alarm
Time
Setting
Zero Calibr-
-ation
Alarm
Time
Setting
Zero Calibr-
-ation
Alarm
Time
Setting
Zero Calibr-
-ation
Alarm
Time
Setting
LEL
H
2
S
CO
O
2
0
20.9
0
0
[ ]
% LEL
[ ]
%VOL
[
]
PPM
[
]
PPM
Calibration
(Calibration)
CH
4
LEL
Zero Calibr-
-ation
Alarm
Time
Setting
Setting
Zero Calibr-
-ation
Alarm
Time
Setting
√
(Setting)
6

long press
OK button
long press
OK button
long press
OK button
long press
OK button
long press
OK button
Step Display
short press
OK button
④The interface of Zero setting operation
CH
4
H
2
S
CO
O
2
0.05
20.9
0
0
[ ]
%VOL
[ ]
%VOL
[
]
PPM
CO
[
]
PPM
[
]
PPM
Zero
CH
4
H
2
S
CO
O
2
0.05
20.9
0
0
[ ]
% VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
Zero
CH
4
H
2
S
CO
O
2
0.00
20.9
0
0
[ ]
% VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
Zero
CH
4
H
2
S
CO
O
2
0.05
20.9
0
0
[ ]
% VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
Zero
CH
4
CO
Set OK.
O
2
0.05
20.9
0
[ ]
% VOL
[ ]
%VOL
Zero
LEL
H
2
S
O
2
1
20.9
0
0
[ ]
%LEL
[ ]
%VOL
[
]
PPM
Zero
LEL
H
2
S
CO
O
2
0 0
0
20.9
0
0
[ ]
% LEL
[ ]
%VOL
[
]
PPM
[
]
PPM
Zero
LEL
H
2
S
CO
O
2
0
0
20.9
0
[ ]
% LEL
[ ]
%VOL
[
]
PPM
[
]
PPM
Zero
LEL
H
2
S
CO
O
2
20.9
0
0
[ ]
% LEL
[ ]
%VOL
[
]
PPM
[
]
PPM
Zero
LEL
CO
Set OK.
O
2
20.9
0
[ ]
% LEL
[ ]
%VOL
Zero
In zero setting interface, press up/down
button to select gas which need to do
zero setting. Selected gas name will
turn green and long press “OK” button,
the background of selected gas will
turn red, long press “OK” button again
to finish this operation with
display of “Set OK”.
CH
4
H
2
S
CO
O
2
0.05
20.9
0
0
[ ]
%VOL
[ ]
%VOL
[
]
PPM
Zero
CH
4
H
2
S
CO
O
2
0.05
20.9
0
0
[ ]
% VOL
[ ]
%VOL
[
]
PPM
Zero
CH
4
H
2
S
CO
O
2
0.05
20.9
0
0
[ ]
% VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
[
]
PPM
CH
4
H
2
S
CO
O
2
0.05
20.9
0
0
[ ]
% VOL
[ ]
%VOL
[
]
PPM
Zero
[
]
PPM
[
]
PPM
Zero
Short press on/off button to exit zero
setting operation
7

Step Display
Step Display
For example:
long press
OK button
long press
OK button
long press
OK button
short press
OK button
LEL
H
2
S
CO
O
2
20.9
0
0
[ ]
% LEL
[ ]
%VOL
[
]
PPM
[
]
PPM
Zero
LEL
H
2
S
CO
O
2
20.9
0
20 20 20
0
[ ]
%LEL
[ ]
%VOL
[
]
PPM
Zero
LEL
H
2
S
CO
O
2
20.9
0
0
[ ]
% LEL
[ ]
%VOL
[
]
PPM
Zero
LEL
H
2
S
CO
O
2
20.9
0
20
0
[ ]
% LEL
[ ]
%VOL
[
]
PPM
[
]
PPM
[
]
PPM
[
]
PPM
Zero
Short press on/off button to cancel
the selected gas zero setting operation.
Notice: The oxygen gas zero setting must use 100%VOL nitrogen gas
For other gases it should be carried out in clean air.
⑤Alarm point setting interface operation
CH4
H2S
CO
1.00
24
10
[ ]
% VOL
[
]
PPM
[
]
PPM
Alarm
CH4
H2S
CO
O2
1.00
20.9
24
10
[ ]
%VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
Alarm
CH4
H2S
CO
O2
1.00
20.9
25
10
[ ]
%VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
Alarm
CH4
H2S
CO
O2
1.00
20.9
24
10
[ ]
%VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
Alarm
CH4
H2S
CO
O2
1.00
20.9
25
10
[ ]
%VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
Alarm
Alarm
25
CH4
CO
Set OK.
O2
1.00
20.9
[ ]
%VOL
[ ]
%VOL
In alarm point setting interface, press
up/down button to select gas name,
selected gas name will turn green and
long press “OK” button, the background
of selected gas will turn red. Press
up/down button to adjust the value,
after finishing the adjusting, long press
“OK” button to complete this
operation with display of “Set OK”
Short press on/off button to exit alarm
point set operation.
O
2
18.0
[ ]
%VOL
[
L
]
8

Step Display
Step Display
press OK
button
long press
OK button
long press
OK button
long press
OK button
⑥Time setting interface operation
Time
Year: 2013
02
18
09
30
Month:
Day:
Hour:
Min:
Time
Year: 2013
02
18
09
30
Month:
Day:
Hour:
Min:
Time
Year: 2013
03
18
09
30
Month:
Day:
Hour:
Min:
Time
Year: 2013
03
18
09
30
Month:
Day:
Hour:
Min:
Time
Year: 2013
03
18
Month:
Day:
Hour
Set OK.
In time setting interface, press “OK” button
to select year, month and date options, it
turns green after being selected. Press
up/down button to adjust the value and
long press “OK” button to complete this
operation with display of “Set OK”
Short press on/off button to return
function interface.
⑦Calibration interface setting operation
CH
4
H
2
S
CO
O
2
0.00
20.9
0
0
[ ]
%VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
CH
4
H
2
S
CO
O
2
0.00
20.9
0
500
0
[ ]
%VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
CH
4
H
2
S
CO
O
2
0.00
20.9
0
[ ]
%VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
CH
4
H
2
S
CO
O
2
0.00
20.9
0
0
[ ]
% VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
Calibration Calibration Calibration
Calibration
CalibrationCalibration
CH
4
H
2
S
CO
O
2
0.00
20.9
500
0
[ ]
%VOL
[ ]
%VOL
[
]
PPM
[
]
PPM
500
CH
4
CO
Set OK.
O
2
0.05
20.9
[ ]
%VOL
[ ]
%VOL
In calibration setting interface, press
up/down to select gas name, selected
gas name will turn green and long press
“OK” button, the background of selected
gas will turn red. After the value is stable
press up/down button to adjust the
display value and make it same with
standard gas concentration value.
Long press “OK” button to complete this
operation with display of “Set OK”
Short press on/off button to exit this
operation.
For example:
( CO standard gas concentration is 500ppm)
9

Step Display
long press
OK button
short press
on/off button
Notice:
• Calibration gas must be standard gas.
⑧Setting interface operation
CH
4
LEL
Setting
√
CH
4
LEL
Setting
√
CH
4
LEL
Setting
√
Zero Calibr-
-ation
Alarm
Time
Setting
In setting interface, user can choose
between methane gas (CH4) and
combustible gas (LEL) according to
practical necessity. Press up/down
button to select and confirm by long
press “OK” button. A tick icon “√”
shows the operation completed.
Short press on/off button to return
function interface.
• This device must do zero setting operation before calibration operation.
• Make sure the flow is stable before connecting calibration gas with this device.
• The calibration couldn’t be done when this device is alarming for low power
• Before calibration, open this device for at least 6 minutes and allow it to run stably.
• Calibration gas flow for methane, combustible gas, carbon monoxide and
oxygen is 300ml/min, it’s 400ml/min for hydrogen sulphide.
• Suggested calibration gas concentrations are: Methane: 1.00%vol;
combustible gas: 20%LEL; CO: 500ppm; H2S: 50ppm.
Oxygen can be calibrated in clean air.
10

7. Attentions of using this device
8. Storage and maintenance
9. Trouble shooting
①Battery of this device shouldn’t be removed in underground places, and DO NOT use different
batteries from standard one.
②Charge this device in ground and safe places.
③The removable catalytic component must be provided by BESANTEK Corporation.
Notice: when maintain this device, Please don’t change any component’s parameter,
specification and model when maintain this device
①If this device couldn’t be turned on, maybe it has no power, please charge it firstly.
② If the value of this device is not zero in clean air,
please carry on zero setting firstly. (except Oxygen )
③If this device couldn’t test accurately, please carry on calibration firstly.
④If this device becomes expired, please purchase a new one.
①This device should be kept by special person, and build a registration system
②This device should be stored in dry and well ventilated places if it’s not used for a long time.
③DO NOT disassemble this device at will, it must be repaired by specialists.
11