Carel BP SE Sensor User manual

Integrated Control Solutions & Energy Savings
User manual
For monitoring environmental conditions
• Temperature
• Humidity
• Light
Wireless sensors


3
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
WARNINGS
CAREL bases the development of its products on decades of experience
in HVAC, on the continuous investments in technological innovations
to products, procedures and strict quality processes with in-circuit and
functional testing on 100% of its products, and on the most innovative
production technology available on the market. CAREL and its subsidiaries
nonetheless cannot guarantee that all the aspects of the product and the
software included with the product respond to the requirements of the nal
application, despite the product being developed according to start-of-the-
art techniques.
The customer (manufacturer, developer or installer of the nal equipment)
accepts all liability and risk relating to the conguration of the product in
order to reach the expected results in relation to the specic nal installation
and/or equipment.
CAREL may, based on specic agreements, acts as a consultant for the positive
commissioning of the nal unit/application, however in no case does it accept
liability for the correct operation of the nal equipment/system.
The CAREL product is a state-of-the-art product, whose operation is specied
in the technical documentation supplied with the product or can be
downloaded, even prior to purchase, from the website www.carel.com.
Each CAREL product, in relation to its advanced level of technology, requires
setup/conguration/programming/commissioning to be able to operate in
the best possible way for the specic application.The failure to complete such
operations, which are required/indicated in the user manual, may cause the
nal product to malfunction; CAREL accepts no liability in such cases.
Only qualied personnel may install or carry out technical service on the
product.
The customer must only use the product in the manner described in the
documentation relating to the product.
In addition to observing any further warnings described in this manual, the
following warnings must be heeded for all CAREL products:
• prevent the electronic circuits from getting wet. Rain, humidity and all
types of liquids or condensate contain corrosive minerals that may damage
the electronic circuits. In any case, the product should be used or stored
in environments that comply with the temperature and humidity limits
specied in the manual.
• do not install the device in particularly hot environments. Too high
temperatures may reduce the life of electronic devices, damage them and
deform or melt the plastic parts. In any case, the product should be used
or stored in environments that comply with the temperature and humidity
limits specied in the manual.
• do not attempt to open the device in any way other than described in the
manual.
• do not drop, hit or shake the device, as the internal circuits and mechanisms
may be irreparably damaged.
• do not use corrosive chemicals, solvents or aggressive detergents to clean
the device.
• do not use the product for applications other than those specied in the
technical manual.
All of the above suggestions likewise apply to the controllers, serial boards,
programming keys or any other accessory in the CAREL product portfolio.
CAREL adopts a policy of continual development. Consequently, CAREL
reserves the right to make changes and improvements to any product
described in this document without prior warning.
The technical specications shown in the manual may be changed without
prior warning.
The liability of CAREL in relation to its products is specied in the CAREL general
contract conditions, available on the website www.carel.com and/or by
specic agreements with customers; specically, to the extent where allowed
by applicable legislation, in no case will CAREL, its employees or subsidiaries
be liable for any lost earnings or sales, losses of data and information, costs of
replacement goods or services, damage to things or people, downtime or any
direct, indirect, incidental, actual, punitive, exemplary, special or consequential
damage of any kind whatsoever, whether contractual, extra-contractual or
due to negligence, or any other liabilities deriving from the installation, use or
impossibility to use the product, even if CAREL or its subsidiaries are warned
of the possibility of such damage.
DISPOSAL
INFORMATION FOR USERS ON THE CORRECT
HANDLING OF WASTE ELECTRICAL AND ELEC-
TRONIC EQUIPMENT (WEEE)
In reference to European Union directive 2002/96/EC issued on 27 January
2003 and the related national legislation, please note that:
1. WEEE cannot be disposed of as municipal waste and such waste must be
collected and disposed of separately;
2. the public or private waste collection systems dened by local legislation
must be used. In addition, the equipment can be returned to the
distributor at the end of its working life when buying new equipment;
3. the equipment may contain hazardous substances: the improper use or
incorrect disposal of such may have negative eects on human health and
on the environment;
4. the symbol (crossed-out wheeled bin) shown on the product or on the
packaging and on the instruction sheet indicates that the equipment has
been introduced onto the market after 13 August 2005 and that it must
be disposed of separately;
5. in the event of illegal disposal of electrical and electronic waste, the
penalties are specied by local waste disposal legislation.
Warranty on materials: 2 years (from the date of production, excluding
consumables).
Approval: the quality and safety of CAREL S.P.A. products are guaranteed by
the ISO 9001 certied design and production system.
Important warning!!!
The rTM SE system devices are
incompatible with the Carel rTM system,
due to an improvement made to the
ZigBeewireless communication
protocol.

4
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010

5
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
Content
1. INTRODUCTION 7
1.1 Wireless monitoring devices ........................................................................ 7
1.2 Codes................................................................................................................. 8
1.3 Terminology.................................................................................................... 9
1.4 Advantages of the wireless system ............................................................. 9
1.5 Type of Carel wireless network (MESH) .................................................. 10
1.6 General features of the system.................................................................. 11
1.7 Using the Router ........................................................................................... 11
1.8 General notes ................................................................................................ 11
1.9 Reference standards..................................................................................... 12
1.10 Battery life....................................................................................................... 12
1.11 List of sensor system variables (alphabetical order)............................. 12
2. BP SE SENSOR (BUILT-IN PROBE) 13
2.1 Functions implemented and supervisor variables available ............... 13
2.2 Sensor configuration.................................................................................... 13
2.3 Sensor activation........................................................................................... 14
2.4 Technical specifications................................................................................ 15
2.5 List of parameters and variables, BP SE Sensor..................................... 16
2.6 Installation notes............................................................................................17
2.7 Physical dimensions ......................................................................................17
2.8 Replacing the battery in the BP SE Sensor...............................................17
2.9 Application examples....................................................................................17
3. EP SE, SA, SI SENSORS AND CI PULSE COUNTER 18
3.1 Parameters and functions ........................................................................... 18
3.2 Description of the acquisition process..................................................... 18
3.3 Device configuration .................................................................................... 18
3.4 Binding procedure........................................................................................ 19
3.5 Resetting the sensor (unbinding).............................................................. 19
3.6 General warnings ......................................................................................... 19
4. EP SENSOR (EXTERNAL PROBE) 20
4.1 Functions implemented...............................................................................20
4.2 Parameters and functions ...........................................................................20
4.3 Technical specifications................................................................................20
4.4 List of parameters and variables, EP SE Sensor ..................................... 21
4.5 EP SE Sensor installation notes..................................................................22
4.6 EP SE physical dimensions .........................................................................22
4.7 EP SE electrical connections .......................................................................22
4.8 Application example.....................................................................................22
5. SA ROOM SENSOR 23
5.1 Functions implemented and supervisor variables available ...............23
5.2 Technical specifications................................................................................24
5.3 List of parameters and variables, SA Sensor...........................................24
5.4 Sensor installation notes .............................................................................25
5.5 Physical dimensions .....................................................................................25
5.6 Application example.....................................................................................25
6. SI INDUSTRIAL SENSOR 26
6.1 Functions implemented and supervisor variables available ...............26
6.2 Technical specifications................................................................................27
6.3 List of parameters and variables, SI Sensor ............................................28
6.4 SI Sensor installation notes.........................................................................28
6.5 Physical dimensions .....................................................................................29
6.6 Application example....................................................................................29
7. CI PULSE COUNTER 30
7.1 Functions implemented...............................................................................30
7.2 Parameters and functions...........................................................................30
7.3 Technical specifications................................................................................30
7.4 List of parameters and variables, CI Pulse Counter .............................. 31
7.5 CI Pulse Counter installation notes...........................................................32
7.6 CI Pulse Counter physical dimensions.....................................................32
7.7 CI Pulse Counter electrical connections...................................................32
7.8 Connection example ...................................................................................32
8. AP ACCESS POINT 33
8.1 Main functions...............................................................................................33
8.2 Parameters and functions ...........................................................................33
8.3 Configuration ................................................................................................33
8.4 Setting the address.......................................................................................33
8.5 Binding procedure........................................................................................34
8.6 Resetting the device .....................................................................................34
8.7 Serial communication parameters............................................................34
8.8 Table of LED status.......................................................................................35
8.9 Technical specifications................................................................................35
8.10 List of Access Point system variables (alphabetical order)..................36
8.11 List of parameters and variables, Access Point versione ..........................
Modbus RTU® .....................................................................................................................................37
8.12 Installation notes...........................................................................................37
8.13 Electrical connections and physical dimensions....................................38
9. RO ROUTER 39
9.1 Parameters and functions .........................................................................39
9.2 Binding the Router to the Access Point....................................................39
9.3 Resetting the device .....................................................................................40
9.4 Table of LED status.......................................................................................40
10. ROUTERS WITH OTHER INTEGRATED FUNCTIONS 41
10.1 EP1 Router-Sensor........................................................................................ 41
10.2 RB Router-Bridge........................................................................................... 41
10.3 RA Router-Actuator.......................................................................................42
10.4 Functions implemented...............................................................................42
10.5 RC Router-Pulse Counter ............................................................................42
10.6 Technical specifications................................................................................43
10.7 List of Router system variables (alphabetical order).............................43
10.8 List of Router parameters .........................................................................44
10.9 List of RA Router-Actuator parameters.....................................................44
10.10 Installation notes...........................................................................................44
10.11 General warnings ........................................................................................46
11. GENERAL NOTES 47
11.1 Notes for correct installation ...................................................................... 47

“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010 6
12. RTM SE HANDHELD 48
12.1 General Features...........................................................................................48
12.2 Operating modes..........................................................................................48
12.3 Main menu.....................................................................................................48
12.4 Scan Energy....................................................................................................48
12.5 Scan Networks...............................................................................................49
12.6 Scan Connection ...........................................................................................49
12.7 Unbinding.......................................................................................................49
12.8 Ping test...........................................................................................................49
12.9 Network commands.....................................................................................49
12.10 “View Mode” menu......................................................................................50
12.11 “Open Network” menu ...............................................................................50
12.12 “Reset One” menu ......................................................................................50
12.13 Password entry menu..................................................................................50
12.14 “Set Passw” Menu – Set Access Point password.................................. 51
12.15 Sensors menu................................................................................................ 51
12.16 List of Sensors................................................................................................ 51
12.17 Set Sensor address .......................................................................................52
12.18 Unbind Sensor...............................................................................................53
12.19 Start screen.....................................................................................................54
12.20 ZigBeehandheld signal meter shutdown..........................................54
12.21 Notes on operation.....................................................................................54
12.22 ZigBeehandheld signal meter electrical specifications...................54
12.23 Physical dimensions....................................................................................54
13. ROUTER-SNIFFER 55
13.1 Router-Sniffer.................................................................................................55
13.2 Technical specifications:...............................................................................55
13.3 Layout..............................................................................................................55
13.4 LED meanings................................................................................................55
14. Z-CONFIG PROGRAM 56
14.5 Layout examples .......................................................................................... 61
15. DIPSWITCH-ID CROSS-REFERENCE TABLE FOR
SENSORS 63
15.1 Dipswitch-ID cross-reference table for sensors......................................63

7
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
1. INTRODUCTION
1.1 Wireless monitoring devices
For the retrot of food refrigeration and room cooling systems, energy
consumption measurement and I/O management via supervisor, CAREL
proposes the rTM SE wireless system (Remote Temperature Monitoring). This
solution guarantees the maximum in terms of:
• Flexibility;
• Functions;
• Reliability;
• Easy operation
• Reduced installation costs;
• Easy commissioning/service;
• Integration with the most common BMS (Building Management Systems);
This solution ensures considerable savings in terms of installation costs
(eliminating the cost of wiring), oering exibility in the layout of supermarkets
and allowing faster retrot installation. Ideal for all installations where electrical
wires cannot be laid, i.e. properties that do not have raised oors or false
ceilings.
The retrot of existing systems is required for compliance with HACCP
standards, for monitoring the systems via remote connections, for
recording events and analysing them for scheduled maintenance.
The CAREL rTM SE system can be used in all industrial and trade businesses
that require the prevention of risks relating to the safety and storage of food
for human consumption, in accordance with the HACCP standards; moreover,
it oers the possibility to manage exible spaces very simply, thus reorganising
the layout of showcases in a supermarket without having an impact on the
wired network (communication and power supply);
The system is a network of wireless sensors tted inside the showcases, easy
to congure and install, connected to a Carel supervisor (PlantVisorPRO or
PlantWatchPRO) for recording the temperature, events and alarm notications.
The data measured and the alarms signalled are saved and can be accessed at
any time, in compliance with EN 12830.
The system can be easily installed on all types of refrigeration unit (showcases or
cold rooms), is independent of the controller installed on the unit and requires
no additional wiring because the devices are wireless and battery powered,
meaning signicant cost savings.
The sensors require no electrical connections as they use a long life battery
(typically 5 to 8 years, depending on the transmission frequency set), a wireless
connection with ZigBee™ technology (mesh) at a transmission frequency of
2.4 GHz authorised for operation in all countries around the world, and are
ready for connection to the most common BMS systems using Modbus®
protocol. The sensors monitor the inputs (temperature, humidity, light and
digital input status) and send the data wirelessly to the Access Point or Router.
Communication between sensors and the Access Point is two way. The
sensors, as well as sending the change in the status of the variables, can also
receive data.
Moreover, a mains powered model has been designed for use in all
applications that require frequent communication (e.g. monitoring oating
suction pressure).
In ambient monitoring applications, the temperature, humidity and light
intensity can all be recorded by simply installing battery powered sensors in
the desired location. The sensors cover a wide range of uses in refrigeration,
air-conditioning and humidication applications.
Many applications are also available for remote I/O management from
supervisors, as the module manages generic I/Os and saves on the cost of
laying cables, without the need for separate power and signal cables.
The wireless devices send the temperature and alarm data wirelessly to the
Router and Access Point, which relay the information to the supervisory system
The CAREL rTM SE system consists of the following components:
• Battery powered devices:
- Temperature sensor from tted inside the showcase, version BP SE (Built-
in Probe);
- Sensor with two external NTC probes and two digital inputs, version EP
SE (External Probe) for showcases and cold rooms;
- Room temperature and humidity sensor for installation in residential
environments, version SA
- Temperature, humidity and light sensor, SI industrial version;
- Pulse counter to be used with the energy meter module congured for
pulse counters, version CI;
• Access Point. Wireless receiver that acquires data from the various sensors
in the ZigBee™ network, making such data available to the supervisor via
Modbus® RTU over RS485. Up to 30 sensors can be directly associated with
each Access Point, or a maximum of 60 if one or more Routers are used. The
supervisor (PlantVisorPRO or PlantWatchPRO) can thus see all the variables
in the rTM SE system;
• Router. To be used when the distance between the Sensors and the
Router exceeds 30 m (relays the wireless signals so as to cover greater
distances between the Access Point and sensors), or if there are more
than 30 Sensors in the network. There can be a maximum of 60 Routers in
the wireless network, 48 of which are visible to the supervisor. The Access
Point automatically assigns a serial address in the order in which these are
“bound”, starting from 200 up to 247. Five versions of Router are available,
which also include other functions:
- Router powered at 230 Vac mains voltage, version RO;
- Router Bridge powered at 12-24 Vac, version RB. Integrates the function
to extend the RS485 network;
- Router Sensor powered at 12-24Vac, version EP1. Integrates the functions
of the battery powered EP SE Sensor);
- Router-Actuator powered at 12-24 Vac, version RA. Integrates the
functions of I/O module or local thermostat;
- Router-Pulse Counter powered at 12-24 Vac, version RC. Integrates the
same functions as the CI battery powered pulse counter;
• Modbus® supervisor system: The rTM SE system is designed to be used
together with Carel PlantVisor PRO or PlantWatch PRO supervisors
Wireless transmission between the various devices uses standard ZigBee™
communication protocol and encryption technology with a Carel private key.
This is an advanced system that has achieved an excellent level of security in
data exchange for wireless communication and is used in many applications.
The CAREL solution uses mesh technology between Access Points and
Routers, ensuring more reliable communication and delivery of the data sent
by the sensor.
Note: ZigBeewireless connection without interoperability.
rTM SE handheld: network analyser used to check the ZigBeewireless signal
level and to open/close the wireless network when binding the devices
(sensors and Routers), including the possibility to set the BP Sensor address
and reset the Router and Access Point. Useful during installation;

8
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
1.2 Codes
Code Model Features Power supply
WS01U01M00 BP SE Sensor Temp. for showcases Battery
WS01W02M00 EP SE Sensor Temp for cold rooms or showcases Battery
(*) WS01G01M00 SA Sensor Room temp./humid. Battery
(*) WS01F01M00 SI Sensor Temp./humid./lux for industrial use Battery
(*) WS01E02M00 CI Pulse Counter Pulse counter for energy modules Battery
WS01AB2M20 AP Access Point ZigBee– RS485 Modbus® Access Point 12-24 Vac/dc
WS01RC1M20 RO Router ZigBeewireless repeater 12-24 Vac/dc
WS01VB2M10 EP1 Router-Sensor Repeater with temp. sensor 12-24 Vac/dc
WS01RB2M20 RB Router-Bridge Repeater with RS485 Modbus® bridge 12-24 Vac/dc
(*) WS01HO2M20 RA Router-Actuator Repeater with I/O module - thermostat 12 Vac/dc
(*) WS01NO2M20 RC Router-Pulse Counter Repeater with Pulse counter (energy modules) 12-24 Vac/dc
(*) available soon
Tab. 0.a
BP SE Sensor EP SE Sensor SA Sensor
SI Sensor Pulse counter CI
Access point RO Router EP1 Router- sensor
RB Router-Bridge RA Router-Actuator RC Router- pulsecounter
Fig. 1.a

9
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
1.3 Terminology
Wireless
Wireless means“without wires”, in contrast to the term wired.
Wireless network
Communications system (series of devices, appliances, methods and
protocols) for the transmission of information via radio, typically radio-
frequency technology used instead of wired connections, making the systems
particularly exible.
ZigBee™
Zigbee™ is a set of specications based on the IEEE-802.15.4 standard for
the creation of Wireless Personal Area Networks (WPAN). Comparable in
some ways to Bluetooth, it stands out for its very low power consumption
and the reduced cost of implementation, despite having a maximum data
transfer speed of 250 kbit/s. ZigBee™ devices, with compact dimensions and
low costs, are designed to work in dedicated self-organised networks (Mesh
networks) and are used in many elds.
1.4 Advantages of the wireless system
Advantages of a wireless network over a wired network
• Mobility of sensors;
• Easy to install and connect the devices;
• Coverage even where obstacles are present;
• Flexibility in the event of structural modications;
• Reduction in wiring costs;
• Robustness.
The advantages of wireless networks can overcome some of the intrinsic limits
in wired systems. Typical network infrastructure features a wired backbone
with wireless access.
Advantages of ZigBee™
• Standard technology;
• Reduced costs;
• Can be used globally;
• Reliable;
• Supports a large number of nodes;
• Easy conguration;
• Long battery life;
• Secure data transmission.
Distance
ConsumptionVelocity
Fig. 1.b
All brands and names shown in the diagram are registered trademarks and
the property of their respective owners.

10
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
Advantages of working at 2.4 GHz
Band of frequencies No. of channels Data parameters Use
Symbol rate Bit rate Mapping
868-868.6 MHz 1 20Kbit/s 20 Kbaud Binary Europe
902-928 MHz 10 40Kbit/s 40 Kbaud Binary North America
2.4-2.4835 GHz 16 250Kbit/s 62.5 Kbaud 16-ary orthogonal Worldwide
The band centred around 2.45 GHz (used in the wireless sensor system for
refrigeration) is the only one that can be used all over the world, without
needing to apply for special licenses. In addition, the ISM band (Industrial,
Scientic and Medical) exploits the full potential of the standard, that is,
can use 16 transmission channels with a bit rate of 250 kbit/s.
Types of nodes
ZigBee™ Access Point - Co-ordinator and Gateway;
- Must be available and on in every network
- Coordinates the creation of the network;
ZigBee™ Router;
- Participates in the delivery of the messages, and must always be on;
- Available in Router-Bridge version for extending a wired local network
(for a list of approved controllers, see chapter on the features of the
Router), and EP1 Router-Sensor version.
ZigBee™ End-Device (sensors);
- Node with limited wireless functions;
- Low power consumption;
- Low cost;
For data communication with the Access Point, the end device uses a“parent”
for eective wireless transmission; this may be a Router or the Access Point
itself.
1.5 Type of Carel wireless network (MESH)
Legenda:
ZigBee»End-Device:
Sensors BP and EP (S) ZigBeeRouter-Bridge (R) ZigBeeCoordinator -
Access point (AP)
RS 485 ModBus
S
S
S
S
SS
RR
S
S
SS
S
R
S
S
S
S
SS
AP
Fig. 1.c
Tab. 0.b

11
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
Example of a Mesh network
The MESH layout, used in the wireless sensor system for refrigeration between
coordinator nodes (access points) and router-bridge devices, ensures a high
tolerance to faults, as if one sensor loses wireless communication, the radio
signal still manages to nd an alternative route to reach the destination.
RAP
RS 485 ModBus
R
R
Fig. 1.d
1.6 General features of the system
Maximum distance between Access Point/Router and Sensors in open eld
(outdoors): 100 m.
Maximum distance between Access Point/Router and Sensors with eld of
sight (indoors): around 30 m (inside rooms and built-up areas).
Transmission frequency: selectable from 2405 to 2480 MHz.
Number of channels available: 16.
Transmission power:
• Access Point 0 dBm
• Router 230Vac +10 dBm
• Router Bridge 0 dBm
• EP1 Router-Sensor +10 dBm
• RA Router-Actuator +3 dBm
• Router-Pulse Counter +10 dBm
• BP SE Sensor +3 dBm
• EP SE Sensor +3 dBm
• SA Room Sensor +3 dBm
• SI Industrial Sensor +3 dBm
• CI Pulse Counter +3 dBm
Wireless protocol: ZigBee™ without interoperability.
Standard: 802.15.4.
Reception sensitivity:
• Access Point -92 dBm
• Router SE 230Vac -97 dBm
• Router Bridge -92 dBm
• Router EP probe1 -97 dBm
• RA Router-Actuator -95 dBm
• Router-Pulse Counter -97 dBm
• BP SE Sensor -95 dBm
• EP SE Sensor -95 dBm
• SA Room Sensor -95 dBm
• SI Industrial Sensor -95 dBm
• CI Pulse Counter -95 dBm
For battery powered devices:
• Maximum current for battery powered devices only: 35 mA, in transmission.
• Current in standby: 1 µA.
Maximum HOP levels: 7 (hops).
Maximum number of wireless network devices:
• 30 for each Access Point (with 1 Router up to 60 units);
• 16 Routers directly connectable to the Access Point up to a maximum of 60
devices on the same network;
• 16 Router directly connectable to each Router up to a maximum of 60
devices on the same network;
Maximum number devices on Modbus® RS485 network:
• 7 Access Point;
• 111 Sensors;
• 60 Routers, max 48 of which monitored by the supervisor;
• On Modbus network in combination with other devices up to max 247
units.
1 234567
RRRRRRR
Access point
Router-Bridge
Max 60 sensors
Fig. 1.e
1.7 Using the Router
When does the Router need to be installed?
The Router is required whenever a direct connection is not possible between
the Access Point and the Sensor; this may occur when:
• The distance between Access Point and Sensor is greater than 30 m MAX
with visibility between the instruments.
• There is no visibility between the Access Point and the Sensor, and/or
there is shielding infrastructure that reduces the wireless communication
distance.
• In addition, the Router is required if the number of Sensors managed
exceeds 30 devices.
In addition, this is used to improve the reliability of the wireless connection,
the Router network can in fact nd an alternative path if one of the direct
connections between the sensors and the access point fails.
Recommended:
Up to 15 sensors 1 Router;
from 16 to 30 sensors 2 Routers;
from 31 to 45 sensors 3 Routers;
from 46 to 60 sensors 4 Routers.
1.8 General notes
The radio range of the devices is around a hundred metres in an open eld,
that is, without any obstacles.
In a closed eld the range varies signicantly based on the type of environment
and the surrounding objects (shelves, furniture, metal walls etc.).
Thick partition walls or reinforced ceilings and oors may represent impassable
obstacles.
The ideal position of the devices, especially the routers, often cannot be
dened theoretically but must be found by trial and error in the actual
installation.
Serial address assignment is valid for all devices
Make sure not to assign the same serial address ID to two devices in the same
wireless network.

12
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
1.9 Reference standards
The Carel wireless sensors have been tested in accordance with the following
standards:
INDUSTRIAL ENVIRONMENT
EN61000-6-4, EN61000-3-2, EN61000-3-3, EN61000-6-2
ETSI EN 301 489-17 V1.2.1, ETSI EN 301 489-1 V1.4.1
RES., COMM. AND LIGHT IND. ENVIRONMENT
EN61000-6-3; EN61000-3-2, EN61000-3-3; EN61000-6-1
Compliant with EN 13485 (Instruments for measuring the temperature of
foodstus)
1.10 Battery life
Transmission time in min. Sensor battery life in years
13
55
10 8
15 8
Tab. 0.cTab. 1.c
The battery life is purely indicative and depends on the cycle transmission
time set and the quality of the wireless connection. If the device does not
communicate correctly with the Access Point (distance or interference
problems) battery life will be reduced due to the continuous attempts to
restore connection to the Access Point/Router.
1.11 List of sensor system variables
(alphabetical order)
Name Description
ADD_HIGH_T_DELAY If when a high temperature is measured the“door open”or
“defrost in progress”signals are present, the device delays
the alarm by the value set for ADD_HIGH_T_DELAY (HR12).
ALM_BATTERY Provides the at battery signal (1 if < 2800 mV).
ALM_GENERAL Provides a general sensor fault signal.
ALM_LONG_DEFROST Provides the alarm status for the Defrost input (1=Alarm);
ALM_PROBE_1 Temperature measurement alarm on probe 1. This may
be caused by a value outside of the maximum range or
by the probe not connected correctly (open or short-
circuited).
ALM_PROBE_2 Temperature measurement alarm on probe 2. This may
be caused by a value outside of the maximum range or
by the probe not connected correctly (open or short-
circuited).
AP_RX_RADIO_LEV Wireless signal level received from the Access Point for the
sensor (see note 1).
AUTO_DELAY Denes a delay time for the evaluation of the type of
showcase when auto-conguration mode is enabled.
AVERAGE_PARAM Weight for calculating the average, as per the formula with
weight M.
AVG_TEMPERATURE Temperature value calculated as the weighted average (in
tenths of a degree °C).
BATTERY_CHARGE Denes the residual charge, counting power consumption
corresponding to the operations eectively carried out.
This can be used, together with the BATTERY_LEVEL value,
for a more complete evaluation of battery charge status.
Full charge mAh.
BATTERY_LEVEL Battery voltage value (mV). The rated value is 3600 mV,
below 2800 mV the battery is discharged.
CMD_PASSW_1 Only used by conguration systems.
CNT_REJOIN Wireless network parameter for internal use
DEFROST_ALM_DELAY Delay time (wait) in minutes before Defrost alarm signal
DEFROST_POL Logical state of the defrost input based on the electrical
state of the contact (open or closed).
DOOR_POL Logical state of the door input according to the electrical
state of the contact (open or closed).
EN_AUTO_CONF Enable automatic conguration mode (1= enabled).
EN_CMD_PW Only used by conguration systems.
EN_DI_DEFROST Enable/disable defrost digital input.
EN_DI_DOOR Enable/disable door digital input.
EN_HI_TEMP_ALM Enable the high temperature alarm signal (if=1), otherwise
the alarm is not measured/signalled. Used for both probes
1 and 2.
EN_SCAFFALE Selects the medium temperature shelf display case
(1=shelf)
FW_VERSION FW revision
HI_TEMP_ALM_1 Provides the status of the high temperature alarm for
probe 1
HI_TEMP_ALM_2 Provides the status of the high temperature alarm for
probe 2
HI_TEMP_TRESHOLD High temperature signal thresholds (in tenths of a degree
°C)
HI_TEMP_TRE-
SHOLD_1
High temperature signal threshold for probe 1. Can be set
in tenths of a degree centigrade
HI_TEMP_TRE-
SHOLD_2
High temperature signal threshold for probe 2. Can be set
in tenths of a degree centigrade
HIGH_TEMP_DELAY Delay (waiting) time in minutes before the high tempe-
rature alarm is actually signalled. Used for both probes 1
and 2
ID_SER_ADDR Sensor serial address, set using the rTM SE handheld or by
switch. Used as the sensor identier
IN_1_STATUS Status of digital input 1
IN_2_STATUS Status of digital input 2
LAST_RX_DELAY Wireless network parameter for internal use
LO_TEMP_ALM Provides the status of the low temperature alarm
LO_TEMP_ALM_1 Provides the status of the low temperature alarm for probe
1
LO_TEMP_ALM_2 Provides the status of the low temperature alarm for probe
2
LO_TEMP_TRESHOLD Low temperature signal threshold (in tenths of a degree
°C). Signal without delays.
LO_TEMP_TRE-
SHOLD_1
Low temperature signal threshold for probe 1. Can be set
in tenths of °C. Signal without delays
LO_TEMP_TRE-
SHOLD_2
Low temperature signal threshold for probe 2. Can be set
in tenths of °C. Signal without delays
MAC_ADDR_0 Unique 32 bit unit identier, LSB. Used to uniquely identify
each sensor
MAC_ADDR_1 Unique 32 bit unit identier, MSB. Used to uniquely identi-
fy each sensor
MACHINE_CODE Peripheral identier for the supervisor
MIN_RSSI_LEVEL Wireless network parameter for internal use
MIRROR_IS Wireless network parameter for internal use
MODE_AUTO_TRESH Denes a threshold in °C below which the procedure
for the automatic recognition of the type of showcase is
activated.
MODE_PARAM Denes the values to be assigned or auto-assigned for the
identication of the eective operating mode. For each
of the four modes, the associated parameters can be set
separately, and are loaded when the mode is activated
NETWORK_ID Wireless network parameter for internal use
OFFS_TEMP Temperature measurement oset, within a maximum of
±9.9 °C.
OFFS_TEMP_1 Calibration oset for probe 1, within a max of ±9.9 C;
OFFS_TEMP_2 Calibration oset for probe 2, within a max of ±9.9 C;
RX_MESSAGE_CNT Wireless network parameter for internal use
RX_MSG_LEVEL Wireless signal level received for the sensor in dBm+100
(see note 1).
TEMPERATURE Instant temperature value (in tenths of a degree °C).
TEMPERATURE_1 Provides the temperature values measured by probe 1. The
temperature reading is in the range from -50°C to +90°C;
TEMPERATURE_2 Provides the temperature values measured by probe 2. The
temperature reading is in the range from -50°C to +90°C;
TIME_STAMP Value expressed in hours:minutes associated with the last
wireless data transmissions received. This can be used to
synchronise the measurements from dierent units with
the same clock. Variable added by the Access Point for
each sensor.
TRANSM_CYCLE Denes the wireless data transmission time to the Access
Point. The value is set in seconds, but must correspond
(rounded o) to a multiple of 60, thus in minutes (see
note 2).
TX_MESSAGE_CNT Wireless network parameter for internal use
TX_POWER Wireless network parameter for internal use
Tab. 0.dTab. 1.d
Note:
• The two values provide an indication of the wireless signal levels seen from
the sensor and the Access Point. The minimum value must be greater than
8, for medium reception from 15 to 30, and excellent for values greater than
30.
• To maximise battery life, the number of transmissions should be limited to
the minimum possible.

13
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
2. BP SE SENSOR BUILTIN PROBE
The BP SE Sensor is designed to be positioned directly inside the showcases, tted using its own fastening bracket. The rear features
metal shielding that, combined with the thermal insulation inside the shell, prevents the formation of frost at the rear of the sensor,
and consequently better thermal insulation from the wall.
Fig. 2.a
2.1 Functions implemented and supervisor
variables available
• Instant temperature measurement performed every minute.
• Measurement ltering with weighted average based on parameter setting
for product temperature simulation.
• Data transmission at settable intervals, in minutes (the parameter aects
battery life).
• Monitoring of temperature thresholds for high temperature (HACCP) or low
temperature (product freezing) alarm signals.
• Automatic mode with preset parameters according to the showcase/
display case (normal, low temperature or open shelf).
• Local mode for Clean showcase status signal. Activating the Clean button
disables the high temperature alarms.
• TimeStamp for recording the instant measurement, expressed in hh:mm.
• Battery level in mV and residual charge in mAh.
• Wireless signal level in dBm +100 (less than 8=low, 15-30=medium, higher
than 30=excellent);
• Temperature alarm status related to the high and low thresholds.
2.2 Sensor conguration
The sensor is supplied with the address ID set to 127 and cannot be used with
the default ID; the range of available addresses is from 16 to 126. To assign the
ID use the rTM SE handheld accessory. For details on the address assignment
procedure see the instructions in the chapter on the rTM SE handheld further
on.
In emergency situations a new ID can be assigned (limited to the range from
16 to 99) using a magnet (e.g. magnetic screwdriver Carel code 0000000722),
as follows:
1. Position the magnet on SW1, holding it in position when the green LED
comes on;
2. The following will be shown in sequence:
- Green LED ON for 2 to 3s then OFF for 3 to 4s;
- Orange LED ON for 3 to 4s;
- Remove the magnet when the LED switches o;
- After a few moments the LED comes on yellow for 1s. This indicates
that programming procedure is active (if no actions are performed,
programming mode ends after 4/5 s, indicated by a double yellow ash,
leaving all the settings unchanged);
3. Move the Clean switch SW2 up and down a number of times equal to the
tens of the serial address being set (e.g. 10, once – 50, ve times). Each
time switch SW2 is moved up the red LED comes on for 1 s, (conrming
stimulation);
4. Subsequently use the magnet to stimulate switch SW1 a number of times
equal to the units (e.g. 1, once – 5, ve times). Each time the magnet moves
over the switch the green LED comes on for 1 s (conrming stimulation).
The order is not important (tens or units rst);
5. After 4/5 s the sensor exits the procedure, with the yellow LED ashing
twice (indicating the end of serial address setting mode);
6. Subsequently the sensor shows the serial address using a sequence of
ashes repeated cyclically three times. To read the codes, see the chapter
“Display sensor serial ID” further on.
7. Moving switch SW2 up interrupts the cycle;
The sensor address has been set and it’s ready to be bound to an Access Point.
The procedure can be performed before or after binding to the Access Point.
Make sure not to assign duplicate serial addresses, also considering other
devices in the network. For further information and explanations on the
procedure, see the rTM SE system installation guide.
Binding procedure
Binding is a special procedure used to associate the sensors with the Access
Point. Once completed, the sensors will send the temperature data measure
wirelessly only to the Access Point dened as its parent. Following this, the
Access Point will forward the data to the Modbus® RTU RS485 serial network.
The binding procedure requires the activation of the Access Point wireless
network and activation of the conguration switch SW1 using a magnet (see
the gure), done by passing the special magnet over magnetic switch SW1
for a few seconds. The LEDs will come on in sequence: green (1s), yellow (4 to
5s), green (6 to 10s). If at the end of the sequence the red LED ashes briey (1
to 2s), binding with the Access Point has failed. If the operation is successful,
successively activating switch SW1 will start manual data transmission,
signalled by the green LED ashing quickly twice.
If the automatic or manual data transmission fails, the red LED will ash briey
after the green LED comes on.
After this operation the sensor will start sending data on the temperature
measured, in the time interval set by parameter. Check that the LED comes on
for a few seconds at regular intervals, based on the transmission time set for
parameter (HR_01 TRANS_CYCLE). When the operation has ended close the
wireless network on the Access Point. The wireless network can be opened
and closed using the rTM SE handheld accessory.
Display sensor serial ID
To check the sensor serial address, proceed as follows:
• Move switch SW2 (CLEAN) up, stimulate SW1;
• The LED starts ashing in sequence. Count the number of the ashes to
calculate the hundreds (Yellow), tens (Green) and units (Red). Removing
the magnet or lowering the button exits the display probe serial address
procedure.
• Move switch SW2 back down.
Yellow Red Green
X 100 X 10 X 1
Hundreds Tens Units
Tab. 2.a
Example
0 yellow ashes 5 red ashes 7 green ashes
057
Sensor address ID=57
Tab. 2.b
Resetting the sensor (maintaining the serial address)
The reset procedure is required when the sensor needs to be moved and
associated with another wireless network (dierent Access Point). This
operation may be required to recongure the sensor in a dierent wireless
network. The value of the serial address remains the same, and after a new
binding operation the sensor is reactivated in the wireless network. To reset
the sensor, proceed as follows:
1. Place the magnet near magnetic switch SW1 (the green LED will come on);
2. Hold the magnet in place until the green LED goes o and the yellow LED
comes on (after approx. 6 to 10 sec.);
3. When the yellow LED comes on, move the magnet immediately away
from the sensor and check that the LED ashes quickly before going o
(RESET COMPLETE).
To check that the sensor has been reset, proceed as follows:
1. Make sure the Access Point wireless network is closed (L1 ashing slowly
1s);
2. Stimulate switch SW1 on the sensor with the magnet;
3. Check that LEDs come on in the following sequence:
- green LED (1 s);
- yellow LED (4 to 5 s);
- green LED (15 s);
- red LED (1 s);
Make sure that there are no sensors with the same serial address in the new
network. If this is the case, assign a new serial address.

14
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
RESET sensor and assign default serial address (=127)
To restore the sensor serial address to the default value, proceed as follows:
1. Place the magnet near magnetic switch SW1, the green LED will come on.
2. Hold the magnet in position until the green LED goes o and the yellow
LED comes on (after approx. 6 to 10 s);
3. Immediately remove the magnet from the sensor and at the same time
move the CLEAN switch (SW2) up, making sure the yellow LED ashes a
few times.
4. Move the CLEAN switch to the OFF position and make sure the yellow LED
completes a rapid sequence of ashes (RESET COMPLETE);
Otherwise repeat the procedure.
To check that the sensor has eectively been reset, proceed as follows:
1. Make sure the Access Point wireless network is closed (L1 ashing slowly
1 s);
2. Stimulate the switch SW1 with the magnet;
3. Check that LEDs come on in the following sequence:
- green LED (1 sec.);
- yellow LED (4 to 5 sec.);
- green LED (15 sec.);
- red LED (1 sec.);
Completing the reset procedure and assigning the default serial address
returns the sensors to the same status as a new device.
To assign a new address, repeat the serial address assignment procedure.
Note:
1. The sensor can only be reset if it has already been bound to an Access
Point.
2. Note that, after resetting the sensor, the number of devices set for the Access
Point remains unchanged. Realignment will occur after a maximum of around
2 hours.
Meaning of the switches and LED signals
CLEAN
SW2
SW1
Led
NTC
Fig. 2.b
Key:
SW1 Internal magnetic conguration switch (above the LED, labelled).
Can be activated by external magnet
SW2 Magnetic CLEAN switch (open = CLEAN MODE)
LED Two-colour red/green (yellow if both are on)
NTC Located inside the case in thermal contact directly with the front wall
The following table describes how the LEDs ash whenever SW1 or SW2 are
stimulated or when data is transmitted.
Action LED sequence (times in s.) Meaning of the signal
Stimulating SW1 /
data transmission
Green ashing (approx. 1s) Communication with
Access Point occurred
correctly
Stimulating SW1 /
data transmission
Green ashing (approx. 1s) red
ON (approx. 0.5s)
Communication with
Access Point NOT
successful
Stimulating SW1 /
data transmission
Green ashing (approx. 1s) OFF
(approx. 1s)red ON (approx.
0.5s)
Communication with
Access Point NOT
successful
Stimulating SW1 Green ON (approx. 1s)yellow
ON (4..5s) green ON (approx.
15s) red ON (approx. 1s)
BP SE Sensor in Reset
status
Binding with Access
Point failed
Stimulating SW1 Green ON (approx. 1s)yellow
ON (4..5s) green ON (6..10s)
OFF
Binding with Access
Point successful
Open CLEAN cover
(SW2)
Red ON (approx. 1s)green ON
(approx. 0.5s)
CLEAN mode activated
Reset procedure Green ON (approx. 2..3s) OFF
(approx. 6..7s)yellow ON
(approx. 2..3s)OFF (approx.
1s)yellow ashing (approx. 1s)
Sensor being reset
Reset procedure
and assign default
serial address
Green ON (approx. 2..3s)OFF
(approx. 6..7s)yellow (approx.
2..3s) yellow ashing (depends
on when the CLEAN door
is closed)OFF (approx.1s)
yellow ashing (approx. 1s)
Reset sensor plus return
serial address to default
value
Tab. 2.c
Note: the LED is two-colour, red and green, which becomes yellow when both
LEDs are on at the same time. There may be dierent shades of yellow due to
dierent tolerance in the brightness of the red and green LEDs.
2.3 Sensor activation
When the sensor is put in SLEEP mode using the rTM SE handheld during
the procedure to assign the serial address (no transmission-minimum power
consumption), the sensor is eectively in standby; nothing is transmitted
until movement of the CLEAN switch is activated (sleep status). Activation is
not reversible, and the sensor will send the temperature measured every 16
min (default value) if the Access Point that the sensor has been bound to is
switched on.
To exit sleep mode, proceed as follows:
• Power up the Access Point;
• Move the CLEAN switch to the ON position (SW2);
• Make sure the red LED comes on for a few seconds.;
• When the red LED comes on immediately the CLEAN switch to the OFF
position;
• The LED on the sensor remains on until it has connected to the Access Point.
If the operation fails, the sensor returns to sleep mode, if however it’s successful
normal operation will resume, with data being sent every 16 min.
Check operation by stimulating the sensor.
Resetting the sensor in sleep mode
If the sensor needs to be reset when in sleep mode (Access Point network
parameters forgotten), proceed as follows:
• Move the CLEAN switch to the ON position (SW2);
• Wait for the red LED to come on;
• Stimulate SW1 continuously while the red LED remains on.
• Keep SW1 stimulated until the LED ashes (yellow) ;
• Remove the magnet from SW1 and move switch SW2 back down.
• Reset completed.
Led
Posizione switch OFF
Switch OFF position
Posizione switch ON
Switch ON position
Fig. 2.c
Parameters and functions
The BP wireless sensor reads the temperature and manages the associated
alarms at one minute intervals.
The data is then transmitted at the intervals set by parameter, according to the
application and the expected battery life. The sensors work most of the time in
low power mode, so as to save battery power. They are activated to make the
measurements and send the data at the preset time.
Activate switch SW1 to send the sensor data manually, or check the connection.
The CLEAN button is used to set cleaning status or deactivate the showcase,
thus disabling the high temperature alarms.
When returning from CLEAN mode, the high temperature alarms are disabled
for a time equivalent to the auto-conguration cycle (AUTO_DELAY).

15
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
The sensor takes individual instant temperature measurements, however can
also provide a weighted average, used to better approximate the product
temperature. The logic for the alarms and all the other functions depends on
the instant temperature measurement.
Wireless communication is activated automatically in the following situations:
• moving the CLEAN mode switch (SW2);
• stimulating the magnetic switch (SW1).
In all other cases, data transmission is dened by the set transmission cycle.
Note:
• The temperature measurement, with the update of the instant and average
values, is performed at 1 minute intervals.
Important: the value is displayed on the supervisor after the set sensor
transmission time .
The average temperature value is calculated using the following formula:
Temp_AVG = (Temp_AVG-1 * (M - 1) + Temp_Ist) / M
Where:
• Temp_AVG-1 Previous average temperature value
• Temp_Ist Instant temperature measurement
• M Average weight (= AVERAGE_PARAM)
The average function also introduces an average measurement delay with a
time constant equal to the average weight value (in minutes).
High temperature alarm function:
Alarm delay
HIGH_TEMP_DELAY
Alarm delay
HIGH_TEMP_DELAY
EN_HI_TEMP_ALM
LO_TEMP_ALM_1
HI_TEMP_ALM_1
Reset counter alarm
Start counter alarm delay
HI_TEMP_TRESHOLD
LO_TEMP_TRESHOLD
Temp. °C
Time
Alarm ON
Fig. 2.d
• When the threshold is exceeded, the alarm is signalled only if this persists
for a time greater than the delay set;
• If the temperature returns within the threshold before the delay time, the
accumulated count is reset;
• The alarm is reset instantly when the temperature returns within the
threshold.
Default values for the modes settable for the MODE_PARAM
parameter
MODE 0
Generic use
MODE
1LOW
showcases
MODE
2MED
showcases
MODE 3MED
shelf cases
High temp. thre-
shold
-15 °C -15 °C +10 °C +8 °C
Low temp. thre-
shold
-40 °C -40 °C -2 °C -2 °C
HACCP delay 180 min. 120 min. 120 min. 120 min.
Average weight 1 (Instant) 16 (16 min.) 8 (8 min.) 12 (12 min.)
Tab. 2.d
Automatic conguration procedure
The automatic recognition procedure is used to recognise the type of
showcase and consequently congure the parameters for the showcase
that the sensor is installed on. The automatic recognition cycle is activated (if
enabled by EN_AUTO_CONF):
• When the temperature falls below the threshold MODE_AUTO_TRESH;
• When returning from CLEAN mode, closing the switch;
• When a previous cycle is completed.
When the AUTO_DELAY time has elapsed, if the following conditions are true:
• Final temperature rise less than 1°C/h;
• Final temperature within a xed band of temperatures for the various types
of showcase:
- medium temp. showcases = from -2°C to + 6°C
- low temp. showcases = less than -10°C.
The MODE_PARAM parameter is given the new value corresponding to the
type of showcase and the associated values for the alarm thresholds, alarm
delay and average weight are loaded;
Note:
• In the event of increases in temperature for low temperature showcases,
the recognition procedure is disabled for 3 times the value of AUTO_DELAY,
to avoid false recognitions.
• The temperature alarms are always enabled, if MODE_PARAM and
consequently the associated parameters are changed, the alarm logic
depends on the new parameters.
• The parameters associated with each mode (0-3) are saved separately and
permanently, and are loaded automatically when the mode is changed.
• The values of the parameters associated with the mode must be set (by the
supervisor) making sure that MODE_PARAM does not change, otherwise
the values transferred may be ignored.
2.4 Technical specications
Power supply 3.6V 2500 mAh lithium battery,“AA”size
Maximum power input 100 mW
Battery life in normal operating
conditions
From 3 to 8 years, depending on the tran-
smission time set.
(CAREL is not responsible for the specied
battery life)
Radio frequency specications Frequency: selectable from 2405 to 2480 MHz
Power transmitted: 0dBm
Wireless protocol: ZigBee
Operating conditions -40T50°C
Storage conditions -20T60°C
humidity range: <80% RH non-condensing
Precision of temperature measu-
rement
± 1 °C -10T30°C;
± 2 °C -30T40°C
Response time to temperature
variations
> 20 minutes
Compliant with EN 13485
Index of protection against atmo-
spheric agents
IP65
Classication according to protec-
tion against electric shock
Can be integrated into class I or class II
appliances
Environmental pollution Normal
PTI of insulating materials 250 V
Period of stress across the insula-
ting parts
Long
Category of resistance to heat
and re
category D (box and cover)
Immunity against voltage surges category 1
Software class and structure Class A
Disposal observe local legislation for the disposal of
electrical material
Product code WS01U01M0 - Wireless sensor ver. BP SE IP65
-40 to 50°C
Accessories WS00BAT000 Battery
WS00B01000 Plastic case only
0000000722 Magnet for activating SW1
Tab. 2.e

16
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
2.5 List of parameters and variables, BP SE
Sensor
Below is the table of supervisor parameters for the BP SE Sensor.
Variable Index Name Description Def. Min Max UoM “Type R/W”
HR0 ‘CMD_PASSW_1’ ‘Command Password (1)’ 0 0 65535 - R/W
HR1 ‘TRANSM_CYCLE’ ‘TX data cycle time’ 960 60 3600 sec R/W
HR2 ‘HI_TEMP_TRESHOLD’ ‘Threshold high Temp.’ -150 -400 500 0,1°C R/W
HR3 ‘LO_TEMP_TRESHOLD’ ‘Threshold low Temp.’ -400 -400 500 0,1°C R/W
HR4 ‘HIGH_TEMP_DELAY’ ‘Delay High Temp. Alarm’ 120 0 254 min R/W
HR5 ‘MODE_AUTO_TRESH’ ‘Threshold Auto Temp.’ 120 0 500 min R/W
HR6 ‘AVERAGE_PARAM’ ‘Parameter Avg-readings’ 16 1 60 - R/W
HR7 ‘AUTO_DELAY’ ‘Delay for AUTO-Cong’ 120 2 254 min R/W
HR8 ‘MODE_PARAM’ ‘Par. MODE for cabinets’ 1 0 3 - R/W
HR9 ‘OFFS_TEMP’ ‘Oset Temperature Measure’ 0 -99 99 0,1°C R/W
HR10 ‘MIN_RSSI_LEVEL’ ‘Minimum rssi level counted (internal use)’ 0 0 99 - R/W
HR11 ‘CNT_REJOIN’ ‘Max counter value before rejoin (internal use)’ 30 1 255 - R/W
IR0 ‘MACHINE_CODE’ ‘Unit type - machine code’ 63 - - - R
IR1 ‘FW_VERSION’ ‘Firmware version (Major/Minor)’ 2051 - - - R
IR2 ‘TX_MESSAGE_CNT’ ‘Total Number of TX radio messages’ 0 0 65535 - R
IR3 ‘RX_MSG_LEVEL’ ‘Radio signal Level’ 0 0 100 dBm+100 R
IR4 ‘ID_SER_ADDR’ ‘Carel_ID Serial_Address DIP-SW value’ - 16 127 - R
IR5 ‘BATTERY_LEVEL’ ‘Battery Level’ - 0 3600 mV R
IR6 ‘AVG_TEMPERATURE’ ‘Temperature average Value’ - -500 1000 0,1°C R
IR7 ‘TEMPERATURE ‘ ‘Temperature Value’ - -500 1000 0,1°C R
IR8 ‘BATTERY_CHARGE’ ‘Counter battery remaining charge’ - 0 65535 - R
IR9 ‘MAC_ADDR_0’ ‘Unit unique identier Mac-Address LSB’ - 0 65535 - R
IR10 ‘MAC_ADDR_1’ ‘Unit unique identier Mac-Address MSB’ - 0 65535 - R
IR11 ‘LAST_RX_DELAY’ ‘Time from last AP Rx message’ - 0 65535 - R
IR12 ‘RX_MESSAGE_CNT’ ‘Counter - AP Rx messages’ - 0 65535 - R
IR13 ‘TIME_STAMP’ ‘Time stamp for Temp. readings (100*hour+minute)‘ - 0 2359 R
IR14 ‘AP_RX_RADIO_LEV’ ‘Radio Lev. for AP Rx messages’ - 0 100 dBm+100 R
IR15 ‘NETWORK_ID’ ‘Network address‘ - 0 65535 - R
IR16 ‘MIRROR_IS’ ‘Mirror Input Status (internal use)’ - 0 65535 - R
CS0 ‘EN_CMD_PW’ ‘Trig. PWD (internal use)’ 0 0 1 - R/W
CS1 ‘EN_HI_TEMP_ALM’ ‘Enable High Temp. Alarm’ 1 0 1 - R/W
CS2 ‘EN_AUTO_CONF’ ‘Enable auto conguration MODE’ 0 0 1 - R/W
CS3 ‘EN_SCAFFALE’ ‘Type of cabinet ( 1= scaale)’ 0 0 1 - R/W
IS0 ‘ALM_BATTERY’ ‘Battery Alarm’ - 0 1 - R
IS1 ‘ALM_GENERAL’ ‘Unit General Alarm’ - 0 1 - R
IS2 ‘ALM_PROBE_1’ ‘Temperature sensor Alarm’ - 0 1 - R
IS3 ‘HI_TEMP_ALM_1’ ‘High Temperature Alarm’ - 0 1 - R
IS4 ‘LO_TEMP_ALM_1’ ‘Low Temperature Alarm’ - 0 1 - R
Tab. 2.f
Key:
HR = Holding register
IR = Input register
CS = Coil Status
IS = Input Status

17
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
2.6 Installation notes
The sensor is installed on the bracket supplied as follows:
1. Fasten the bracket to the wall with two screws, supplied together with the
sensor, considering that the unit being installed is a radio device, and thus
taking the necessary precautions;
2. Couple the sensor to the bracket, making sure it clicks and locks into place.
N.B. To remove the sensor from the bracket, lift the release spring using a
suitable screwdriver and lift the sensor.
click
Fig. 2.e
2.7 Physical dimensions
7.527 83.9
71.634
SW1
SW2
NTC
LED
44
50
Fig. 2.f
2.8 Replacing the battery in the BP SE
Sensor
The case of the BP SE wireless sensor has been designed to provide high
protection. When opening the two plastic shells to replace the battery, the
locking catches may be damaged or break. Consequently, the spare battery
is supplied together with a new case. Take maximum care when removing
the electronic board from the old shell and placing it in the new one, so as to
not damage the electronic components. Make sure battery polarity is correct.
Remove the product label from the old case and place it on the new one.
Rules for disposing of the battery
Do not dispose of the product as municipal waste; it must be disposed of
through specialist waste disposal centres.
The product contains a battery that must be removed and separated from the
rest of the product.
Improper use or incorrect disposal of the product may negative eects on
human health and on the environment.
The public or private waste collection systems dened by local legislation
must be used for disposal.
In the event of illegal disposal of electrical and electronic waste, the penalties
are specied by local waste disposal legislation.
2.9 Application examples
Supermarket showcases
Fig. 2.g
Fig. 2.h
Example of supermarket layout and installation connections
RS 485 ModBus
Router/Bridge
Fig. 2.i

18
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
3. EP SE, SA, SI SENSORS AND CI PULSE COUNTER
3.1 Parameters and functions
The wireless devices read the temperature and manage the associated
alarms at intervals set by the transmission time parameter, according to
the application and the expected battery life. The sensors work most of the
time in low power mode, so as to save battery power. Press the button or
stimulate the magnetic switch to send the sensor data manually, or check the
connection.
3.2 Description of the acquisition process
The devices acquire all the values before sending their status to the Access Point.
Consequently, the device sampling interval is equal to the transmission time.
3.3 Device conguration
Select the desired network address using the 8 dipswitches (0=OFF; 1=ON) as
shown in the table. The possible sensor addresses are from 16 to 126.
Address Dipswitch Notes
12345678
0..15 x x x x x x x x address not allowed (*)
16 0 0 0 0 1 0 0 0
17 1 0 0 0 1 0 0 0
18 0 1 0 0 1 0 0 0
19 1 1 0 0 1 0 0 0
20 0 0 1 0 1 0 0 0
...127 0 1 1 1 1 1 1 1 Reserved. Do not use
128,...199 1 1 1 0 0 0 1 1 address not allowed (*)
200...256 x x x x x x x x address not allowed (*)
Tab. 3.a
For the complete list see the table at the end of the manual.
(*) The address may be set however the device cannot connect to the Access
Point/Router. Pressing the button the LED ashes quickly in sequence to
indicate an invalid address.
EXAMPLE required sensor address setting 117:
Decimal value: 117
Conversion to binary notation:(MSB) 0111 0101 (LSB)
Reverse the value of the string (10101110) and assign dipswitches from (LSB)
1 to 8. (MSB).
Dipswitch
1 2 3 456 7 8
x x x xxx x x
0 0 0 010 0 0
1 0 0 010 0 0
0 1 0 010 0 0
1 1 0 010 0 0
0 0 1 010 0 0
0 1 1 111 1 1
1 1 1 000 1 1
x x x xxx x x
Tab. 3.b
This section provides all the information common to the devices whose serial ID is congured by dipswitch. Refer to the specic instructions for each device.
EP SE SA Sensor SI Sensor CI Pulse Counter
Fig. 3.a

19
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
3.4 Binding procedure
Binding is a special procedure used to associate the sensors with the Access
Point. Once completed, the sensors will send the temperature data measure
wirelessly only to the Access Point dened as its parent. Following this, the
Access Point will forward the data to the Modbus® RTU RS485 serial network.
Before performing this operation, make sure that the sensor serial address has
been set.
After having opened the domain on the Access Point (see the instructions in
the chapter on the Access Point), proceed as follows on the sensor:
Remove the protection from the contact on the battery to power up the
device;
Check that the LED comes on for a few seconds with brief ashes.
Press the button once or activate the magnetic switch. Pressing it again
activates a procedure to check the quality of the wireless signal (see the
chapter“Analysing wireless signal quality”);
LED L1 on the sensor remains on until connection to the Access Point is
complete, ashing for around 10s, then L1, L2 and L3 ash together for a few
seconds (wireless network connection).
The procedure for analysing wireless signal quality then starts for around 1
minute. The following come on in sequence:
1. L1 Indicates wireless transmission has occurred;
2. L1-L2 Indicates the signal has been received by the Access Point;
3. L3 ashes from 1 to 3 times, based on the quality of the wireless signal;
- 1 ash, wireless connection with minimum signal strength;
- 2 ashes, wireless connection with medium signal strength;
- 3 ashes, wireless connection with excellent signal strength;
Button T1 is connected in parallel with the magnetic switch. The case does not
need to be opened to stimulate the sensor for communication
Note: if LED L1 ashes once instead of remaining on, it means that the sensor
has already been bound to an Access Point. In this case, reset the sensor (see
Resetting the devices)
The Access Point shows that connection has been made by LED L3 coming
on for around 1s., even if another node in the network is sending a message.
Check the conguration: the sensor will be correctly bound if whenever the
button is pressed or the magnetic switch is activated, the LEDs come on for a
1 min sequence.
• L1, on for 1s;
• L1-L2, on for 1s;
• L3, ashes from 1 to 3 times, based on the quality of the wireless signal;
- 1 ash, wireless connection with minimum signal strength;
- 2 ashes, wireless connection with medium signal strength;
- 3 ashes, wireless connection with excellent signal strength;
For the EP SE Sensor, in normal operation LED L1 ashes for 1s every 20s. In
general, for the other devices, the LED comes on whenever data is sent, and
consequently based on the device transmission time.
Important: the sensor binding operation may fail if:
- the distances are high and/or there is infrastructure that does not allow
communication between the devices (see the example of sensor S2 in Figure
4.c);
- the maximum limit of sensors allowed for the Access Point has been reached
(max 30). In this case, an additional Router-Bridge is needed for up to a
maximum of 60 sensors.
3.5 Resetting the sensor (unbinding)
The reset procedure is required when the sensor needs to be moved and
associated with another wireless network (dierent Access Point). This
operation may be required to recongure the sensor in a dierent wireless
network. The value of the serial address remains the same, unless the
conguration dipswitches are moved. After a new binding operation the
sensor is reactivated in the wireless network. To reset the EP SE Sensor, proceed
as follows:
1. Remove the battery (press the button to discharge any residual energy in
the circuit) and replace the battery in its socket (LEDs L1, L2, L3 come on
at the same time, then ash quickly and switch o).
2. Immediately after the LED have switched o (within a few seconds) press
button T1 until the pairs of LEDs L1-L3 and L2 ash alternately.
3. Release the button. LEDs L1, L2, L3 will ash briey and then switch o.
To make sure the sensor has eectively been reset:
1. Make sure the Access Point wireless network is closed (L1 ashes slowly
1s).
2. Press button T1 on the sensor and make sure LED L1 comes on and
remains on for around 20 sec.
WARNING: The sensor has been unbound (reset) and maintains the same
network address assigned.
To change the address, remove the battery, move dipswitches 1 to 8, and
replace the battery.
Note:
1. The sensor can only be reset if it has already been associated with an
Access Point;
2. Resetting the sensor does not delete the space reserved inside the Access
Point, which will continue to maintain the data saved inside. Note that,
after resetting the sensor, the number of devices set for the Access Point
remains unchanged. Realignment will occur after a maximum of around
2 hours.
Important: pay careful attention to avoid duplicate assignment of network
serial addresses, so as to avoid overlapping temperature values.
The sensor is supplied with the battery already tted, and with the positive
pole insulated by a protective lm; this must be removed after assigning the
network serial address.
3.6 General warnings
When replacing the battery, strictly observe the following instructions.
The battery may explode if replaced with another of an incorrect type. Dispose
of the used batteries according to the standards in force;
Install the sensor with the cable gland facing downwards;
Replacing the battery
Remove the cover, remove the battery, and replace with another of the same
type. Close the cover again.
Rules for disposing of the battery
Do not dispose of the product as municipal waste; it must be disposed of
through specialist waste disposal centres.
The product contains a battery that must be removed and separated from the
rest of the product.
Improper use or incorrect disposal of the product may negative eects on
human health and on the environment.
The public or private waste collection systems dened by local legislation
must be used for disposal.
In the event of illegal disposal of electrical and electronic waste, the penalties
are specied by local waste disposal legislation.

20
ENG
“Wireless probes”+0300030EN - rel. 1.0 - 16.07.2010
4. EP SENSOR EXTERNAL PROBE
The EP SE Sensor is designed to be tted inside showcases or cold rooms, and can house two external passive NTC temperature probes
10K@25°C (Beta(25/85) = 3435K) and two digital inputs to be used to monitor door and defrost status, or used as generic inputs.
Fig. 4.a
4.1 Functions implemented
• Instant temperature measurement performed every minute, probe 1;
• Instant temperature measurement performed every minute, probe 2;
• Data transmission at a settable interval in minutes (this aects battery life);
• Monitoring of temperature thresholds for high temperature (HACCP) or low
temperature (product freezing) alarm signals.
Main variables available to the supervisory system
• Probe 1 temperature;
• Probe 2 temperature;
• Battery level in mV ;
• Wireless signal level in dBm +100 (8 = low signal, 15 to 30 = medium signal,
greater than 30 excellent).
• Temperature alarm status related to the high and low thresholds.
• Data transmission interval;
• Enable high temperature alarm;
• TimeStamp for recording the instant measurement, expressed in hh:mm;
4.2 Parameters and functions
The EP SE wireless sensors read the temperature and manage the associated
alarms at intervals set by the transmission time parameter, according to the
application and the expected battery life. The sensors work most of the time in
low power mode, so as to save battery power. Press the button on the sensor
to send the sensor data manually, or check the connection.
The main parameters and functions of the sensor are:
-Data transmission activation:
-Wireless transmission is activated in the following conditions:
• Change in status of the digital inputs, door and defrost;
• Change in status of the digital inputs, door and defrost;
• Temperature probe fault alarms;
• Briey pressing the button.
In all other cases, data transmission is dened by the set transmission time.
Logical state of DOOR_POL and DEFROST_POL variables
The following table shows the logical state of the input based on the electrical
state of the contact (open or closed).
Contact state Polarity Logical state of DOOR input
OPEN 1 Door CLOSED
CLOSED 1 Door OPEN
OPEN 0 Door OPEN
CLOSED 0 Door CLOSED
Tab. 4.a
DEFROST_IN_STATUS and DOOR_IN_STATUS = Provide the current logical
state of the two digital inputs.
0 = Door CLOSED 1 = Door OPEN
0 = Defrost NOT Active 1 = Defrost Active
High temperature alarm function:
Temp °C
Time
Alarm ON Alarm ON
HI_TEMP_TRESHOLD
High temperature event
started normal condition
High temperature event
started with open door
LO_TEMP_TRESHOLD
LO_TEMP_ALM_1
HI_TEMP_ALM_1
DOOR
ALARM
Start counter alarm delay
Reset counter alarm
Alarm delay
HIGH_TEMP_DELAY
Alarm delay
HIGH_TEMP_DELAY ADD_HIGH_T_DELAY
Fig. 4.a
The gure shows the high temperature alarm function:
1. when the threshold is exceeded, the alarm is signalled only if this persists
for a time greater than the delay set;
2. if the temperature returns within the threshold before the delay time, the
accumulated count is reset;
3. the alarm is reset instantly when the temperature returns within the threshold.
Enabling and disabling the digital inputs
Management of the door and defrost inputs can be enabled or disabled using
EN_DI_DOOR for the door input and EN_DI_DEFROST for the defrost input.
If not enabled, the inputs are inactive (0). Even when disabled, however, the
logical state of the digital inputs can be identied by reading the value of
IN_1_STATUS for IN_1 (1=Active, 0=Not active) and IN_2_STATUS for input IN_2
(1=Active, 0=Not active). By default the door and defrost inputs are enabled.
The high temperature alarm is activated after the delay if the door input is
active. For correct operation of the temperature alarm signals, the status of the
door and defrost must always be read as inactive, even if the two inputs are
not used. To return both to normal conditions, the value of the two polarity
states can be set to 1 for the door (DOOR_POL) and defrost (DEFROST_POL), or
alternatively the two inputs can be jumpered if not used.
Enable and disabling the analogue inputs
EN_NTC_1 and EN_NTC_2 enable and disable the probe inputs. The probe
inputs can be enabled and disabled using parameters EN_NTC_1 and EN_
NTC_2 respectively. If a probe input is disabled, the temperature reading is
equal to 0°C; in this event, the probe alarm fault is not managed and remains
constantly inactive (0). By default the probe inputs are enabled.
4.3 Technical specications
Power supply 3.6V 2500 mAh lithium battery,“AA” size
Maximum power input 100 mW
Battery life in normal operating
conditions
From 3 to 8 years, depending on the transmis-
sion time set. (CAREL is not responsible for the
specied battery life)
Radio frequency specications Frequency: selectable from 2405 to 2480 MHz
Wireless protocol: ZigBee
Power transmitted: 0 dBm
Operating conditions -40T50°C
Storage conditions -20T60°C
humidity range: <80% RH non-condensing
Precision of temperature measu-
rement
± 1 °C -10T30°C; ± 2 °C -30T40°C
Response time to temperature
variations
> 20 minutes
Compliant with EN 13485
Index of protection against atmos-
pheric agents
IP65
Classication according to protec-
tion against electric shock
Can be integrated into class I or class II applian-
ces
Environmental pollution Normal
PTI of insulating materials 250 V
Period of stress across the insula-
ting parts
Long
Category of resistance to heat and
r e
category D (box and cover)
Immunity against voltage surges category 1
Software class and structure Class A
Disposal observe local legislation for the disposal of
electrical material
Product code WS01W02M00 - Wireless sensor ver. EP SE batte-
ry powered 2NTC- 2DI -50 to 90°C
Accessories WS00BAT000 - Battery, single packge
0000000722 Magnet for activating SW1
Tab. 4.b
Note: the index of protection is maintained only if a single cable is used for
power and RS485 communication with an outside cross-section of less than
8 mm.
This manual suits for next models
10
Table of contents
Other Carel Security System manuals
Popular Security System manuals by other brands

PNI
PNI Safe House PG300 user manual

gefran
gefran 2400 Guide to configuration and network installation

Vega Absolute
Vega Absolute M200 user manual

Medpage
Medpage FIRE CENTURION FC-CH instruction sheet

SeNd Technology
SeNd Technology GSMWhitebox2 user guide

Moore Industries
Moore Industries SPA HLPRG user manual