BodiMetrics O2 Vibe User manual

!
!
!
!
!
!
!
BodiMetrics!
O2Vibe!Sleep%&%Fitness%Monitor!
!
!
!
!
1.!Introduction*and*Intended*Use!................................................................!1!
2.""""""Getting"Started!.........................................................................................!4!
3.!!"#$%&'(&)#*+&,-++.&/&0#1$+""&23$#134!...................................................!6!
4.!O2!Vibe%Care%and%Maintenance!.............................................................!10!
5.!Troubleshooting!.....................................................................................!12!
6.!Limited'Warranty!...................................................................................!13!
7.!Specifications!.........................................................................................!13!
!
!
User’s&Manual!
!!Tracking(Yo u r % Oxygenation*in*a*Smart&Way
!
TM
!

1!
1. Introduction!and!Intended!Use!
Thank&you&for&choosing&the&O2Vibe%Sleep%&%Fitness%Monitor.!The$ O2# Vibe!
can$ be$ used!to# measure,# display# and# store#blood# oxygen# saturation# (SpO2)"
and$pulse$rate%of%adults%and%teenagers(during'both'motion'and'non-motion&
conditions.)It)is)intended)for)personal!use$only.!!
!
The!data!and!results!provided!by!this!device!are!for!tracking!useful!data!
during!sleep!and!exercise!to!promote!healthy!living!and!support!fitness!
goals.!!This!is!not!a!medical!device!and!cannot!be!directly!used!for!
diagnosis!or!treatment!of!any!disease.!!Always!consult!a!physician!if!you!
are!concerned!about!your!health!or!wish!to!better!understand!how!to!
change!your!diet,!exercise!or!sleep!habits.!
!
Please&read&these&instructions&carefully&before&using$this$product$and$keep$
the$manual%and%O2#Vibe!in#a#safe#place.!
!!
1.1 !Warnings!and!Cautions!
•Do#not#use#this#device&during&MRI&examination.!
•The$O2#Vibe!has$no$alarm,$do$not$use$when$alarms$are$required.!
•Please&DO!NOT!squeeze!the!sensor&or&apply&excessive!force!to!the!
sensor!or!cable&as&shown&below&to&ensure&longer&use&and&
accuracy&of&recordings.!
!
!Never!submerge!the!device!in!water!or!other!liquids.!Do!not!
clean!the!device!with
!
acetone!or!other!volatile!solutions.!
!Consult!your!doctor!immediately!if!you!experience!symptoms!
that!could!indicate
!
a"health"issue.!
!Do!not!self-diagnose!on!the!basis!of!this!device!data$ and$
reporting!without!consulting* your!doctor.!!Specifically,!do!not!
start%or%modify%any%treatment%plan%prescribed%by%a%physician.!
!Use!only!USB$cables'and!sensors!or#other!accessories!provided(
with%your%O2%Vibe%or%purchased%directly%from%BodiMetrics.!
!Prolonged!continuous!SpO2!monitoring!may!increase!the!risk!of!
undesirable+ changes+ in+ skin+ characteristics+ such+ as+ irritation,+
reddening,( blistering( or( burns( on( your( fingers( or( wrist.((
Discontinue!use$immediately$if$any$of$these$changes$occur.!

2!
!Check!the!thumb& ring!sensor!site!every!6-8!hours!to!determine!
the!positioning,)circulation)and)skin)sensitivity)of)the)area.))User)
sensitivity(may(vary(depending(on(user’s(health(or(skin(condition.((
For$users$with$poor$peripheral$blood$circulation$or$sensitive$skin$
please&inspect&wrist&band&and&sensor&sites&more&frequently.!
!This!device!is!designed!to!determine!the!arterial!oxygen!
saturation!percentage!of
!
functional!hemoglobin.!Factors!that!
may!degrade!pulse!oximeter!performance!or
!
affect!the!
accuracy!of!the!measurement!include!the!following:!
oexcess ambient light
oexcessive motion
oblood flow restrictors - i.e. blood pressure cuffs
omoisture in the sensor
oimproperly applied sensor
oincorrect sensor type
opoor pulse quality
ovenous pulsations
oanemia or low hemoglobin concentrations
ocarboxyhemoglobin
omethemoglobin
odysfunctional hemoglobin
1.2 How$O2#Vibe!Sleep%&!Fitness'Monitor'Works!
The! O2# Vibe!Sleep% and% Fitness% Monitor% (O2% Vibe)% measures% oxygen%
saturation,!pulse&rate,"steps"and"motion."The"O2#Vibe!works&by&shining&two&
beams&into&the&small&blood&vessels&or&capillaries&of&the$finger,(reflecting(the(
amount'of'oxygen'in'the'blood'and'displaying'the'measurement!on#the#O2#
Vibe’s# screen.""An"internal"accelerometer"tracks"the"steps"and"activity"of"the"
wearer.!
!
1.3 Guide&to&Symbols(!
Symbol
Meaning
Application*part*type*BF
Manufacturer!
FCC
In#conformity# with# FCC# rules# part# 15C,#Operation# is# subject# to#
the$ condition$ that$ this$ device$ does$ not$ cause$ harmful$
interference.!

3!
MRI$unsafe,$Presents!hazards!in!all!MR!environments!as#device#
contains(strongly(ferromagnetic(materials.!
Indicate) separate) collection) for) electrical) and) electronic)
!"#$%&!'()*+,,,-.!
IP22
Against!ingress' of' solid' foreign' objects' ≥12.5mm' diameter,'
Against(dripping!(15°%tilted)!
Follow%operating%instructions%!
Caution!!
SpO2
Percent'oxygen'saturation!
Heart&rate&–!Sleep%mode!
Heart&rate&–!Fitness'mode!
Sleep%mode!
Fitness'mode!
Bluetooth(connected'to'O2'Vibe%APP!
Insert'the'ring%sensor!plug!into%the%device!
Slide&your&thumb&or&finger'into'the'ring%sensor!
SN!
Serial'number!–!found&on&the&underside!of#the#device!
No#alarm#system!

4!
Getting!Started!
!
2.1 Model&O2!Vibe!–!What’s'in'the'box?!
The$model$O2Vibe$package$includes$the$following$items:!
l1"Model&O2Vibe!
l1"Pulse"Oximeter"ring"sensor!
l1"USB"micro"D!charging(cable!
l1!Model&O2Vibe&User’s&manual!
Confirm(that(the(items(listed(are(included(in#the#package.#If#any#item#on#this#
list%is%missing%or%damaged,%contact%your%distributor.!
!
2.2 Turning'your'O2'Vibe!ON/OFF/Changing+Modes!
Press%and%hold%the%button%for%1%second%to%power%ON%the%device%or%change%
screens.'Press'and'hold'button'for'3'seconds'to'power'OFF'the'device.!You $
change'screens'and'modes'as'well'using'this'button$by!pressing!quickly(for(
new$menus$or$functions.!
!
!
!
Note:&To"reset"your"O2"Vibe"device,"please"hold"ON/OFF%button&down!for$8!
seconds'continuously.!
2.3 Charging!
!
!!!!!!Caution!The$ O2# Vibe# device& cannot& be& used& for& any& measurement&
during'charging.!!Please&fully%charge'the$battery$for$ three!hours&before!
first! use.! ! Fully! discharging! the! battery! before! recharging!your% battery!
fully!will!extend!the!overall'life'of'your'O2'Vibe'battery.!
!!
To"charge"your%O2%Vibe!battery:!
1. Connect! the!small% end% of% the!USB$ charging$ cable$ to$ the$ multi-
Clasp!
Wrist&Band!
Main%Unit!
ON/OFF/Menu)
Button!
Multi-functional!USB$Connector!
for!External)SpO2)Cable)and)Charging)Cable!
!

5!
functional*connector*on#the#O2#Vibe#device&as#shown#below.!
! ! !
2. Connect! the! larger! end! of! the! USB$ charging$ cable$ to$ a!personal)
computer!USB$port.!
!
Note:& Manufacturer& recommends& ONLY& charging& your& device&
using&a&personal&computer&and&NOT$directly$to!an#AC#outlet.!
!
2.4 O2#Vibe!requires'a'companion'O2'Mobile'APP!
Before&using&the!O2#Vibe!for$the$first$time,$please&download!and$install!the$
O2!Vibe% APP!from% iTunes% or% Google% Play% APP# Stores" (do# not$ download'
BodiMetrics+or+BodiMetrics+Plus+apps+as+they+support+other+products).!!O2#
Vibe%does%not%work%with%Windows%Mobile%devices.!!!
!
Android'app!users%only;%if%a%PIN%is%requested,%please%use%default%“8888”.!
!!!
Note:"The$O2$Vibe$does!not$work$with$Windows$based$mobile$devices.!
2.5 Pairing$ O2$ Vibe$ to$ O2# Vibe# mobile# APP!via!
Bluetooth!
The$O2$Vibe$device$has$a$built$in$Bluetooth$radio$that$connects$with$your#
O2# Vibe# mobile# APP# to# store# Sleep# or# Fitness# recordings;' Update& device&
software;* display* exercise* metrics* via* mobile* Dashboard;* and* set* SpO2*
Threshold)%,)Vibration!On/Off!and$Intensity$using$your$O2#Vibe#mobile#APP!
when%O2%Vibe%device%is#connected!via$Bluetooth.!
!
In#Sleep%Mode,!you$do$not$need$to$connect$to$O2#Vibe#mobile#APP!via$
Bluetooth.!!!Connecting)during)Sleep)Mode)to)your)O2#Vibe#mobile#APP!will$
reduce&your&battery&life&during&overnight!recording.!!Typical(Sleep(recording(
time%is%10%hours%before%you%will%need%to%recharge%your%O2%Vibe%device!
battery.!
!!!!
In# Fitness!Mode,!you$ do$ not$ need$ to$ connect$ to$ O2# Vibe# mobile# APP!via$
Bluetooth(unless(you(want(to(display(your(real(time(SpO2%,(HR(and(Steps(
on# your# mobile# app# Dashboard# during# exercise.# # Your" O2" Vibe" will" record"
your% exercise% metrics% when% in% Fitness% Mode% even% when% you% are% not%
connected' to' the' O2# Vibe# mobile# APP!via$ Bluetooth.$ $ Typical$ Fitness$

6!
recording) time) is) 5) hours) before) you) will) need) to) recharge) your) O2) Vibe)
battery.!!
2.6 How$to$Set-up#your#O2#Vibe#Device#Settings!
After& Bluetooth!is# paired,( follow( the( instructions( in( the( APP( to( set( your(
target!number' of' step' or' Goal," switch" ON/OFF!Vibration," adjust" the"
vibration)Intensity!and$the$lowest$SpO2$Threshold!!"#$%&'()*+),-./)(0)+#()&1)
your%O2#Vibe!Wrist&Pulse&Oximeter.!
2.7 Data$Storage!
Your% O2% Vibe% stores% 4% recordings,% Sleep% or% Fitness,% so% it% is% recommended%
you$ synch$ your$ O2$ Vibe$ device$ to$ your$ O2# Vibe# mobile# APP!at# minimum#
every% other% day.% % The# O2# Vibe# will# delete# the# oldest!recording) when) you)
record& your% 5th!recording.* * All* recordings* on* your* O2* Vibe* device* will*
automatically* sync* and* transfer* to* your* O2# Vibe# mobile# APP!when% you%
connect&via&Bluetooth.!
3. Using!O2!Vibe!Sleep!&!Fitness!Monitor!
3.1 Prior%to%Use!
Before& using& O2# Vibe," see# Warnings( and( Cautions!on# page# 1,# and# pay!
attention!to!the!following!points!in!order!to!obtain!precise!measurements.!
!The$finger$inserted$in$SpO2!sensor&must&be&clean&to&ensure&proper&
reading.!
!Any$ of$ the$ following$ conditions$ may$ cause$ inaccurate$
measurements,*including*but*not*limited'to:!
oFlickering*or*very*bright*fluorescent+lighting;!
oPoor$blood$circulation;!
oLow$hemoglobin;!
oHypotension,+ severe+ vasoconstriction,+ severe+ anemia+ or+
hypothermia;!
oAny$ tests$ recently$ performed$ on$ you$ that$ required$ an$
injection(of(intravascular(dyes.!
!The"O2"Vibe!may$not$work$if$you$have$poor$circulation.$Rub$your$
finger'to'increase' circulation,'reposition!your% finger% or!move% the%
sensor& to& another& finger.!!Your% thumb% is% best,% then% your% index%
finger'on'down'to'your'smallest'finger.!
!The$ O2# Vibe!measures' oxygen$ saturation$ of$ functional$
hemoglobin.+High+levels+of+ dysfunctional+hemoglobin+(caused+by+
sickle' cell' anemia,' carbon' monoxide,' etc.)' could' affect' the'

7!
accuracy&of&the&measurements.!
!Field& strengths& from& fixed& transmitters,& such& as& base& stations& for&
radio& (!"##$#%&'!(&)#"**+, -"#"./(0"*, %0), #%0), 1(23#", &%)3(*4,
amateur' radio,' AM' and' FM' radio' broadcast' towers,' and' TV'
broadcast)towers)may)affect)accuracy.!
3.2 How$to$Wear$O2$during$Recording*or"Displaying!
1. Connect' the' external' SpO2!sensor& to& the& multi-
functional* connector!port% on% the% side% of% your%
O2#Vibe#device.#Slide#the!SpO2!ring%sensor!onto$
your%left%hand%finger!as#pictured.#Thumb#finger'
is!recommended(for(best(results.!
2. Press%ON/OFF%button%to%turn%on!device,!after!a!
few!seconds,(your(SpO2(reading(will(appear(on(
the$O2#Vibe#screen!display!and$O2#Vibe#mobile#
APP!if# connected' via' Bluetooth.' ' Real-time%
display(uses$the$O2#Vibe#mobile#APP!Dashboard)
function.!
3. If# you$ don’t$ see$ your$ readings!the$ signal' strength' may$ be$ low;!
reposition%or%switch'to'another'finger!&!try$again.!
4. Remember& do& not& squeeze& your& ring&
sensor& forcibly& as& shown& here!as# it# may#
damage&sensor.!
!
3.3 Turning'ON'and'Using'your'O2'Vibe'for'First'Time!
To# turn# ON# your# O2# Vibe# press# the# ON/OFF# button!on# the# side% of% the% O2%
Vibe% device!for$ approximately$ 1-2" seconds." " The" first" screen" you" see" will"
depend%on%whether%or%not%you%have%the%ring%sensor%plugged%into%device.!
!
Function!
Sleep!Mode!
Fitness'Mode!
O2#Vibe#screen#icon!
!
!
!
Parameters!
SpO2,!HR,!
Movement!
SpO2,!HR,$
Movement,)Steps!
Data!measuring!interval!
Every&5!to!10#secs!
self-adapting!
Every&10!secs!
Bluetooth!radio!
OFF!
ON#or#OFF!

8!
Normal'recording'time!
>10!hours!
>5!hours!
!
After& approximately& 5!seconds' your!measurement) results!will$ begin$
showing(if(you(do(not(perform!any$action.!
!
!
If#you$O2$Vibe$is$not"measuring!results,"the"pulse"rate"symbol!!!!!will$flash;$
try$adjusting$the$ring$sensor$on$your$thumb.!
!
!
!
While& in!Sleep% Mode," if!the$ sensor& ring& plug% is# disconnected!from% your!
device!or#the$sensor$slips$from$your$finger,$you$will$see$the$O2$device$begin$
a!countdown( from( 10( seconds!and$ icons$ showing$ you$ to$ reinsert$ the$ ring$
sensor&into&your&device&or&readjust&the&ring&sensor&on&your&finger.!
!
!
!
If# you# don’t# put# on# the# sensor# or# insert# the# plug# into# the# device# in# 10#
seconds,( it( will( stop( measurement( automatically( and( show( your% O2# score!
and$how$many$times!(“Drops”)!your%SpO2%"was!lower&than&Threshold!you$
set.!After& 2!minutes( the$ O2$ Vibe!will$ automatically' turn% OFF.!
!
!
! !
In#Fitness!Mode!if#the#sensor#ring#plug#is#disconnected#from#your#device#or#
the$sensor!ring!slips%from%your%finger,!you$will$see$the$O2$device$begin$a$
countdown(from!10"seconds!and$icons$showing$you$to$reinsert$the$ring$
sensor&into&your%device%or%readjust%the%ring%sensor%on%your%finger.%%After%2%
minutes(the(O2(Vibe(will(automatically(turn(OFF.!
!
!
In# Fitness!Mode," you" can" press" the" button" to" change" screen" to" show" the"
number'of'steps!you$have%taken"at"any"time.!
!
!
!
Device!clock!
Sync%device&clock&with&your&phone!
When%your%device%is%connected%to%App,%the%device%clock%will%be%synced%with%

9!
your% phone% automatically.% You% may% need% to% sync% the% clock% under% these%
conditions:) For) the) first) use;) No) use) for) a) long) time;) After) firmware)
updated.!
!
3.4 How!Sleep%&%Fitness!Modes"Work!
The$device$has!2"operating!modes,"Sleep%Mode!and$Fitness'Mode,!pressing!
the$button$on#the#side#of#your#O2#Vibe#will#switch#between#these!2"modes."!
Function!
Sleep!Mode!
Fitness!Mode!
Parameters!
SpO2,!HR,!
Movement!
SpO2,&HR,#
Movement,)Steps!
Data!measuring!interval!
Every&5!to!10#secs!
self-adapting!
Every&10!secs!
Bluetooth!radio!
OFF!
ON#or#OFF!
Normal'recording'time!
>10!hours!
>5!hours!
3.4.1 Sleep%Mode!
When%your%O2%Vibe%device&is&in&Sleep%Mode!there$are$2!operating*status:!
1. Normal&Mode!and!
2. Hibernation!Mode!
!
When%you%turn%ON%your%O2%Vibe%device!with%the%thumb%ring%sensor%plugged%
in#you#will#be#in#Normal#Mode!where%your%screen& shows& SpO2,&and&Heart&
Rate%measurements.%%To%change%your%mode!press%the%button%or%if#you#wait#
approximately, 2!minutes( the$ O2$ Vibe$ screen$ will$ go$ “dark”$ and$ into$
Hibernation) Mode!automatically* to* eliminate* bright* light* while* you* are*
sleeping.!
!
While&in!Hibernation)Mode!you$may$press$the$button$to!change'the$screen&
display( between( measurements" and# Sleep$ Mode!screen." " The# screen# will#
automatically*go*“dark”!when%there%is%no#user#input#for#3"seconds.!
!
Your% O2% Vibe% is% recording% your% measurements% after% 2% minutes% in% either%
Normal'or'Hibernation'Modes'and'your'results'will'display'on'your'O2'Vibe!
device!once% you% stop% recording& by& turning& OFF& your& O2& Vibe& device& or&
remove& the& thumb& ring& sensor& for& more& than& 10& seconds.& & You& may& also&
review&a&detailed&report&of&your&results'on#the#O2#mobile#APP#after&you&sync&
your% recording% to% your!Apple% or% Android% smartphone," iPad" or" tablet!via$
Bluetooth!-see#Pairing!Bluetooth!on#page#5.!

10!
3.4.2 Fitness!Mode!
When%you%turn%ON%your%O2%Vibe%device!with%the%thumb%ring%sensor%plugged%
in#you#will#be#in#Normal#Mode#where#your#screen#shows#SpO2,#and#Heart#
Rate%measurements.%% To%change%to# Fitness' Mode!press%the$ button$ on# the#
side%of%your%O2%Vibe%to%show%[insert%Fitness%“legs”%icon]%screen.!
!
To# display# real# time# fitness# m easurements# on# your# sm artphone,# iPad# or#
tablet& open& your& O2# Vibe# mobile# APP!and$ the$ device$ will$ automatically$
connect&via&Bluetooth!to#your#O2#Vibe#device.""To"see!your!real-time%SpO2,&
HR#and#Steps#on#your!O2#mobile!APP#select!the$Dashboard$Menu$on$your%
O2#mobile#APP#while#connected#in#Fitness'Mode.!!!
!
Your% O2% Vibe% is% recording% your% measurements% after% 2% minutes% in% Fitness'
Mode!and$ your$ results!will$display$ on$ your$O2$Vibe$device$once$you$stop$
recording) by) turning) OFF) your) O2) Vibe) device) or) remove) the) thumb) ring)
sensor&for&more&than&10"seconds.""You"may"also"review"a"detailed"report"of"
your% results% on% the% O2% mobile% APP% after% you% sync% your% recording$ to$ your$
Apple% or% Android% smartphone,% iPad% or% tablet% via% Bluetooth!-see# Pairing'
Bluetooth!on#page#5.!
4. O2!Vibe!Care!and!Maintenance!
With% proper% care," maintenance!and$ following$ usage$ instructions$ the#
expected' useful' life' of' your' O2' Vibe' device& is!5!years!and$ ring% sensor% is% 2!
years.!!A" one" year!limited!warranty' is' provided' by' the' manufacturer' so'
check%the$limited'warranty'document"at"www.bodimetrics.com!and$register$
your%warranty.!
•Clean!the!device!weekly#by!carefully!swabbing!the$device$surface$with$
a"soft"cloth"or"cotton"swab"with"rubbing"alcohol."!
•Do#not#pour#alcohol#or#other#cleaning#solvent#directly)onto!or#into#the#
device;!its$openings$or$buttons.!
lDo#not#drop#this#device#or#subject#it#to#forceful!impact.!
lDo#not$place$this$device$in$pressure$vessels$or$gas$sterilization$device.!

11!
lDo#not#attempt&to&disassemble&the$device.!
lDo#not#store#the#device$in$the$following$locations:!
olocations)where!device!is#exposed#to#direct#sunlight!
olocations)with)high$temperatures(or(levels%of%moisture!
olocations)in!combustible+environments"or"heavy"contamination!
olocations)near)sources)of)water)or)fire;)or)!
olocations)that)are)subject) to)strong)electromagnetic) influences.!
!
! !

12!
5. Troubleshooting!
Problem!
Possible!Cause!
Possible(Solution!
Device!does!not!
turn!ON!
Battery&may&be&
low!
Charge'battery'for$at$least$1$
hour%and%try%again!
Device!may$be#
damaged!
Please!contact&your&local&
distributor)or)BodiMetrics)at)
(844)%744-8800,$option$2!
The!APP!cannot!
find!O2#Vibe%
device!
The$O2$Vibe$
device&is¬$
communicating*or*
close&enough&to&a&
!"#$%&'()*+,-.#/,
or#tablet!O2#APP!
Press%the%button%for%8%
seconds'to'reset#your#device."
Sync!device!to#O2#APP!within&
2"minutes!after&powering&ON!
or#after#measurement#is#
complete.!!
Make%sure%your%O2%APP!is#
within&reasonable%range%of%
your%O2%Vibe%device.!
SpO2!or#heart!
rate%does%not%
display!
Sensor!positioned)
incorrectly!
Reposition!sensor&on&your&
thumb&or&try&another&finger&
starting'with'Index'finger.''
The$thumb%is%recommended.!
Finger'is'cold!
Warm%finger%by%rubbing!
Finger'is'wet!
Dry$finger$and$sensor!
Excessive(hand%
motion!
Reduce&motion!
Error$4!
Defective(thumb(
ring%sensor!
Reset%O2%Vibe%device%by%
holding(the(ON/OFF(button(
for$8!seconds'continuously!
to#reset#your$device$to$see$if$
Error$4$is$cleared.!
!
Contact'Support'at#(844)#
744-8800,$option$2$for$
replacement*sensor.!
!!

13!
6. Limited!Warranty!
The$ product$ is$ warranted$ to$ be$ free$ from$ defects$ in$ materials$ and$
workmanship, within, the$ one$ year$ warranty' period' when' used' in'
accordance(with$the$provided$instructions.$Warning:!Only!have!the!device!
repaired! by! authorized!service' centers." " " Repairs" or" parts" replaced" by"
unauthorized, service, centers, will, invalidate, your, warranty.!!Contact'
BodiMetrics+ toll+ free+ (844)+ 744-8800,$ option$ 2$ for$ warranty$ repairs$ or$
replacement.!
7. Specifications!
Classifications
EC Directive
MDD, 93/42/EEC
R&TTE, 1999/5/EC
ROHS 2.0, 2011/65/EU
Degree protection against electrical
shock
Type BF
Environmental
Item
Operating
Storage
Temperature
5 to 40°C
-25 to 70°C
Relative humidity (noncondensing)
10% to 95%
10% to 95%
Barometric
700 to 1060 hPa
700 to 1060 hPa
Degree of dust & water resistance
IP22
Physical
Weight
Less than 50 g (main unit)
Display
0.69’’ OLED
Connector
Micro D connector
Wireless connectivity
Built-in Bluetooth
Power Supply
Charge input:
DC 5V ±10%
Battery type
Rechargeable lithium-polymer battery
Battery run time
Sleep Mode: > 10 hours
Activating Mode: > 5 hours
Charge time
Less than 2 hours to 90%
Battery cycle life
500 cycles
!
! !

14!
!
SpO2
Standards
Meet standards of ISO 80601-2-61
Measurement accuracy verification: The SpO2accuracy has been verified in human
experiments by comparing with arterial blood sample reference measured with a CO-
oximeter. The pulse rate accuracy has been verified by ECG Signal Emulator. Pulse
oximeter measurement are statistically distributed and about two-thirds of the
measurements are expected to come within the specified accuracy range compared to
CO-oximeter measurements.
SpO2range
70% to 100%
SpO2Accuracy (Arms)
80-100%:±2%, 70-79%:±3%
PR range
30 to 250 bpm
PR accuracy
±2 bpm or ±2%,whichever is greater
Sleep Recording
Recording time
Up to 10 hours
Data storage
SpO2, pulse rate
Measurement summary
O2 score, Total duration, <90% STAT, Average
saturation, Lowest saturation, Summary(No abnormal
detected, blood oxygen drop detected, Unable to
analyze)
Pedometer
Range
0 to 99999 steps
Vibrator
Event
Steps, low oxygenation
Threshold
Oxygenation: 80% to 90%, adjustable
Steps: 0 to 99999, adjustable
Mobile APP
Operating system
IOS 9.0 or above, Android 4.0 or above
IOS Capability
iPhone 4s and models launched subsequently; iPad 3
and models launched subsequently;
Android Capability
Mobile phone or pad with Bluetooth 2.1 or above
Functionality
Data export, data review, data sharing
!

!
!
!
!
!
!
!
!
!
!
!
!
!
Distributed*in*USA*by*BodiMetrics,*LLC!
www.bodimetrics.com!
!
!
!
!
!
!
!
!
!
!
!
!
PN:255-00917-00!!Version:!A!
Contents'of'this'manual'are'subject'to'changes'without'prior'notice.'!
©Copyright+2016!Shenzhen&Viatom&Technology&Co.,&Ltd.&All&rights!reserved.'!
!
!
!O2#Vibe#Sleep#&#Fitness#Monitor!
Knowing'and'Tracking'Your'Health!
!!!!!
Tracking(Yo u r % O xygenation*in*a*Smart&Way!