Brunson 771 Guide

Models 771, 75, & 76-RH
Optical Tooling Transits
Instrument
Company
Operation and
Field Adjustment Manual

Helping the World Measure
Since 1927
www.brunson.us
Copyright 2009-2011 Brunson Instrument Company
Kansas City, Missouri
8000 E. 23rd Street
Kansas City, MO 64129
Tel: 816.483.3187
Fax: 816.241.1945

TableofContents
Thankyoufor
purchasing
aBrunsonTransit.
Rememberthatour
customersupportdoes
notstopaftershipmentof
aproduct—wearehereto
helpyouwithany
measurementchallenges
thatyoumayhave.
Chapter1:Introduction1
WhatisaTransit?1
WhatDoesaTransitDo?3
TheThreeDifferentTransits4
Chapter2:GettingStartedWithYourTransit7
HandleYourTransitWithCare7
RemovingFromtheCase8
UsingtheShiftingCenter8
RoughLeveling8
PreciseLeveling9
TheReticle11
HowtoProperlyFocus11
FindingInfinityFocus11
PointingtheTelescope12
InstallingLightingAccessories12
Chapter3:MeasuringTechniques15
BuckingIn15
Autoreflection17
Autocollimation19
Collimation23
Collineation26
ReticleProjection27
MeasuringWithMicrometers28
Chapter4:FieldCalibrationChecks35
Plate(“Bull’sEye”)Vial37
MicrometerChecks38
“DoubleCenter”Check40
HorizontalAxisRunout41
VerticalSpindleRunout42
HorizontalCentering43
HorizontalCollimation44
PlumbLineCheck46
Cross‐AxisTelescopeCollimation49
PegTest51


1Operation and Adjustment Manual
Inthischapterwewilladdressthefollowingtopics:
Whatisatransit?
Whatarethecomponents?
Whatdoesatransitdo?
ThedifferencebetweenthethreeBrunsontransits
Themostimportantthingaboutatransitisthatithasatelescopeon
agimballingmechanism,soitcanrotatebackandforthhorizon‐
tally(azimuthdirection),butitcanalsorotateupanddown
(elevationdirection).
Almosteveryoneisfamiliarwithasurveyor’stransit.However,it
isworthexplainingthegeneraldifferencebetweenourtransitsand
thoseusedbysurveyors.OpticalToolingtransitshaveanumberof
modificationstomakethemdeadlyaccuratewhenshootingover
distanceswhich,toasurveyor,wouldseemshort(lessthanacou‐
pleofhundredfeet).Inaddition,wedonotneedtotakereadings
intwotelescopepositions(directandreverse)inordertoachieve
thisaccuracy.Thisispossiblebecauseourtransitshaveseveral
specialcharacteristics:
a)Theyhaveanextremelystraightlineofsight.Asyoufocusthe
telescopefromneartofar,thelineofsighttravelsinanextremely
straightline.Thatis,pictureputtinganimaginary,
weightless,tautstringdownthecenterofthetelescope,
stretchingstraightoutacoupleofhundredfeet—the
stringwouldalwaysbeinthecenterofthecrosshairsas
youfocusfromneartofarandbackagain.
b)Atransit’sthreemajoraxesareallmutuallyorthogonal
(allatrightanglestoeachother).Therearetworota‐
tionalaxesonyourtransit,andoneaxiswhichdoesnot
rotate(seeFigure1.2).Thefirstrotationalaxisisthever‐
ticalaxis.Thespindleandbearingswhichcreatethever‐
ticalaxisarehousedinthetransit’sbase.Theentireup‐
perportionofthetransit(standardsandtelescope)rotate
horizontallyaroundthisaxis.Thesecondrotationalaxis
isthecross‐axis.Thisisthehorizontalaxisaroundwhich
thetelescoperotates.Thethirdaxisisthelineofsight.
c)Lastly,allthreeofthemajoraxesmeetatoneprecise
point.Justbecausetheaxesareallmutuallyorthogonal
PleaserefertoFigure1.1
forageneralorientationto
componentterminology.
Chapter1‐Introduction
Note:Thismanualappliestothe
followingBrunsonproducts:
771and771‐HJigTransits
75and75‐HOpticalTransit
Squares
76‐RH,76‐RH190,and76‐
RH190MTelescopicTransit
Squares
Atthistime,themodels771,771‐H,
75,and75‐Harenolongerpro‐
ducedbythefactory.However,we
haveincludedtheminthismanual
duetotheirpastpopularityandthe
factthatmanyoftheseinstruments
remaininthehandsofalignment
technicianswhomayfindthisinfor‐
mationuseful.
Whatisatransit?

Model 771, 75, and 76-RH190 Transits2
1.Levelingscrews(4)
2.“Spider”base
3.Platevial
4.Telescope(eyepieceend)
5.Cross‐axistelescopeeyepiece
6.Focusknob
7.Telescope(objectiveend)
8.190‐xOpticalMicrometer
9.Cross‐axistelescopeobjectivelens
10.Verticaltangentclamplock
11.Coincidencelevel
12.Cross‐axistelescopelightingadapter
13.Standards
14.Horizontaltangentclamplock
15.Baseplate
16.Horizontaltangentadjustmentscrew
17.Plate
18.Verticaltangentadjustmentscrew
19.Shiftingcenter
Figure1.1
OrientationtoComponents

3Operation and Adjustment Manual
doesn’tnecessarilymeanthatthelineswouldallintersectatthe
sameexactpointinspace.However,onourtransits,theselinesdo
intersect.Thisisanimportantcharacteristicwhenaccuracyispara‐
mount,particularlyatshorterdistances.Ofcourse,youcan’tsee
thisintersectionpoint–itisburiedinthemiddleofthetelescope.
Thesearethethreedistinctcharacteristicsthatourtelescopeshave
whichallowthemtomakeveryprecisemeasurementsoverrela‐
tivelyshortworkingdistances.However,thereisonemorevery
distinctivecharacteristicwhichseparatesOpticalToolingtransits
fromallothers–theopticalcross‐axis.Our76‐RH190hasaninfin‐
ity‐focusedtelescope,completewithasecondeyepieceandsecond
setofcrosswires,housedinthecrossaxis.Thiscross‐axistelescope
isusedinanumberofoperationstosetthemaintelescopeperpen‐
diculartoanothertelescopeoramirror(whichmaybemountedon
arotatingshaftorotherreferenceplane).
Atransitisaveryflexiblemeasuringtool.Itcanestablishvertical
orhorizontalopticalplaneswhichareextremelyflat.Itcanestab‐
lishopticallineswhichareextremelystraight,extendinginanydi‐
rection.Oncetheseplanesandlinesareestab‐
lished,youcanusethemasreferencesfrom
whichtomeasureotherpointsofinterest.You
candeterminewhethermachinecomponents
areparallel,flat,level,square,orstraight.This
meansyoucanevaluatethegeometricrela‐
tionshipsrelatedtothealignmentofalmost
anypart,structuralcomponent,machine,tool,
surface,substrate,wall–orjustaboutany‐
thingelse.Averyfewexamplesofthiswould
be:
Alignmentofbearingsandbearing
journals.
Rollparallelisminpaperandmetals
mills,orprintingpresses
Machinetoolevaluation:bedlevel‐
ing,bedflatness,geometricintegrity
ofspindlesandothercomponents
Layoutformachining
Crankcasealignment
Straightnessofenginebores,turbine
cases,torpedotubes,lineshafts,etc.
Checkrelationshipsofrightangle
planes
TheOpticalToolinginstrumentsareextremely
flexible.Youcanusethemtomakejustabout
anymeasurementorgeometricevaluationthat
Figure1.2
Sowhatdoesatransitdo?

Model 771, 75, and 76-RH190 Transits4
youcandesigninyourmind.
Asweindicatedearlier,the76‐RH190istheonlycurrentlypro‐
ducedOpticalToolingtransit.Althoughthismanualwaswritten
specificallyforthe76‐RH190,muchoftheinformationitcontainsis
applicabletotheothertransitsaswell.
Thedifferencesbetweenthe771‐x,75‐x,and76‐xtransitsrelate
solelytothewayinwhichthecross‐axisisconstructedandused.
Alltransitssharethesamebase,standardassembly,bearings,and
telescopeoptics.Onalltransits,thetelescoperotatesinavertical
planearoundahorizontalaxis(the“cross‐axis”).Thecross‐axisis
perpendiculartothelineofsight.
Themostbasictransitisthe771or
771‐H.Thesetransitshavenoop‐
ticsinthecrossaxis.Theyare
usedas“jig”transitstoestablish
straightlinesandplanes.How‐
ever,theyhavenocapabilityto
makeanaccuraterightangleturn.
Onestep“up”isthemodel75or
75‐H(ontheleftinFigure1.3).
Thesetransitshaveacrossaxis
whichishollowandcanbeseen
through.Ononesideofthecross‐
axisisaclearglasswindow,and
ontheothersideisapartiallyre‐
flectivemirror.The75(‐H)does
everythingthatthe771(‐H)can
do,butcanalsobeusedincon‐
junctionwithanotherinstrument
toproduceaveryaccurateright
angle.
Themostflexibletransit(andthe
onlyonecurrentlyproduced)is
themodel76‐RH190.Thecross‐
axisofthistransitcontainsatele‐
scopethatispermanentlyfocusedatinfinityandisequippedwith
reticlelighting.Arightangleeyepieceisprovidedasstandard
equipmentonthecross‐axistelescope.The76‐RH190maybeused
toestablishstraightlines,horizontalorverticalplanes,andcreate
accuraterightangleswithrespecttoothertransits,alignmenttele‐
scopes,mirrors,orotherphysicaltargets.Whenthecrosswiresof
thetransit’scross‐axistelescopearesetonthoseofareferencetele‐
scopeormirror,themaintelescopeofthetransitwillsweepaplane
atpreciserightanglestothecross‐axisreferenceline.
The75‐Htransit
(above,left)hasbeendis‐
continuedbutmanyare
stillinuse,andmostof
theinformationcon‐
tainedinthismanual
stillapplies.
Figure1.3
Thethreedifferenttransits

5Operation and Adjustment Manual
The“‐H”inthevariousmodelnumbersreferstoa“Hollow”verti‐
calaxis.Transitsthathavethe‐Hdesignationallowyoutolook
straightdownthroughahollowverticalspindleandtoseetargets
orcollimatorswhicharedirectlybelowthebaseoftheinstrument.
Whydoweusefour“feet”(levelingscrews)insteadofthree?
Weareoftenaskedthisquestion,sincemosttheodoliteshavethreeleveling
screws.Thereasonisthatwhenadjustingfourlevelingscrews,theinstrument
heightisnotaffected.
Whenthreelevelingscrewsareused,twoof
themmustnecessarilyformapivotaxis,and
thethirdisusedtoraiseorloweronesidetotilt
theinstrument.Itgivestheeffectshownatleft.
Thereddotindicatestherelativeinstrument
heightbetweentwopositionsoftheinstrument
(tiltedandnottilted).
However,whenfourlevelingscrewsareused,
theinstrumentposition(again,reddot),effec‐
tively,doesnotchange.Thisgivestheeffect
shownatright.Onesidegoesupbythesame
amountastheothersidegoesdown,andthe
instrumentpivotsatitscenter,notononeside
ortheother.
Thisisjustanotherexampleofthewaythatwehavecraftedourinstrumentsto
bemoreaccurateatshorterranges.Overlongerdistances,minorchangesinin‐
strumentheightarenotimportant.Butwhenmeasuringveryaccuratelyover
shorterdistances,avoidingchangesininstrumentheightcanbecomecrucial.

Model 771, 75, and 76-RH190 Transits6

7Operation and Adjustment Manual
Inthischapter,wewilldiscuss
Notethattheseproductsareprecisioninstrumentsandshouldal‐
waysbehandledwithcare.
Donotforceanyofthescrews.Afirm(butnottight)ten‐
siongivesthebestresults.
Keepyourinstrumentcleananddry.Protectitfromthe
weatherifusedoutdoors.
Allowyourinstrumentto“soak”foratleast4hoursin
thetemperatureinwhichitwillbeused,sothatshifts
duetothermalchangesareminimized.Ifyouaremov‐
ingthetransitacrossthermaldifferentialsofmorethan
20°F,allowittosoakfor8hours.Theinstrumentshould
beoutofthecaseduringthistimeforbetteraircircula‐
tion.
Whentakingasighting,or
readingoneoftheseinstru‐
ments,removeyourhands,
andmakesureanyother
pressureisalsoremoved.
Keepyourinstrumentcalibrated.
Maketherecommendedfieldchecks
(seeChapter4)aswellasperforminga
completecalibrationatleastannually.
Notethatthereisawarningprintedon
thesideofyourtransitwhichsays
“Caution—DoNotPlungeTelescope
FromFlange”.Plungingthetelescope
meanstorotateitintheverticalplane,
asyoumightwhenyouwanttolook
furtherupordown.Specifically,
plungingreferstoreversingthedirec‐
tionofthetelescopebyrotatingit180°
Chapter2‐GettingStartedwithYourTransit
Howtohandleyourtransit
Removingitfromthecase
Usingtheshiftingcenter
Roughleveling
Preciseleveling
TheReticle
Focusingtheeyepieceand
telescope
FindingInfinityfocus
Pointingthetelescope
Installinglightingaccesso‐
ries
Figure2.1
Handleyourtransitwithcare

Model 771, 75, and 76-RH190 Transits8
intheverticalplane.The“flange”isthecomponentuponwhichthe
warningisprinted.Thissimplymeansnottorotatethetelescope
bygrabbingtheflangeandtwistingit.Torotatethetelescopeprop‐
erly,putyourhandonthetelescopebarrelitself.
Loosenboththehorizontalandverticaltangentclampscrews.
Graspthestandardsfirmlywithonehand,andthebasewiththe
otherhand.Liftgentlyfromcaseandplaceonastandorothersup‐
port.Whilemaintainingaholdonthestandardswithonehand,
rotatethebasesothatthebottomplateisfirmlythreadedontoany
3½”–8externalthreadedmountorstand.Whenreplacingthe
transitintothecase,slightlytightenthehorizontalandverticaltan‐
gentclamplockscrews—butonlydosogently.
Aspecialcenteringmount,differentthanmostofourinstrument
adapters,maybeemployedtophysicallyplacethehorizontalcen‐
terofthetransitinaknownlocation.Theshiftingcenterofthe
transitextendsaboveandbelowthebaseplate(visiblebetweenthe
levelingscrewsandalsofrombelowasa2”diameterringcentered
onthebottomofthebaseplate).Thisshiftingcenter(radius=1±
0.0001”)fitsintospecialmountshavinganaccuratelyboredinner
diameter.Ifthetransitistobelocatedononeofthesemounting
rings,thelevelingscrewsshouldbeloosenedslightlysothebase
platecanmovearoundinrelationtotheshiftingcenter.Re‐tighten
thelevelingscrewsonlyafterbaseplateistightenedonthemount.
Oncemountedonastandorotherstableinstrumentmount,thefol‐
lowingprocesswill“rough‐level”thetransit,meaningthatthetran‐
sitwillbebroughtroughlytoplumb,andbewithinrangeofthe
precisionlevelvialmountedalongthetelescopeaxis.
Toroughlevel,loosenanytwoadjacentlevelingscrewsandturn
thefourarmʺspiderʺbasesothattwodiagonallyopposedleveling
screwsareinlinewiththeprincipalsightingdirectionthatwillbe
used.Figure2.2illustratesthis,withthespiderbaselinedup
“square”withthetelescopeandcrossaxis.Re‐tightentheleveling
screws.
Acircular“bull’seye”vialrestsonthehorizontal“plate”abovethe
basebutbelowthetelescopeaxis(Figure2.3).Wecallthisthe
“platevial”.Centerthisbubbleinthefollowingmanner:Puta
Removingfromthecase
Usingtheshiftingcenter
Roughleveling

9Operation and Adjustment Manual
thumbandforefingeroneachoftwodiagonallyopposedleveling
screws.Tomovethebubble,turnthelevelingscrewsatthesame
time,andinoppositedirections(thumbsmovingtogetheror
thumbsmovingapart).Thebubblewillmoveinthedirectionof
movementofyourleftthumb.Movethebubblesothatitisinthe
middleoftheplatevial,andatleastinlinewiththeinscribedcircle
(itprobablywon’tbeexactlycentered).Nowmoveyourhandsto
theothersetoflevelingscrews.Usingthesameprocess,movethe
bubbledirectlyundertheinscribedcircleinthecenteroftheplate
vial.Youmayhavetorepeatthisentireprocessacoupleoftimesto
getthebubblecentered.Keepalightbutfirmtensiononthescrews
allthetime.Neverputheavytensiononlevelingscrews.
Nowrotatethetransit180°aboutitsverticalspindle.Thebubble
shouldstaycenteredundertheinscribedcircle.Ifitdoesn’t,the
vialisoutofadjustment(refertoChapter4ofthismanualforthe
calibrationproceduretocorrectthis).
Topreciselylevel(sometimescalled“precisionplumb”)thetransit
usingthecoincidencevial,rotatethestandardsofthetransituntil
thetelescopeisinlinewithonepairofdiagonallyopposedleveling
screws.Plunge(rotate)thetelescopesothatthecoincidencevialis
ontopofthetelescope,andlevelthetelescopebyeye(Figure2.4).
Toassistyou,lookatthesideofthecoincidencevialandposition
thebubbleroughlyinthecenterofitstravel.Gentlytightenthe
Figure2.2
Figure2.3
Preciseleveling

Model 771, 75, and 76-RH190 Transits10
verticaltangentlock.
Thebubbleinthecoincidencevialop‐
erateslikeanyotherlevelvial,butthe
operatorviewsa“foldedimage”of
bothendsofthebubblesimultane‐
ously.Ifthetelescopeistilted,thetwo
bubbleimagesmoveinoppositedirec‐
tions.However,whenbothendsof
thebubblearebrought“intocoinci‐
dence”(Figure2.5),thevialisdead
level.Youwillseesomethingsimilar
toFigure2.5bylookingintothevial’s
turret.Sincethehumaneyeisex‐
tremelyskillfulatdetectingeventiny
breaksinotherwisecontinuouslines,
thissystemmakesreadingthelevelextremelysensitiveandaccu‐
rate.
Rotatetheturretonthecoincidencevialsothatyoucanconven‐
ientlyviewthesplitimageofthebubbleinside.Areflectorhaving
onewhitesideandonemirroredsideisprovidedtohelpilluminate
theimage.Positionthereflectorforoptimalbubbleillumination.
1.Usingtheverticaltangentadjustmentscrew,bringthe
splitimageofthebubbleintocoincidence(Figure2.5).
2.Rotatethestandards180°sothatthetelescopeisnow
pointingintheoppositedirection.Ifthebubbleisoff
center,removeone‐halftheerrorwiththetwoleveling
screwswhicharelocateddirectlyunderthetelescope.
Removetheotherhalfoftheerrorusingtheverticaltan‐
gentscrew.Rememberthatyoucanseethebubbleby
lookingintothesideofthevialitselfratherthanintothe
turretwindow—thismayhelpyoudeterminewhichway
thebubblemustmove.
3.Nowrotatethestandards90°sothatthetelescopeislo‐
cateddirectlyovertheotherpairoflevelingscrews.
Bringthebubbletocenter,usingonlythetwoleveling
screwsunderthetelescope.
4.Fromthisposition,rotatethestandards180°.Again,if
thebubbleisoffcenter,removeone‐halftheerrorwith
theverticaltangentscrew,andtheremainingone‐half
errorwiththetwolevelingscrewswhicharelocatedun‐
derthetelescope.
5.Repeatsteps2‐4,alternatingovereachpairofleveling
screws,untilthebubbleremainsincoincidenceinallfour
“compasspoint”positions.
Figure2.4
Notethatthetransit’sverti‐
calspindlemaybebrought
intoplumb(exactlyvertical)
evenifthecoincidencevial
itselfisnotcalibratedprop‐
erlywithrespecttothetele‐
scopebarrel.
Figure2.5

11Operation and Adjustment Manual
ThetelescopeofeachOpticalToolinginstrumentcontainsareticle–
orcrosshair–whichdefinesthecenterofthelineofsight.When
youlookintoatelescope,youseethereticleʺsuperimposedʺupon
theimageofwhateveryouhavethe
telescopefocusedon.Thereticleis
splitintofourquadrants;twohavea
singleʺwireʺ(filar),theothertwohave
adoubleʺwireʺ(bi‐filar).Thiscon‐
figurationwasoriginallyadoptedbe‐
causethehumaneyeisextremely
goodatdetectingtinydifferencesin
alignmentbetweencloselyspacedob‐
jects.Therefore,thispatternmakesit
mucheasiertovisuallyalignthereti‐
clewithotherimages(ex.,opticaltar‐
gets,scales,andreticleimagesre‐
flectedfrommirrors).
Focusingsoundseasy,butit’ssurprisinghowmanypeopledon’t
understandtheeffectsofsomethingcalledparallax.Here’showto
properlyfocusyourinstrument.First,pointthetelescopeatasheet
ofwhitepaperorotherlightcoloredbackground.Don’tworry
aboutfocusingthetelescopeusingthefocusknob,theimagecan
remainblurry.Instead,usetheknurledringontheeyepiecetofo‐
custhereticlesothatitissharpandblack.
Nowpointthetelescopeatatargetofinterestandfocustheimage
usingthefocusknobonthetelescope.Slightlyshiftyourheadleft
andrightwhilelookingatthereticleasitis“superimposed”over
theimageofthetarget.Ifthereticleappearstofloatslightlytothe
leftandrightoverthetarget,parallaxispresentinthesystem.This
occurswheneitherthetelescope’smainopticalsystemortheocular
(eyepiece)system,orboth,arenotfocusedpreciselyontheplaneof
thereticle.Parallaxcanleadtoerroneousmeasurementsifnotre‐
moved.Toremovetheparallax,repeatthetwostepsagain,being
carefultoachievecriticalfocusineachstep.
Therearemanyoccasionswhenitwillbeimportanttoputyour
telescopeatinfinityfocus(e.g.,autocollimatingyourtelescopeona
mirror).Infinityisimportantbecause,fromanopticalstandpoint,
lightrayscomingfroman“infinitedistance”areallparalleltoeach
TheReticle(crosshairs)
Howtoproperlyfocus
Thefilar/bi‐filarreticle
imageasseenwhen
lookingintothetele‐
scope.
Figure2.6
Findinginfinityfocus

Model 771, 75, and 76-RH190 Transits12
other,orcollimated.
Thefocusknobonthetelescopehasan
indicatorofthedirectionyoumustturn
ittofindinfinityfocus,butyoucan’t
justturnitallthewayinthatdirection
andexpecttothetelescopetobefo‐
cusedatinfinity.
Theeasiestwaytofindtrue“infinity
focus”istoholdanopticallyflatmirror
(suchasourmodel185)rightup
againsttheobjectiveendofthetele‐
scope.Turnonthetransit’sinternalillumination(see“Installing
lightingaccessories”below).Turnthefocusknoballthewayinthe
directionofinfinity(indicatedontopofthefocusknob).Then,turn
theknobslowlyintheoppositedirection,andwatchcloselyforthe
reflectedimageofthereticle.Itcangobyprettyquickly,sogoing
slowlyisimportantwhilelookingthroughthetelescope.Onceyou
havefocusedthetelescopesothatyoucanclearlyseethereflected
reticleimage,youknowthatthetelescopeis,bydefinition,focused
atinfinity.
Releaseboththehorizontalandverticaltangentclamplocks,and
movetheinstrumentbyhand.Precisepointingisaccomplishedby
re‐lockingbothclampswhenyougetclosetothetarget,andthen
turningthetangentadjustmentscrewsforpreciseplacementofthe
reticle.Bothtangentadjustmentscrewsareoutfittedwithwobble
pinswhichallowforsmoothtranslationoftheaxes.Thefinalset‐
tingshouldbemadebyturningthetangentscrewinaclockwise,or
pushing,direction.Thiswilleliminateanypotentialsubsequent
movementduetofrictionofthespring‐loadedplungerswhichop‐
posethetangentadjustmentscrews.
Forcertainproceduresthatwewilldiscussinthismanual,(e.g.,col‐
limation,autocollimation,andreticleprojection)itisnecessaryto
illuminatethereticleofyourtransitfromtheinside.Amodel76‐x
transithastworeticles(oneinthemaintelescopeandoneinthe
cross‐axistelescope)whichmayneedtobeilluminated,depending
uponwhatyouaredoing.
Illuminatingthecross‐axistelescopereticle.Yourtransitcomes
equippedwithalightsourceforthecross‐axistelescope(model76‐
Thereflectedimageofa
reticle(here,located
aboveandtotherightof
center)willappearonly
atinfinityfocus.
Figure2.7
Installinglightingaccessories
Pointingthetelescope

13Operation and Adjustment Manual
xonly).ThisLEDlightsourcefitsintothereceptacle(red
arrowinFigure2.8).Thesupplytransformerworkswith
either110Vor220Vsources.Installationorremovalofthe
lightsourceissimpleandquick.Simplyinstallthetwo‐
wayadapteronthecross‐axislightingreceptacle,andpush
theLEDlamphousinginuntilitclicksintoplace.Thelight
sourcehasanin‐lineswitchforeasycontrolofthelight.
Rememberthatthefocusofthecross‐axistelescopeisper‐
manentlyfixedatinfinity.Thistelescopeismeanttobe
usedasacollimatororautocollimator.Wewilldiscussthis
more,lateroninthemanual.
Illuminatingthemaintelescopereticle.Whenyouwishto
usethemaintelescopeforcollimation,autocollimation,orreticle
projection,youmustinstallamodel196‐1combinationeyepiece
unit.Todothis,removetheshortsectionofthemaintelescopebar‐
rellocatedjustinfrontoftheeyepiece(SeeFigure2.9),andreplace
itwiththe196‐1eyepieceadapter.Notethatthe196‐1hasa
knurledlockringwhichallowsyoutotightentheadapterinany
givenorientationonthetelescopebarrel.
Beforeinstallingthiseyepiece
adapteronthetelescope,make
suretheknurledlockring(red
arrowinFigure2.10)issecured
againsttheadapter.Itshouldbe
“bottomedout”onthethreads
ratherthanlooseonthethreads
orneartheopenendofthe
adapter.Thenthreadthe
adapterintothemaintelescope
barrelasfarasyoucan.Next,
backofftheadaptertothede‐
siredradialorientation(lamp
receiverpointinguporpointing
down)andlockitinplaceby
turningtheknurledlockring
towardthefrontofthetele‐
scope.
Next,removetheeyepieceassemblythatisstillcon‐
nectedtotheshortsectionofthetelescopebarrelwhich
wasremoved,andinstallitonthe196‐1eyepiece
adapter.
Toinstallthelight,removethecapfromthelampre‐
ceiver(undertherightthumbinFigure2.10).Thenin‐
stallthetwo‐wayadapterbythreadingitsknurledring
ontothelampreceiver.PlugintheLEDlightsource
untilitclicksintheadapter.
Figure2.8
Figure2.9
Note:Toinstallarightangleview‐
ingeyepieceonthemaintelescope
(convenientforveryhighangle
shots),youmaywishtoinstalla193
‐Lratherthanthe196‐1.Theseac‐
cessoriesareinstalledinbasically
thesameway,byfirstremovingthe
shorttubesectioninfrontofthe
eyepiece.

Model 771, 75, and 76-RH190 Transits14
Laterinthismanual,wewill
discusssomeoftheoptical
toolingtechniqueswhichre‐
quirebacklightingofthereti‐
cleusingoneoftheseadapt‐
ers.
Figure2.10

15Operation and Adjustment Manual
Therearemanyjobsthatyoucandowithyourtransit,whichmay
bepurchasedwitha190‐typemicrometerinastandardconfigura‐
tion.Youcanevaluateanumberofalignment‐relatedcharacteris‐
ticsofmachineryorotherstructures,suchasstraightness,plumb,
level,flatness,andothers.However,toperformcertainOptical
Toolingprocedures,otheraccessoriesarerequired.Wewillmen‐
tionsomeoftheseaccessoriesaswediscusstherelatedtopics.
Inthischapterwewilldiscussthefollowingtopics:
ThisdiscussionismeanttomakeyouawareofsomeoftheOptical
Toolingproceduresthatyoucanuseinalignmentandmeasure‐
mentwork.Thismanualisnotagoodsubstituteforourtraining
classes,inwhichweofferhands‐ontrainingoftheseandmany
otherOpticalToolingtechniques.
We’llexplainwhatthesetermsmean,andhowtoaccomplishthem
usingyourtransit.
Inmanyalignmentjobs,itisnecessarytoputthelineofsightof
yourtransitdirectlyonalinecreatedbytwoknownreferencetar‐
gets.Forexample,inFigure3.1(showingaviewfromoverhead),
sometargetshavebeeninstalledinthefloor.Thesetargetsrepre‐
sentalinewhichisparalleltothecenterlineofamachineonwhich
youneedtoaligncomponents.Inpreparationfordoingthisalign‐
mentjob,itcouldbenecessarytobringtheinstrumentinlinewith
thesetwofloormonuments.
NoteinFigure3.1,thetransitmustnotonlyberotatedsothatitis
paralleltothelinedefinedbythetargets,butitmustalsobeshifted
sideways.“Buckingin”isarepetitiveprocessinwhichyouadjust
boththeinstrument’sangleaswellasitslateralpositionsothatthe
transitisplaceddirectlyinlinewithtwotargets.Ortobemoreac‐
curate,sothatthetransit’sverticalspindleliesintheverticalplane
whichalsorunsthroughthecenterofthefloortargets.
Notethatangularadjustmentsareeasyenoughtomakeusingthe
horizontaltangentadjustmentscrewonthetransititself.However,
Chapter3‐MeasuringTechniques
“Buckingin”
Autoreflection
Autocollimation
Collimation
Collineation
Reticleprojection
Measuringwithmicrometers
“Buckingin”

Model 771, 75, and 76-RH190 Transits16
Alwayssetthetelescopeon
thefartargetbymak‐
ingangularmoves
usingthehorizontal
andverticaltangent
screw(s).
Alwayssetthetelescopeon
theneartargetby
makinglateralmoves
usingalateralslideor
precisionlift.
lateralshiftsoftheentireinstrumentarenotaseasy.Togreatly
simplifythis,werecommendtheuseofalateralslidemountedbe‐
tweenthetransitandthestandonwhichitismounted.
Stepstobuckthetransitintothetargetline:
1.Roughlypositionthetransit.Setuptheinstrument
standnominallyinlinewiththetwotargets.
2.Roughlevel.Roughlevelthetransitasdescribedin
Chapter2“RoughLeveling”.
3.Preciselevel.Formanyapplications,preciseleveling
yourtransitisrequiredpriortobuck‐in(coveredin
Chapter2).Technicallyisitpossibletobuckintotargets
withoutpreciseleveling,butthetransitwillnotendupin
theverticalplanedefinedbythosetargetsunlessthetran‐
sititselfispreciselyplumb.Afulldiscussionof
thisfactisbeyondthescopeofthismanual,butis
coveredinmoredetailinourOpticalTooling
classes.
4.Settofartarget.Focusthetelescopeonthefar
targetandsetthecrosswiresdirectlyoverthattar‐
getusingthehorizontaland/orverticaltangent
adjustmentscrews.Ifthetransitisbetweenthe
targets(asinFigure3.2),simplyassignoneofthe
targetstobethe“fartarget”,andsetonit.
5.Settoneartarget.Withoutmovingthehori‐
zontaltangentscreworturningthestandardsof
thetransit,turnthetelescopedownwardandfocus
onthesecondtarget.Notethatifthetransitislo‐
catedbetweenthetargets(asinFigure3.2),youwill
havetoturnthetelescope“through”thestandards
topointintheoppositedirection.Regardlessof
whetheryouareinasituationlikeFigure3.1or
Figure3.2,thissecondtargetwillinvariablybeoff
toonesideofthefieldofviewunlessyouarein‐
crediblylucky.Forpurposesofillustration,let’s
saythattheobservedlateraldistancetothenear
targetis“L”(seeFigure3.1or3.2).Nowyoumust
movethetransitlaterallytowardtheneartarget
usingthecrossslide—buthowmuch?The
amountthatyoumovethetransitdependsupon
thegeometryofthesetup.Theequationtocalcu‐
latetheproperamountofshiftis:
TransitLateralShift=Lx(D2/D1)
where
D1=Distancebetweentargets
Figure3.1
This manual suits for next models
6
Table of contents
Other Brunson Measuring Instrument manuals