manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Camlab
  6. •
  7. Measuring Instrument
  8. •
  9. Camlab TRUEscience ION Iodide ISE User manual

Camlab TRUEscience ION Iodide ISE User manual

Some More Hints
Ensure that the temperature of all standards and
samples are the same to reduce errors.
Using a magnetic stirrer for laboratory analysis is
recommended but not essential. It is however
important to ensure the solution is homogeneous and a
stirrer will help to achieve this.
NOTE: DO NOT stir during measurements.
Prior to sample measurement ensure that the
electrode is thoroughly rinsed with deionised water.
It is worth performing this rinse twice given the
possibility of carryover being greatest in high
concentration solutions. Before reading sample values it is
a good idea to rinse with sample in a seperate beaker.
Prepare standards by serial dilution.
Make sure your electrode is conditioned by leaving
the tip in the lowest concentration standard for 1
hour prior to analysis.
Methods of Analysis
Direct Potentiometry is described under ‘Operation’.
This method is simplified by using a direct reading ion
meter. There are several other methods, which are
useful.
Known Addition: An incremental technique where
the potential of the sample solution is measured
followed by addition of a small volume of a higher
concentration standard solution.
Sample Addition: An incremental technique where
the potential of a dilute standard solution is
measured followed by the addition of a small volume
of more concentrated sample. .
End Point Titration: I E’s are ideal end point
indicators and will produce a significant potential
change at the equivalence point. The Ion in question
must be contained in the titrand or the titrant and
must therefore be in excess or absence at the end point.
www.truescience.co.uk www.truescience.co.uk
ion
7 male connector can be connected directly to the
TRUEscience cap or used with an adapter cable for
other meters using a 7 Female connector
ee our website for details.
The TRUEscience cap is ideal for measuring up to six different parameters at
the same time. It simply clips onto your beaker and can measure pH,
Redox,Dissolved oxygen or specific ions simultaneously on an Android Tablet
Camlab Ltd, Norman Way Industrial Estate, Over,
Cambrid e, CB24 5WE, En land
© Camlab Ltd - Version V1.0/2018
[email protected]
+44 (0) 1954 233 144
1141001 - Iodide ISE
Instruction Manual
Specifications
Overall length 155 mm
Body Diameter 12 mm
Cap Diameter 16mm
Connector 7 Male
Resistance at 25°C < 2.5 Meg Ohm
Concentration Range 0.6 to 127,000 ppm
lope -54 to -59 mV/decade
Potential Drift Operating 1 mV per day
pH range 2 to 12
Temperature range 5 to 50°C
Endpoint time Typically 10 to 30 seconds
Interferences: Ions with
coefficients above 0.001 ulphide and Cyanide
www.truescience.co.uk [email protected]
+44(0) 1954 233 144
Operation
 Connect the electrode to the meter being used for
analysis.
 Prepare a series (at least 2) of standards that
bracket the expected sample concentration. This is
best done by serial dilution of the stock solution.
Ideally standards should be a decade in
concentration apart e.g. 1, 10, and 100 ppm.
 Dispense 50 ml of each standard into analytically
clean beakers. (100 to 150 ml size is perfect)
 Add I AB in the appropriate ratio. As a guide with
sample concentrations in the 1 to 1000 ppm range 1ml
of a 5 Molar I AB to 50 ml sample is satisfactory
 Rinse the electrode with deionised water and blot
dry with a lint free cloth and place in the lowest
standard. When the reading is stable record the
mV value.
 Repeat step  for all subsequent standards
proceeding from lowest to highest.
 Use an ion meter to create a calibration curve or
plot a calibration curve on semi log paper using
mV values on the linear Axis and concentration on
the log scale. Note that towards the end of the scale
the curve is not linear.
 Rinse the electrode in deionised water and blot
dry. Place the electrode in the sample and record
the stable mV value.
 Using the calibration curve determine the
unknown sample concentration.
NOTE: The TRUEscience app allows you to save
sets of tandard Dilutions  and creates a calibration
curve  using your readings  . This curve will then
be used to calculate your sample readings  and can
then be referred to at any point in the future. It also keeps
track of the batch of tandard  and I AB’s  used.
Iodide ISE
The TRUEscience ion selective electrode has a solid-
state PVC Polymer membrane with an integral
dri-tek reference.
The electrode is designed for the detection and analysis
of Iodide ions in aqueous solutions and is suitable for use
in the field, in the laboratory, and in on-line
analysers
Installation
Connect the I E directly to the TRUEscience MART cap
or to a mV or ion meter using an appropriate adapter
cable. Remove the black protective cap and keep it in a
safe place.
The I E can be used immediately but pre soaking for 5
minutes in a 100ppm Iodide solution is recommended.
The ionic strength of the standards and solutions should
be kept constant between all standards and samples.
This is achieved by the simple addition of a Ionic trength
Adjustment Buffer(I AB). odium Nitrate is ideal.
A typical addition would be 1 ml of 5M I AB to 50 ml of
standard and sample. For low level measurements,
below 50 ppm in relatively pure samples, no I AB is
needed.
No temperature correction is necessary however
standards and samples should be measured at the same
temperature.
Begin calibration from the lowest concentration
standard to avoid cross contamination. Calibration should
cover the anticipated range of the samples.
Rinse tip with deionised water between
measurements.
Avoid strongly acidic or alkaline samples, strong deter-
gents and organic solvents.
Storage and Maintenance
After use rinse with deionised water, wipe clean with a
tissue or lint free cloth, replace protective cap and store dry in
its box.
If performance becomes sluggish, rinse with dilute
detergent, then rinse with deionised water and immerse the
tip in a 1000ppm Iodide solution for 1 hour.
Tips For Successful Analysis
TRUEscience I E’s are designed to be used with the
TRUEscience MART Cap but can be used with any pH/mV
meter or Ion meter. You will require an adapter cable with an
7 female to the connector type for your meter. This is
usually BNC or DIN and these cables are available from your
TRUEscience distributor.
Meters with a 0.1 mV resolution are recommended whilst
dedicated Ion meters will provide direct concentration
readouts saving time and effort in constructing calibration
curves and performing calculations. Your TRUEscience
distributor can advise on the most suitable meter.
Magnetic stirrer/stirrer bars are recommended for
laboratory analysis. Please operate at the lowest constant
speed available.
emi-logarithmic (4-cycle) graph paper is required for prepar-
ing calibration curves when you are using a mV meter.
Required Solutions
Distilled or deionised water will be required to prepare
tandards, I ABs and to rinse the electrode between
measurements.
1000 ppm Stock Standard solution. Used for
preparation of tandards. (Prepared by customer)
ISAB. Used to adjust the Ionic strength of all standards and
samples. Typical addition is 1 ml of I AB to 50ml of all stan-
dards and samples.

This manual suits for next models

1

Other Camlab Measuring Instrument manuals

Camlab MS-H280-Pro User manual

Camlab

Camlab MS-H280-Pro User manual

Camlab Water CW2050 User manual

Camlab

Camlab Water CW2050 User manual

Camlab CW6120 User manual

Camlab

Camlab CW6120 User manual

Camlab CW8200 User manual

Camlab

Camlab CW8200 User manual

Camlab CW/6230 User manual

Camlab

Camlab CW/6230 User manual

Camlab CW/6210 User manual

Camlab

Camlab CW/6210 User manual

Camlab CW6110 User manual

Camlab

Camlab CW6110 User manual

Camlab Lovibond MD 610 User manual

Camlab

Camlab Lovibond MD 610 User manual

Popular Measuring Instrument manuals by other brands

Biospherical Instruments Inc MER-2040 Software user manual

Biospherical Instruments Inc

Biospherical Instruments Inc MER-2040 Software user manual

Aquametro CALEC ST Mounting and operating instructions

Aquametro

Aquametro CALEC ST Mounting and operating instructions

Powerline S4BL instruction manual

Powerline

Powerline S4BL instruction manual

EL-CELL ECD-3 user manual

EL-CELL

EL-CELL ECD-3 user manual

Mi-Sol WS4000 instruction manual

Mi-Sol

Mi-Sol WS4000 instruction manual

AquaMetrix Aqua-Check Operator's manual

AquaMetrix

AquaMetrix Aqua-Check Operator's manual

FRO­NIUS Smart Meter 50kA-3 quick guide

FRO­NIUS

FRO­NIUS Smart Meter 50kA-3 quick guide

Avtech Room Alert User's guide & reference manual

Avtech

Avtech Room Alert User's guide & reference manual

Sofar solar CHINT DDSU666 Series Quick installation guide

Sofar solar

Sofar solar CHINT DDSU666 Series Quick installation guide

Pro-tec PLM-8000 Operation manual

Pro-tec

Pro-tec PLM-8000 Operation manual

PerkinElmer DSC 4000 Installation and hardware guide

PerkinElmer

PerkinElmer DSC 4000 Installation and hardware guide

FARM SCAN Jackal v2 General description

FARM SCAN

FARM SCAN Jackal v2 General description

USFilter DEPOLOX 3 PLUS manual

USFilter

USFilter DEPOLOX 3 PLUS manual

Check-line MINI-MAX Operation manual

Check-line

Check-line MINI-MAX Operation manual

KEPCO 52U manual

KEPCO

KEPCO 52U manual

Audio Control SA-3052A Operation manual

Audio Control

Audio Control SA-3052A Operation manual

Unit UT311 operating manual

Unit

Unit UT311 operating manual

Unit UT256A operating manual

Unit

Unit UT256A operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.