manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Camlab
  6. •
  7. Measuring Instrument
  8. •
  9. Camlab Water CW2050 User manual

Camlab Water CW2050 User manual

Optics: LED:λ= 528 nm
Battery: 9 V block battery (life = approx. 600 tests)
Auto-OFF: auto unit switch-off approx. 5 minutes after a
key was last pressed
Ambient conditions
:
5-40°C
30-90% rel. humidity (non-condensing)
CE:
DIN EN 55 022, 61 000-4-2, 61 000-4-8,
50 082-2, 50 081-1, DIN V ENV 50 140, 50 204
Technical data
Operation
CW2050 Colorimeter for Chloride
User messages
Zero
Test
METHOD
METHOD
METHOD
METHOD
0.0.0
RESULT
RESULT
Zero
Test
÷Err
EOI
-Err
LO BAT
ON
OFF
r1
Chloride 0,5 - 25 mg/l
0.0.0
RESULT
RESULT
Zero
Test
r1
Switch the unit on using the ON/OFF switch
The display shows the following:
Select analysis using the MODE key:
r1 →r2 →r1 .......... (Scroll)
The display shows the following:
Pour 1 ml of the water sample into a clean vial and fill to
the 10 ml mark with deionized water. Replace the cap
tightly and place the vial in the sample chamber with the
∇vial marking aligned with the∆housing marking.
Press the ZERO/TEST key.
The method symbol flashes for approx. 3 seconds.
The display shows the following:
After zero calibration is completed, remove the vial from
the sample chamber.
The characteristic coloration starts to appear after the
addition of the reagent tablet(s).
Replace the cap tightly and place the vial in the sample
chamber with the ∇and ∆marks aligned.
Press the ZERO/TEST key.
The method symbol flashes for approx. 3 seconds.
The result appears in the display.
Repeating the analysis:
Press the ZERO/TEST key once again.
New zero calibration:
Press the MODE key until the desired method symbol
appears in the display again.
Light absorption too great. Reason - e.g. soiled lens.
Measuring range exceeded or excessive turbidity.
Result outside bottom measuring range limit.
Replace 9 V battery immediately; no further analysis
possible.
Perform zero calibration (see “Operation”).
Adjust the sample temperature to 20°C ±2 °C.
Add one CHLORIDE T1 tablet straight from the foil to the
prepared vial. Crush using a clean stirring rod and mix
until all particles of the tablet have dissolved. Add one
CHLORIDE T2 tablet straight from the foil to the same
sample, crush and mix gently to dissolve. Ensure all
particles are dissolved – Chloride causes an extremly finely
distributed Turbidity with a milky appearance. Replace
the cap tightly and place this vial in the sample chamber
with the ∇and ∆marks aligned.
Wait for a reaction time of 2 minutes!
(do not remove or shake the vial during this time)
Press the ZERO/TEST key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l Cl
–
.
Tolerance: ±2,5 mg/l.
Chloride 5 - 250 mg/l
The display shows the following:
Pour 1 ml of the water sample into a clean vial and fill to
the 10 ml mark with deionized water. Replace the cap
tightly and place the vial in the sample chamber with∇
and ∆marks aligned.
Press the ZERO/TEST key.
The method symbol flashes for approx. 3 seconds.
The display shows the following:
Adjust the sample temperature to 20°C ±2 °C.
Add one CHLORIDE T1 tablet straight from the foil to the
prepared vial. Crush using a clean stirring rod and mix
until all particles of the tablet have dissolved. Add one
CHLORIDE T2 tablet straight from the foil to the same
sample, crush and mix gently to dissolve. Ensure all
particles are dissolved – Chloride causes an extremly finely
distributed Turbidity with a milky appearance. Replace
the cap tightly and place this vial in the sample chamber
with the ∇and ∆marks aligned.
Wait for a reaction time of 2 minutes!
(do not remove or shake the vial during this time)
Press the ZERO/TEST key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l Cl
–
.
Tolerance: ±25 mg/l.
RESULT
RESULT
Zero
Test
r2
Zero
Test
METHOD
METHOD
0.0.0
r2
METHOD
METHOD
Mode
METHOD
METHOD
RESULT
Troubleshooting: Guidelines for photometric measurements
1. Vials, caps and stirring rods should be cleaned thoroughlyafter each
analysis to prevent errors being carried over. Even minor reagent
residues can cause errors in the test results. Use the brush provided for
cleaning.
2. The outside of the vial must be clean and dry before starting the analysis.
Clean the outside of the vials with a towel. Fingerprints or other marks
will be removed.
3. Zero calibration and test must be carried out with the same vial as there
may be slight differences in optical performance between vials.
4. The vials must be positioned in the sample chamber for zero calibration
and test with the ∆-mark on the vial aligned with the∇−mark on the
instrument.
5. Always perform “Zero calibration” and “Test” with closed vial lid.
6. Bubbles on the inside walls of the vial lead to incorrect measurements.
To prevent this, remove the bubbles by swirling the vial before performing
the test.
7. Avoid spillage of water or reagent solution in the sample chamber. If
water should leak into the photometer housing, it can damage electronic
components and cause corrosion.
8. Soiling of the lens (LED and photosensor) in the sample chamber leads
to incorrect measurements.
Check - and if necessary clean - the light entry surfaces of the sample
chamber at regular intervals. Clean using a moist cloth and cotton buds.
9. Always add the reagent tablets to the water sample straight from the
foil without touching them with your fingers.
10. Large temperature differentials between the photometer and the
operating environment can lead to incorrect measurement due to the
formation of condensate in the area of the lens or on the vial (e.g).
11. To avoid errors caused by stray-light do not use the instrument in bright
sunlight.
RESULT
RESULT
METHOD
METHOD
Zero
Test
Zero
Test
Zero
Test
Zero
Test
Zero
Test
• Replacement Reagents
Chloride T1 tablet pk 100
Ref: TT/51.59.00
Chloride T2 tablet pk 100
Ref: TT/51.59.20
• Notes
1. High concentrations of electrolytes and organic compounds have different
effects on the precipitation reaction.
2. Ions which also form deposits with silver nitrate in acidic media, such as
bromides, iodides and thiocyanates, interfere with the analysis.
3. Highly alkaline water should - if necessary - be neutralised using nitric acid
before the reaction is performed using the reagent tablet.
• Method notes
Observe application options, analysis regulations and matrix effects of methods.
Reagent tablets are designed for use in chemical analysis only and should be
kept well out of the reach of children.
If necessary, request safety data sheets.
Ensure proper disposal of reagent solutions.
Mode
Camlab CW2050 Chloride 2/5/06 3:21 pm Page 1
Bediener-Hinweise
E 10
E 70
E 71
Recommended calibration value
Chloride: between 10 and 20 mg/l Cl
–
Technical changes without notice
Printed in Germany 05/00
Calibration factor “out of range”
Factory calibration not OK / deleted
User calibration not OK / deleted
The unit can be reset to delivery condition (factory
calibration) as follows:
Press MODE and ZERO/TEST together andhold
depressed .
Switch the unit on using the ON/OFF key. Release MOD
E
and ZERO/TEST keys after approx. 1 second.
The following messages appear in the display in
alternating mode:
The unit is in delivery condition.
(SEL stands for Select)
or:
The unit operates with a calibration performed by the
user. (If the user calibration is to be retained, switch the
unit off using the ON/OFF key.)
Factory calibration is activated by pressing the MODE
key. The following messages appear in alternating mode
in the display:
Switch the unit off using the ON/OFF key.
User calibration : cAL
Factory calibration : CAL
Zero
Test
Mode
On
Off
CAL
SEL
SEL
CAL
Mode
On
Off
cAL
SEL
Calibration mode
On
Off
Mode
CAL
r1
CAL
RESULT + x
RESULT + x
Mode
Zero
Test
0.0.0
Zero
Test
RESULT
RESULT
CAL
Zero
Test
On
Off
: :
Note
CAL
cAL
METHOD
METHOD
METHOD
METHOD
CAL
Press MODE key andkeep it depressed .
Switch unit on using ON/OFF key.
Release MODE key after approx. 1 second.
The display shows the following in alternating mode:
Perform zero calibration (see "Operation").
Press the ZERO/TEST key.
The method symbol flashes for approx. 3 seconds.
The display shows the following in alternating mode:
Place the calibration standard to be used in the sample
chamber with the ∆and ∇marks aligned.
Press the ZERO/TEST key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display, alternating with CAL.
If the result displayed corresponds with the value of the
calibration standard (within the tolerance quoted), exit
calibration mode by pressing the ON/OFF key.
Otherwise, pressing the MODE key once increases the
displayed value by 1 digit.
Pressing the ZERO/TEST key once decreases the
displayed value by 1 digit.
Pressing the relevant key until the displayed value equals
the value of the calibration standard.
By pressing the ON/OFF key, the new correction factor
is calculated and stored in the user calibration software.
Confirmation of calibration (3 seconds).
It is not necessary to make a calibration of the r2-range as the
software refer to the calibration of the r1-range.
Factory calibration active.
Calibration has been set by the user.
Correct filling of the vial
correct wrong
10 ml 10 ml
Camlab Water CW2050
Chloride Colorimeter
Reference CW/20.71.00
Camlab Limited , Camlab House, Norman Way, Over
Cambridge CB4 5WE
Tel: 01954 233110 • Fax: 01954 233101
www.camlab.co.uk
camlab
Choice, Service and Trust
Technical changes may occur without notice
Mode
Zero
Test
Zero
Test
Zero
Test
Mode
Mode
Zero
Test
Mode
Camlab CW2050 Chloride 2/5/06 3:21 pm Page 2

Other Camlab Measuring Instrument manuals

Camlab TRUEscience ION Iodide ISE User manual

Camlab

Camlab TRUEscience ION Iodide ISE User manual

Camlab CW6120 User manual

Camlab

Camlab CW6120 User manual

Camlab CW/6230 User manual

Camlab

Camlab CW/6230 User manual

Camlab MS-H280-Pro User manual

Camlab

Camlab MS-H280-Pro User manual

Camlab CW6110 User manual

Camlab

Camlab CW6110 User manual

Camlab Lovibond MD 610 User manual

Camlab

Camlab Lovibond MD 610 User manual

Camlab CW/6210 User manual

Camlab

Camlab CW/6210 User manual

Camlab CW8200 User manual

Camlab

Camlab CW8200 User manual

Popular Measuring Instrument manuals by other brands

Envertech EVB300 User's operating manual

Envertech

Envertech EVB300 User's operating manual

Omntec PROTEUS B installation manual

Omntec

Omntec PROTEUS B installation manual

Ground Control CP4 installation guide

Ground Control

Ground Control CP4 installation guide

LR-Cal FLUID 100-N operating manual

LR-Cal

LR-Cal FLUID 100-N operating manual

Humboldt FD-4A product manual

Humboldt

Humboldt FD-4A product manual

Horex 22 117 16 operating instructions

Horex

Horex 22 117 16 operating instructions

LaserLiner GreenLine-Laser 2 P operating instructions

LaserLiner

LaserLiner GreenLine-Laser 2 P operating instructions

Laser Quantum Ventus with smd12 operating manual

Laser Quantum

Laser Quantum Ventus with smd12 operating manual

Withings Aura quick start guide

Withings

Withings Aura quick start guide

PT W80158 owner's manual

PT

PT W80158 owner's manual

Ametek 5830 user manual

Ametek

Ametek 5830 user manual

Barnant Company 559-3800 operating manual

Barnant Company

Barnant Company 559-3800 operating manual

MFJ MFJ-834 instruction manual

MFJ

MFJ MFJ-834 instruction manual

Zoetis vetscan SA user manual

Zoetis

Zoetis vetscan SA user manual

Mecmesin Compact Force Gauge operating manual

Mecmesin

Mecmesin Compact Force Gauge operating manual

REMA DSD05 Series owner's manual

REMA

REMA DSD05 Series owner's manual

Bender ISOMETER isoEV425 manual

Bender

Bender ISOMETER isoEV425 manual

yellow jacket 40813 user manual

yellow jacket

yellow jacket 40813 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.