Operation!
!
Once installed, the MB2-RSH operates in exactly the same way as a conventional MB2
bioaerosol sampler.!
!
For operation instructions, please refer to the MicroBio Manual, document number
P0001W003.!
!
!
!
Cleaning and bio-decontamination!
!
The control unit and the remote sampling head may both be cleaned using conventional
sanitising wipes, such as Tuffie-5 or 70% IPA.!
!
As the remote sampling head will normally be fixed within a compartment, it is suitable for
use with Hydrogen Peroxide Vapour (H2O2 vapour) bio-decontamination.!
!
Tests have been carried out by Bioquell (UK) Ltd on sensitive electronic equipment to
determine effects of such processes. The conditions of these tests were:!
!
!Gas Concentration! ! 1000 ppm!
!Conditioning time! ! 20 minutes!
!Gassing time!!!30 minutes! !
!Aeration time!!!150 minutes!
!
As the MicroBio MB2-RSH has a fan within the remote head it is recommended that during
the decontamination process the MicroBio MB2-RSH is set to sample 2000 litres several
times so that the H2O2 vapour may be drawn through the head, fan and exhaust for
optimum decontamination.!
!
Throughout the trials, Bioquell demonstrated that bio-decontamination with H2O2 vapour
does not appear to be detrimental in any way, effect operational aspects or aesthetics of
sensitive electronic equipment.!
!
As the control panel will not normally be exposed to bio-contaminants as the remote head,
cleaning using sanitising wipes may suffice, but it is also suitable for use within H2O2
vapour environments.!
!
For further information on Bioquell H2O2 vapour Material Compatibility please refer to their
document BDS-3-MATCOMP-V3.2 available from their website www.bioquell.com or
contact them on +44 (0)1264 835835.