COMPENSATION PROCEDURE AND CLEANING
When necessary, the probe and accessories can be cleaned with a
damp cloth using water and a mild detergent.
Adequate drying time should be allowed before use.
WARNING
Dry the probe completely before attempting to make voltage
measurements.
For an accurate indication of the measuring signal, it is necessary to adjust the capacity
of the probe to the input capacity of the oscilloscope. For this purpose the probe has an
adjusment screw in the handling section. Connect the probe to the oscilloscope and the
tip to the reference calibration signal of the oscilloscope.
Turn the adjustment scew until the oscilloscope shows an exact square wave.
SPECIFICATIONS, PERFORMACE AND COMPLIANCE
SAFETY HANDLING AND OPERATION
Probe must be fully connected to the measurement instrument prior to connection to
the device under test Probe must be disconnected from the device under test prior to
disconnecting from the measurement instrument.
Barriers are provided to indicate hazardous areas that cannot be touched during operation.
Voltage Ratings
400 V Maximum input voltage of 400 V AC+ DC
300 V Maximum input voltage of 300 V AC+ DC
Category Ratings
CAT II Suitable for connection to category two circuits.
CAT III Suitable for connection to category three circuits.
CAT IV These probes are not suitable for attachment to CAT IV circuits.
When using the Hook Tip it is safe to use the rib provided for actuation of the hook tip.
Passive Probe
P500 User Manual Guide Passive Probe
P500 User Manual Guide
ACCESSORIES
Accessories Voltage Ratings
Alligator Ground Lead 400 V CAT II, 300 V CAT III
Hook Tip 400 V CAT II, 300 V CAT III
BNC Adapter Not Applicable
Rf Ground Not Applicable
Screwdriver Not Applicable
Tip Cover Not Applicable
Screwdriver
Alligator Ground
RF Ground
Tip Cover
Hook tip
BNC Adapter