Document Number: 0610440-25-0001 (Rev: D)
Type 4-131
VELOCITY VIBRATION TRANSDUCER
Made in the United States of America
DESCRIPTION
The 4-131 is a seismic mass type velocity transducer designed for measuring vertical vibrations at low frequencies and high
temps up to +700°F (+371°C) or +900°F (+482°C). A seismic mass magnet is suspended by springs and a coil attached to the
case. The output signal results from relative movement between the magnet and coil when the case vibrates. This air damped
system operates above its natural frequency so the output is proportional to velocity.
SPECIFICATIONS
Sensitivity:
@ +77°F (+25°C*), 100 Hz, 1 IPS (25.4 mm/sec) RMS into a 10,000 ohm resistive load
Model No. Sensitivity Connection Method
4-131-0001:
4-131-0103:
4131-0104:
4131-0105:
4-131-0106:
4-131-0109:
4-131-0116:
4-131-0120:
4-131-0123:
4-131-0125:
4-131-0128:
4-131-0141:
4-131-0148:
135 mV/ IPS (5.31 mV/mm/s)
150 mV/ IPS (5.91 mV/mm/s)
145 mV/ IPS (5.71 mV/mm/s)
145 mV/ IPS (5.71 mV/mm/s)
100 mV/ IPS (3.94 mV/mm/s)
135 mV/ IPS (5.31 mV/mm/s)
200 mV/ IPS (7.87 mV/mm/s)
*135 mV/ IPS (5.31 mV/mm/s)
*135 mV/ IPS (5.31 mV/mm/s)
135 mV/ IPS (5.31 mV/mm/s)
*145 mV/ IPS (5.71 mV/mm/s)
150 mV/ IPS (5.91 mV/mm/s)
135 mV/ IPS (5.31 mV/mm/s)
Std. 173960 / 169500-XXXX
Std. 173960 / 169500-XXXX
Int. Cable, 20 ft., prepped leads
Int. Cable, 6 ft., prepped leads
Std. 173960 / 169500-XXXX
Int. Cable, 15 ft., prepped leads
Std. 173960 / 169500-XXXX
Int. Cable, 21 ft., prepped leads
Std. 173960 / 365195-XXXX
Int. Cable, 21 ft., prepped leads
Int. Cable, 21 ft., prepped leads
Int. Cable, 35 ft., prepped leads
Int. Cable, 50 ft., prepped leads
* 900°F (+482°C) version
Frequency Range: 15 – 2000 Hz
Mounting Orientation: Sensitive axis is in longitudinal direction
Amplitude Range: 0.07 inch (1.78 mm) pk-pk maximum
Acceleration Range: 0.02 to 50 g peak
Sensitivity Shift w/ Position: 10% maximum
Temperature Range: -65°F to +700°F optional +900°F (-54°C to +371°C optional +482°C)
Thermal Coefficient of Sensitivity: 0.02%/°F from +77°F (0.04%/°C from +25°C)
Shock: 50 g
Damped Resonant Frequency: 15 Hz nominal
Excitation: Self-generating
Insulation Resistance: 100K ohm minimum
Polarity: Pin 1 is positive when case is moved away from base.
Weight: 5.8 oz (164 g) nominal
Accessories Included: Calibration Record
Optional Accessories: Cable w/connector (P/N 169500-XXXX)
Connector (P/N 173960)
INSTALLATION
Rigidly attach the transducer to the equipment under test using four screws and lock washers. Be sure to tighten the screws
securely. Connect one end of cable assembly to the transducer and the other end to the vibration monitoring device.
CAUTION: Never apply in excess of 45VDC across the terminals.
REPAIR
For further information contact:
Customer Service
CEC Vibration Products
746 Arrow Grand Circle
Covina, CA 91722
Phone: 626-938-0200
Fax: 626-938-0202
Document Number: 0610440-25-0001 (Rev: D)
COIL
TRANSDUCER
SHIELD
SIG
SIG
SIGNAL
CONDITIONER
COM
DIFFERENTIAL
INPUT
COIL
TRANSDUCER
SHIELD
SIG
COM
SIGNAL
CONDITIONER
SINGLE ENDED
INPUT
SENSITIVE AXIS
OF 4-131
SENSOR
0°
90°
180°
270°
12 o'clock
6 o'clock
BAD
OK
OK

Other cec Transducer manuals

cec 4-137 User manual

cec

cec 4-137 User manual

cec 4-126 User manual

cec

cec 4-126 User manual

Popular Transducer manuals by other brands

SIAP+MICROS t003 TRH User manual and maintenance

SIAP+MICROS

SIAP+MICROS t003 TRH User manual and maintenance

celesco MTA Series quick start guide

celesco

celesco MTA Series quick start guide

Humminbird TROLLING MOTOR TRANSDUCER installation guide

Humminbird

Humminbird TROLLING MOTOR TRANSDUCER installation guide

Camille Bauer 542-41210 operating instructions

Camille Bauer

Camille Bauer 542-41210 operating instructions

Mianyang Weibo Electronic WBV411S07 user manual

Mianyang Weibo Electronic

Mianyang Weibo Electronic WBV411S07 user manual

MAMAC SYSTEMS PR-276 technical information

MAMAC SYSTEMS

MAMAC SYSTEMS PR-276 technical information

Raymarine RVM-100 installation instructions

Raymarine

Raymarine RVM-100 installation instructions

SSET CE-IJ03 SK-0.5 Series instructions

SSET

SSET CE-IJ03 SK-0.5 Series instructions

MBS EMBSIN 101 I operating instructions

MBS

MBS EMBSIN 101 I operating instructions

Allen-Bradley PowerFlex 4M installation instructions

Allen-Bradley

Allen-Bradley PowerFlex 4M installation instructions

M-system 18PV instruction manual

M-system

M-system 18PV instruction manual

Ashcroft A2X instruction sheet

Ashcroft

Ashcroft A2X instruction sheet

S+S Regeltechnik Thor II Series Operating Instructions, Mounting & Installation

S+S Regeltechnik

S+S Regeltechnik Thor II Series Operating Instructions, Mounting & Installation

S+S Regeltechnik PREMASREG 7100 VA ID Series Operating Instructions, Mounting & Installation

S+S Regeltechnik

S+S Regeltechnik PREMASREG 7100 VA ID Series Operating Instructions, Mounting & Installation

Ashcroft ZL91 Installation and maintenance sheet

Ashcroft

Ashcroft ZL91 Installation and maintenance sheet

Dwyer Instruments IS626 Series Installation and operating instructions

Dwyer Instruments

Dwyer Instruments IS626 Series Installation and operating instructions

Lumel P30H user manual

Lumel

Lumel P30H user manual

promesstec UMU 500 user manual

promesstec

promesstec UMU 500 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.