CEM DT-9935 User manual


Page
Dual Display LCR Meter Instruction Manual
Contents
Safety Summary
Safety Guidelines
Functional description
Introduction
Features
Front Panel Overview
Front Panel Display Descriptions
Powering Instrument
Installing Battery
Low Battery Indication
Operation Instructions
Accuracy Specification
Supplemental Information
Selecting Test Frequency
Selecting Series or Parallel Mode
Accuracy Discrepancies
Guard Terminal
.........................................
........................................
....................................
..........................................
..............................................
....................................
......................
.....................................
.....................................
................................
...................................
................................
...............................
.............................
.....................
..............................
......................................
1.
2.
3.
3-1.
3-2.
4.
4-1.
5.
5-1.
5-2.
6.
6-1.
7.
7-1.
7-2.
7-3.
7-4.
3
4
5
5
5
6
7
8
8
8
9
15
20
20
21
22
23

1.Safety Summary
The following safety precautions apply to both operating and
maintenance personnel and must be observed during all phases
of operation, service, and repair of this instrument.
Do not operate the instrument in the presence of flammable
gases or fumes. Operation of any electrical instrument in such an
environment constitutes a definite safety hazard.
Instrument covers must not be removed by operating personnel.
Component replacement and internal adjustments must be made
by qualified maintenance personnel.
Do not install substitute parts or perform any unauthorized
modifications to this instrument. Return the instrument to
distributor for service and repair to ensure that safety features
are maintained.
and statements, such as the following
examples, denote a hazard and appear throughout this manual.
Follow all instructions contained in these statements.
A statement calls attention to an operating procedure,
practice, or condition, which, if not followed correctly, could result
in injury or death to personnel.
A statement calls attention to an operating procedure,
practice, or condition, which, if not followed correctly, could result
in damage to or destruction of part or all of the product.
Do Not Operate In an Explosive Atmosphere
Keep Away From Live Circuits
Do Not Substitute Parts or Modify The Instrument
Warnings And Cautions
Warning Caution
Warning
Caution
3
Dual Display LCR Meter Instruction Manual

4
2.Safety Guidelines
This meter is for indoor use, altitude up to 2,000 m.
The warnings and precautions should be read and well
understood before the instrument is used.
When measuring in-circuit components, first de-energize the
circuits before connecting to the test leads.
Discharge capacitor before testing.
Use the meter only as specified in this manual. Otherwise, the
protection provided by the meter may be impaired.
The power for the meter is supplied with six standard 1.5V
batteries.
Do not measure a capacitor that is not fully discharged.
Connecting a charged or partially charged capacitor to the input
terminals will damage the instrument.
When measuring within a circuit, the circuit must be de-energized
before connecting the test leads.
When used in a dusty environment, the instrument should be
wiped and cleaned regularly.
Do not leave the instrument exposed to direct heat from the
sun for long periods of time.
Before removing the cover, ensure that the instrument is
disconnected from any circuit and is powered OFF.
To ensure that you use this device safely, follow the safety
guidelines listed below:
Caution
•
•
•
•
•
•
Dual Display LCR Meter Instruction Manual

3.Functional Description
The LCR meter could measure Inductance/Capacitance/
Resistance with secondary parameters including dissipation factor
(D), quality factor(Q), phase angle(θ), equivalent series/parallel
resistance(ESR or Rp). The meter is fully auto ranging operation
for AC impedance & DC resistance measurement. It means the
user could measure the L/C/R components directly at AUTOLCR
smart mode without changing the function key. User could also
select the target test frequencies of 100Hz/120Hz/1kHz/10kHz
/100kHz depending on DUT(device under test) type. Components
could be measured in series or parallel mode according to the DUT
impedance automatically.
Dual LCD display
Auto LCR smart check and measurement
Series/Parallel modes are selectable
Ls/Lp/Cs/Cp with D/Q/
3-1 Introduction
Features3-2
•
•
•
•
•
•
•
•
•
•
•
θ/ESR parameters
Support DCR mode 200.00Ω~200.0MΩ
Five different test frequency are available: 100/120/1k/10k/
100k Hz
est AC signal level: 0.6mVRMS typ.
Test range: (ex. F=1kHz)
200.00 μH ~ 2000.0 H
2000.0 pF ~ 2.000 mF
20.000 Ω ~ 200.0 MΩ
Multi-level battery voltage detector
Support Backlight & Buzzer sound driver
Primary Parameters Display:
DCR:DC Resistance
Ls:Serial Inductance
Lp:Parallel Inductance
Cs:Serial Capacitance
Cp:Parallel Capacitance
T
L:
C:
R:
5
Dual Display LCR Meter Instruction Manual

θ
Rs:Serial Resistance
Rp:Parallel Resistance
Second Parameter Display:
:Phase Angle
ESR:Equivalence Serial Resistance
D:Dissipation Factor
Q:Quality Factor
•
4.Front Panel Overview
6
Dual Display LCR Meter Instruction Manual
Discharge Capacitor Before Testing
D/Q
ESR
SER
PAL
REL
%
FREQ HOLD
SORTING
ENTER
CAL FUNC APO
Digital LCR Meter
020406080100%
GUARD
GUARD GUARD
On all inputs
30V
MAX
HOLD Cal APO Auto LCR
Range
Ls Lp
Cs Cp
Rs Rp
DCR
Sorting
To l 0 .12 0 5 % 8 0 %
12 0 10 0 K H z
ESR
RP
DQ
Mk
F
F
Mk
P
P
%
SETUP
1
2
3
4
5
6
7
8
16
9
10
11
12
14
15
13

4-1 Front Panel Display Descriptions
1.LCD Display
2.Mode (Auto LCR/L/C/ACR/DCR) selection button
3.Calibration mode selection button
4.Sorting mode button
5.Secondary Display mode (for dissipation factor(D), quality
factor (Q), phase angle (θ), equivalent series resistance (ESR),
equivalent parallel resistance(Rp) measurement) selection and
the modify sorting value ( ) button
6.Test Frequency selection button
7.Relative mode and the modify sorting value ( ) button
8.Power ON/OFF button
9.Enter modify sorting value mode button
10.APO (Auto power off) button
11.Modify sorting value ( ) button
12.Confirm and select the value user need to modify in sorting
mode
13.Hold Display mode button
14.Parallel or Series measurement method selection and the
modify sorting value ( ) button
15.Back light button
16.Input sockets (banana jack inputs) and terminals for positive,
negative, and guard (see “Guard Terminal” in “SUPPLEMENTAL
INFORMATION” section for details)
7
Dual Display LCR Meter Instruction Manual

5.Powering Instrument
Before beginning to operate the instrument, a power source
is necessary for it to turn on. Installing Battery
The LCR meter use battery to provide power to the instrument
so that it can be portable. It use six standard 1.5V size batteries.
Place the meter upside down. Open up the back-flip stand, and
locate the screw that tightens the battery compartment cover.
Use a screwdriver to unscrew and remove the cover.
Insert six 1.5V batteries into compartment. Note the positive
(+) and negative (-) terminals as indicated inside the battery
compartment. Be sure to insert the battery with matching
polarity.
Place the battery compartment cover piece. Place screw at the
bottom of the cover piece and tighten down with a screw driver.
Push and hold down the button for 2 seconds to turn on
the instrument.
The LCR meter has a battery indicator to notify the user
when to replace battery. When it displays ,the battery
voltage is below normal working voltage. In this case,accuracy
of the meter will also decrease. It is recommended that the
battery be replaced as soon as possible before continuing
operation.See "Installing Battery" for instructions.
5-1 Installing Battery
2 Low Battery Indication
•
•
•
•
5-
8
Dual Display LCR Meter Instruction Manual

6.Operation Instructions
Push key function allowed to be active will be marked as "
When LCR power on,all of the LCD segments will be ON for 2
seconds. Then the default initialization process will be started.
The default mode is AUTOLCR smart mode and the default test
frequency is 1 kHz.When the PWR_KEY is pushed during power
-on mode,the meter will enter power-off mode.The LCD will show
the "OFF" state before the whole system enters the power off
status.
In order to extend the battery life, except of using external
power supply, APO feature will be helpful. It can be enabled or
disabled APO function by press the APO button and the LCD will
show whether the function is enable or not. When all function
keypads do not be pushed or impedance range switching detected
within 5 minutes, the system will launch the alarm buzzer beep
at three times before the auto power-off status. During the period
of alarm, the meter will be kept in operation by pushing any
function key again. If any key is not in operation further, the
system power will be off.
If the function keypad available is pushed,the buzzer output
beeps once.If the function keypad not available is pushed,the
buzzer beeps twice.
◆"
•
•
•
Power ON/OFF
Auto power off
Buzzer
Keypads
AUTOLCR
L
C
ACR
DCR
FUNC HOLD Dqθ S/P BKLIT SORT REL%
FREQ
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
◆
9
Dual Display LCR Meter Instruction Manual

•
•
•
Backlight
Battery detect
Primary impedance with secondary parameter test mode
When user push the button, the backlight will be active.
Push the key again to disable the backlight. When the
backlight is active to last for 60 seconds, the backlight will be
disabled automatically.
The meter will detect the battery multi-level voltages periodically.
The LCD indicators of battery life will be disappeared according
to the decreasing of battery voltage.
When AUTO/L/C/R function selection key is pushed, the
main test mode could be selected sequentially: Auto-LCR mode
Auto-L mode Auto-C mode Auto-R mode DCR mode Auto-
LCR mode. The default test mode is Auto LCR mode which could
check the type of impedance smartly and enter to the L/C/R
measurement mode automatically. The secondary parameter will
follow the L/C/R measurement. It means that (L + Q) , (C+D) ,
(R+Q) are combined in one group respectively. When Auto-L
or Auto-C mode is selected, the impedance measurement is auto
ranging. The primary LCD display will show the inductance or
capacitance of DUT (device under test). The secondary LCD display
will show the quality or dissipation factor. The D/Q/
→→→→→
2
2
θ/ESR value
can also be shown by pressing the button. When Auto-R
(ACR mode) or DCR mode is selected, the secondary parameter
is omitted.
When Auto-LCR mode is active,the secondary parameter
will show the equivalent resistance in parallel mode (Rp) to
replace the D factor if the C measured value of DUT is less
than 5pF.
Auto-LCR mode only.During Auto-R mode or DCR mode,
the secondary parameter is not available.
Note:
Note:
FUNC
D/Q
ESR
10
Dual Display LCR Meter Instruction Manual

Discharge Capacitor Before Testing
D/Q
ESR
SER
PAL
REL
%
FREQ HOLD
SORTING
ENTER
CAL FUNC APO
Digital LCR Meter
020406080100%
GUARD
GUARD GUARD
On all inputs
30V
MAX
HOLD Cal APO Auto LCR
Range
Ls Lp
Cs Cp
Rs Rp
DCR
Sorting
To l 0.12 0 5 % 8 0%
12 0 10 0 K H z
ESR
RP
DQ
Mk
F
F
Mk
P
P
%
SETUP
Figure 2–Device under test display
The LCR meter offers the option to select between parallel or
series measurement mode. Depending on which mode is selected,
the method to measure the component will be different.
Additionally, one measurement mode may provide better
accuracies over the other measurement mode depending on the
type of component and the value of the component to be tested.
For more details, refer to the “SUPPLEMENTAL INFORMATION”
section.When any L/C/R functional mode is selected, the default
measurement in series or parallel mode is auto selected and the
AUTO segment will be shown on LCD display. It depends on the
total equivalent impedance measured. If the impedance is larger
than 10k
• Series/Parallel mode select
Ω, parallel mode is set and Lp/Cp/Rp is shown on the
display. If it is less than 10kΩ, series mode is set and Ls/Cs/Rs
is shownon the display. When the button is pressed, the
impedance measurement will be set in series mode or in parallel
mode sequentially. The LCD indicators for LS/LP/CS/CP/RS/RP
symbols will be indicated by related LCR measurement mode
setting.
SER
PAL
11
Dual Display LCR Meter Instruction Manual
(DUT)

CAL
REL
%
HOLD
•
•
•
Hold mode
The data hold function allows the user to freeze the display when
pressed,holding the measured value until data hold is turned off.
To use data hold, press the button once. The "HOLD"
indicator will display on the screen when data hold is active.
To disable the data hold,press again.The "HOLD" indicator
will disappear on the screen,and meter will remain in normal
operation mode.
Press the button reserve the current DUT readings (DCUR)
on primary display as a reference value (DREF) and the "
Turn On Data Hold
Turn Off Data Hold
Relative mode
Δ"
indicator will be active . The secondary display will show the
percentage of relative value REL%. The REL% = (DCUR – DREF)
/ DREF * 100%. Press the button again to show the reference
value DREF on primary display and the "Δ" segment will be blinking.
The percentage range is -99.9%~99.9%. When the relative value
is larger than double of reference value (DREF), the "OL%"
indication will be shown on the secondary display. Press and hold
downthe button for 2 seconds to exit the relative mode.
In order to improve the accuracy of high/low impedance, it is
recommended to do OPEN/SHORT calibration mode before
measurement. Press and hold down the button for 2 seconds
to enter calibration mode. The calibration procedure: OPEN
ready OPEN calibration(30s) SHORT ready
SHORT calibration(30s) . During open or short calibration
processing, the 30-second countdownwill be shown on LCD
panels. If the calibration
Calibration mode
1 2
→→ → → →
→
procedure is finished, the PASS or FAIL
symbol will shown on the primary display. If PASS symbol for
both OPEN and SHORT modes, the calibration data will be saved
HOLD
REL
%
REL
%
CAL
CAL
CAL
12
Dual Display LCR Meter Instruction Manual

SORTING
after push CAL key again.
ready means the input sockets or terminals have
nothing connected
ready means put a shorting bar or a short piece of
conductive metal (i.e. paper clip) across the "+" and "-"
input sockets or terminals.
OPEN
SHORT
The sorting mode could help the user to make a quick sort for
a bunch of components.Select the primary measurement mode
(L/C/R) based on the type of components to be measured.Insert
the component to be used as the "standard" reference value.
Another words, insert a known "good" component that will be
used for testing against all other components. Press button to
enter to the sorting mode.The sorting mode cannot be activated
unless the meter senses a component is connected to either the
input sockets or terminals.When sorting mode is activated,the
reference value,range and the tolerance settings can be modified.
Figure 3-Open Calibration (left) and Short Calibration (right)
• Sorting mode
13
Dual Display LCR Meter Instruction Manual
CalCal

FREQ
SORTING
The setting process:
range setting(use / ) reference value setting (use
/ / / ) tolerance setting(use / ) sorting mode
The tolerance range setting selection: +0.25% +0.5% +1%
+2% +5% +10% +20% +80%-20%. The default
tolerance is +1%.
In the sorting mode,the primary display to show PASS or FAIL
status depends on whether the impedance measured exceeds
tolerance range.The current measurement result will be shown
on the secondary display. Press the button to exit this mode.
If the component to be measured is a capacitor, be
sure that the capacitor is fully discharged BEFORE inserting it
into the input sockets or terminals. For large capacitors, it may
take longer periods of time for a full discharge. Inserting a
charged or partially charged capacitor into the meter’s input
sockets or terminals may produce an electric hazard and may
also damage the instrument, making it unusable.
When the button is pressed, the test frequency will be
changed sequentially. There are five different test frequencies
(100Hz/120Hz/1kHz/10kHz/100kHz) can be selected. The test
frequency can affect the accuracy of the results depending on what
frequency is selected and what type and value of a component
is being measured or tested.
For details on selecting the optimal test frequency for
measurement, refer to the " SUPPLEMENTAL INFORMATION "
section.
→→→
→→ →
→→
→→→→→
WARNING:
Test Frequency Select•
SETUP ENTER
ENTER
ENTER
14
Dual Display LCR Meter Instruction Manual

6-1 Accuracy Specification
Notes:
Inductance @ Ta =18 ~ 28 (De)C
Measurement performed at the test socket.
Measurements performed after correct open and short
calibration.
DUT and test leads must be properly shielded to guard if
necessary.
Q value is the reciprocal of DF.
Accuracies based within 10% to 100% of full scale of range;
values outside of range should be used as reference only.
— means parallel or series measurement mode.
Frequency = 100 Hz/120 Hz
Frequency = 1kHz
O
•
•
•
•
•
•
Range
20.000mH
200.00mH
2000.0mH
20.000H
200.00H
2000.0H
20.000kH
Resolution
1uH
0.01mH
0.1mH
1mH
0.01H
0.1H
0.001kH
Lx Accuracy
1.5% 10d
1.4% 15d
1.5% 15d
1.6% 10d
1.3% 10d
2.0% 15d
2.5% 15d
±
±
±
±
±
±
±
DF Accuracy
1.5% 50d
1.4% 50d
1.5% 50d
1.6% 50d
1.3% 50d
2.0% 50d
2.5% 0d
±
±
±
±
±
±
±
Measurement
Mode
Series
Series
Series
---
Parallel
Parallel
Parallel
Range
2000.0uH
20.000mH
200.00mH
2000.0mH
20.000H
200.00H
2000.0H
Resolution
0.1uH
1uH
0.01mH
0.1mH
1mH
0.01H
0.1H
Lx Accuracy
1.3% 10d
1.2% 10d
1.2% 10d
1.5% 15d
1.5% 15d
2.0% 10d
2.5% 15d
±
±
±
±
±
±
±
DF Accuracy
1.3% 50d
1.2% 50d
1.2% 50d
1.5% 50d
1.5% 50d
2.0% 50d
2.5% 50d
±
±
±
±
±
±
±
Measurement
Mode
Series
Series
Series
---
Parallel
Parallel
Parallel
15
Dual Display LCR Meter Instruction Manual

Frequency = 10kHz
Range
200.00uH
2000.0uH
20.000mH
200.00mH
2000.0mH
20.000H
Resolution
0.01uH
0.1uH
1uH
0.01mH
0.1mH
1mH
Lx Accuracy
1.8% 10d
1.5% 10d
1.2% 10d
1.5% 15d
2.0% 10d
2.5% 15d
±
±
±
±
±
±
DF Accuracy
1.8% 50d
1.5% 50d
1.2% 50d
1.5% 50d
2.0% 50d
2.5% 50d
±
±
±
±
±
±
Measurement
Mode
Series
Series
Series
---
Parallel
Parallel
Range
20.000uH
200.00uH
2000.0uH
20.000mH
200.00mH
Resolution
0.001uH
0.01uH
0.1uH
1uH
0.01mH
Lx Accuracy
2.5% 10d
1.5% 10d
1.3% 15d
2.0% 15d
2.5% 15d
±
±
±
±
±
DF Accuracy
2.5% 50d
1.5% 50d
1.3% 50d
2.0% 50d
2.5% 50d
±
±
±
±
±
Measurement
Mode
Series
Series
Series
Parallel
Parallel
Frequency = 100kHz
Capacitance@Ta=18~28 (De)
Frequency = 100 Hz/120 Hz
OC
Range
20.000nF
200.00nF
2000.0nF
20.000uF
200.00uF
2000.0uF
20.00mF
Resolution
1pF
0.01nF
0.1nF
1nF
0.01uF
0.1uF
0.01mF
Cx Accuracy
2.5% 10d
1.2% 10d
0.9% 10d
1.0% 15d
1.2% 10d
2.5% 10d
5.0% 10d
±
±
±
±
±
±
±
DF Accuracy
2.5% 50d
1.2% 50d
0.9% 50d
1.0% 50d
1.2% 50d
2.5% 50d
5.0% 50d
±
±
±
±
±
±
±
Measurement
Mode
Parallel
---
---
Series
Series
Series
Series
16
Dual Display LCR Meter Instruction Manual

Frequency = 1kHz
Range
2000.0pF
20.000nF
200.00nF
2000.0nF
20.000uF
200.00uF
2000uF
Resolution
0.1pF
1pF
0.01nF
0.1nF
1nF
0.01uF
1uF
Cx Accuracy
3.5% 15d
1.0% 10d
0.9% 10d
1.0% 10d
1.2% 15d
2.5% 10d
4% 20d
±
±
±
±
±
±
±
DF Accuracy
3.5% 50d
1.0% 50d
0.9% 50d
1.0% 50d
1.2% 50d
2.5% 50d
4% 50d
±
±
±
±
±
±
±
Measurement
Mode
Parallel
---
---
Series
Series
Series
Series
Frequency = 10kHz
Range
200.00pF
2000.0pF
20.000nF
200.00nF
2000.0nF
20.000uF
200.0uF
Resolution
0.01pF
0.1pF
1pF
0.01nF
0.1nF
1nF
0.1uF
Cx Accuracy
3.0% 8d
1.0% 10d
0.9% 10d
0.8% 10d
1.0% 8d
2.0% 8d
4.5% 15d
±
±
±
±
±
±
±
DF Accuracy
3.0% 50d
1.0% 50d
0.9% 50d
0.8% 50d
1.0% 50d
2.0% 50d
4.5% 50d
±
±
±
±
±
±
±
Measurement
Mode
Parallel
---
---
Series
Series
Series
Series
Frequency = 100kHz
Range
200.00pF
2000.0pF
20.000nF
200.00nF
2000.0nF
Resolution
0.01pF
0.1pF
1pF
0.01nF
0.1nF
Cx Accuracy
2.5% 15d
1.0% 8d
1.8% 8d
1.5% 10d
2.5% 15d
±
±
±
±
±
DF Accuracy
2.5% 50d
1.0% 50d
1.8% 50d
1.5% 50d
2.5% 50d
±
±
±
±
±
Measurement
Mode
Parallel
Parallel
Parallel
Series
Series
17
Dual Display LCR Meter Instruction Manual

Resistance @ Ta =18 ~ 28 (De)
OC
Frequency = 100 Hz/120 Hz
Range
200.00Ω
2.0000kΩ
20.000kΩ
200.00kΩ
2.0000MΩ
20.000MΩ
200.0MΩ
Resolution
0.01Ω
0.1Ω
1Ω
0.01kΩ
0.1kΩ
1kΩ
0.1MΩ
Rx Accuracy
1.2% 10d
0.8% 5d
0.9% 5d
0.7% 3d
1.0% 5d
2.2% 10d
2.5% 10d
±
±
±
±
±
±
±
Measurement Mode
---
---
---
---
---
---
—
Frequency = 1kHz
Range
20.000Ω
200.00Ω
2.0000kΩ
20.000kΩ
200.00kΩ
2.0000MΩ
20.000MΩ
200.0MΩ
Resolution
1mΩ
0.01Ω
0.1Ω
1Ω
0.01kΩ
0.1kΩ
1kΩ
0.1MΩ
Rx Accuracy
1.2% 10d
0.8% 5d
0.8% 3d
0.7% 3d
1.0% 5d
1.5% 10d
1.8% 10d
6.0% 50d
±
±
±
±
±
±
±
±
Measurement Mode
---
---
---
---
---
---
---
—
Frequency = 10kHz
Range
20.000Ω
200.00Ω
2.0000kΩ
20.000kΩ
200.00kΩ
2.0000MΩ
20.00MΩ
Resolution
1mΩ
0.01Ω
0.1Ω
1Ω
0.01kΩ
0.1kΩ
0.01MΩ
Rx Accuracy
1.5% 10d
0.8% 10d
0.9% 5d
0.8% 3d
1.0% 5d
2.5% 10d
2.8%10d
±
±
±
±
±
±
Measurement Mode
---
---
---
---
---
---
—
18
Dual Display LCR Meter Instruction Manual

Frequency = 100kHz
Range
20.000Ω
200.00Ω
2.0000kΩ
20.000kΩ
200.00kΩ
2.000MΩ
Resolution
1mΩ
0.01Ω
0.1Ω
1Ω
0.01kΩ
1kΩ
Rx Accuracy
2.3% 10d
1.5% 5d
0.8% 20d
0.8% 20d
1.5% 10d
2.5% 30d
±
±
±
±
±
±
Measurement Mode
---
---
---
---
---
—
DC Resistance @ Ta =18 ~ 28 (De)
Frequency = 100Hz/120Hz/1kHz/10kHz/100KHz
OC
Range
200.00Ω
2.0000kΩ
20.000kΩ
200.00kΩ
2.0000MΩ
20.000MΩ
200.0MΩ
Resolution
0.01Ω
0.1Ω
1Ω
0.01kΩ
0.1kΩ
1kΩ
0.1MΩ
±
±
±
±
±
±
±
Rx Accuracy
1.8% 10d
0.6% 20d
0.6% 10d
0.5% 3d
1.5% 5d
2.0% 5d
2.5% 5d
±
±
±
±
±
±
±
Measurement Mode
---
---
---
---
---
---
—
D value Accuracy @ Ta =18 ~ 28 (De)
OC
Freq. / Z
100/120Hz
1kHz
10kHz
100kHz
0.1- 1Ω
0.030
0.030
0.030
0.040
±
±
±
±
1–10Ω
0.010
0.010
0.010
0.030
±
±
±
±
10–100kΩ
0.009
0.009
0.009
0.010
±
±
±
±
100k–1MΩ
0.010
0.010
0.009
0.010
±
±
±
±
1M–20ΜΩ
0.020
0.020
0.010
0.020
±
±
±
±
20Μ−200MΩ
0.040
0.090
0.040
0.040
±
±
±
±
Freq. / Z
100/120Hz
1kHz
10kHz
100kHz
0.1- 1Ω
0.65
0.65
0.65
1.27
±
±
±
±
O
O
O
O
1–10Ω
0.36
0.36
0.36
0.65
±
±
±
±
O
O
O
O
10 – 100kΩ
0.23
0.23
0.23
0.49
±
±
±
±
O
O
O
O
100k – 1MΩ
0.45
0.45
0.45
0.65
±
±
±
±
O
O
O
O
1M – 20ΜΩ
0.65
0.65
1.35
1.35
±
±
±
±
O
O
O
O
20Μ−200MΩ
1.35
3.63
N/A
±
±
O
O
D value Accuracy @ Ta =18 ~ 28 (De)
OC
19
Dual Display LCR Meter Instruction Manual

7.Supplemental Information
This section provides supplemental information for user
consideration when operating the LCR meters.Some
recommendations and explanations are provided to help aid in
the use of some functions and features,in which can help the user
gain optimal and accurate measurement results.
Test frequency can greatly affect the results of measurement
reading,especially when measuring inductors and capacitors.This
section provides some recommendations and suggestions to
consider.
When measuring capacitance selecting,the right frequency is
important in obtaining the most accurate measurement results.
Generally,a 1 kHz test frequency is used to measure capacitors
that are 0.01
7-1 Selecting Test Frequency
Capacitance
μF or smaller.For capacitors that are 10 μF or larger,a
lower frequency of 120 Hz is used.Following this trend,high test
frequencies are best for testing very low capacitance components.
For large capacitance components,low frequency would be optimal.
For example,if the capacitance of the component is to be in the
mF range,than selecting 100 Hz or 120 Hz for test frequency
would give much better results.The results will also be obvious
because if the same component was tested with 1 kHz or 10 kHz,
the measured readings may look erroneous on the display.
In all cases,it is best to check with the manufacturer’s data
sheet in order to determine the best test frequency to use for
measurement.
Typically,a 1 kHz test frequency is used to measure inductors
that are used in audio and RF circuits.This is because these
components operate at higher frequencies and require that they
be measured at higher frequencies such as 1 kHz or 10 kHz.
However,a 120 Hz test signal is used to measure inductors that
Inductance
20
Dual Display LCR Meter Instruction Manual
Table of contents
Other CEM Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Brookfield
Brookfield CAP 1000+ operating instructions

Fieldpiece
Fieldpiece SMAN SM480VINT Operator's manual

CEMB
CEMB N130 quick guide

Agilent Technologies
Agilent Technologies 490 Micro GC Installation and configuration

TPI
TPI The Value Leader 716N manual

KERN
KERN ORD 45BM, ORD 92BM, ORD 85BM, ORD 92HM, ORD 1SM, ORD 3SM, ORD 2WM, ORD 1PM, ORD 2UM, ORD 5UM, ORD 6 US, ORD... operating instructions