manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Clements
  6. •
  7. Water Pump
  8. •
  9. Clements HiVac SUC 84109 User manual

Clements HiVac SUC 84109 User manual

CLEMENTS
HiVac
High
Vacuum
High Flow
Suction
Pump
SUC
84109
proven
through
performance
User
Manual
Manual
No,
SUC
91005
414
Issue
17
44
CLEMENTS
Safety
Thank
you
for
purchasing
this
Clements
HiVac
High
Suction
Pump.
For
your
safety
it
is
imperative
that
this
unit
only
be
operated
by
authorised
personnel
in
accordance
with
the
instructions
as
described
in
this
manual.
Operated
in
this
way,
the
HiVac
High
Suction
Pump
will
provide
the
standard
of
service
specified.
Due
to
continual
improvements
in
product
design,
the
HiVac
High
Suction
Pump
may
vary
in
detail
from
the
descriptions
in
this
manual.
In
the
event
of
further
questions
please
contact
your
local
distributor
or
BMDi
TUTA
Healthcare
direct.
User
Manual
.
HiVac
High
Suction
Pump
Manual
Number
SUC
91005
414
Issue
17
Copyright
©
2010
BMDi
TUTA
Healthcare
Pty
Ltd
The
information
in
this
manual
was
originated
by,
and
is
the
exclusive
property
of
BMDi
TUTA
Healthcare
Pty
Ltd.
It
is
furnished
for
customer
information
only,
and
is
not
an
authorisation
or
licence
to
make
this
product
or
to
furnish
this
information
to
others.
Level
17,
275
Alfred
St
i
BMDi
TUTA
Healthcare
Pty
Ltd
North
Sydney
NSW
2060
Australia
Phone:
+612
9466
5300
Fax:
+61
2
9922
7165
CLEMENTS
Contents
Classifications
ee
4
Specifications
..............5
ини
5
Transportation
and
Storage
Mere
ek
6
Description
eee
8
Installation
and
Operation..............................
9
Accessories
and
Spares
...................................
13
Maintenance
................
니 니 니 이 니 이 니
아 이
아 아 아 이 이 이
이
의
15
Troubleshooting
....................
i.
i
17
Diagnostic
Setup...
19
Disassembly
and
Repair..................................
21
Emissions
Guidance
Table
201
.......................
23
Immunity
Guidance
Table
202
es
24
Immunity
Guidance
Table
204
.......................
25
Warraniy..........................................
26
Contents
|
3
~
CLEMENTS
Identification
SUC
84109
HiVac
High
Vacuum
/
High Flow
Suction
Pump
Intended
Use
To
provide
a
continuous
vacuum
source,
within
the
stated
operating
vacuum
range,
for
the
aspiration
of
fluids
and
particulate
matter
in
medical
procedures
carried
out
by
clinically
trained
and
authorized
personnel.
Uninterrupted
Availability
This
equipment
is
mains
powered.
If
patient
care
requires
the
uninterrupted
availability
of
a
source
of
vacuum,
the
clinical
facility
must
make
provision
for
either
a
backup
power
supply
or
standby
suction
units
which
are
battery
or
manually
powered.
Classifications
GMDNS
Code
36776
GMDNS
Term
Suction
Unit,
High
Vacuum
/
High
Flow
GMDNS
Synonym
Aspirator
Device
Class Class
Ila
Electric
Shock
Class
I
Equipment
Protection
Additional
Defibrillation-proof
Type
BF
Applied
Part
Electric
Shock
Protection
Sterilisation
Not
supplied
in
sterile
state
Anaesthetic
NOT
Category
AP
Rating
NOT
Category
APG
Operation
Mode
Continuous
operation
Classifications
人
—
은
~
CLEMENT
Specifications
HiVac
SUC
84109
、
Power
Rating
220/240
V
50Hz
230
V
60Hz
1.6
A
>
1.6
A
Vacuum
Range
|0
to
-98kPa
0
to
-98
kPa
0
to-735
mm
Hg
0
to
—735
mm
Hg
Max
Flow
Rate
|32
litre
/
min
36
litre
/
min
(free
air)
Motor
Thermally
protected
PSC
motor
.
Pump
Oil
Less
twin
piston
Ambient
Continuous
operation
in
ambient
tempera-
Requirement
tures
of
5°C
to
35°C.
Filters
Inlet:
disposable
bacteria
filter
Exhaust:
clean,
bleached
cotton
wool
(5g)
Internal:
disposable
hydrophobic
filter
Collection
Jar
|
2.x
2
litre
autoclavable,
shatter
resistant
poly-
carbonate
jars
with
non-static
polypropylene
bungs
and
overflow
shut-off
valves.
Regulator
Needle
Valve
Gauge
Bourdon
tube
type
Dual
Scale,
mm
Hg
/
kPa.
Accuracy
+3%
FSD
Gauge
Range
“Toto
-760
mm
Hg
graduated
at
20
mm
Hg
O
to
-100
kPa
graduated
at
5
kPa
Unit
Width
406mm,
Depth
305mm,
Height
800mm
Dimensions
Weight
16.5kg
Standard
25°C,
Sea
Level,
100kPa
Conditions
Environment
Indoor
Specifications
5
44
CLEMENTS
Transportation
and
Storage
Environmental
conditions
for
transportation
and
storage
are
shown
in
the
following
table.
In
addition,
for
vehicular
transportation,
the
unit
should
be
packed
in
its
original
packaging
and
kept
upright.
-
Parameter
Minimum
Maximum
Temperature
10°C
40°C
Humidity
60%
RH
95%
RH
Barometric
Pressure
|
700
mBar
1060
mBar
Warning
Symbols
Legend
The
warning symbols
marked
on
the
equipment
and
their
meanings
are
shown
as
follows.
/
|
Attention,
consult
accompanying
documents
Dangerous
voltage
inside
4 上
Defibrillation-Proof
Type
BF
applied
part
6
Transportation
and
Storage
CLEMENTS
Electromagnetic
Interference
This
equipment
generates,
uses
and
can
radiate
radio
frequency
energy
and,
if
not
installed
and
used
in
accordance
with
the
instructions,
may
cause
harmful
interference
to
radio
communications.
However,
:
there
is
no
guarantee
that
interference
will
not
occur.
i’
a
particular
installation.
If
this
equipment
does
cause
hafmful
interference
to
radio
or
television
reception,
which
can
be
determined
by
turning
the
equipment
off
and
on,
the
user
is
encouraged
to
try
to
correct
the
interference
by
one
or
more
of
the
following
measures:
*
Reorient
or
relocate
the
receiving
antenna.
e
Increase
the
separation
between
the
equipment
and
the
receiver.
+
Connect
the
equipment
into
an
outlet
on
a
circuit
different
from
that
to
which
the
receiver
is
connected.
«
Consult
the
dealer
or
an
experienced
radio/TV
technician
for
his
or
her
help.
Changes
or
modifications
to
the
equipment
not
expressly
approved
by
Clements
could
void
the
user’s
authority
to
operate
the
equipment.
==
Electromagnetic
Interference
7
LE
CLEMENTS
Description
The
Clements
HiVac
High
Suction
Pump
is a
mobile,
enclosed
suction
pump
designed
for
high
suction
and
high
flow.
The
main
feature
of
this
unit
is
the
low
maintenance
piston
motor
and
pump
assembly
which
can
generate
suction
strengths
from
O
to
-735
mm
Hg
(-98
kPa)
with
flow
rates
up
to
32
litres
of
Free
Air
Per
Minute
(FAPM),
depending
on the
model
and
power
supply.
~
The
unit
is
mounted
on
large
rubber
wheels
which
make
it
easy
to
move
to
any
part
of
a
hospital.
The
unit
also
has
a
storage
compartment
at
the
rear
to
house
additional
tubing,
handpieces
and
other
items
needed
when
using
the
pump.
Main
Components
of
Unit
了
CLEMENTS
Installation
and
Operation
Installation
、
The
Clements
HiVac
High
Suction
Pump
is
supplied
with:
*
2x2
litre
Collection
Jar
with
Bung
Assembly
*
ix
Single
Use
Handpiece
*
Yellow
Suction
Tubirg
i
Important
Remove
strapping
from
motor
before
starting.
Ensure
power
is
disconnected
before
opening
unit.
WARNING
This
unit
can
generate
high
vacuums.
It
must
not
be
used
for
applications
which
require
low
vacuum.
Always
check
patient
requirements
before
applying
the
Clements
HiVac
High
Suction
Pump.
'
Power
Switch
Exhaust
Filter
Vacuum
Gauge
Handle
Regulator
Control
Knob
Tube
Storage
Hooks
OPO
NT
A
Inlet
Connection
Storage
Compartment
如
|
上
wii
Pump
(Internal)
For
applications
which
require
low
vacuum,
please
contact
your
distributor
for
information
about
the
Clements
Low
Suction
range
of
products.
Connecting
The
Collection
Jars
Because
the
unit
is
fitted
with
two
Collection
Jars
it
can
either
be
connected
in
a
single-jar
arrangement
or
a
double-jar
arrangement.
With
the
single-jar
arrangement,
only
one
jar
is
in
use
at
a
time,
with
the
second
jar
on
standby.
As
one
jar
is
filled
it
is
disconnected,
so
that
it
can
be
emptied,
and
the
other
jar
is
connected.,
The
double-jar
arrangement
allows
both
jars
to
be
in
use
at
the
same
time,
thus
providing
longer
periods
between
changeover.
4
8
Description
Installation
and
Operation
9
14
CLEMENTS
Single-Jar
Arrangement
Connect
This
Jar
ube
To
The
~
Outlet
Patient
Jar
Inlet
Overfiow
Cut
Spare
Off
Valves
In
Patient
Both
Jars
Tube
\
This
Jar
This
Jar
In
Use
On
Standby
To
connect
the
Collection
Jars
in a
single-jar
arrangement,
use
the
following
procedure
and
refer
to
the
block
diagram
below.
1.
Place
two
clean
jars
in
the
holders,
ensuring
that
both
jars
are
fitted
with
an
Overflow
Cut
Off
Valve.
2.
Connect
the
SUCTION
outlet
of
the
jar
to
be
used
first
to
the
Bacteria
Filter
inlet
on the
main
unit.
3.
Connect
the
PATIENT
inlet
of
the jar
to
be
used
first
to
the
Suction
Tube
leading
to
the
patient.
4.
When
the
first
jar
is
full,
disconnect
it
and
use
steps
2
and
3
to
connect
the
second
jar.
Hydrophobic
Shut
Off
Valve
Filter
(Internal)
Blue
Bacteria
Filter
(External)
γ
#
To
Patient
LY
>
Overflow
Cut
Off
Valves
in
Both
Jars
Suction.
PUMP
Jar
in
Jaron
ni
Use
Standby
10
Installation
and
Operation
CLEMENTS
Pump
Double-Jar
Inlet
Arrangement
Jar
PERA
Connect
This
Outlet
=
Tube
To
The
Al
Patient
Jar
Inlet
=
Beth
Jars
In
Use
Spare
Patient
0
:
Tube
Ověrflow
Cut
Off
Valve
{in
this
Jar
Only)
To
connect
the
Collection
Jars
in
a
double-jar
arrangement,
use
the
following
procedure
and
refer
to
the
block
diagram
below.
1.
Place
two
clean
jars
in
the
holders,
ensuring
that
only
one
jar,
Jar
1,
is
fitted
with
an
Overflow
Cut
Off
Valve.
Connect
the
SUCTION
outlet
(with
the
Overflow
Cut
Of
Valve)
of
Jar
1
to
the
Bacteria
Filter
inlet
on
the
main
unit.
Connect
the
PATIENT
inlet
of
Jar
1
to
the
SUCTION
outlet
of
the
jar
without
an
Overflow
Cut
Off
Valve,
Jar
2.
Connect
the
PATIENT
inlet
of
Jar
2
to
the
Suction
Tube
leading
to
the
patient.
Hydrophobic
Shut
Filter
(Internal)
Blue
Bacteria
Filter
(External)
Off
Valve
a
To
Patient
NO
Overflow
Cut
Off
Valve
Suction
Pump
Unit
Jar
2
a
Installation
and
Operation
il
LE
CLEMENTS CLEMENTS
Operation
If
the
motor
thermal
overload
trips,
switch
the
power
off at
the
|
mains
switch.
This
will
prevent
unexpected
re-starting
of
the
To
operate
the
Pump,
follow
the
steps
listed
below.
pump
when
the
thermal
overload
automatically
resets
upon
1.
Place
the
pump
on
a
level
surface
with
a
15cm
clearance
cooling.
from
walls
or
other
objects.
2.
Install
the
Collection
Jars
and Suction
Tubing
in
either
the
singlejar
or
double-jar
arrangement
as
described
in
the
Connecting
The
Collection
Jars
section
on
page
8.
elf
it
is
suspected
that
fluid
may
have
entered
the
pump,
:
either
by an
external
spill
or
by
a
jar
overflow,
the
unit
must
1
be
taken
out
of
service
untilithe
fault
has
been
fixed
and
its
.
continued
electrical
safegy
has
been
confirmed.
The
service
department
must
also
be
notified
of
any
potential
biohazard.
3.
Connect
the
unit
to
the
mains
power
supply
and
switch
the
power
on
at
the
mains
supply
outlet.
4.
Press
the
Power
Switch
down
into
the
‘|’
(ON)
position
and
the
Motor
will
run.
5.
Occlude
the
Suction
Tubing
and
adjust
the
Regulator,
by
turning
the
knob,
to
obtain
the
required
vacuum
level,
as
indicated
on
the
Vacuum
Gauge.
.
Accessories
and
Spares
6.
Check
that
there
is
no
loss
of
vacuum
in
the
Suction
Tubing
and
fittings.
Connect
the
catheter
to
the
patient.
Check
the
Collection
Jars
to
ensure
that
suction
is
being
SUC
80301
2
litre
Collection
Jar
for
Suction
Pum:
applied
correctly.
(Jar
Only)
Pe
・
SUC
80350
Metal
Handpiece
with
Finger
Valve
and
Multi
.
Please
Note:
Although
the
Collection
Jars
are
fitted
with
Overflow
Cut
Off
Valves,
the
fluid
levels
should
be
monitored
to
prevent
spill-over
inte
the
pump
unit.
The
Hydrophobic
Filter
fitted
to
the
inlet
of
the
pump
acts
as
a
single
use
shut-off
valve
to
protect
the
pump.
It
must
be
replaced
if
fluid
has
spilled
over
into
the
pump
unit.
Always
start
the
pump
with
Suction
Tubing
unoccluded.
Always
clear
any
blockages
before
re-starting
the
pump.
The
pump
motor
is
protected
by
a
thermal
overload
mechanism
that
will
stop
the
pump
if
the
temperature
exceeds
a
pre-set
level.
The
mechanism
will
then
automatically
restart
the
Pump
after
the
motor
has
cooled
to
the
appropriate
temperature.
12
Installation
and
Operation
-
Accessories
SUC
80297
001
Yellow
Suction
Tubing
(20
metre
roll)
Hole
Tip
SUC
80404
001
Disposable
Plastic
Handpiece
(Pack
of
10)
SUC
80490
..
|Metal
Handpiece
with
Multi
Hole
Tip
SUC
84100
162
Blue
Bacteria
Filter
(Set
of
12)
SUC
84200
033
Bung
Assembly
with
Overflow
Cut
Off
Valve
for
2
litre
Collection
Jar
+
SUC
91005
414
User
Manual
for
the
HiVac
High
Suction
Pump
κ
Accessories
and
Spares
13
aa
CLEMENTS
Spares
MUL
94003
702
|
Fuse
T5.0
Amp
250V
MUL
94004000
|
Power
Switch
SUC
37043
Cotton
Wool
for
Suction
Pumps
Exhaust
Filter
(Pack
of
10,
5
grams
each)
SUC
80330
High
Vacuum
Gauge
0-100kPa
SUC
84100
049
Exhaust
Filter
Assembly
for
Suction
Pumps
SUC
84100121
Resilient
Motor
Mount
(Pack
of
4)
SUC
84100
405
Hydrophobic
Shut-Off
Valve
Filter
(Pack
of
5)
SUC
84100
505
Footswitch
Kit
for
Suction
Pumps
with
(Switch,
Pneumatic
Actuator)
SUC
84109
002
Motor
and
Pump
for
Hi
Vac/Hi
Flo
SUC
84200
501
Seal
for
Collection
Jar
Bung
(Set
of
2)
SUC
84200
502
Suction
Inlet
Nipple
for
2
litre
Jar
Bung
(Set
of
4)
SUC
84200
503
Kit
Grey
Bung
Float
Chamber
(Set
of
2)
SUC
84200
518
Washer
for
Overflow
Valve
(Pack
of
4)
SUC
84200
519
Blue
Male
Nipple
for
Bacteria
Filter
{Pack
of
10).
SUC
84200
520
Blue
Female
Nipple
for
Bacteria
Filter
(Pack
of
10)
SUC
89107
Kit
for
Pump
Vacuum
Controller
(Needle
Valve
Type)
SUC
89240
069
Regulator
Knob
Replacement
Kit for
Suction
Pumps
(Regulator
Knob,
Screw)
14
Accessories
and
Spares
~
=
-
CLEMENTS
Maintenance
、
IMPORTANT
There
are
no
user-serviceable
components
inside.
Maintenance
must
be
carried
out
by
qualified
personnel
only.
“To
keep your
Clements
HiVae
‘High
Suction
Pump
in
good
condition,
the
following
maintenance
is
recommended.
After
Each
Operation
Blue
Bacterial
Filter
The
blue
bacterial
filter
is
an
important
device
to
help
protect
the
pump
from
moisture
and
bacteria
taken
from
patient
airways.
Clements
recommend
that
the
filter
be
changed
every
month
or
when
the
unit
is
used
on
infectious
patients
or
when damp
or
discoloured.
Housing
Wipe
clean
with
a
damp
soapy
cloth.
Do
not
immerse
or
allow
liquid
to
enter
the
housing.
Suction
Tubing
Suction
Tubing
should
be
sterilised
with
sterilants
determined
by
local
protocols
and
compatible
with
PVC
tubing.
Suction
Tubing
may
be
autoclaved
to
a
maximum
of
121°C
for
20
minutes.
At
higher
temperatures
the
Suction
Tubing
will
discolour
and
lose
shape.
Collection
Jars
Place
jars
upright
or
up-side-down
in
autoclave:
not
on
their
sides.
Autoclave
to
a
maximum
of
136°C
for
10
minutes.
Do
not
use
phenolic
solutions
as
disinfecting
agents
in
polycarbonate
jars.
Bung
Disconnect
float
cage
and
float
from
Bung.
Place
all
items
upright
or
up-side-down
in
autoclave:
not
on
their
sides.
Autoclave
all
items
to
a
maximum
of
136°C
for
10
minutes.
Note:
Do
not
use
abrasive
cleaning
agents
as
this
may
damage
plastic
surfaces
and
reduce
the
expected
performance
of
an
item.
i
Maintenance
15
44
CLEMENTS
After
Every
100
Hours
or
2
Months
of
Operation
.
Replace
the
Exhaust
Filter:
1.
Remove
the
Exhaust
Filter
from
the
rear
of
the
unit.
2.
Unscrew
the
caps
and
replace
the
cotton
wool.
The
filter
takes
5
gram
of
bleached
and
teased-out
cotton
wool.
+
Check
all
Suction
Tubing
and
replace
if
it
is
perished,
soft
or
discoloured.
Also
check
the
tubing
nipples.
.
*
Check
the
seal
rings
on
the
Bungs
and
replace
them
if
they
are
hard,
cracked
or
perished.
Check
the
fit
of
the
Bung
in
the
Collection
Jar.
«
Check
the
Overflow
Cut
Off
Valve
seals
and
replace
them
if
they
are
perished
or
damaged.
IMPORTANT
The
preceding
maintenance
should
also
be
carried
out
immediately
after
use
with
infectious
patients.
Disposal
and
Recycling
:
Waste
Materials
The
contents
of
the
collection
jars,
suction
tubing,
bacteria
filter,
internal
hydrophobic
shut-off
filter,
or
the
cotton
wool
in
the
exhaust
filter
may
contain
biohazard
wastes.
Handle
using
safe
handling
procedures,
which
may
include
the
use
of
rubber
gloves
and
eye
protection,
and
dispose
of
according
to
local
protocols
for
biohazard
materials.
Recycling
At
the
end
of
their
service
life,
the
pump
and
accessories
should
be
dismantled
if
necessary,
and
returned
to
a
local
materials
recycling
centre.
16
Maintenance
-
CLEMENTS
Troubleshooting
Fault
Check
Rectify
No
Suction
But
Tubing
Blocked
Clear
or
Replace
Motor
is
Working
Overflow
Cut
Off
Clear,
Replace
or
Valve
Closed
Empty
Jar
Hyd:Sphobic
Shut-
|Replace
Off
Valve
Filter
Hydraphobic
Shut-
Blocked
„
(Off
Valve
Filter
Inlet
Blocked
Clear
Vacuum
Gauge
Defective
Replace
Vacuum
Gauge
Regulator
Blocked
Clear
or
Replace
Bung
Not
Sealing
Adjust
or
Fit
New
Seal
Regulator
at
Zero
Adjust
Setting
No
Suction:
Mains
Fuse
Blown
|Replace
Motor
Not
Working
Faulty
Mains
Power
|
Replace
Lead
Power
Not
Turned
Turn
On
On
Replace
Unit
Fuses
Blown
Thermal
Overload
Activated
Wait
For
Unit
To
Cool
Down
No
Suction
But
Humming
Can
Be
Heard
From
Motor
Blockage
In
Pump
Dismantie
And
Clean
Pump
Motor
Windings
Replace
Motor
Troubleshooting
17
14
CLEMENTS
-
‘CLEMENTS
Fault
Check
Rectify
Not
Enough
Suction
Regulator
Setting
Too
Low
Adjust
To
Correct
Setting
Leaks
In
Tubing
or
Fittings
Tighten
Connections
Blockage
In
Tubing
|
Clear
or
Replace
or
Fittings
“
Hydrophobic
Shut-
|Replace
Off
Valve
Filter
Hydrophobic
Shut-
Blocked
Off
Valve
Filter
Excessive
Noise
Pump
Worn
Replace
Pump
Coming
From
Pump
Pump
Incorrectly
|
Assemble
Correctly
Assembled.
Pump
Runs
Thermal
Overload
Is
|Disconnect
Mains
Intermittently
Cutting
In
and
Out
¡Power
And
Wait
For
Unit
To
Cool
Down
Replacing
the
Hydrophobic
Shut
Off
Valve
Filter
Transparent
Filter
Housing
hs
To
Pump
Filter
Media.
—
To
Patient
To
ensure
there
is
no
entry
of
fluid
into
the
pump
the
Hydrophobic
Shut
Off
Valve
Filter
MUST
be
fitted
as
shown
above.
18
Troubleshooting
Setup
for
Diagnostic
and
Performance
Testing
Note
that
the
following
arrangements
are
not
used
for
actual
suctioning
applications.
They
are
specified
to
remove
unnecessary
variations
when
diagnosing
faults
and
as
a
standard
setup
for
performance
measurement.
Vacuum
Check
è
.
ῃ
Max|
4.
Unoccluded
-
zero
check
©
©
acuum
Gauge
Disconnect
all
items
from
inlet
and
with
pump
switched
off,
confirm
that
gauge
reads
zero.
A
non-zero
reading
indicates
a
faulty
gauge.
Set
vacuum
control
knob
to
maximum,
switch
on
pump
and
confirm
that
gauge
reads
zero.
A
non-zero
reading
indicates
obstruction
in
internal
tubing
or
connections.
T
x
©
©
Max
2.
Occluded
-
vacuum
check
Vacuum
Gauge
Switch
on
pump
and
ocelude
inlet.
Note
maximum
vacuum
reading.
Max
©
3.
Occluded
-
gauge
check
Vacuum
Calibrated
Gauge
Vacuum
Gauge
Connect
a
calibrated
vacuum
gauge
directly
to
inlet
and
repeat
maximum
vacuum
reading.
Confirm
that
pump
gauge
reads
within
the
specified
tolerance.
q
Diagnostic
Setup
19
fa
CLEMENTS
Flow
Check
1.
Occluded
-
leak
check
Jar
SUC
81099
(1
litre)
or
SUC
80312
(2iitre)
Tubing:
SUC
80297
(8x13mm)
Vacuum
F-
Gauge
Connect
pump
as
shown
in
diagram.
Set
vacuum
control
knob
to
maximum,
and
switch
on
pump.
Occlude
jar
inlet
and
con-
firm
that
pump
achieves
the
same
maximum
vacuum
as
in
the
previous
vacuum
check
setup
.
Any
difference
indicates
leaks
in
jar
or
connections.
2.
Unoccluded
-
flow
check
Max
ο
©
Vacuum
Gauge
Jar
SUC
81099
(1
litre)
er
SUC
80312
(2
litre)
Tubing:
SUC
80297
(8x13mm)
Calibrated
Collection
Flowmeter
Jar
Connect
pump
as
shown
in
diagram.
Set
vacuum
control
kneb
to
maximum,
and
switch
on
pump.
Connect
jar
inlet
to
calibrated
flow
meter
and
note
flow
reading.
If
flow
is
significantly
below
specification,
check
internal
tubing
and
pump
itself
(or
internal
shutoff
valve
if
fitted).
20
Diagnostic
Setup
‘CLEMENTS
Disassembly
and
Repair
Dismantling
The
Pump
The
steps
for
dismantling
the
Pump
are
as
follows.
1.
Disconnect
the
unit
from
the
mains
power
supply.
2.
Remove
Collection
Jars-and
Suction
Tubing.
3.
Remove
the
4
screws
from
the
base
plate
tabs
at
the
rear
of
the
unit
and
push
the
housing
forward
off
the
base
plate.
4.
Remove
the
four
cylinder
head
screws
from
each
head.
Remove
the
cylinder
heads.
5.
Check
O-ring,
pistons
and
replace
if
worn,
cracked
or
perished.
6.
Clean
the
inside
of
the
cylinder
head.
Check
flapper
valves
and
replace
if
worn,
cracked
or
perished.
7.
Reassemble
pump.
Disassembly
and
Repair
21
aa
CLEMENTS
Fuse
Replacement
The
Pump
is
fitted
with
two
cartridge
fuses
located
in
fuse
holders
that
are
accessible
from
outside
the
unit.
To
replace
the
fuses
use
the
following
procedure.
1.
Disconnect
the
unit
from
the
mains
power
supply:
2.
Locate
the
two
fuse
holders
which
are
mounted
on the
rear
side
of
the
unit
at
the
lower
right-hand
side.
3.
Lift
off
the
fuse
holder
cover.
4.
Remove
the
old
fuses
and
inspect
them.
If
they
are
blown,
replace
them
with
new
fuses.
=
5.
Replace
the
fuse
holder
cover.
6.
Connect
the
unit
to
the
mains
power
supply,
switch
the
power
on,
and
ensure
that
the
unit
is
operating
correctly.
Electrical
Wiring
Diagram
IEC
INLET
MAIN
SWITCH,
ェ
ーーーーーーーーー
1
|
BROWN
RED
4
ESA
Co
一
I
i 1
E
O
| |
NOT
ㅣ
No
ol
Επ
LIGHT
BLUE
|
GREEN
/
YELLOW
2X
15.0
FUSES
GREEN
/
YELLOW
PSC
MOTOR
22
Fuse
Replacement
CLEMENTS
EMC
Information
Tables
per
EN60601-1-2:2001-09
Section
6.
‘2001
Medical
electrical
equipment
-
Part
1-2:
General
ity
-
in
accordance
with
EN
60601-
requirements
for
safety
-
Collateral
standard:
Electrornagnetic
compatil
Requirements
and
tesis
1)
"Medical
Electrical
Equipment
needs
special
precautions
regarding
EMC
and
needs
to
be
installed
and
put
inte
service
according
to
the
EMC
information
provided
in
the
Accompanying
Documents”
{the
following
tables).
2)
“Portable
and
Mobile
RF
Communications
Equipment
can
affect
Medical
Electrical
Equipment.”
è
3}
“The
equipment
or
system
should
not
be
used
adjacent
to
or
stacked
with
other
equipment
and
that
if
adjacent
or
stacked
use
is
necessary,
the
equipment
or
system
should
bé
observed
to
verify
normal
operation
in
the
configuration
in
which
it
Is
used.”
The
following
tables
provide
information
regarding
the
EMC
characteristics
of
this
Medical
Electrical
Equipment.
7
nce
levels
of
immunity
referred
te
in
tables
202 and 204
have
not
been
arrived
at
by
testing,
but
by
declaration
on
the
basis
that
the
HiVac
suction
pump
contains
no
electronic
components
and
is
inherently
immune
to
the
specified
levels
of
electromagnetic
disturbance.
The
performance
of
alt
functions
of
the
HiVac
suction
pump
are
considered
essential
performance
for
the
purpose
of
electromagnetic
immunity.
Table
201
—
Guidance
and
manufacturer’s
declaration
-
electromagnetic
emissions
-
all
equipment
and
systems
x
Guidance
and
manufacturer's
declaration
-
electromagnetic
emissions
The HiVac
suction
pump
is
intended
for
use
in
the
electromagnetic
environment
specified
below.
The
customer
or
user
of
the
HiVac
suction
pump
should
assure
that
it
is
used
in
such
an
environment.
Emissions
Test
Compliance
|
Electromagnetic
environment
-
guidance
Harmonie
Emissions
|
Class
A
The
HiVae
suction
pump
is
suitable
for
use
in
IEC
61000-3-2
ail
establishments,
including
domestic
establishments
and
those
directly
connected
to
the
public
low-voltage
power
supply
network
Voltage
fluctuations
/ |
Complies
that
supplies
b
igs
used
for
domestic
flicker
emissions
purposes.
IEC
61000-3-3
RF
emissions
Complies
The
HiVac
suction
pump
is
not
suitable
for
CSIPR
14-1
interconnection
with
other
equipment.
->
Emissions
Guidance
Table
201
23
th
CLEMENTS
Table
202
—
Guidance
and
manufacturer's
declaration
-
electromagnetic
immunity
-
for
all
equipment
and
systems
Guidance
and
manufacturer's
declaration
-
electromagnetic
immunity
The
HiVac
suction
pump
fs
intended
for
use
in
the
electromagnetic
environment
specified
below.
The
customer
or
user
of
the
HiVac
suction
pump
should
assure
that
if
is
+2
KV
line(s)
to
earth
#2
KV
line(s)
to
earth
used
in
such
an
environment.
Immunity
Test
TEC
61000
Compliance
Electromagnetic
test
level level
environment
-
guidance
(not
tested)
*
>
Electrostatic
+6
kV
contact
#6
kV
contact
Floors
should
be
wood,
charge
(ESD)
|
+8
kV
air
+8
KV
air
concrete
or
ceramic
tile.
If
IEC
61000-3-4
floors
are
covered
with
synthetic
material,
the
relative
humidity
should
be
at
least
30%.
Electrical
fast
42
kV
for
power
|
12
kV
for
power
|
Mains
power
quality
should
be
transientíburst
supply
lines
supply
lines
that
of
a
typical
commercial
or
TEC
61000-4-4
+1
KV
forinput!
[51
KV
for
input!
|
hospital
environment.
output
lines
output
lines
Surge
21
kV
line(s)
to
|+1kV
line(s)
to
{
Mains
power
quality
should
be
IEC
61000-4-5
line(s)
ine(s)
that
of
a
typical
commercial
or
hospital
environment.
Voltage
dips,
short
|
<5
%
Ur <5
%
Ur
Mains
power
quality
should
be
interruptions
and
|
(795%
dip
in
Ur
|
(795%
dip
in
Ur
|
that
of
a
typical
commercial
or
voltage
variations
|
for
0.5
cycle}
for
0.5
cycle)
hospital
environment.
The
unit
on
power
supply
|
40
%
Ur 40
%
Ur
shut
off
during
the
>95%
dip
input
lines
(60%
dip
in
Uy
|
(60%
dip
in
U;
|
[for 5
sec
disturbarice.
If
the
IEC
61000-4-11
for
5
cycles)
for
5
cycles)
user
of
the
HiVac
suction
70%
Ur
70%
Ur
pump
requires
continued
(30%
n
Ur
|
(30%
dip
in
U;
|
operation
during
power
mains
for
25
cycles)
for
25
cycles)
interruptions,
it
is
.
<5
%
Ur <5
%
Ur
recommended
that
the
HiVac
(795%
dip
in
Ur
|
(795%
dip
in
Ur
|
suction
pump
be
powered
for
5
sec)
for
5
sec)
from
an
uninterruptible
power
supply
or
battery.
Power
frequency
{3
A/m
3
Alm
Power
frequency
magnetic
|
(50/60Hz)
fields
should
be
af
levels
magnetic
field
©
characteristic
of
a
typical
IEC
6100048
location
commercial
ør
hospital
environment.
24
Immunity
Guidance
Table
202
CLEMENTS
Table
204
—
Guidance
and
manufacturer's
declaration
-
electromagnetic
immunity.
-
for
equipment
and
systems
that
are
NOT
life-supporting.
Guidance
and
manufacturer’s
declaration
-
electromagnetic
immunity
The
HiVac
suction
pump
is
intended
for
use
in
the
electromagnetic
environment
ーー
specified
below.
The
customer
or
user
of
the
HiVac
suction
pump
should
assure
that
itis
used
in
such
an
environment.
Immunity
Test
|
IEC
61000
Compliance
Electromagnetic
environment
-
test
level
evel
guidance
(not
tested)
*
Portable
and
mobi
coramunications
equipment
should
be
used
no
closer
to
any
part
of
the
HiVae
suction
pump,
including
cables,
than
the
recommended
separation
distance
calculated
from
the
equation
applicable
to
the
frequency
of
the
transmitter.
Recommended
separation
distance
Conducted
RF
3
Vims
3
Vims
d=[35M,NP
IEC
61000-4-6
150
kHz
to
80
MHz
Radiated RF
3Vim
*.
İsim
d=
[3.5/E,
VP
80
MHz
to
800MHz
1ЕС
61000-4-3
80MHZ
to
=
[7E,]VP
800
MHz
to
2.5GHz
2.5
GHz
where
P
is
the
maximum
output
power
rating
of
the
transmitter
in
watis
(W)
according
to
the
transmitter
manufacturer
and
dis
the
recommended
separation
distance
in
metre
(m).
Field
strengths
from
fixed
RF
transmitters,
as
determined
by
an
electromagnetic
site
survey,
should
be
less
than
the
compliance
level
in
each
frequency
range.
Interference
may
oceur
in
the
vicinity
of
equipment
marked
with
the
following
symbol
D
Immunity
Guidance
Table
204
25
CLEMENTS
Warranty
BMDi TUTA
Healthcare
Medical
Equipment
Pty
Limited
("BMDi
TUTA
Healthcare")
warrants
that
this
product
is
free
from
defects
in
workmanship
and
materials
for
a
period
of
12
months
(3
months
for
batteries)
from
the
date
of
shipment
by
BMDi TUTA
Healthcare
or
its
authorised
agent
to
the
purchaser.
Subject
to
the
conditions
of
this
warranty,
if
the
product
fails
to
operate
for
any
reason
within
the
warranty
period
and
the
product
is
returned
to
the
place
of
purchase
at
the
purchaser's
expense,
BMDi TUTA
Healthcare
will
repair
or
replace
the
product
free
of
charge.
Ha
valid
warranty
claim
is
made
within
30
days
from
the
date
of
shipment,
then
BMDi
TUTA
Healthcare
will
also
reimburse
the
purchaser
for
reasonable
freight
costs
în
returning
the
product
to
the
place
of
purchase.
Conditions
of
Warranty
1.
The
product
must
be
returned
to
the
place
of
purchase
with
proof
of
purchase.
2.
This
warranty
is
only
available
to
the
original
purchaser
of
the
product.
3.
The
product
must
not
have
had
its
serial
number
removed,
defaced
or
changed,
its
casing
opened,
its
power
supply
altered
or
have
been
tampered
with
in
any
other
way.
4.
This
warranty
does
not
cover
:
e
inadequate
or
incorrect
site
preparation;
*
improper
installation;
*
connection
to
the
wrong
voltage;
e
failure
of
the
product
due
to
misuse;
e
the
use
or
operation
of
the
product
outside
of
the
physical,
electrical
or
environmental
specifications
of
the
product;
e
use
in
a
manner
or
environment
in
which
the
product
is
not
designed
to
be
used;
»
improper
adjustment,
calibration
or
operation
by
the
purchaser;
ㆍ
the
use
of
accessories
inchiding
consumables,
hardware
or
software
which
were
not
manufactured
or
approved
in
writing
by
BMDi
TUTA
Healthcare;
26
Warranty
it
CLEMENTS
10.
11.
12.
13.
e
any
modifications
of
the
product
which
were
not
authorised
in
writing
by
BMDi
TUTA
Healthcare;
*
any
contamination
or
leakages
caused
or
induced
by
the
purchaser;
and
國
*
inadequate
or
improper
maintenance
of
the
product,
This
warranty
does
not
cover
normal
wear
and
tear.
BMDi TUTA
Healthcare
will
not
be
responsible
for
damage
or
loss
caused
during
shipping.
x
In
Australia,
apart
from
any
warranties
implied
by
the
Trade
Practices
Act
1974
all
other
warrantfés
expressed
or
implied
and
whether
arising
by
virtue
of
statute
or
otherwise
are
hereby
excluded.
Outside
Australia,
all
other
warranties
expressed
or
implied
and
whether
arising
by
virtue
of
statute
or
otherwise
(including
any
warranties
implied
by
the
Vienna
Convention)
are
hereby
excluded.
BMDi
TUTA
Healthcare’
obligations
under
this
warranty
are
limited
to
the
repair
or
replacement
of
the
product,
within
the
terms
of
this
warranty
and
the
total
lability
of
BMDi TUTA
Healthcare
for
loss
or
damage
of
every
kind
whether
arising
pursuant
to
the
terms
of
the
sale
of
the
product
or
otherwise
in
connection
with
the
product
is
limited
to
the
amount
paid
by
the
purchaser
to
BMDi TUTA
Healthcare
for
the
product.
Apart
from
any
liability
imposed
by
Part
VA
of
the
Trade
Practices
Act,
BMDi TUTA
Healthcare
accepts
no
other
liability
for
any
loss
or
damage
occasioned
(including
consequential
loss
or
damages)
in
any
way
as
a
result
of
the
use
of
the
product.
The
warranty
does
not
extend
to
cover
damage
to
the
following
parts
as
they
are
inherentiy
prone
to
wear
:
+
motor
brushes
This
warranty
does
not
extend
to
cover
corrosion
due
to
any
cause
nor
to
any
damage
to
painted
or
anodised
surfaces.
BMDi TUTA
Healthcare
will
give
the
purchaser
the
benefit
of
any
manufacturer's
warranty
in
respect
of
any
components
in
the
product
which
were
not
manufactured
by
BMDi TUTA
Healthcare,
if
such
a
manufacturer's
warranty
is
available.
Warranty
27
sá

Other Clements Water Pump manuals

Clements SUC84602 User manual

Clements

Clements SUC84602 User manual

Clements SUC 81030 User manual

Clements

Clements SUC 81030 User manual

Clements SUC 82015 User manual

Clements

Clements SUC 82015 User manual

Popular Water Pump manuals by other brands

Einhell Royal RGP 1300 Niro operating instructions

Einhell Royal

Einhell Royal RGP 1300 Niro operating instructions

Packo HP Series instruction manual

Packo

Packo HP Series instruction manual

Xylem LOWARA ResiBoost MMW Installation, operation & maintenance manual

Xylem

Xylem LOWARA ResiBoost MMW Installation, operation & maintenance manual

GORMAN-RUPP 02K11-B Installation, operation, and maintenance manual with parts list

GORMAN-RUPP

GORMAN-RUPP 02K11-B Installation, operation, and maintenance manual with parts list

Sealey TP94 instructions

Sealey

Sealey TP94 instructions

Enerpac PUJ-1400B Service instructions

Enerpac

Enerpac PUJ-1400B Service instructions

VERDER VERDERAIR VA-E50 Operation

VERDER

VERDER VERDERAIR VA-E50 Operation

GORMAN-RUPP PUMPS PRIME-AIRE PA6B Series MAINTENANCE AND REPAIR WITH TROUBLESHOOTING

GORMAN-RUPP PUMPS

GORMAN-RUPP PUMPS PRIME-AIRE PA6B Series MAINTENANCE AND REPAIR WITH TROUBLESHOOTING

GORMAN-RUPP PUMPS 13D-L70W FT4 Installation, operation, and maintenance manual with parts list

GORMAN-RUPP PUMPS

GORMAN-RUPP PUMPS 13D-L70W FT4 Installation, operation, and maintenance manual with parts list

TotalPond 52325 quick start guide

TotalPond

TotalPond 52325 quick start guide

salmson MINI SUBSON Installation and operating instructions

salmson

salmson MINI SUBSON Installation and operating instructions

Rem Air System replacement pump instruction manual

Rem

Rem Air System replacement pump instruction manual

Silverline UK-917615 Original instructions

Silverline

Silverline UK-917615 Original instructions

Samoa 150 000 Parts and technical service guide

Samoa

Samoa 150 000 Parts and technical service guide

webtrol EZ Series Maintenance & Repair Manual

webtrol

webtrol EZ Series Maintenance & Repair Manual

HEISSNER P1100E-00 Instructions for use

HEISSNER

HEISSNER P1100E-00 Instructions for use

Dover HydroSystems Superdos 15 0.3 operating manual

Dover

Dover HydroSystems Superdos 15 0.3 operating manual

Wilo TOP-Z 20/4 EM Installation and operating instructions

Wilo

Wilo TOP-Z 20/4 EM Installation and operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.