Activity Display Screen/Visual Aid
4. Removethecapfromtheone-time-use21glancet,
place it against the skin, and push down rmly against
thengertolancethesurfaceofskin.Note:smaller
gauge lancets should not be used.
5. The blood should bleed freely. If it doesn’t, gently
squeezethengeroneachsideofthewoundto
get it started. Lower your hand below your heart if
necessary to produce a “bead” sized drop of blood.
6.
Hold sample transfer tube below bulb. DO NOT
TOUCHORSQUEEZEBULB.Withtubehorizontal
touchtiptobeadofblood.Letcapillaryactionllcap-
illaryportionoftubeuntilbloodowstops.Squeeze
ngertogenerateadditionalbloodifrequiredtocom-
pletelyllcapillaryportionofsampletransfertube.
7. Movengerstoatsidesofbulbbeingsuretocover
air hole. Rest hand on meter. Insert tip into sample
application well of test strip, touching tip down on
stripatashinggreenlight.Gentlysqueezebulbuntil
bloodleavestube(donotproduceairbubbles).Keep
pressureonbulbandpullawayfromteststrip(avoids
back suction). You must apply blood to the test
strip within 15 seconds of lancing the nger.
Themeterwillknowthatyouhaveaddedthesample
and the display will then read: “TESTING, PLEASE
WAIT”. Testing time is patient’s actual clotting time.
8. When testing is complete, the meter beeps once
and the results (INR and Prothrombin Time (PT) in
seconds) appear on the screen.
Recordtheresults.Removetheteststripandproperly
dispose along with the transfer tube and lancet. Turn
othemeter.
NOTE: ReadSelf-TestUser’sManualforcompleteinstructionsforuse,indications,warnings,precautionsand
troubleshooting tips. If any error messages are displayed, refer to the “Troubleshooting” section of the User
ManualorcallCoaguSenseTechnicalServiceat1-866-903-0890.