Comet System P8510 User manual

COMET SYSTEM
www.cometsystem.cz
P8510 intelligent thermometer
P8511 intelligent single-channel transducer
P8541 intelligent four-channel transducer
USER MANUAL

IE-SNC-P85x1-11
2
©Copyright: COMET System, Ltd.
Is prohibited to copy and make any changes in this manual, without explicit agreement of
company COMET System, Ltd. All rights reserved.
COMET System, Ltd. makes constant development and improvement of their products.
Manufacturer reserves the right to make technical changes to the device without previous
notice. Misprints reserved.
Manufacturer is not responsible for damages caused by using the device in conflict with
this manual. To damages caused by using the device in conflict with this manual can not be
provide free repairs during the warranty period.
This user manual describes device with firmware version 4-5-5.x. For devices with older
firmware version 4-5-1.x please read manual IE-SNC-P85x1-10.

IE-SNC-P85x1-11
3
Table of contents
Introduction........................................................................................................................................................4
General safety rules .....................................................................................................................................4
Device description and important notices...............................................................................................5
Getting started....................................................................................................................................................6
What is needed for operation ....................................................................................................................6
Mounting the device....................................................................................................................................6
Device settings .............................................................................................................................................8
Checking functions....................................................................................................................................10
Device setup .....................................................................................................................................................11
Setup using web interface.........................................................................................................................11
Setup using TSensor software .................................................................................................................18
Factory defaults..........................................................................................................................................18
Communication protocols .............................................................................................................................20
Website........................................................................................................................................................20
SMTP –sending e-mails...........................................................................................................................20
SNMP ..........................................................................................................................................................21
Modbus TCP ..............................................................................................................................................22
SOAP...........................................................................................................................................................23
Syslog...........................................................................................................................................................24
SNTP ...........................................................................................................................................................24
Troubleshooting...............................................................................................................................................25
I forgot the device IP address..................................................................................................................25
I can not connect to the device...............................................................................................................25
Error or n/a is displayed instead the measured value .........................................................................26
Error2 is displayed on all channels instead the measured value.........................................................26
I forgot the password for setup...............................................................................................................26
Factory defaults..........................................................................................................................................26
Technical specifications..................................................................................................................................27
Dimensions.................................................................................................................................................27
Basic parameters ........................................................................................................................................28
Operating terms .........................................................................................................................................30
End of operation .......................................................................................................................................30
Technical support and service .................................................................................................................30
Preventive maintenance............................................................................................................................30
Optional accessories........................................................................................................................................31

IE-SNC-P85x1-11
4
Introduction
This chapter provides basic information about device. Before starting please read
this manual carefully.
P8510 Intelligent Thermometer or P8511 and P8541 Transducers are used to measure
temperature or relative humidity. Temperature can be displayed in °C or °F. Relative
humidity have unit %RH. Communication with the device is realized via Ethernet network.
P8510 Thermometer has compact design and measures the temperature in place of
installation. To P8511 Transducer is designed to connect one probe. To P8541 Transducer
is possible connect up to fours probes. Temperature or humidity probes are available as
optional accessories.
General safety rules
The following summary is used to reduce the risk of injury or damage the device.
To prevent injury, please follow instructions in this manual.
The device can be services only by a qualified person. The device contains no
serviceable parts inside standard means.
Don’t use the device, if it doesn’t work correctly. If you think, that the device is
not working correctly, let check it by qualified service person.
It’s forbidden to use the device without the cover. Inside the device can be a dangerous
voltage and may be risk of electric shock.
Use only the appropriate power supply adapter according to manufacturer specifications
and approved according to relevant standards. Make sure, that the adapter does not have
damaged cables or covers.
Connect the device only to network parts approved according to relevant standards.
Connect and disconnect the device properly. Don’t connect or disconnect Ethernet cable
or probes, if the device is powered.
The device may be installed only in prescribed areas. Never expose the device to higher or
lower temperatures than is allowed. The device has not improved resistance to moisture.
Protect it from dripping or splashing water and do not use at areas with condensation.
Don’t use device in explosive environments.
Chapter
1

IE-SNC-P85x1-11
5
Don’t stress the device mechanically.
Device description and important notices
This chapter contains informations about basic features. Also there are
important notices concerning to functional safety.
Values from the device can be read using an Ethernet connection. The following formats
are supported:
Web pages with user changeable look and XML files
Modbus TCP protocol
SNMPv1 protocol
SOAP protocol
The device can also be used to check measured values and if the limit is exceeded, device
sends warning messages. Possible ways to sending warning messages:
Sending e-mails up to 3 e-mail addresses
Sending SNMP traps up to 3 configurable IP addresses
Displaying the alarm status on web page
Sending messages to Syslog server
The device setup can be made by the TSensor software or web interface. TSensor software
can be free downloaded from the manufacturer’s website. Also you will find there latest
firmware for your device. Do not upload to your device firmware which is not designed for
it. Unsupported firmware can damage your device.
Device does not support powering over Ethernet cable (PoE). PoE splitter must be used.
Compatible PoE splitter can be purchased as optional accessories. Splitter must have 5V
output with approximately 1W.
Reliability of warning messages delivering (e-mail, trap, syslog), depends on
actual availability of necessary network services. The device should not be used
for critical applications, where malfunction could cause to injury or loss of
human life. For highly reliable systems, redundancy is essential. For more
information please see standard IEC 61508.
Never connect the device directly to the Internet. If it is necessary connect the
device to the Internet, properly configured firewall must be used. Firewall can
be partially replaced with the NAT.

IE-SNC-P85x1-11
6
Getting started
Here you can find informations necessary to put newly purchased equipment to
operation. This procedure is only informative.
What is needed for operation
To install the unit you need to the following equipment. Before installation check if it’s
available.
P8510 Thermometer or P8511 Transducer, P8541 Transducer
power supply adapter 5V/250mA (or compatible PoE splitter)
RJ45 LAN connection with appropriate cable
free IP address in your network
for P8511 Transducers one probe. For P8541 Transducer up to 4
temperatures probes type DSTR162/C, DSTGL40/C, DSTG8/C or relative
humidity probe DSRH
Mounting the device
check if the equipments from previous chapter are available
install the latest version of TSensor software. This software is used to all
device settings.
TSensor software can be free downloaded from the manufacturer’s website.
Software can be also supplied on CD. Device configuration can be made
using web interface. For web configuration is not TSesnor software necessary.
contact your network administrator to obtain following informations for the
connection to the network:
IP address:
_____._____._____._____
Gateway:
_____._____._____._____
Netmask:
_____._____._____._____
Chapter
2

IE-SNC-P85x1-11
7
check if there is no IP address conflict when you connect the device into
network for the first time. The device has from factory set the IP address to
192.168.1.213. This address must be changed according to informations from
the previous point. When you installing several new devices, connect them to
the network one after another.
connect probes to P8511 or P8541 Transducer
connect the Ethernet connector
connect the power adapter 5V/250mA
LEDs on LAN connector should blink after connecting the power
P8510 Thermometer connection:
P8511 Transducer connection:

IE-SNC-P85x1-11
8
P8541 Transducer connection:
Connect via PoE splitter:
Device settings
run configuration software TSensor on your PC
switch to an Ethernet communication interface
press button Find device...
the window shows all available devices on your network

IE-SNC-P85x1-11
9
click to Change IP address to set new address according to network
administrator instructions. If your device is not listed, then click Help! My
device wasn’t found! Then follow the instructions. MAC address is on
product label. The device is factory set to IP 192.168.1.213.
gateway may not be entered if you want to use the device only in local
network. If you set the same IP address which is already used, the device will
not work correctly and there will be collisions on the network. If the device
detects a collision of IP address then reboot is performed automatically.
after changing IP address device is restarted and new IP address is assigned.
Restart of the device takes about 10 seconds.
connect to device using TSensor software and check the measured values. If
P8511 or P8541 Transducer values are not displayed, it’s necessary to find
probes using button Search probes (Find probes).
set the other parameters (alarm limits, SMTP server, etc.). Settings are saved
after click on button Save changes.

IE-SNC-P85x1-11
10
Checking functions
The last step is to check measured values on the device website. In the address bar of the
web browser, enter the device IP address. If the default IP address was not changed, then
insert http://192.168.1.213.
Displayed web page lists actual measured values. If the web pages are disabled, you can see
text Access denied. If the measured value exceeds the measurement range or probe is not
correctly installed, then is shown Error message. If the channel is switched off, the web site
displayed n/a instead of the value.

IE-SNC-P85x1-11
11
Device setup
This chapter describes basic device configuration. There is a description of
settings using web interface.
Setup using web interface
Device can be setup using web interface or TSensor software. Web interface can be
managed by the web browser. Main page will be shown when you insert device address
into address bar of your web browser. There you find actual measured values. Page with
history graphs is shown when you click to tile with actual values. Access to device setup is
possible via tile Settings.
Chapter
3

IE-SNC-P85x1-11
12
General
Device name can be changed using item Device name. Measured values are stored into
memory according History storage interval field. After changing of this interval all
history values will be cleared. Changes must be confirmed by Apply settings button.
Network
Network parameters can be obtain automatically from DHCP server using option Obtain
an IP address automatically. Static IP address is configurable via field IP address. It is
not necessary setup Default gateway while you use device inside one subnet only. Option
Standard subnet mask sets network mask automatically according A, B or C network
class. Subnet mask field must be set manually, when network with non standard range is
used. Periodic restart interval enables to restart device after selected time since device
start.

IE-SNC-P85x1-11
13
Alarm limits
For each measurement channel is possible to set upper and lower limits, time-delay for
alarm activation (0 –65535s) and hysteresis for alarm clearing.
Example of setting the limit to the upper alarm limit:
In Point 1 the temperature exceeded the limit. From this time, the time-delay is counting.
Because at point 2 the temperature dropped below the limit value before the time delay
expired, alarm was not set.
In Point 3 the temperature has risen over limit again. During the time-delay the value does
not drop below the set limit, and therefore was in Point 4 caused alarm. At this moment
were sent e-mails, traps and set alarm flag on website, SNMP and Modbus.
The alarm lasted up to Point 5, when the temperature dropped below the set hysteresis
(temperature limit –hysteresis). At this moment was active alarm cleared and e-mail send.
When alarm occurs, alarm messages will be sent. In case of power failure or device reset
(e.g. changing the configuration) will new alarm state evaluated and new alarm messages
will be send.

IE-SNC-P85x1-11
14
Channels
Channel can be enabled or disabled for measuring using item Enabled. Channel can be
renamed (max. 14 characters) and it is possible select unit of measured value according
connected probe type. When channel is not used, it is possible copy to it one of other
channels –option Clone channel. This option is not available at fully occupied device.
Find sensors button starts searching for connected probes. All changes must be
confirmed using Apply settings button. History values are cleared after changing of
channel settings.
SOAP protocol
SOAP protocol can be enabled by option SOAP protocol enabled. IP address of the
destination http server is configurable via SOAP server IP address filed. For setup of
destination server port can be used option SOAP server port. Target place of the http
server can be set via SOAP server target place. Device send SOAP message according
selected Sending interval. Option Send SOAP message when alarm occurs sends
message when an alarm on channel occurs or alarm is cleared. These SOAP messages are
sent asynchronously to selected interval.

IE-SNC-P85x1-11
15
Email
Email sending enabled option allows email features. It is necessary set address of the
SMTP as IP address into SMTP server IP address field. Default port of the SMTP server
can be changed using item SMTP server port. SMTP authentication can be enabled using
SMTP authentication option. When authentication is enabled Username and Password
must be set.
For successfully email sending it is necessary insert Email sender address. This address is
usually same as username of the SMTP authentication. Into fields Recipient 1 to
Recipient 3 it is possible set address of email recipients. Option Short email enable
sending emails in short format. This format is usable when you need to forward emails into
SMS messages.
When option Alarm email repeat sending interval is enabled and there is active alarm on
channel, then emails with actual values are sent repeatedly. Info email sending interval
option enables sending emails at selected time interval. CSV history file can be sent
together with the repeat/info emails. This feature can be enabled by Alarm and Info
emails attachment option.
It is possible to test email function using button Apply and test. This button save a new
settings and send a testing email immediately.

IE-SNC-P85x1-11
16
Modbus a Syslog protocols
ModbusTCP and Syslog protocol settings are configurable via menu Protocols. Modbus
server is enabled by default. Deactivation is possible via Modbus server enabled option.
Modbus port can be changed via Modbus port field. Syslog protocol can be enabled using
item Syslog enabled. Syslog messages are sent to IP address of the Syslog server - field
Syslog server IP address.
SNMP
For reading values via SNMP it is necessary to know password - SNMP read community.
SNMP Trap can be delivered up to three IP address - IP address of the Trap recipient.
SNMP Traps are sent at alarm or error state on the channel. Trap feature can be enabled by
option Trap enabled.

IE-SNC-P85x1-11
17
Time
Time synchronization with SNTP server can be enabled by Time synchronization
enabled option. IP address of the SNTP is necessary to set into SNTP server IP address
item. SNTP time is synchronized at UTC format, and due to be necessary set
corresponding time offset - GSM offset [min]. Time is synchronized every 24 hours by
default. Option NTP synchronization every hour decrease this synchronization interval
to one hour.
WWW and security
Security features can be enabled by the Security enabled option. When security is enabled
it is necessary to set administrator password. This password will be required for device
settings. When secured access is required even to actual values reading it is possible to
enable User account only for viewing. Port of the www server can be changed from the
default value 80 using filed WWW port. Web pages with actual values are refreshed
according to Web refresh interval field.

IE-SNC-P85x1-11
18
Setup using TSensor software
TSensor software is an alternative to web configuration. Some less important parameters
are configurable only by the TSensor software.
Parameter MTU size can reduce size of the Ethernet frame. Lowering of this size can
solve some communication problems mainly with Cisco network infrastructure. Sensor
software can set offset of values at temperature probes. At DSRH humidity probe is
possible set correction of the humidity and temperature.
Factory defaults
Factory defaults button set the device into factory configuration. Network parameters (IP
address, Subnet mask, Gateway) are left without changes.
Network parameters are changed while you close jumper inside the device. After jumper
closing it is necessary connect power supply. Factory defaults have no effect to user
correction inside probe.
Factory parameters settings:
Parameter
Value
IP address of SMTP server
0.0.0.0
SMTP server port
25
Alarm email repeat sending interval
off
Info email repeat sending interval
off
Repeat email attachment
off
E-mail recipients addresses
cleared
E-mail sender
sensor@IP
SMTP authentication
off
SMTP user/SMTP password
cleared
E-mail sending enabled
off
IP addresses SNMP traps recipients
0.0.0.0

IE-SNC-P85x1-11
19
Password for SNMP reading
public
Sending SNMP Trap
off
Website refresh interval [sec]
10
Website enabled
yes
Website port
80
Security
off
Administrator password
cleared
User password
cleared
Modbus TCP protocol port
502
Modbus TCP enabled
yes
History storage interval [sec]
60
SOAP server IP address
0.0.0.0
SOAP destination port
80
SOAP server target place
cleared
SOAP sending interval [sec]
60
SOAP protocol enabled
off
Syslog server IP address
0.0.0.0
Syslog protocol enabled
off
SNTP server IP address
217.31.205.226
GMT offset [min]
0
NTP synchronization every hour
off
SNTP synchronization enabled
off
MTU
1400
Periodic restart interval
off
Demo mode
off
Turn on all channels
yes
Upper limit
50
Lower limit
0
Hysteresis –hysteresis for alarm clearing
1
Delay –time-delay of alarm activation [sec]
30
Unit on the channel
°C or %RH according used probe
Channel name
Channel X (where X is 1 to 5)
Device name
PoE Ethernet Thermometer, etc.

IE-SNC-P85x1-11
20
Communication protocols
Short introduction to communication protocols of the device. To use some
communication protocols is necessary software, which can use the protocol. This
software is not included. Detailed description of protocols and application notes
can be found in the manual appendix.
Website
The device supports displaying of measured values, history graphs and configuration using
web browser. History graphs are based on HTML5 canvas. Web browser must support this
feature for proper function of graphs. Firefox, Opera, Chrome or Internet Explorer 9 can
be used. If the device has IP address 192.168.1.213 type into your browser
http://192.168.1.213. Using TSensor software or web interface can be set automatic
webpages refresh in interval. The default value is 10sec. Actual measured values can be
obtained using XML file values.xml.
Values from history can be exported in CSV format. History storage interval can be set
using TSensor software or web interface. History is erased after every reboot of the device.
Reboot of the device is performed when the power supply is disconnected and also after
configuration change.
The device allows you to customize the design of web pages. Details of how to change the
design, you can find in the manual appendix. Manual appendix can be obtained from the
manufacturer’s website.
SMTP –sending e-mails
When measured values are over the set limits, the device allows send e-mail to a maximum
of 3 addresses. E-mail is send when alarm condition on the channel is cleared or a
measuring error occurs. It is possible to set repeat interval for email sending. For correct
sending of e-mails it is necessary to set IP address of SMTP server. SMTP authentication is
supported but SSL not. Standard SMTP port 25 is used by default. SMTP port can be
changed. Contact your network administrator to obtain configuration parameters of your
SMTP server. E-mail sent by the device can not be answered.
Chapter
4
This manual suits for next models
2
Table of contents
Other Comet System Thermometer manuals