Contec Combi 08A pulse User manual

SOUTHEASTERN MEDICAL SUPPLY, INC
Advanced Blood Pressure and SpO2
All manuals and user guides at all-guides.com
all-guides.com

1
User Manual
Blood Pressure Monitor
Model CONTEC08A
■To assure the correct use of the product safety measures, please carefully read user manual
before using.
■ After reading, please validly keeping to refer and consult at any moment.
All manuals and user guides at all-guides.com

Contents
Safety Precautions: .........................................................................................................................1
Main Unit........................................................................................................................................3
Button Functions.............................................................................................................................4
External Interfaces..........................................................................................................................4
Dry Battery/AC Adapter Installation ..............................................................................................5
Dry Battery Installation...........................................................................................................5
Using The AC Adapter............................................................................................................5
Setting The Date And Time.............................................................................................................6
Unit.................................................................................................................................................7
User Switch.....................................................................................................................................7
Applying the Arm Cuff...................................................................................................................8
BP Measurement.............................................................................................................................9
Accurate Measurement Way...................................................................................................9
BP Measurement.....................................................................................................................9
Memory Function .........................................................................................................................10
Review The Memory Value ..................................................................................................10
Delete Memory Values.......................................................................................................... 11
Alarm Function.............................................................................................................................12
SPO2 Measurement Function(Separate Sale)...............................................................................12
Monitoring Procedure...................................................................................................................13
Maintenance and Cleaning............................................................................................................14
Installation Of The Software.........................................................................................................15
Demand Of Editor..................................................................Error! Bookmark not defined.
Installation Of Software........................................................................................................15
Error Message...............................................................................................................................16
TROUBLESHOOTING................................................................................................................17
Clean And Maintenance................................................................................................................18
Keeping.........................................................................................................................................18
Specification .................................................................................................................................19
All manuals and user guides at all-guides.com

1
Safety Precautions:
Before use, carefully read "Safety Precautions" for a correct use.
To prevent users suffered hurt or damnification due to improper use, see "Safety Precautions", and use
this product properly.
For safety reasons, be sure to comply with safety precautions.
Note
If not to use correctly, it exists that a potentially hazardous situation which may result in injury to the user
or patient or damage to the equipment or other property.
Note
Self-diagnosis and treatment using measured results may be dangerous.
Follow the instructions of your physician.
Contact your physician for specific information about your blood pressure.
Please hand measurement results to the doctor who know your health to accept diagnosis.
For severe blood circulation disorder or arrhythmia patients, please use the device under the guidance of a
doctor.
Otherwise it may lead to acute hemorrhage, or measurement error as a result of squeezed arm.
This device is intended for use in measuring blood pressure and pulse rate.
Do not use for any other purpose.
Otherwise it may cause accident or holdback.
Please use special cuff.
Otherwise it is possible that measurement result is incorrect.
Do not disassemble or attempt to repair the unit or components without permission.
Otherwise it can not measure correctly.
All manuals and user guides at all-guides.com

2
Operation for AC Adapter (Separate Sale)
Note
Please use sold separately dedicated AC adapter.
Otherwise it may cause trouble.
Sold separately dedicated AC adapter be sure to use the AC 220V. Be sure to use a separate socket.
Otherwise it may cause electric shock or injury.
When there is breakage of sold separately dedicated AC adapter plug or wire ,
please immediately pulled the plug from the socket.
Otherwise it may cause electric shock or injury.
Do not plug or unplug the adapter power cord with wet hands.
Otherwise it may cause electric shock or injury.
Operation for battery
Note
Please use 4 "AA" size manganese or alkaline batteries, do not use batteries of other types.
Otherwise it may cause fire.
New and old batteries, different kinds batteries can not be confusion.
Otherwise it may cause battery leakage, heat, rupture, and damage to blood pressure.
+ and - polarities of the batteries must match the polarities of the battery compartment as indicated.
When the batteries power exhausts, replace with four new batteries at the same time.
Please take out the batteries when you do not use the device for a long time(more than 3 months).
Otherwise it may cause battery leakage, heat, rupture, and damage to blood pressure.
If battery fluid should get in your eyes, immediately rinse with plenty of clean water.
Contact a physician immediately.
Otherwise it will cause blindness or other hazards.
If electrolyte of the batteries immodestly glues on the skin or the clothes,
immediately rinse with plenty of clean water.
Otherwise it may hurt the skin.
Advice
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Do not inflate before the cuff wrapps around the arm
Do not inflect the cuff and the air tube forcibly
This device is intended for using in measuring blood pressure and pulse rate in the adult、pediatric and
neonatal population.
All manuals and user guides at all-guides.com

3
Main Unit
The production is in the package . Open the package and confirm whether the production is whole.
Adult Cuff Dry Battery
Accessories:
(Specification: limb circumference 22-32cm (middle part of upper arm), please choose suited cuff when
measuring pediatric or other).)
USB Data Line Software CD User Manual
Separate Sale:
AC Adapter CMS0106
InputAC 100-240V 50/60Hz AC 150mA
Output DC 6.0V±0.2V 1.0A
Spo2 Probe Spo2-10UC15
Spo2 Measurement Range 35%~ 100 %
Measurement Accuracy 70% ~ 100% 2 %
Pulse Measurement Range 30 ~ 250bpm
UP/DOWN Button
MEMORY
AC Adapter Jack
MENU Button
USB Plug Jack
Cuff Plug
Cuff Air
USER SWITCH
ON/OFF Button
START/STOP Button
Display
All manuals and user guides at all-guides.com
all-guides.com

4
Button Functions
All the operations to the Electronic Blood Pressure Monitor are through the buttons. The names of the
buttons are above them. They are:
【ON/OFF】Hold the button to start or close the device.
【START/STOP】Press to inflate the cuff and start a blood pressure measurement. When
measuring, press to cancel the measurement and deflate the cuff.
The three buttons correspond with the hint in the LCD screen downside, pressing any
button will carry on corresponding function, eg:【MENU】【 ENTER】【 LIST】etc.
Up and down buttons respectively carry on the functions of moving the cursor up and down,
changing the parameters and switching the status.
External Interfaces
Note
Please hold the air plug to remove the NIBP cuff.
①Cuff Socket
left side
The right side of the instrument is USB socket and AC adapter socket.
①USB socket
②AC adapter socket right side
All manuals and user guides at all-guides.com

5
Dry Battery/AC Adapter Installation
The production can use dry battery or AC adapter as power source.
Dry Battery Installation
①Press the▲indicator on the battery cover and slide the cover off in the direction of the arrow.
②Install 4 "AA" size dry batteries so the +(positive) and 一(negative) polarities match the polarities of
the battery compartment as indicated.
③Replace the battery cover.
Icon “ ”:the batteries power will exhaust.
Replace with four new batteries(the same sort) at the same time.
Turn the unit off before replacing the batteries.
Note
Dispose of the batteries according to applicable local regulations about environmental.
Using The AC Adapter
①Connect Monitor and the AC Adapter. Insert the AC Adapter Plug into the AC Adapter Jack on the
right side of the monitor.
②Plug the AC Adapter into a 220VAC outlet.
Note
①Hold and pull the Housing to remove the AC Adapter from the electrical outlet. Do not remove by
pulling on the cord.
②R em ove th e A C A d a p ter p l u g f r om th e m on i tor.
Please be sure to use dedicated AC adapter (Model:CMS0106).
Note
When adapter and dry battery are both used at the same time ,the battery power will not be
consumed. Switch adapter and battery as power supply when the device is off, otherwise the device
may shutdown due to power failure.
All manuals and user guides at all-guides.com

6
Setting The Date And Time
It is necessary to set date and time after turning on the device .
The Blood Pressure Monitor automatically stores measurement results with date and time.
If dry battery power exhausts or removed ,then, after the device turned on,the date resumes from the
last setting value and the time from 00:00:00,set the date and time again.
The Blood Pressure monitor stores the measure results of three users automatically, and up to 100 items for
every user. The results can be uploaded to PC via USB and processed with the PC software. If the date and
time are set correctly, the date and time when measuring will be correct in the memory ,otherwise it may
not be correct.
Note
Correctly use data upload function:
1. Turn on the device to enter the main interface shown as the follow:
2.Press 【MENU】button to enter 【SYSTEM MENU】and select【SYSTEM TIME】item in system
menu. The current time will be displayed:
3.Press【UP】or 【DOWN】buttons to set date and time.
4.After setting, select 【CONFIRM】item and press【ENTER】button to confirm the setting value. If you
do not want to change the time, select 【EXIT】item and press【ENTER】button to return the previous
menu.
All manuals and user guides at all-guides.com

7
Unit
There are two units: "mmHg" and "kPa".
The default is: "mmHg".
To switch "mmHg" and "kPa" units, enter the 【SYSTEM SETUP】submenu in 【SYSTEM MENU】,
and complete switching in【UNIT】item.
User Switch
The Blood Pressure monitor stores the measure results of three users automatically, and up to 100 items for
every user.
Userscan press 【USER】button in main interface to switch user. Or press 【USER PURVIEW】item in
【SYSTEM SETUP】menu to switch user.
Note
When the 【USER PURVIEW】is set to be 【ALL】, current user can be switched under main
interface; when set to a certain user, it will not be able to switch.
All manuals and user guides at all-guides.com

8
Applying the Arm Cuff
The measurement can be carried out by applying the cuff on left or right arm.
Remove tight-fitting clothing from your upper arms.
Carry out the operation in a room with comfortable temperature.
When measuring, take the thick clothes off instead of rolling up the sleeves.
In order to measure accurately,pay attention to applying the cuff properly (left arm).
①Make sure the air plug is securely inserted in the main unit.
②Stretch cuff into a barrel for the arm can conformable enter into the barrel.
③Arm penetrate throught the cuff, the air tube of the cuff will pass the top of your palm.
④Apply the cuff to your upper arm. The color marker is on the inside center of your arm and make the
air tube aligned with your middle forger.
⑤The bottom of the cuff should be approximately 2cm~3cm above your elbow.
⑥Be fixed with cloths, and wrapped tight cuff, the arm and the cuff should not have gaps.
All manuals and user guides at all-guides.com
all-guides.com

9
BP Measurement
Accurate Measurement Way
Measurement in quiet and relaxing state.
1.Place your arm on a table.
2.The cuff is level with your heart.
3.The palm of the hand is up, and the body relax.
Try to measure your blood pressure at the same time each day with the same arm
and the same pose for consistency.
The high and low location of cuff will cause changes in measurement results.
Do not touch BP monitor, cuff and windpipe during measurement.
Measurements should be taken in a quiet place and the body is in relaxing state.
Remain still 4~5 minutes before measurement.
Relax the body, do not let the muscle activity.
Do not talk and movement during the measurement.
Wait 4~5 minutes between measurements
Do not use a cellular phone near the device
BPMeasurement
Press 【START/STOP】button to take a measurement.
Start measurement measurement result
During measurement, please keep correct pose and quiet state, do not move.
Stop Measurement
During measurement , if you wish to stop measuring , press 【START/STOP】button to stop and deflate.
Confirm Measurement Value
Measurement value can be automatically stored.([using memory function] refer to 10 page)
*Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your
physician.
All manuals and user guides at all-guides.com

10
Note
Wait 4-5 minutes between measurements.
When repeated measurements, because the arm appears congestion, it may not get correct blood pressure
measurement. After the blood flow, take a measurement once again.
When some factors affect the measurement results in measurement process, error messages hints will
appear on the screen, you can obviate the malfunction and restart a measurement.
In no physiological alarm state, press any button to carry on the corresponding button function; in audio
alarm state, press any button (except 【ON/OFF】button) to clear up the audio alarm.
④Tak e off the cuff, hol d 【ON/OFF】button to turn the monitor off.
*The monitor will automatically turn off after two minutes in which there is no operation to the device,
even if you forget to turn the power off.
Memory Function
The monitor is designed to store and display the blood pressure, pulse rate values and the date and time
when measured, which are up to 100 items. If there have been 100 items stored, when the 101
measurement have been taken, the earliest results are deleted.
Review The Memory Value
1.In the main interface (interface when boot-strap), press 【MEMORY】button to review the most recent
measurement values in large-print with the serial number from 1 to 100.
2.Press 【UP】/【DOWN】button to circularly switch the former measurement values.
*The right figure shows that there is no Measurement results.
All manuals and user guides at all-guides.com

11
3.Press 【LIST】button to switch to data table interface.
4.Press 【TREND】button to display trend interface.
5.Finish displaying the measurement values
Press 【EXIT】to return the main interface or hold 【ON/OFF】button to turn the power off.
*The monitor will automatically turn off after two minutes in which there is no operation to the device,
even if you forget to turn the power off.
Delete Memory Values
Users can delete all values stored in the memory of each user separately instead of deleting a special item.
1.Press 【MENU】button to enter 【SYSTEM MENU】, select 【DELETE DATA】item and enter its
interface, in which select the user whose data will be deleted. All measurement results of selected user will
be deleted after confirm
正确
2、Finish Operation
Select 【CONFIRM】or 【EXIT】to return the previous menu, or hold 【ON/OFF】button to turn the
power off.
*The monitor will automatically turn off after two minutes in which there is no operation to the device,
even if you forget to turn the power off.
All manuals and user guides at all-guides.com

12
Alarm Function
Alarms are classified into two categories: technical alarm and physiological alarm.
PhysiologicalAlarm
User can press 【MENU】button to enter system menu, select 【ALARM SETUP】item to enter its
interface, and then set alarm on-off and the high and low alarm limits, when blood pressure is higher than
the high limit or lower than the low limit, the physiological alarm will occur.
Press any button to cancel the alarm in physiological alarm state and this method brings no affection to the
next alarm. To close alarm in【ALARM SETUP】will disable the alarm function until switch alarm on.
TechnicalAlarm
When power is about to exhaust and alarm is on , then the alarm will occur. This alarm can not be
cancelled unless being closed or the power replaced.
SPO2 Measurement Function(Optional)
Please pay attention to:
Warning
◎Pulse oximeter can overestimate the SPO2 value in the presence of Hb-CO, Met-Hb or
dye dilution chemicals.
◎ES (Electrosurgery) equipment wire and SpO2 cable must not be tangled up.
◎Do not put the sensor on extremities with arterial catheter or venous syringe.
◎Do not perform SPO2 measuring and NIBP measuring on the same arm at one time, because
obstruction of blood flow during NIBP measuring may adversely affect the reading of
SPO2 value.
Note
◎Make sure the nail covers the light window.
◎the wire should be on the backside of the hand.
◎SPO2 value is always displayed in the fixed place.
All manuals and user guides at all-guides.com

13
Warning
◎Check if the sensor cable is in normal condition before monitoring. After unplugging the SpO2
probe cable from the socket, the interface will return.
◎Do not use the SpO2 probe once the package or the sensor is found damaged. Instead, you shall
return it to the vendor.
◎Prolonged and continuous monitoring may increase jeopardy of unexpected change of dermal condition
such as abnormal sensitivity, erubescence, vesicle, repressive putrescence, Particularly in newborns or in a
Perfusion disorders and changes or immature skin form of the patient. According to skin quality change,
correct optical path alignment and attachment methods to regularly check the place of SPO2 probe, and
change the attachment position when the quality of skin decline. More frequent examinations may be
required for different patients.
Monitoring Procedure
1. Attach the sensor to the appropriate site of the patient finger as following figure.
place SPO2
2.Plug the connector of the SPO2 probe cable into the USB socket in the lower right of the monitor. The
main interface will switch to SPO2 interface. This operation brings no affection to other functions.
Measurement Limitations
During operation, the accuracy of oximeter readings can be affected by:
High-frequency electrical noise, including noise created by the host system, or noise from external
sources, such as electrosurgical apparatus connected to the system.
Intravascular dye injections.
Excessive patient movement.
External light radiation.
Improper sensor installation or incorrect contact position of the patient.
SPO2 probe temperature (optimal temperature between 28°C and 40°C).
Placement of the SPO2 probe on an extremity that has a blood pressure cuff, arterial catheter, or
intravascular line.
All manuals and user guides at all-guides.com
all-guides.com

14
Significant concentrations of dysfunctional hemoglobin, such as carboxyhemoglobin and
methemoglobin.
SpO2 too low.
Bad circular injection of the part being measured.
It is required to use SPO2 probe which is provided by our company, contact with our sale department
when changes SPO2 probe.
Maintenance and Cleaning
Warning
Take cuff off the power source before cleaning the monitor or the SPO2 probe.
Caution
Do not subject the sensor to autoclaving.
Do not immerse the SPO2 probe into any liquid.
Do not use any SPO2 probe or cable that may be damaged or deteriorated.
Cleaning:
Use a cotton ball or a soft mull moistened with hospital-grade ethanol to wipe the surface of the SPO2
probe, and then dry it with a cloth. This cleaning method can also be applied to the luminotron and
receiving unit.
The cable can be cleaned with 3% hydrogen dioxide, 70% isopropanol, or other active reagent.
However, connector of the SPO2 probe shall not be subjected to such solution.
All manuals and user guides at all-guides.com

15
Installation Of The Software
Computer requirements
Pentium IV 1.8G or more
Operation System:Windows XP Pro or newer
EMS memory: 256M or more
Hard Disk: 40G or more
Display: 17 inch or more
CD-ROM
USB: 2.0 or newer
Resolution of printer: 600 DPI
Installation Of Software
1.Place the CD-ROM in the CD-ROM compartment located on your computer.
2.If Auto Play for CDs is enabled, place CD in reader and follow instructions when they appear in the
screen; otherwise follow install instructions below:
1.Open Windows Explorer.
2.Click on the root CD-ROM directory.
3.Double click file Contec08A_Setup.EXE.
4.Follow the instructions in the screen.
Refer to "Software Help"for details about the operation method of the PC software.
All manuals and user guides at all-guides.com

16
Error Message
Error message will be displayed in the screen if there is something wrong when measuring. The causes and
solutions are shown as follows:
Error Message Causes Solutions
Self-test failure
System failure Function abnormal Please contact us
Loose cuff Cuff is not connected
correctly. Correctly connect cuff (refer to 8 page)
Air leakage Cuff plug fall off Make sure the cuff plug is securely
inserted in the windpipe (refer to 8 page)
Air pressure error Air pressure error Keep arm, body still, measure again
Weak signal The pulse signal is too weak
or the cuff is loose. Correctly connect cuff (refer to 8 page)
overpressure Cuff is blocked or squeezed Correctly connect cuff (refer to 8 page)
Excessive movement
Over range
Saturated signal
The signal extent is too big
owing to the arm or body
moving or other reasons when
measuring Keep arm, body still, measure again
Time out It takes too much time
All manuals and user guides at all-guides.com

17
TROUBLESHOOTING
Abnormal
Phenomenons
Causes Solutions
BP measurement
values too high or
too low.
Cuff is not connected correctly. Correctly connect cuff (refer to 8
page)
Talk or move arms when
measuring
Keep quiet and restart a
measurement
The turnup clothing presses the
arm
Take off the clothing which presses
the arm, and restart a measurement
No pressure
Cuff leakage Buy a new cuff
The cuff windpipe is not correctly
connected with cuff Correctly connect
Cuff deflates in
short time Loose cuff Correctly apply cuff
It can not carry on measurement when press the
measurement button
Switch on the power once again
and restart a measurement
Power off suddenly
when inflating
No use for a long time, the power
of batteries can be exhausted
owing to the changed temperature
Replace all four batteries with new
ones.
Hold the on/off
button but can not
start the device
Power of batteries can be
exhausted
Replace all four batteries with new
ones.
The battery polarities is reversed
Check the battery installation for
proper placement of the battery
polarities.
Other phenomenon
Switch on the power once again
and restart an operation.
Replace the batteries.
If no, please contact us.
All manuals and user guides at all-guides.com
Other manuals for Combi 08A pulse
1
Table of contents
Other Contec Blood Pressure Monitor manuals