INTENDED USE 
The CONTOUR®NEXT EZ blood glucose monitoring system 
is an over the counter (OTC) device utilized for self-testing
by persons with diabetes at home for the quantitative
measurement of glucose in whole blood, is for single-patient use 
only, and should not be shared. The system is intended for 
self-testing outside the body (in vitro diagnostic use) and is to be 
used as an aid to monitor the eff ectiveness of diabetes control. 
The CONTOUR NEXT EZ blood glucose monitoring system is 
indicated for use with fresh fi ngertip capillary whole blood 
samples. The clinical utility of this device is to aid in monitoring 
the eff ectiveness of your diabetes control program. 
The CONTOUR NEXT EZ blood glucose monitoring system is 
not intended for the diagnosis of or screening for diabetes 
mellitus and is not intended for use on neonates.
CONTOUR®NEXT test strips are intended for self-testing by 
persons with diabetes for the quantitative measurement of 
glucose in whole blood samples. 
The CONTOUR®NEXT control solutions are aqueous glucose 
solutions intended for use in self-testing by people with diabetes 
as a quality control check to verify that the CONTOUR NEXT EZ 
system is working properly.
IMPORTANT SAFETY INFORMATION 
• Read your CONTOUR NEXT EZ meter user guide, 
the lancing device insert and all instructional materials 
provided in your meter kit before testing. Please follow all 
instructions for use and care exactly as described to help 
avoid inaccurate results.
• Used test strips and lancets are possible biohazards and 
must be disposed of as medical waste. 
Cleaning Your Meter
Supplies Needed for Cleaning: 
• Clorox Germicidal Wipes containing 0.55% sodium 
hypochlorite (bleach)
• Paper towels 
1. Clean the meter with Clorox Germicidal Wipes until all soil is 
removed. Do not allow cleaning solution to run into the meter 
through areas such as around the buttons or the meter’s test 
strip or data ports.
2. Dry as necessary with a clean paper towel. 
Disinfecting Your Meter
Supplies Needed for Disinfecting: 
• Clorox Germicidal Wipes containing 0.55% sodium 
hypochlorite (bleach)
• Paper towels 
• Timing device
1. Before disinfecting, clean the meter as described above in 
Cleaning Your Meter. 
For proper disinfection, you must keep all meter surfaces wet 
for 60 seconds.
Signs of Deterioration 
The cleaning and disinfecting directions provided should 
not cause any damage or degradation to the external case, 
buttons or display. 
Your CONTOUR NEXT EZ meter has been tested for 260 cycles 
of cleaning and disinfection (one cycle per week for 5 years). 
This device has been demonstrated to withstand 5 years of 
cleaning and disinfection without damage.
You should contact Customer Service for assistance if your 
device malfunctions for any cause or if you notice any changes 
in the meter case or display.
Clorox Germicidal Wipes are available for purchase online 
at http://www.offi  cedepot.com or http://www.amazon.com, 
or for more information contact Customer Service at 
1-800-348-8100 (Monday through Sunday, from 8:00 am 
through 12:00 Midnight, Eastern Time).
For more information see: 
“FDA Public Health Notifi cation: Use of Fingerstick Devices on 
More than One Person Poses Risk for Transmitting Bloodborne 
Pathogens: Initial Communication” (2010) 
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/
ucm224025.html
“CDC Clinical Reminder: Use of Fingerstick Devices on More 
than One Person Poses Risk for Transmitting Bloodborne 
Pathogens” (2010) http://www.cdc.gov/injectionsafety/
Fingerstick-DevicesBGM.html 
Using a new Clorox Germicidal Wipe, carefully wipe all outer 
surfaces of your meter until wet. 
Do not allow cleaning solution to run into the meter through areas 
such as around the buttons or the meter’s test strip or data ports.
Test Strip Port
Buttons
2. Dry all meter surfaces and test strip port using a clean paper 
towel if needed.
Cleaning and Disinfecting Your Lancing Device 
Refer to the lancing device insert provided with your kit 
for detailed instructions for cleaning and disinfecting the 
lancing device. 
Your Meter Display 
The picture below shows all the symbols that appear on your 
meter display. To view, with the meter off , press and hold  or 
. All display segments will appear for 10 seconds. If you need 
more than 10 seconds to check the display, press and hold 
or again.
Icon What It Means 
Day/month format. 
The batteries are low and need to be changed, and 
will remain on display until batteries are changed. 
 Appear if the meter is in 12 hour setup. 
Indicates sound is turned on. 
Thank you for choosing the CONTOUR NEXT EZ blood 
glucose monitoring system! We are proud to be your 
partner in helping you manage your diabetes.
Your Meter Features
Easy: The CONTOUR NEXT EZ blood glucose monitoring system 
is about simplicity. You will notice its simple approach the fi rst 
time you see it. 
Automatic: Your CONTOUR NEXT EZ meter is designed with 
No Coding technology, to automatically code itself every time 
you insert a CONTOUR NEXT test strip. It will even tell you when a 
test strip is underfi lled. 
Programmed: Your CONTOUR NEXT EZ meter has been 
preset at the factory with the standard date and time for your 
convenience.
• Your meter has been preset and locked to display results in 
mg/dL (milligrams of glucose per deciliter). In some countries, 
the meters are preset to display units in mmol/L (millimoles of 
glucose per liter);
• Results in mg/dL will never have a decimal point;
• Results in mmol/L will always have a decimal point.
Example: or 
Check your display to ensure that results are being displayed in 
mg/dL. If they are not, contact Customer Service. See Contact 
Information.
Customizable: The CONTOUR NEXT EZ meter allows you to set 
personal choices for some features. For more information, see 
section 3, Setup and Use.
Important Safety Information
• Always keep the test strips in the 
original bottle. Tightly close the bottle 
immediately after removing a test 
strip. The bottle is designed to keep 
the test strips dry. Exposure to room 
humidity from leaving the bottle open or 
not storing them in the original bottle can 
damage your test strips. This could lead 
to inaccurate test results.
• Always wash your hands well with soap and water and 
dry them well before and after testing, handling the 
meter, lancing device, or test strips.
• Test strips are for single use only. Do not use a test 
strip that appears damaged or has been used.
• Your test strip is designed to easily “sip” the blood 
into the sample tip. Do not drop blood directly on the fl at 
surface of the test strip.
• Do not press the test strip against your fi nger when 
testing. This may block the sample tip.
• Let the meter and test strips adjust to the temperature 
where you are testing. Whenever the meter is moved 
from one location to another, allow approximately 
20 minutes for the meter to adjust to the temperature of 
a new location before performing a blood glucose test. 
The meter has been designed to give accurate results at 
temperatures between 41°F and 113°F.
• Check the expiration dates on your test strips and 
control solution. It is important to not use the test strips or 
control solution if the expiration date printed on the bottle 
label and carton has passed. For the control solution do 
not use if it has been six months since you fi rst opened the 
bottle. It will help to write the six month discard date on the 
control solution label.
• For more information about cleaning and disinfecting your 
meter or lancing device see:
“FDA Public Health Notifi cation: Use of Fingerstick Devices 
on More than One Person Poses Risk for Transmitting 
Bloodborne Pathogens: Initial Communication” (2010) 
http://www.fda.gov/MedicalDevices/Safety/
AlertsandNotices/ucm224025.html
“CDC Clinical Reminder: Use of Fingerstick Devices 
on More than One Person Poses Risk for Transmitting 
Bloodborne Pathogens” (2010) http://www.cdc.gov/
injectionsafety/Fingerstick-DevicesBGM.html
 TESTING 
Important Safety Information 
• Have all the materials you will need ready before you 
begin testing. This includes your CONTOUR NEXT EZ meter, 
the CONTOUR NEXT test strips, the lancing device and lancets. 
You will also need CONTOUR NEXT control solution to run a 
quality control check. CONTOUR NEXT control solutions are 
sold separately if not included in the meter kit.
• The lancing device is intended for self testing by a single 
patient. It must not be used on more than one person 
due to the risk of infection. 
• Use a new lancet each time you test because it is no 
longer sterile after use.
• Do not perform a blood glucose test when the 
CONTOUR NEXT EZ
 meter is connected to an external device, 
e.g., computer.
• Use only approved equipment from the manufacturer or 
certifi ed body such as UL or TUV. 
• The CONTOUR NEXT EZ blood glucose monitoring system has 
not been evaluated using capillary blood samples obtained 
from sites other than the fi ngertip.
• Use only CONTOUR NEXT test strips with your 
CONTOUR NEXT EZ system.
• Use only CONTOUR NEXT control solution with your 
CONTOUR NEXT EZ system.
The following steps are shown in the proper order for 
performing a blood glucose test.
Preparing the Test Strip 
1. Wash and dry your hands well before testing. 
2. Remove a CONTOUR NEXT test strip from the bottle. 
3. Hold the test strip with the gray 
square end facing up. 
4. Insert the gray square end into the 
test strip port on the meter. 
 The meter will turn on. An image of a test strip with a fl ashing 
blood drop will appear letting you know the meter is 
ready to test. 
Preparing the Lancing Device
The lancing device graphics are for illustration purposes 
only. Your lancing device may look diff erent. Refer to 
your lancing device insert for detailed instructions on 
preparing the lancing device. 
WARNING: Potential Biohazard 
The lancing device provided with your kit is intended 
for self-testing by a single patient. It must not be used 
on more than one person due to the risk of infection.
WARNING
• Dispose of the used lancet as medical waste or as 
advised by your health care professional. 
• Do not reuse lancets. Use a new lancet each time 
you test.
Getting Ready to Test
1. Remove the endcap from the lancing device.
2. Loosen the round protective cap on 
a lancet by rotating the cap ¼ turn, 
but do not remove the cap.
3.  Insert the lancet fi rmly into the 
lancing device until the lancet 
comes to a complete stop. 
4. Twist off  the round protective lancet 
cap. Save the cap for disposing of 
the used lancet. 
5. Replace the endcap. 
6. Rotate the endcap dial to adjust 
the puncture depth. The amount of 
pressure applied to the puncture 
site also aff ects the puncture depth. 
Getting the Blood Drop and Testing 
1.  Press the lancing device fi rmly 
against the puncture site and press 
the release button. 
2.  Stroke your hand and fi nger towards 
the puncture site to form a drop of 
blood. Do not squeeze around the 
puncture site.
Suggested drop size
3. Test immediately after a good blood 
drop has formed.
Fingertip Testing
4. Immediately touch the tip of the test 
strip to the drop of blood. The blood 
is pulled into the test strip through 
the tip.
Hold the tip of the test strip in the 
blood drop until the meter beeps.
Do not press the tip against the skin or place the blood on top of 
the test strip or you could get inaccurate results or errors. 
WARNING
• The meter and lancing device are for single patient 
use. Do not share them with anyone including other 
family members! Do not use on multiple patients!1,2
• All parts of the kit are considered biohazardous and 
can potentially transmit infectious diseases, even 
after you have performed cleaning and disinfection.
• Always wash your hands well with soap and water 
and dry them well before and after testing, handling 
the meter, lancing device, or test strips. 
For complete instructions on cleaning and disinfecting 
your meter and lancing device, see Cleaning and 
Disinfection in section 2, Testing. 
If the fi rst blood drop is not enough, the meter may beep twice 
and display a strip underfi ll screen image  .
You have 20 seconds to apply more blood to the same strip. If 
you are not able to apply enough blood within that time, an E2 
error code will be displayed. Dispose of the used test strip as 
medical waste and start over with a new strip.
5. After a beep, you will see the meter 
count down 5 seconds and your blood 
test result is displayed and automatically 
stored in the meter memory. Please 
do not touch the test strip during the 
countdown as this may result in an error. 
NOTE: You can set your meter to use meal marker options. 
See Set-up for Basic (L-1) and Advanced (L-2) Modes in 
section 3, Setup and Use. If you want to mark your result 
you will need to do it before removing the test strip.
6. Remove the test strip to turn your meter 
off . Dispose of the used test strip as 
medical waste.
7. Always wash your hands well with soap 
and water before and after testing, 
handling the meter, lancing device, or 
test strips. 
Test Results
Expected Values
Blood glucose values will vary depending on food intake, 
medication dosages, health, stress, or activity. Non-diabetic 
plasma glucose concentrations should be < 100 mg/dL in the 
fasting state and < 140 mg/dL in the post-prandial state.3You 
should consult with your health care professional for 
glucose values specifi c to your needs. 
CAUTION:
• If your blood glucose reading is under 50 mg/dL, or you 
see (less than 20 mg/dL) on the meter display, follow 
medical advice immediately.
• If your test result is above 250 mg/dL, or you see 
 (greater than 600 mg/dL) on the meter display, call 
your health care professional as soon as possible.
• Always consult your health care professional before 
changing your medication based on CONTOUR NEXT EZ 
test results.
Ejecting and Disposing of the Used Lancet 
1. Do not use your fi ngers to remove 
the lancet from the lancing device.
The lancing device provided with 
your kit has an automatic lancet 
ejection feature.
2. Refer to the separate lancing device insert, if provided 
with your kit, for instructions on automatic ejection of 
the lancet. 
Do not reuse lancets. Use a new lancet each time you test.
WARNING
• Dispose of the used lancet as medical waste or as 
advised by your health care professional.
• Do not reuse lancets. Use a new lancet each time you 
test.
WARNING: Potential Biohazard 
• Always dispose of test strip and lancet as medical 
waste or as advised by your health care professional.
• All products that come in contact with human blood 
should be handled as if capable of transmitting 
infectious diseases.
• Please refer to your lancing device package insert for 
instructions on how to properly remove and dispose 
of the lancet.
Cleaning and Disinfection
Your CONTOUR NEXT EZ meter should be cleaned and 
disinfected once a week. For the meter, use only Clorox 
Germicidal Wipes containing 0.55% sodium hypochlorite 
(bleach), which has been proven to be safe to use with the 
CONTOUR NEXT EZ meter. Refer to the lancing device 
insert provided with your kit for detailed instructions for 
cleaning and disinfecting the lancing device.
Cleaning is the removal of visible dirt and debris, but does not 
reduce the risk for transmission of infectious diseases. Your 
CONTOUR NEXT EZ system should be cleaned of dirt and debris 
once a week.
Disinfecting (if performed properly) reduces the risk of 
transmitting infectious diseases. Your meter and lancing device 
should be disinfected once a week.
NOTE: If the meter is being operated by a second person 
who is providing testing assistance to you, the meter and 
lancing device should be disinfected prior to use by the 
second person. 
Control Solution Testing
CAUTION: Use only CONTOUR NEXT control 
solution with your CONTOUR NEXT test strips. 
Using any other control solution may cause 
inaccurate results. 
Quality Control
• Shake the control solution bottle well, about 
15 times before every use.
• Unmixed control solution may cause 
inaccurate results.
You should perform a control test when:
• Using your meter for the fi rst time.
• You open a new bottle or package of test strips.
• You think your meter may not be working properly.
• You have repeated, unexpected blood glucose results. 
WARNING
Shake the control solution well before testing. 
WARNING
• Do not calibrate your continuous glucose monitoring 
device from the control result.
• Do not calculate a bolus based on a control result. 
CAUTION: Check the expiration date 
on the test strip bottle and the expiration 
and discard dates on the control solution 
bottle. DO NOT use expired materials. 
Level 1 and Level 2 control solutions are available and sold 
separately if not included in the meter kit. You can test your 
CONTOUR NEXT EZ meter with control solution when the 
temperature is 59°F – 95°F. Always use CONTOUR NEXT control 
solutions. Other brands could present incorrect results. If you 
need help locating CONTOUR NEXT control solutions, contact 
Customer Service. See Contact Information.
1. Wash and dry your hands well before and after testing.
2. Remove a test strip from the bottle and fi rmly snap the lid 
closed. 
3. Hold the test strip with the gray square 
end facing up.
4. Insert the gray square end into the test 
strip port on the meter. 
The meter will turn on. An image of a 
test strip with a fl ashing blood drop will 
appear letting you know the meter is 
ready to test.
5. Shake the control solution bottle well, 
about 15 times before every use.
6. Remove the bottle cap and use a tissue to wipe away any 
solution around the bottle tip before dispensing a drop.
7. Squeeze a small drop of control 
solution onto a clean nonabsorbent 
surface. Do not apply control 
solution to your fi ngertip or to the 
test strip directly from the bottle. 
8. Immediately touch the tip of the test strip to the drop of control 
solution. 
9. Hold the tip in the drop until the meter beeps. 
10. The meter will count down for 
5 seconds and the control test result 
will display on the meter. 
The meter will automatically recognize 
and mark the control result for you.
NOTE: The does not indicate the control test result is 
within the specifi ed range. It only marks it as a control 
solution test. It will not be included in your averages.
11. Compare your control test result to 
the ranges provided on the test strip 
bottle or carton.
To turn your meter off , simply remove the 
test strip. Dispose of the used test strip 
as medical waste. 
The result will be automatically marked 
and stored in the meter memory. Control 
results will not be included in your blood 
glucose averages or in the 7 day HI-LO 
summary.
CAUTION: If the result you get does not fall within the range 
listed on the test strip bottle label or carton, there may be 
a problem with the test strips, the meter or your testing 
technique.
If your control test result is out of range, do not use your 
CONTOUR NEXT EZ meter for blood glucose testing until you 
resolve the issue. Contact Customer Service. See Contact 
Information.
IMPORTANT: To ensure accuracy of your blood glucose test 
results and accuracy of your control test results, use only 
CONTOUR NEXT control solutions.
 SETUP AND USE 
Setting the Time, Date and Sound 
Your CONTOUR NEXT EZ meter allows you to set the date, time, 
and sound as you like them. 
• You can view 12 or 24 hour time; 
• month/day or day.month; and 
• choose to have your meter beep or not. 
To get into set-up mode when the meter is off , press and hold 
M for 3 seconds. A complete display will briefl y appear, then 
numbers will fl ash in the top left corner of the display. 
Setting the Time
With the 12H fl ashing on the display, 
press or to select the 12H (with AM 
and PM) or 24H setting.
Press Mto set.
The number in the hour position will 
now fl ash. Press  or until you reach 
the current hour.
Press Mto set.
The minute will now fl ash. Press  or 
to select the current minute. 
Press Mto set.
 SETUP AND USE 
3 
 SETUP AND USE 
If you selected a 12 hour time setting, the 
AM or PM will now fl ash. Press  or to 
select AM or PM. 
Press Mto set.
Setting the Month and Day Format
m/d will now fl ash on the display. Press 
 or to select the format you wish 
(m/d for month/day format or d.m for 
day.month format). 
Press Mto set.
Setting the Date
The year will be fl ashing. Press  or 
to select the current year.
Press Mto set.
The month will now fl ash. Press  or 
to select the month. 
Press Mto set.
The day will now fl ash. Press  or to 
select the day. 
Press Mto set.
 
Setting the Sound
The fl ashing bell symbol will now appear 
on the display along with the word .
This option controls whether you want to 
hear the beep sound during testing.
To turn the beep sound off  during testing, 
press or to display .
Press Mto set.
After setting the sound,  will fl ash. 
To keep your meter in Basic Mode, 
press Mto set. To change your meter to 
Advanced Mode, press or to change 
to , and press Mto set.
If you are using Basic Mode (L-1 will 
appear on your display), you have 
completed your meter set-up.
Set-up for Basic (L-1) and 
Advanced (L-2) Modes
Your new CONTOUR NEXT EZ meter is pre-set for the Basic Mode 
and off ers the same easy test procedure in either the Basic or 
Advanced Mode. 
BASIC MODE (L-1) ADVANCED MODE (L-2) 
• 7 day HI and LO summary
• 14 day average
• 480 test result memory
• 7 day HI and LO summary
• 7, 14, and 30 day averages
• 480 test result memory
• Pre- and post-meal 
markers
• 30 day pre- and post-meal 
averages
• Selectable post-meal 
reminders (2.5, 2.0, 1.5, 
1.0 hours)
• Personal HI and LO test 
settings
BASIC MODE (L-1) ADVANCED MODE (L-2) 
Your meter is pre-set to Basic 
Mode. If you have set your 
meter to the Advanced Mode, 
and want to return to Basic 
Mode: 
1. Press Mto turn your meter 
on. Wait to see the fl ashing 
test strip.
2. Press and hold or for 
3 seconds.
3. L-2 will fl ash. 
4. Press or to change to 
L-1.
5. Press Mto set.
To set your meter to 
Advanced Mode, and take 
advantage of meal markers 
and other functions available 
to you: 
1. Press Mto turn your meter 
on. Wait to see the fl ashing 
test strip.
2. Press and hold or for 
3 seconds.
3. L-1 will fl ash. 
4. Press or to change to 
L-2. 
5. Press Mto set.
Personalizing LO, HI and Reminder 
Settings
If you selected Advanced Mode 
(L-2 on your display) you can change 
your meter’s personal low and 
high blood glucose settings. By 
personalizing these settings, your 
CONTOUR NEXT EZ meter will help you 
track these HI and LO results. 
The LO setting will fl ash (default 
72 mg/dL). To change this setting, press 
 or until you reach your desired LO 
setting, and then press Mto set. 
(The range of values for LO is 
60 mg/dL – 90 mg/dL.)
Your HI blood glucose setting will now 
fl ash (default 180 mg/dL). To change, 
press or until you reach your 
desired HI setting. Press Mto set. 
(The range of values for HI is 
100 mg/dL – 250 mg/dL.) 
 
Post-Meal Test Reminder Set-up
To change the reminder time for post-
meal testing, press or to display 
the options available. Select 2.5, 2.0, 
1.5, or 1.0 hour(s) to set your post-meal 
reminder (default is 2.0 hours). 
Press Mto set.
You have now completed your 
CONTOUR NEXT EZ meter set-up for 
Advanced Mode!
Using the Markers
If you have set up your meter in 
Advanced Mode, your meter will 
display three markers after you test 
your blood. 
Pre-meal = Marks a test result 
that was taken before a meal.
Post-meal = Marks a test result 
that was taken after a meal. 
Logbook = The logbook marker can be used to mark any 
result as being unique or diff erent in some way, and for which 
you may want to make notes in your logbook. Seeing a result 
with this symbol next to it in the memory will help remind you 
that there is more information recorded about this test result.
Please see the following panels to use these options.
If you do not want to mark the test result
Remove the used test strip to automatically save the test result 
in the meter memory with no markers. This will also turn the 
meter off .
 GETTING STARTED 
Your CONTOUR®NEXT EZ meter 
Your CONTOUR NEXT test strip 
 GETTING STARTED 
1
 GETTING STARTED 
Table of Contents
1 GETTING STARTED
Your Meter and Test Strip 
Meter Display and Symbols
Meter Features
2 TESTING
Prepare the Test Strip and Lancing Device
Fingertip Testing
Understand Test Results
Clean and Disinfect Your Meter
Test with Control Solution
3 SETUP AND USE
Set Time, Date, and Sound
Set Basic (L-1) and Advanced (L-2) Modes
Customize LO, HI, and Reminder Settings
Mark Your Test Results
Transfer Results to a Computer
View Basic and Advanced Mode Results
4 TECHNICAL, SERVICE & CARE
Error Codes 
Replace the Batteries
Meter Care
High and Low Blood Glucose
Meter Specifi cations
Customer Service and Manufacturer Warranty
Order Meter Kit Supplies
WARNING
• The meter and lancing device are for single patient 
use. Do not share them with anyone including 
other family members! Do not use on multiple 
patients!1,2
• All parts of the kit are considered biohazardous 
and can potentially transmit infectious diseases, 
even after you have performed cleaning and 
disinfection.
• Always wash your hands well with soap and 
water and dry them well before and after testing, 
handling the meter, lancing device, or test strips. 
• For complete instructions on cleaning and 
disinfecting your meter and lancing device, see 
Cleaning and Disinfection in section 2, Testing. 
• This device is not intended for use in healthcare or 
assisted-use settings such as hospitals, physician 
offi  ces, or long-term care facilities because it has 
not been cleared by FDA for use in these settings, 
including for routine assisted testing or as part of 
glycemic control procedures.
• Use of this device on multiple patients may lead to 
transmission of Human Immunodefi ciency Virus 
(HIV), Hepatitis C Virus (HCV), Hepatitis B Virus 
(HBV), or other bloodborne pathogens.
CONTACT INFORMATION
Customer Service is available Monday through Sunday, 8:00 am 
through 12:00 Midnight, Eastern Time. If this is a medical emergency, please 
contact your health care professional or dial 911. To leave a message for Customer 
Service, remain on the line. One of our representatives will return your call as soon 
as possible.
1-800-348-8100
Distributed by: 
Ascensia Diabetes Care US Inc. 
5 Wood Hollow Road 
Parsippany, NJ 07054 
www.diabetes.ascensia.com
For information on patents and related licenses, 
see: www.patents.ascensia.com.
Ascensia, the Ascensia Diabetes Care logo, Clinilog, Contour, Glucofacts, 
and the No Coding logo are trademarks and/or registered trademarks of 
Ascensia Diabetes Care Holdings AG. 
All other trademarks are the property of their respective owners. 
While all characters are displayed, 
it is very important to verify that 
 is fully displayed. If there 
are missing segments, see Error 
Codes and Symbols in section 4,
Technical, Service & Care. This may 
aff ect the way you see your results.   
Icon What It Means 
Top area on the display presenting date and time. 
Pre-meal result marker. 
Post-meal result marker. 
 Identifi es a Logbook entry. 
Indicates a reminder has been set. 
Center area on the display that shows either your 
test result or any error codes. 
The meter is ready to test. 
Control solution result marker. 
Indicates it is too cold or too hot for your meter to 
test accurately.
Indicates Average.
Results are displayed as mg/dL or mmol/L. 
The 7 appears when viewing 7 day HI-LO summary. 
• Examine product for missing, damaged, or broken parts. 
If the test strip bottle is open inside the new box of strips, 
do not use those strips. Contact Customer Service for 
replacement parts. See Contact Information.
Blood Glucose Monitoring System
USER GUIDE
Uses only CONTOUR®NEXT
blood glucose test strips.
Online meter registration available at 
www.diabetes.ascensia.com or call our 
Customer Service.
90007891_CntrNxtEZ_UG_Pamphlet_EN_F1B1_v1 December 9, 2020 6:29 AM
90007891 Rev. 11/20
© 2020 Ascensia Diabetes Care Holdings AG. 
All rights reserved.
TURN OVER TO SEE BACK
2
23
30
16
9
21
28
14
7
19
26
12
5
3
24
31
17
10
1
22
29
15
8
20
27
13
6
4
18
25
32
11
DATE:
CLIENT:
DESCRIPTION:
SGK JOB#:
SKU#:
DIMENSIONS:
PRINTER SPEC:
RELEASE: V-1 LANGUAGE(S): ENGLISHINTERNAL: R-X
BAN#:90007891  Rev. 11/20
Cntr Nxt EZ Mtr UG Pamphlet - US
Alt#1 - 12/09/20 unmask
December 09, 2020
Ascensia Diabetes Care
Cntr Nxt EZ Mtr UG Pamphlet - English